Fricke gel dosimeters for the measurement of the anisotropy function of a HDR Ir-192 brachytherapy source Mauro Carrara 1, Stefano Tomatis 1, Giancarlo.

Slides:



Advertisements
Similar presentations
How is radiotherapy given Radiotherapy can be given in two ways: from outside or inside the body. External radiotherapy is the most common method of treatment.
Advertisements

The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
Cirrone G. A. P. , Cuttone G. , Raffaele L. , Sabini M. G
Chapter 4 Radiation Dosimeters
Introduction to x-rays
Giorgio Russo National Research Council, Institute of Bioimaging and Molecular Imaging (IBFM) Fondazione Istituto San Raffaele G. Giglio di Cefalù Istituto.
Estimation of TLD dose measurement uncertainties and thresholds at the Radiation Protection Service Du Toit Volschenk SABS.
14th GEANT4 User and Collaboration Workshop - Catania, Oct. 15/ B.Caccia, G.Frustagli, M.Mattia, S.Valentini Istituto Superiore di Sanita' e INFN,
Determination of Concentration Using Spectrophotometry
Alexander Klyachko, IU Cyclotron, PTCOG51, Seoul, South Korea, May 17-19, 2012 A.V. Klyachko 1, D.F. Nichiporov 1, L. Coutinho 2, C.-W. Cheng 2, 3, M.
Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warsaw, Poland 8th ECMP, Athens, Dosimetry audits in radiotherapy.
A general purpose dosimetric system for brachytherapy
P. Pérez, V. Galván, G. Castellano & M. Valente.  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol.
Lic. Pedro Antonio Pérez.  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol Orange added Fricke.
Introduction A Fricke gel was the first dosimetric system to offer the possibility of making 3-D radiation dose measurements in a totally non-destructive.
MONTE CARLO RADIATION DOSE SIMULATIONS AND DOSIMETRY COMPARISON OF THE MODEL 6711 AND I BRACHYTHERAPY SOURCES Mark J. Rivard Department of Radiation.
1 M.G. Pia et al. The application of GEANT4 simulation code for brachytherapy treatment Maria Grazia Pia INFN Genova, Italy and CERN/IT
Prof.Dr. Gehan Mosaad. At the end of this lecture the student must be able to:  Define electrotherapy, electric current and electromagnetic spectrum.
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Spectrophotometry: An Analytical Tool. PGCC CHM 103 Sinex IoIo I Cell with Pathlength, b, containing solution light source detector blank where I o =
The brachytherapy advanced example Susanna Guatelli (CERN/INFN)
Lecture 2b. Electromagnetic Spectrum Visible range: = nm Ultraviolet: = nm Low energyHigh energy.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Results The measured-to-predicted dose ratio criteria used by the RPC to credential institutions is , however for this work, a criteria of
Determination of Concentration Using Spectrophotometry
Introduction to Nuclear Medicine
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Introduction A modification of the Fricke solution system to make it suitable for low-level dosimetric studies was proposed about three decades ago [1].
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
Computational Model of a Capillary: The effect of geometry on hemoglobin measurements Eugene Joh Medical Biophysics 6-Week Project Supervisor: Dr. Dan.
M.G. Pia et al. Brachytherapy at IST Results from an atypical Comparison Project Stefano Agostinelli 1,2, Franca Foppiano 1, Stefania Garelli 1, Matteo.
Milestones M5 and M6 Mid-Term Review, October 2013M. Silari – ARDENT overview1 M5: comparison of detector technologies (WP1) M6: choice of detector technology.
16 th International Congress on Neutron Capture Therapy June 14-19, 2014 Helsinki, Finland Study of suitability of Fricke-gel-layer dosimeters for in-air.
G. Bartesaghi, 11° ICATPP, Como, 5-9 October 2009 MONTE CARLO SIMULATIONS ON NEUTRON TRANSPORT AND ABSORBED DOSE IN TISSUE-EQUIVALENT PHANTOMS EXPOSED.
Ye, Sung-Joon, Ph.D. Ove, Roger, M.D., Ph.D.; Shen, Sui, Ph.D.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Investigation of 3D Dosimetry for an Anthropomorphic Spine Phantom R. Grant 1,2, G. Ibbott 1, J. Yang 1, J. Adamovics 3, D Followill 1 (1)M.D. Anderson.
Congresso del Dipartimento di Fisica Highlights in Physics –14 October 2005, Dipartimento di Fisica, Università di Milano An application of the.
1 5-9 October th ICATPP, Como, Italy S. Maltezos NITROGEN MOLECULAR SPECTRA OF AIR FLUORESCENCE EMULATOR USING A LN 2 COOLED CCD S. Maltezos, E.
1 UV-Vis Absorption Spectroscopy Lecture Measurement of Transmittance and Absorbance: The power of the beam transmitted by the analyte solution.
Brachytherapy and GYN malignancy
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Determining Radiation Intensity
Determination of Concentration Using Spectrophotometry
Goddard February 2003 R.Bellazzini - INFN Pisa A new X-Ray Polarimeter based on the photoelectric effect for Black Holes and Neutron Stars Astrophysics.
Investigation into the radiochromic (FXG) gel dosimeter: stability and uncertainty in optical measurements M A Bero, S J Doran and W B Gilboy Department.
1 Neutron Effective Dose calculation behind Concrete Shielding of Charge Particle Accelerators with Energy up to 100 MeV V. E Aleinikov, L. G. Beskrovnaja,
TLD POSTAL DOSE QUALITY AUDIT FOR 6MV AND 15MV PHOTON BEAMS IN RADIOTHERAPY CLINICAL PRACTICE Sonja Petkovska 1, Margarita Ginovska 2, Hristina Spasevska.
Measurement of Radiation: Instrumentation&Techniques Part Three
Absorption Spectroscopy CHEM 251 Week of November 1 st, 2010 Alexis Patanarut.
Purpose N-isopropylacrylamide (NIPAM) polymer gel dosimeters were employed to verify the dose distribution of clinical intensity modulated radiation therapy.
Methods Conclusions References ResultsIntroduction After all tests were performed, the optimal tolerance value was This tolerance value had an overall.
Utilizing Spectrophotometry in Life Science Date: Subject:
J Cho, G Ibbott, M Kerr, R Amos, and O Mawlawi
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
6th European ALARA Network Workshop Madrid, oct. 23rd 2002
Dr: Mohamed Afifi By Lecturer Radiological Science
Compact readout unit for the EP-RADMON radiation monitoring sensors
Development and characterization of the Detectorized Phantom for research in the field of spatial fractionated radiation therapy. D. Ramazanov, V. Pugatch,
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
Distributed Optical Fibre Radiation and Temperature Sensing at High Energy Accelerators and Experiments Update on installation of DOFRS in PS-Booster.
Lecture 2b Beer’s Lambert Law.
Resident Physics Lectures (Year 1)
Resident Physics Lectures (Year 1)
By: Vi Nhan Nguyen University of Arizona
Geant4 at IST Applications in Brachytherapy
Spectrophotometry A method to determine concentration of a species exploiting the absorption of EMR.
Presentation transcript:

Fricke gel dosimeters for the measurement of the anisotropy function of a HDR Ir-192 brachytherapy source Mauro Carrara 1, Stefano Tomatis 1, Giancarlo Zonca 1, Grazia Gambarini 2,3, Giacomo Bartesaghi 2,3, Chiara Tenconi 2, Annamaria Cerrotta 1, Carlo Fallai 1, 1 Fondazione IRCCS Istituto Nazionale dei Tumori di Milano 2 Dipartimento di Fisica, Università degli Studi di Milano 3 Istituto Nazionale di Fisica Nucleare, Milano

Brachytherapy (from the Greek brachios, meaning “short”) is a form of radiotherapy where one or more sealed radioactive sources are placed inside or next to the area requiring treatment. Brachytherapy

HDR-Brachytherapy Ir-192 source (initial activity: 10Ci) In high dose rate (HDR) brachytherapy a single sealed source (usually Ir-192) is adopted to deliver radiation to the target with a dose-rate of at least 12 Gy/h.

The anisotropy function Treatment planning systems (TPS) are adopted in clinical practice to optimize source dwell times and positions inside the catheters, with the aim of conforming the prescribed dose to the target volume. TPS dose calculation algorithm:

The anisotropy function The anisotropy function F(r,θ) has been recommended to account for the dose distribution anisotropy that results adopting sealed sources. This function is implemented in the AAPM dose calculation formalism and is widely adopted by the currently available treatment planning systems. Radiation propagation is anisotropic due to: source self-absorbance absorbance by the capsule F(r,θ)

To develop and verify a method based on Fricke gel layer dosimetry for the characterization of the anisotropy function F(r,θ) of an Ir-192 source. Purpose of the work

Fricke gels are radiochromic tissue-equivalent dosimeters that can be established in form of layers. Fricke gel layer dosimeters FGLDs are prepared in our laboratory by infusing a ferrous sulphate solution and the metal-ion indicator Xylenol Orange (XO) in a tissue- equivalent gel matrix. Exposure to ionizing radiation of a FGLD produces a conversion of ferrous ions Fe 2+ into ferric ions Fe 3+ and the complex XO-Fe 3+ causes visible light absorption around 585nm, with yield proportional to the absorbed dose. Fe 2+ Fe 3+ RADIATION Fricke gel layer dosimetry (FGLD)

Optical imaging is performed by means of visible-light transmittance analysis. Transmitted light images are detected with a CCD camera. CCD Camera Gel-layer Plane uniform light source Computer FGLD image acquisition system

with Δ(OD) related to the absorbed dose D Δ(OD) = (OD) ir – (OD) ni = log 10 (I ni / I ir ) (OD) ni = -log 10 (I 0 / I ni ) I0I0 I ni Before irradiation (ni) I0I0 I ir After irradiation (ir) (OD) ir = -log 10 (I 0 / I ir ) FGLD analysis

FGLD CHARACTERISATION

tissue-equivalent phantom Catheter FGLD Irradiation set-up: FGLD characterisation Source

Lack of linearity at doses lower than 400 cGy Saturation at doses higher than 2800 cGy FGLD characterisation

Response saturation for dose- rates higher than 400 cGy/min FGLD characterisation

At doses lower than 400 cGy, the  (OD) response to the dose is not linear At doses lower than 400 cGy, the  (OD) response to the dose is not linear At doses higher than 2800 cGy and at dose-rates higher than 400 cGy/min, the  OD) response to the dose saturates At doses higher than 2800 cGy and at dose-rates higher than 400 cGy/min, the  OD) response to the dose saturates The dose response is independent on the energy over most of the adopted energy range The dose response is independent on the energy over most of the adopted energy range Each single dosimeter may require a specific calibration Each single dosimeter may require a specific calibration FGLD characterisation

FGLD calibration procedure I.II cGy k I. irradiation: elimination of the non-linear behaviour at low doses II. irradiation: determination of the sensitivity factor k A double 400cGy pre-irradiation of each FGLD resulted to be the optimal procedure for calibration. Applying this procedure, the obtained calibration curves were straight lines crossing the origin. D [cGy]

ANISOTROPY FUNCTION MEASUREMENT

Irradiation set-up: FGLD Tissue-equivalent phantom x y z Ir-192 source A series of measurements of the same irradiation set-up composed of a tissue-equivalent phantom and a FGLD with a built-in plastic catheter were performed. The plastic catheter permitted us to convey the source directly inside the dosimeter. Anisotropy function measurement

The delivered dose was of 1500cGy at 10mm distance from the source. Images of the irradiated dosimeters were acquired and elaborated with a dedicated software developed in Matlab®. Anisotropy function measurement

The measured anisotropy function at radial distances of 15mm and 20mm at angles between 15° and 165°: Anisotropy function measurement

The measured anisotropy function at radial distances of 25mm and 30mm at angles between 15° and 165°: Anisotropy function measurement

The measured anisotropy function at radial distances of 15mm and 20mm at angles between 15° and 165°: Anisotropy function measurement

The measured anisotropy function at radial distances of 25mm and 30mm at angles between 15° and 165°: Anisotropy function measurement

The measured anisotropy function at radial distances of 35mm, 40mm, 45mm and 50mm at angles between 15° and 165°: Anisotropy function measurement

 At short distances from the source, percentage differences between tabulated and measured data are almost always smaller than 3%.  At increasing distances, data become more scattered due to the low doses delivered to the dosimeter.  Measurements with a higher irradiation time will be performed to achiever better accuracy at higher distances from the source  A FGLD results to be an accurate tool for the Ir-192 anisotropy function measurement. This instrument could be easily adopted in QA protocols to verify the anisotropy function for each newly installed Ir-192 source. Discussion and Conclusions

Thank you for your attention Vancouver Island 2009