Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences.

Slides:



Advertisements
Similar presentations
IN THE NAME OF GOD.
Advertisements

Comparison of intraocular lens power calculation formulas with LENSTAR LS 900® Tsunekazu Hamada Hamada Eye Clinic, Osaka, Japan Disclosure statement of.
TOPOGRAPHY ORBSCAN S.A.A. Mortazavi MD.
IN THE NAME MY GOD 1. Phacoemulsification in long and short eyes Dr. memarzadeh MD 1387/11/ Dr memarzadeh MD ophthalmologist 2.
Corneal topography orbscan
إعداد : أ. بلسم فهد صوفي،،،المصدر:محاضرات د.حنان 1 Ultrasound in Medicine Ch.4 Ultrasound pictures of the body.
Presented by: Ray T. Oyakawa, M.D. Sharper Vision Centers A MEDICAL GROUP Comparison of PalmScan and IOL Master axial lengths measurements No financial.
Accuracy of Predicted Refractive Error in Resident-Performed Cataract Surgery Using Partial Coherence Interferometry Nickolas P. Katsoulakis, M.D., Paul.
A-scan Biometry Introduction:-
Ultrasound is a high frequency sound that you cannot hear, but it can be emitted and detected by special machines. Ultrasound travels freely through fluid.
Development of an Apple iPhone IOL Calculator Application for Cataract Surgery Rony Gelman, MD 1, Richard E. Braunstein, MD 1 1 Department of Ophthalmology,
IOL power calculation following keratorefractive surgery
Yüksel Totan, Ramazan Yaĝcı, Zeynel Arslanyılmaz, Uĝurcan Keskin The authors have no financial interest.
Financial Disclosure I have a financial interest with the following companies: Abbott Medical Optics Alcon Calhoun Vision NuLens Optimedica Optivue  
Eltutar, Kadir; Akcetin, Tulay A.; Ozcelik, N. Demet Istanbul Education and Research Hospital Department of Ophthalmology The authors state that they have.
IOLs & Biometry Alireza Peyman, MD Isfahan University of Medical Sciences.
INTRA-OCULAR LENS POWER CALCULATION IN POST- REFRACTIVE SURGERY ASIAN EYES USING THE HAIGIS-L FORMULA Dr Daphne Han, FRCS Dr Wei-Han Chua, FRCS Dr Peter.
Choosing the Proper Power for the IOL
IOL power calculation after refractive surgery A. Peyman, MD.
Personalised eye modelling for customised intraocular lens designs Matthew Sheehan, Eamonn O’Donoghue, Conor Sheil and Alexander Goncharov Photonics Ireland,
The Human Eye Written for Physics 106 Friday, Nov. 7, 2008 A good web site for learning much about the anatomy of the human eye is :
JAIRO E. HOYOS, MD MELANIA CIGALES, MD Instituto Oftalmológico Hoyos Barcelona - SPAIN.
PIGGYBACK IMPLANTATION OF FOR HIGH HYPEROPIA 24 MONTHS FOLLOW UP JL. FEBBRARO I.VIELPEAU, F. KRIFA, S.BERBACHE, S. GANEM Department of Ophthalmology S.
Hong Kong Eye Hospital Biometry Audit 2012 SN60WF IOL Dr. Rose Chan
Phacoemulsification in eyes with previous anterior chamber phakic IOL surgery Walton Nosé, MD, PhD 1,2 Adriana dos Santos Forseto, MD 1 Mariana Ávila,
Hong Kong Eye Hospital Ms Frenchy Chiu Dr Victoria Wong IOL master
Seyed Javad Hashemian MD Eye Research Center Rasoul Akram Hospital Tehran University of Medical Sciences No financial interest
Hong Kong Eye Hospital Biometry Audit 2011 SN60WF IOL Dr. Rose Chan Resident, Hong Kong Eye Hospital.
Biometric Accuracy in High Hypermetropes and Myopes
Phacoemulsification some Basic Ideas…
Psychology 4051 The Eye/Optics. Light The stimulus for vision is light energy-a small part of the spectrum of electromagnetic radiation. Other organisms.
Optical Coherence Tomography for Retinal Imaging ECE 172A Julio Flores.
Artemis tm II VHF Ultrasound for Sizing the Visian ICL ASCRS 2008 Philip C. Roholt, MD N. Canton, Ohio, USA The author has no proprietary interest in products.
INTRAOCULAR LENS POWER CALCULATION BY IMMERSION A-SCAN BIOMETRY VERSUS CONTACT A-SCAN BIOMETRY MEASUREMENTS BEFORE CATARACT SURGERY Burak Bilgin**, M.D.,
Phakic posterior chamber IOLs 1.Cells involved in crystalline lens and capsular bag opacification 2.Evolution of designs of PPCIOLs 3.Relevant aspects.
Paul “Butch” Harton, MD Harbin Clinic Eye Center Rome, GA USA ASCRS Boston Poster, 2010 The Author has no financial interest in the subject matter of this.
Endothelial Dystrophy Association with Hyperopia Glenn W. Thompson, M.D. Timothy P. Page, M.D. Authors have no financial interest.
Phoebe D. Lenhart, M.D. 1, Amy K. Hutchinson, M.D. 1, Michael J. Lynn, M.S. 2, Scott R. Lambert M.D. 1 1 Department of Ophthalmology, Emory University,
EVALUATION OF REFRACTIVE OUTCOMES AFTER IOL IMPLANTATION
Edmondo Borasio Moorfields Eye Hospital Dubai Results of BESSt 2 biometry and Borasio Myopic/Hyperopic Regression (BMR / BHR) IOL POWER CLUB – 9 th ANNUAL.
Partial Coherence Interferometry Failure Rate in a Teaching Hospital Leslie A. Wei 1,2, BA, Nickolaus P. Katsoulakis 2, MD, Theodoros Filippopoulos 3,
Adriana S. Forseto1, MD Walton Nosé1,2, MD
Accutome A-scan Plus Connect ® Visual Aid
IOL Calculations Based on Partial Biometry in Humanitarian Missions Joseph Schmitz, MD Kimberly Davis, MD, FACS Scott McClatchey, MD The authors have no.
Sherman W. Reeves, MD, MPH 1,3,4 ; Jacob A. Kozisek, OD 1,2 ; Noumia Cloutier-Gill, OD 1,2 ; David R. Hardten, MD 1,2,3,4 Accuracy of Scheimpflug Imaging.
Comparison of Central Corneal Thickness, Anterior Chamber Depth, and Central Corneal Power Measurements between Two Scheimpflug Imaging Systems Yuichi.
Comparision of ultrasound A- scan with optical A-scan Dr.Somshekar.N Consultant Refractive Services Narayana Nethralaya, Bangalore No Finanacial Interest.
Comparison of visual function following piggyback implantation of Acrysof ReSTOR intraocular lenses with Tecnis multifocal ZM900 intraocular lenses. Rodrigo.
COMPARISON AND EVALUATION OF OCULAR BIOMETRY USING A NEW NON-CONTACT OPTICAL LOW COHERENCE REFLECTOMETER David Goldblum 1,3, Kaspar Rohrer 1, Rudolf Waelti.
Simulated Experiments on IOL Power Calculation Using Anterior Segment OCT Dong Hyun Jo, M.D., 1,2 Mee Kum Kim, M.D., 1,2 Won Ryang Wee, M.D. 1,2 1 Department.
Nebraska Laser Eye Associates Mark Johnston MD FRCSC
Poster# P36 Comparison of Techniques of Measuring Anterior Chamber Depth and Corneal Curvature in Indian Eyes Sujata Das, MS, FRCS Anjula Kumari, B.Optom.
Sergio Kwitko, Diane Marinho, Samuel Rymer Ophthalmology Department - Hospital de Clínicas de Porto Alegre Federal University of Rio Grande do Sul - Porto.
Intraocular Lens Outcomes: Comparison of Technologies and Formulas Carolina Eyecare Physicians, LLC Research Assistant Professor of Ophthalmology Storm.
Marilyn Watkins-Ramdin, COMT ROUB
Volume 12, Issue 2, Pages (April 2008)
Premium IOL May be one way to achieve the visual goals of selected patients I describe some pearls for premium IOL implantation to help ensure that we.
Corneal Pachymetry in Prediction of Refraction After Cataract Surgery
Visual Aid.
Refractive outcomes of intraoperative wavefront aberrometry versus optical biometry alone for intraocular lens power calculation Zina Zhang MD1, Logan.
The authors have no financial interest.
Aug 11, 2017 Comparison of predictive accuracy on Partial coherence interferometry (PCI) and swpt-source optical coherence tomographry (SS-OCT) Choun-Ki.
Effect of Axial Length Measurement Method on Refractive Outcomes of Cataract Surgery: Real World Comparison of Partial Coherence Interferometry and Immersion.
MP Weikert, M Shirayama, L Wang, DD Koch
Early Experience with Descemet’s Stripping Automated Endothelial Keratoplasty Combined with Phacoemulsification: Clinical and Refractive Outcome University.
IN THE NAME OF GOD.
Deepak Vayalambrone FRCOphth
Biometry before catarct surgery
SUBLUXATION LENS, A NO-RING APPROACH
Michael Goodman, Alexandra Paul and Andrew Hsu
Presentation transcript:

Biometry: Long & Short eyes Alireza Peyman, MD Isfahan University of Medical Sciences

Normal Range: Axial length  Biggest source of error in IOL power calculations. 96% of axial lengths fall within the range 21.0 to 25.5 mm 60% of AL is between 22.5 and 24.5 mm

Ultrasounds are calibrated with average velocities for normal length eyes. These velocities are incorrect for short eyes, causing significant measurement errors. Accurate measurement of axial length in hyperopic eyes is especially important since any error is greatly magnified in proportion to the length of the eye. Immersion or LASER biometry can provide superior results in these cases.

In a hyperopic eye with axial length of 20 millimeters each mm error in measurement cause 3.75 D post operative refractive error.

Difference in axial length of more than 0.3mm requires confirmation Most individuals have similar axial lengths in each eye except: –unilateral refractive error –coloboma –Staphyloma

In the axial myopia, the presence of a staphyloma should be considered, especially if there is difficulty obtaining a distinct retinal spike during A-scan. an immersion A/Bscan approach to axial length measurement in the setting of a posterior staphyloma

nearly all eyes with pathologic myopia have some form of posterior staphyloma. For highly myopic eyes (axial > 28mm) a B-scan should be carried out to determine the presence or otherwise of staphylomata

Long eye, foveal deep

Very long eye, posterior pole staphyloma

Buphthalmic globe: very long eye (37·5 mm): deep anterior chamber.

Megalocornea: average length eye (23·1 mm): deep anterior chamber (5·2 mm).

The Ascan vector is adjusted to pass through the middle of the cornea as well as the middle of the anterior and posterior lens echoes to assure that the vector will intersect the retina in the region of the fovea. IOL Master permits accurate measurements when posterior staphylomata are present. Since the patient fixates along the direction of the measuring beam.

Calculation formula: The SRK/T and Holladay formulae worked best overall. SRK I and II were found to be least accurate No statistical difference was found between SRK/T, Hoffer Q and the Holladay formulae

Axial LengthFormula <22 mmHoffer Q or SRK/T 22 – 24.5 mmSRK/T or Holloday, Haigis > 24.6SRK/T

When using IOL Master do not forget For AL < 22 do immersion A scan For AL > 30 do A and B scan

Silicone oil filled vitreous cavity: eye measures 49 mm on B-scan, but actual axial length is 34·3 mm.

Thank you for your attention