Light Source: Globar Silicon Carbide Rod (5mm diameter, 50 mm long) Heated electrically to 1300 – 1500 K Positive temperature coefficient of resistance.

Slides:



Advertisements
Similar presentations
FT-IR Instrument 1. 2 i)Dispersive spectrometers or ii)Fourier transform spectrometers. Most commercial instruments separate and measure IR radiation.
Advertisements

Training at Stockholm University ( ) Martynas Lelis.
The Wave Nature of Light
Wavelength Selection Devices
 PART Absorption Spectrometer Dr. S. M. Condren SourceWavelength SelectorDetector Signal Processor Readout Sample.
FOURIER TRANSFORM INFRARED SPECTROSCOPY
Chapter 16 Infrared Absorption Spectroscopy An IR spectrum contains information about the functional groups in a molecule, and this is used to uniquely.
Chem. 133 – 4/1 Lecture. Announcements I Lab –Returning Term Project Proposals (I expect to be getting boxes of supplies ready) –Returning Set 2 Period.
Describing a Real Source 1) Identify m of real source and adjust T in to line up m 2) The ratio of: 3) Measure T w ( ) to calculate  ( ) Ingle and Crouch,
Photodiodes Photons incident on the depletion layer induce a current.
Prismswww.edmundoptics.com Right AngleEquilateralPentagonalDove.
EM Radiation Sources 1. Fundamentals of EM Radiation 2. Light Sources 3. Lasers.
Infrared Spectroscopy Near Infrared: 770 to 2500 nm Near Infrared: 770 to 2500 nm 12,900 to 4000 cm -1 Mid Infrared: 2500 to 50,000 nm (2.5 to 50  m)
Interference and Diffraction
Describing a Real Source 1) Identify m of real source and adjust T in to line up m 2) The ratio of: 3) Measure T w ( ) to calculate  ( ) Ingle and Crouch,
Infrared Radiation 780 nm m Near, Mid and Far
Photomultiplier Tube m = k 8–19 dynodes (9-10 is most common).
FT-IR Instrument. Components Source Michelson Interferometer Sample Detector.
Rotational Spectra Simplest Case: Diatomic or Linear Polyatomic molecule Rigid Rotor Model: Two nuclei joined by a weightless rod J = Rotational quantum.
Photodiodes Ingle and Crouch, Spectrochemical Analysis Photons incident on the depletion layer induce a current. In most cases, best response in the NIR.
FT-IR Instrument. Components Source Michelson Interferometer Sample Detector.
Overlapping Orders Douglas A. Skoog and James J. Leary, Principles of Instrumental Analysis, Saunders College Publishing, Fort Worth, d(sin  + sin.
Transducers Converts one type of energy into another. Light  Electrical (current, voltage, etc.) What characteristics should we look for in a transducer?
Introduction to Infrared Spectrometry Chap 16. Quantum Mechanical Treatment of Vibrations Required to include quantized nature of E From solving the wave.
Are you getting the concept? If the average irradiance from the Sun impinging normally on a surface just outside the Earth’s atmosphere is 1400 W/m 2,
Fourier Transform IR Spectroscopy Saptarshi Basu.
Lecture 3 INFRARED SPECTROMETRY
Introduction and Applications of Infrared Spectrometry
Neutron Scattering 102: SANS and NR
Infrared Spectroscopy. Theory of Infrared Absorption Spectroscopy IR photons have low energy. The only transitions that have comparable energy differences.
Surface Characterization Techniques Topics: –Contact Angle Analysis –Light Microscopy –X-ray Photoelectron Spectroscopy (XPS) –Fourier-Transform Infrared.
10/11/ ENGINEERING RESEARCH CENTER FOR S TRUCTURED O RGANIC P ARTICULATE S YSTEMS RUTGERS UNIVERSITY PURDUE UNIVERSITY NEW JERSEY INSTITUTE OF TECHNOLOGY.
Pharmacy 325 Infrared (IR) Spectroscopy Dr. David Wishart Rm Ph Hours: anytime after 4 pm.
1.1 Introduction Energy of IR photon Near IR 12, cm -1 Mid IR cm -1 Far IR cm -1 Energy of IR photons insufficient to cause.
Advanced Spectroscopy 3. Infrared Spectroscopy. Revision 1.What molecular or structural features give rise to absorption of infrared (IR) radiation? covalent.
Fourier Transform Infrared (FTIR) Spectrometer Subhashree Mishra ATMS Grad Student, UNR W. P. Arnott Physics, UNR Introduction to Atmospheric Instrumentation.
Infrared Spectroscopy Chapters Chapter 7- I.
Molecular Vibrations
1 Introduction to Spectroscopic methods Spectroscopy: Study of interaction between radiation (or other forms of energy) and matter (a branch of science).
Chapter 12 Infrared Spectroscopy Jo Blackburn Richland College, Dallas, TX Dallas County Community College District  2006,  Prentice Hall Organic Chemistry,
Introduction to Spectrochemical Methods
The basics of in-situ Electrochemcial FTIR spectroscopy
INFRARED SPECTROSCOPY
Spectrophotometer.
Infrared Microspectroscopy
For 3 rd year students- Biophysics Department – Faculty of Science – Cairo University Prepared by: Dr. Heba Mohamed Fahmy and Dr. Alaa Mamdouh.
By Madhu Gyawali Narayan Adhikari Yadab Paudel FTIR Measurements on November 06,08 at Reno Nov 18, 08 PHYS 749 UNIVERSITY OF NEVADA, RENO,NV.
Fourier Transform IR Spectroscopy. Absorption peaks in an infrared absorption spectrum arise from molecular vibrations Absorbed energy causes molecular.
Infrared Spectroscopy (IR) Fourier Transform Infrared (FTIR)
Infrared Microspectroscopy A training guide for using light microscopy and infrared spectroscopy to analyze materials.
Fourier Transform Infrared (FTIR) Spectrometer Subhashree Mishra ATMS Grad Student, UNR W. P. Arnott Physics, UNR Introduction to Atmospheric Instrumentation.
Chem. 133 – 3/29 Lecture. Announcements I Grading – Set 2 lab reports (return?) – Term project proposals (still working on) – UV-Visible seemed to work.
Dr. S. M. Condren Chapter 16 An Introduction to Infrared Spectroscopy.
Optical Instruments II Instruments for Imaging the Retina.
 FT-IR stands for Fourier Transform Infrared, the preferred method of infrared spectroscopy. In infrared spectroscopy, IR radiation is passed through.
Fourier Transform IR (FTIR)
INFRARED SPECTROSCOPY Pramod K Singh School of Basic Sciences &Research Sharda University, Greater Noida, INDIA.
ATTENUATED TOTAL REFLECTION SPECTROSCOPY Outline Introduction to ATR technology Evanescent Waves Crystal Composition Data Collection Advantages & Disadvantages.
LAB. S4 Identification of Drugs Using Attenuated Total Reflectance Fourier Transform Infrared Spectroscopy (ATR-FT-IR Spectroscopy)
UV/VIS SPECTROSCOPY.
UNIT-3 ADVANCES IN METROLOGY
Optical Characterization Techniques
Spectrophotometer Dr . S. Jayakumar.
Infrared absorption sepctroscopy (FT-IR)
Learning Outcomes By the end of this topic students should be able to:
Infrared Spectroscopy
Learning Outcomes By the end of this topic students should be able to:
Infrared Sampling Methods
Lab Equipment.
Presentation transcript:

Light Source: Globar Silicon Carbide Rod (5mm diameter, 50 mm long) Heated electrically to 1300 – 1500 K Positive temperature coefficient of resistance Electrical contact must be water cooled to prevent arcing Ingle and Crouch, Spectrochemical Analysis

Sample Preparation for IR Spectroscopy Ingle and Crouch, Spectrochemical Analysis

Liquid Samples: Cell Thickness Ingle and Crouch, Spectrochemical Analysis

Window and Cell Materials Ingle and Crouch, Spectrochemical Analysis

Solvents Pretsch/Buhlmann/Affolter/Badertscher, Structure Determination of Organic Compounds

Suspension Media for Solid Samples Pretsch/Buhlmann/Affolter/Badertscher, Structure Determination of Organic Compounds

Interferences Pretsch/Buhlmann/Affolter/ Badertscher, Structure Determination of Organic Compounds

Fourier Transform IR Spectrometer Ingle and Crouch, Spectrochemical Analysis

Michelson Interferometer Split source into two beams whose path lengths can be varied periodically to give interference patterns.

Michelson Interferometer

Interferograms Douglas A. Skoog and James J. Leary, Principles of Instrumental Analysis, Saunders College Publishing, Fort Worth, 1992.

Fourier Transform of the Interferogram Gives the Spectrum Ingle and Crouch, Spectrochemical Analysis

Resolution: Jacquinot’s Advantage: Resolution is not limited by aperture width. Can increase light throughput by fold. Multiplex Advantage (aka Fellget’s Advantage): Signal from all are continuously monitored, increasing S/N. S/N for average of n measurements: Advantages of FT-IR over Dispersive IR

How long would it take to capture a dispersive IR spectrum from 500 – 5000 cm -1 with 3 cm -1 resolution elements if you could scan one resolution element every 0.5 seconds? If you want to increase your S/N by a factor of 2, how long will you have to scan? If you collect FT-IR for the same length of time, what is the theoretical S/N advantage? Are you getting the concept?

Evanescent Waves for TIR Microscopy Increased → Increased d p

ATR-FTIR Spectroscopy (Attenuated Total Reflection FTIR) Ingle and Crouch, Spectrochemical Analysis ZnSe or Ge

Schematic of a FTIR Imaging Spectrometer Koenig, J.L.; Wang, S.-Q.; Bhargava, R., Anal. Chem. 2001, 73, 361A.

Focal Plane Array Colarussa, P. et al. Appl. Spectroscopy. 1998, 52, 106A. Square array of MCT pixels Indium electrical contact between each MCT contact and readout electronics layer Total area ~ 1 mm 2 Total time < 10 sec

IR Array Detectors Colarussa, P. et al. Appl. Spectroscopy. 1998, 52, 106A.

FTIR Imaging: Dissolution of a Polymer (a) Two materials are brought into contact and allowed to diffuse across the interface. The diffusion region develops over time. (b) An image is obtained by monitoring radiation passing through the sample at a direction perpendicular to the diffusion direction. (c) Concentration profiles for three different functional groups obtained from the same sample as in (b) (diffusion of a liquid crystal monomer at 265 K for 3 h). Koenig, J.L.; Wang, S.-Q.; Bhargava, R., Anal. Chem. 2001, 73, 361A.

FTIR Imaging: Dissolution of a Polymer FTIR images of the dissolution of poly(alpha-methylstyrene) with a mole weight of 31,000 as a function of time with a mixture of the good solvent methyl isobutyl ketone and the nonsolvent (C 6 D 12 ). Koenig, J.L.; Wang, S.-Q.; Bhargava, R., Anal. Chem. 2001, 73, 361A.

FTIR Microspectroscopy: Tissue Imaging Colarussa, P. et al. Appl. Spectroscopy. 1998, 52, 106A.

FTIR Microspectroscopy: Tissue Imaging Colarussa, P. et al. Appl. Spectroscopy. 1998, 52, 106A.

Other Uses for IR Array Detectors Colarussa, P. et al. Appl. Spectroscopy. 1998, 52, 106A.

Other Uses for IR Array Detectors Colarussa, P. et al. Appl. Spectroscopy 1998, 52, 106A.