Principles of Biomedical Systems & Devices 0909.504.04 / 0909.402.02 WEEK 10: Flow & Volume of Blood.

Slides:



Advertisements
Similar presentations
Cardiac Output Prof. K. Sivapalan 2013 Cardiac output.
Advertisements

Foundations of Medical Ultrasonic Imaging
Flow and Volume of Blood Dept. of Biomedical Engineering YOUNHO HONG.
There are three major types of Blood flow Doppler patterns: 1) Continuous Wave 2) Pulse Wave 3) Color Flow DOPPLER ULTRASOUND.
Review 29:008 Exam 3. Ch. 12 Electrostatic Phenomena.
إعداد : أ. بلسم فهد صوفي 1 Ultrasound in Medicine Ch.3 Ultrasound pictures of the body.
Maxwell’s Equations The two Gauss’s laws are symmetrical, apart from the absence of the term for magnetic monopoles in Gauss’s law for magnetism Faraday’s.
Chapter 8. Measurement of Flow and Volume of Blood John G. Webster
إعداد : أ. بلسم فهد صوفي،،،المصدر:محاضرات د.حنان 1 Ultrasound in Medicine Ch.4 Ultrasound pictures of the body.
Chapter 5 Part Three Blood Pressure and flow by Ibrhim AlMohimeed BMTS /3/2013.
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
ECE 501 Introduction to BME
Copyright © Cengage Learning. All rights reserved. 8 Further Applications of Integration.
1 Sinusoidal Waves The waves produced in SHM are sinusoidal, i.e., they can be described by a sine or cosine function with appropriate amplitude, frequency,
Absorption and emission processes
Fiber-Optic Communications James N. Downing. Chapter 2 Principles of Optics.
Ultrasound Imaging Atam Dhawan.
The Doppler Effect The Doppler effect is the shift in frequency of a wave that occurs when the wave source, or the detector of the wave, is moving. Applications.
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Closed Conduit Measurement Techniques ä Pipeline systems ä pipe networks ä measurements.
FURTHER APPLICATIONS OF INTEGRATION
Electromagnetic Flowmeter Why use flow meters What is a magnetic flow meter – How do they work – How are the different from other meters Items that will.
1 Syafruddin Hasan. 2 Systems Concepts Automated systems that control flow rates or volume employ the following components: Automated systems that control.
1 Syafruddin Hasan. 2 Systems Concepts Automated systems that control flow rates or volume employ the following components: Automated systems that control.
Flow Sensors.
Ultrasonic Testing This technique is used for the detection of internal surface (particularly distant surface) defects in sound conducting.
Ultrasound – Physics & Advances
DOPPLER SONOGRAPHY By Dr/ Dina Metwaly.
Doppler Radar From Josh Wurman Radar Meteorology M. D. Eastin.
4.4.1 Wave pulse: a wave pulse is a short wave with no repeated oscillations Progressive wave: a wave that moves through a medium transferring energy as.
Rudolf Žitný, Ústav procesní a zpracovatelské techniky ČVUT FS 2010 Flowrate Experimental methods E EXM5 Some pictures and texts were copied from.
Physics Subject Area Test WAVES LIGHT & OPTICS.
© From J. G. Webster (ed.), Medical instrumentation: application and design. 3 rd ed. New York: John Wiley & Sons, Figure 8.1 Several methods of.
ELECTRICAL CIRCUIT ET 201 Define and explain characteristics of sinusoidal wave, phase relationships and phase shifting.
Instructor: Lichuan Gui
생체계측 II Report # 송성진 Medical Instrumentation II.
Waves. The Nature of Waves What is a mechanical wave?  A wave is a repeating disturbance or movement that transfers energy through matter or space 
The Hong Kong Polytechnic University Optics II----by Dr.H.Huang, Department of Applied Physics1 Light Waves Nature of Light: Light can be viewed as both.
EE 5340/7340, SMU Electrical Engineering Department, © Carlos E. Davila, Electrical Engineering Dept. Southern Methodist University slides can be.
Display of Motion & Doppler Ultrasound
1 Physics 211 – lecture 28: Sound Waves Sound Waves - mechanical longitudinal waves  Sound waves come from periodic pressure variations moving along in.
Resident Categorical Course

Chapter 13 - Sound 13.1 Sound Waves.
Display of Motion & Doppler Ultrasound
Lesson 10 The Fields of Accelerating Charges. Class 29 Today we will: learn about threads and stubs of accelerating point charges. learn that accelerating.
Sarah Gillies Ultrasound Sarah Gillies
Waves Superposition and Standing Waves The Electromagnetic Spectrum Pulse-Echo Techniques Refraction Polarisation Diffraction.
Today Course overview and information 09/16/2010 © 2010 NTUST.
Chapter 8 Charges in Magnetic Fields. Introduction In the previous chapter it was observed that a current carrying wire observed a force when in a magnetic.
Chapter 11 Vibrations and Waves.
Laws of Thermodynamics Gibb’s Free Energy Applications of Ultrasound X-Ray imaging
Blood Flow Detection > by the Doppler principle > by Electromagnetic principles © D. J. McMahon 2014 rev
Ultrasound Learning Objectives: Describe the properties of ultrasound;
Physics Mrs. Dimler SOUND.  Every sound wave begins with a vibrating object, such as the vibrating prong of a tuning fork. Tuning fork and air molecules.
Electromagnetic Blood Flow Meters Measures instantaneous pulsatile flow of blood Works based on the principle of electromagnetic induction The voltage.
Magnetic Induction 1Physics is Life. Objectives To learn how magnetic fields can produce currents in conductors To understand how this effect is applied.
Electromagnetic Induction
Range & Doppler Accuracy by Ramya R.
TWI Ultrasonic Testing Part 1.
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN
Cardiac Output And Hemodynamic Measurements
Biomedical Instrumentation
Ultrasound.
Measurement of Flow and Volume
Properties of sound.
MCQs (1) – Questions Ultrasound propagates through the medium as a transverse wave Ultrasound velocity is equal to the product of frequency and wavelength.
Figure 8.1 Several methods of measuring cardiac output In the Fick method, the indicator is O2; consumption is measured by a spirometer. The arterial-venous.
Principles of cardiovascular measurement I and II
SPECTROPHOTOMETRY Applied Chemistry.
Copyright © Cengage Learning. All rights reserved.
Presentation transcript:

Principles of Biomedical Systems & Devices / WEEK 10: Flow & Volume of Blood

Measurement of Flow & Volume of Blood  A measurement of paramount importance: concentration of O 2 and other nutrients in cells  Very difficult to measure  Second-class measurement: blood flow and changes in blood volume  correlate well with concentration  Third-class measurement: blood pressure  correlates well with blood flow  Fourth class measurement: ECG  correlates adequately with blood pressure  How to make blood flow / volume measurements? Standard flow meters, such as turbine flow meters, obviously cannot be used!  Indicator-dilution method: cont./rapid injection, dye dilution, thermodilution  Electromagnetic flowmeters  Ultrasonic flowmeters / Doppler flowmeters  Plethysmography: Chamber / electric impedance / photoplethysmography

Indicator Dilution with Continuous Injection  Measures flow / cardiac output averaged over several heart beats  Fick’s technique: the amount of a substance (O 2 ) taken up by an organ / whole body per unit time is equal to the arterial level of O 2 minus the venous level of O 2 times the blood flow  Blood flow, liters/min (cardiac output) Consumption of O 2 (mL/min) Arterial and venous concentration of O 2 (mL/L of blood) Change in [] due to continuously added indicator m to volume V

Fick’s technique  How is dm/dt (O 2 consumption) measured?  Where and how would we measure C a and C v ? (Exercise)

Indicator Dilution with Rapid Injection  A known amount of a substance, such as a dye or radioactive isotope, is injected into the venous blood and the arterial concentration of the indicator is measured through a serious of measurements until the indicator has completely passed through given volume.  The cardiac output (blood flow) is amount of indicator injected, divided by average concentration in arterial blood.

Indicator – Dilution Curve After the bolus is injected at time A, there is a transportation delay before the concentration begins rising at time B. After the peak is passed, the curve enters an exponential decay region between C and D, which would continue decaying alone the dotted curve to t 1 if there were no recirculation. However, recirculation causes a second peak at E before the indicator becomes thoroughly mixed in the blood at F. The dashed curve indicates the rapid recirculation that occurs when there is a hole between the left and right sides of the heart.

An Example

Dye Dilution  In dye-dilution, a commonly used dye is indocyanine green (cardiogreen), which satisfies the following  Inert  Safe  Measurable though spectrometry  Economical  Absorption peak is 805 nm, a wavelength at which absorption of blood is independent of oxygenation  50%of the dye is excreted by the kidneys in 10 minutes, so repeat measurements is possible

Thermodilution  The indicator is cold – saline, injected into the right atrium using a catheter  Temperature change in the blood is measured in the pulmonary artery using a thermistor  The temperature change is inversely proportional to the amount of blood flowing through the pulmonary artery

Measuring Cardiac Output Several methods of measuring cardiac output In the Fick method, the indicator is O 2 ; consumption is measured by a spirometer. The arterial-venous concentration difference is measure by drawing simples through catheters placed in an artery and in the pulmonary artery. In the dye-dilution method, dye is injected into the pulmonary artery and samples are taken from an artery. In the thermodilution method, cold saline is injected into the right atrium and temperature is measured in the pulmonary artery.

Electromagnetic Flowmeters  Based on Faraday’s law of induction that a conductor that moves through a uniform magnetic field, or a stationary conductor placed in a varying magnetic field generates emf on the conductor: When blood flows in the vessel with velocity u and passes through the magnetic field B, the induced emf e measured at the electrodes is. For uniform B and uniform velocity profile u, the induced emf is e=BLu. Flow can be obtained by multiplying the blood velocity u with the vessel cross section A.

Electromagnetic Flowmeter Probes Comes in 1 mm increments for 1 ~ 24 mm diameter blood vessels Individual probes cost $500 each Made to fit snuggly to the vessel during diastole Only used with arteries, not veins, as collapsed veins during diastole lose contact with the electrodes Needless to say, this is an INVASIVE measurement!!! A major advantage is that it can measure instantaneous blood flow, not just average flow

Ultrasonic Flowmeters  Based on the principle of measuring the time it takes for an acoustic wave launched from a transducer to bounce off red blood cells and reflect back to the receiver.  All UT transducers, whether used for flowmeter or other applications, invariably consists of a piezoelectric material, which generates an acoustic (mechanical) wave when excited by an electrical force (the converse is also true)  UT transducers are typically used with a gel that fills the air gaps between the transducer and the object examined

Near / Far Fields  Due to finite diameters, UT transducers produce diffraction patterns, just like an aperture does in optics.  This creates near and far fields of the UT transducer, in which the acoustic wave exhibit different properties  The near field extends about d nf =D 2 /4λ, where D is the transducer diameter and λ is the wavelength. During this region, the beam is mostly cylindrical (with little spreading), however with nonuniform intensity.  In the far field, the beam diverges with an angle sinθ=1.2 λ/D, but the intensity uniformly attenuates proportional to the square of the distance from the transducer Higher frequencies and larger transducers should be used for near field operation. Typical operating frequency is 2 ~ 10 MHz.

UT Flowmeters Zero-crossing detector / LPF Determine direction High acoustic impedance material

Transit time flowmeters Effective velocity of sound in blood: velocity of sound (c) + velocity of flow of blood averaged along the path of the ultrasound (û) û=1.33ū for laminar flow, û=1.07ū for turbulent flow ū: velocity of blood averaged over the cross sectional area, this is different than û because the UT path is along a single line not over an averaged of cross sectional area Transit time in up/down stream direction: Difference between upstream and downstream directions

Transit Time Flowmeters The quantity ∆T is typically very small and very difficult to measure, particularly in the presence of noise. Therefore phase detection techniques are usually employed rather then measuring actual timing.

Doppler Flowmeters  The Doppler effect describes the change in the frequency of a received signal, with respect to that of the transmitted signal, when it is bounced off of a moving object.  Doppler frequency shift Speed of sound in blood (~1500 m/s) Angle between UT beam and flow of blood Speed of blood flow (~150 cm/s) Source signal frequency

Doppler Flowmeters

Problems Associated with Doppler Flowmeters  There are two major issues with Doppler flowmeters  Unlike what the equations may suggest, obtaining direction information is not easy due to very small changes in frequency shift that when not in baseband, removing the carrier signal without affecting the shift frequency becomes very difficult  Also unlike what the equation may suggest, the Doppler shift is not a single frequency, but rather a band of frequencies because Not all cells are moving at the same velocity (velocity profile is not uniform) A cell remains within the UT beam for a very short period of time; the obtained signal needs to be gated, creating side lobes in the frequency shift Acoustic energy traveling within the beam, but at an angle from the bam axis create an effective ∆θ, causing variations in Doppler shift Tumbling and collision of cells cause various Doppler shifts

Directional Doppler  Directional Doppler borrows the quadrature phase detector technique from radar in determining the speed and direction of an aircraft.  Two carrier signals at 90º phase shift are used instead of a single carrier. The +/- phase difference between these carriers after the signal is bounced off of the blood cells indicate the direction, whereas the change in frequency indicate the flowrate

Directional Doppler (a) Quadrature-phase detector. Sine and cosine signals at the carrier frequency are summed with the RF output before detection. The output C from the cosine channel then leads (or lags) the output S from the sine channel if the flow is away from (or toward) the transducer. (b) Logic circuits route one-shot pulses through the top (or bottom) AND gate when the flow is away from (or toward) the transducer. The differential amplifier provides bi-directional output pulses that are then filtered.