Valery Ray Particle Beam Systems & Technology, Methuen, USA High-Throughput Milling of HAR Vias with Concentrated Gas Delivery PBS&T FIB User GroupISTFA 2005 San Jose, CA
5/20/2015 FIB User Group, ISTFA Concepts of Concentrated Gas Delivery Commercial “Shroud” concentrator “Cupola” concentrator
5/20/2015 FIB User Group, ISTFA Imaging through Shroud Concentrator: FOV 20, 5.8pA Beam, Dielectric Sample Images taken through Shroud gas concentrator (left) have lower S/N then images taken with Penta nozzle (right).
5/20/2015 FIB User Group, ISTFA Main Chamber Vacuum as Function of Gas Pressure with “Shroud” Concentrator Factor of ~35 pressure reduction in main chamber of FIB system CCIG, Torr
5/20/2015 FIB User Group, ISTFA Comparison of 5µm via Milling in SiO2 with XeF 2, Penta (2 Torr) vs. Shroud (0.8 Torr) Dose, nC Time, mm:ss Mill box, µm Factor ~ 3.5 throughput enhancement
5/20/2015 FIB User Group, ISTFA Endpointing with Gas Concentrator: Secondary Electron Signal Secondary electron endpoint of 5µm deep via made by 0.3x0.3µm mill box, 5.8 pA beam current, and 0.8 Torr XeF 2 gas pressure, shroud gas concentrator. 0V bias 2V bias
5/20/2015 FIB User Group, ISTFA Endpointing with Gas Concentrator: Sample Absorbed Current Stage current endpoint of 5µm deep via made by 0.3x0.3µm mill box, 5.8 pA beam, and 0.8 Torr XeF 2 gas pressure.
5/20/2015 FIB User Group, ISTFA Indirect Comparison of Via Profile: Penta vs. Shroud (Beehive) FOV 5µm Vias milled by Beehive are expected to have straighter profile then vias milled by Penta
5/20/2015 FIB User Group, ISTFA Summary Concentrated gas delivery is viable for milling of multiple access vias in FIB Circuit Edit. Provides enhanced throughput and reduced pressure in main chamber. Endpoint is reliable with sample absorbed current and secondary electrons.