Instrumentation in Environmental Physics -- Is Factory Calibration Reliable? Instrumentation in Environmental Physics -- Is Factory Calibration Reliable?

Slides:



Advertisements
Similar presentations
Micrometeorology Surface Layer Dynamics and Surface Energy Exchange
Advertisements

Using Flux Observations to Improve Land-Atmosphere Modelling: A One-Dimensional Field Study Robert Pipunic, Jeffrey Walker & Andrew Western The University.
A thermodynamic model for estimating sea and lake ice thickness with optical satellite data Student presentation for GGS656 Sanmei Li April 17, 2012.
Soil temperature and energy balance. Temperature a measure of the average kinetic energy of the molecules of a substance that physical property which.
W1188 Progress Report Tyson E. Ochsner USDA-ARS St. Paul, MN Jan. 3, 2005.
Near Surface Soil Moisture Estimating using Satellite Data Researcher: Dleen Al- Shrafany Supervisors : Dr.Dawei Han Dr.Miguel Rico-Ramirez.
Distance observations
Blackbody radiation and solar radiation 1)More on Blackbody Radiation; Black body radiation 2) Magnitude of Solar radiation.
Meteorological Instrumentation Setups and Measurements in Mica Creek Watershed Wenguang Zhao and Russell J. Qualls Idaho NSF EPSCoR Project.
Atmospheric phase correction for ALMA Alison Stirling John Richer Richard Hills University of Cambridge Mark Holdaway NRAO Tucson.
Card 1. MODTRAN Card deck/Tape5_Edit Tutorial Explanation of Parameters & Options.
Using Scatterometers and Radiometers to Estimate Ocean Wind Speeds and Latent Heat Flux Presented by: Brad Matichak April 30, 2008 Based on an article.
Theoretical Study & Modelling of the Intensity Distribution of VLF Signals Tamal Basak [1], Sujay Pal [1], S K Chakrabarti [1,2] 1] S N Bose National Centre.
University of Bayreuth Department of Micrometeorology Thomas Foken 3 rd COPS and GOP Workshop, Hohenheim, April 2006 Boundary-Layer Measurements.
Boundary Layer Notes 5 Observational Techniques. Sources: Kaimal & Finnegan, Atmospheric Boundary Layer Flows: their structure and measurement, Oxford.
A Basic Introduction to Boundary Layer Meteorology Luke Simmons.
Water Resources in a Changing Climate: NSF ESPCoR VI Hydroclimatology V. Sridhar, Xin Jin, David Hoekema, Sumathy Sinnathamby, Muluken Muche R. Allen,
Matt Robinson Thomas Tolman.  Describes Convective Heat Transfer  Needed for all external and internal flow situations.
Lecture #17 Boundary Layer Measurements  Boundary layer Thickness  * Displacement Thickness  Momentum Thickness.
July 2001Zanjan, Iran1 Atmospheric Profilers Marc Sarazin (European Southern Observatory)
D EDICATED S PECTROPHOTOMETER F OR L OCALIZED T RANSMITTANCE A ND R EFLECTANCE M EASUREMENTS Laetitia ABEL-TIBERINI, Frédéric LEMARQUIS, Michel LEQUIME.
1 Vegetation dynamics and soil water balance in a water-limited Mediterranean ecosystem on Sardinia, Italy Nicola Montaldo 1, John D. Albertson 2 and Marco.
Assessing the Water and Energy Balances at the BERMS Flux Towers, 1999 to 2005 A. G. Barr 1, G. van der Kamp 1, T. A. Black 2, J. H. McCaughey 3, R. Granger.
Optical Measurements of Aerosols for MIRAGE-MEX University of Iowa Bill Eichinger John Prueger Piotr Lewandowski Heidi Holder.
Characterization of Aerosol Physical, Optical and Chemical Properties During the Big Bend Regional Aerosol and Visibility Observational Study (BRAVO) Jenny.
Evaporation What is evaporation? How is evaporation measured? How is evaporation estimated? Reading: Applied Hydrology Sections 3.5 and 3.6 With assistance.
SENSIBLE HEAT FLUX ESTIMATION USING SURFACE ENERGY BALANCE SYSTEM (SEBS), MODIS PRODUCTS, AND NCEP REANALYSIS DATA Yuanyuan Wang a, Xiang Li a,b a, National.
LL-III physics-based distributed hydrologic model in Blue River Basin and Baron Fork Basin Li Lan (State Key Laboratory of Water Resources and Hydropower.
High-Resolution Climate Data from Research and Volunteer Observing Ships: A Strategic Intercalibration and Quality Assurance Program A Joint ETL/WHOI Initiative.
Ground-based energy flux measurements for calibration of the Advanced Thermal and Land Application Sensor (ATLAS) Eric Harmsen, Associate Professor Dept.
NRES 408/808 Instrumentation Demonstration -- Radiation Instrumentation Mark Mesarch 229D Hardin Hall or
TURBULENT FLUX VARIABILITIES OVER THE ARA WATERSHED Moussa Doukouré, Sandrine Anquetin, Jean-Martial Cohard Laboratoire d’étude des Transferts en Hydrologie.
Variation of Surface Soil Moisture and its Implications Under Changing Climate Conditions 1.
Analysis of Turbulence Development in the Morning
2012 IAL Conference A novel technique to measure the total evaporation and its components during sprinkler irrigation Jasim Uddin Rod Smith Nigel Hancock.
1 NOAA-UPRM COOP Program in Atmospheric Sciences and Meteorology, Department of Physics, University of Puerto Rico at Mayagüez, Mayagüez, PR Yaítza.
Meteorology meets Astronomy : open discussion 1.Usefullness of atmospheric mesoscale modelling for astrophysical applications - to forecast astrophysical.
Oceanography By Brenda Owens Picture from
ATM 301 Lecture #9 (6.1, Appendix D) Surface Energy Fluxes and Balance Major Surface Energy Fluxes Review of Radiation Solar Radiation Thermal Radiation.
Dona Aña ET Studies By A. Salim Bawazir, Ph.D. and J. Phillip King, Ph.D., P.E Civil and Geological Engineering Department NMSU, Las Cruces, NM This project.
Evaporation Slides prepared by Daene C. McKinney Reading: Applied Hydrology Sections 4.1 and 4.2 Quotation for today (from Socrates) "There is only one.
Ground-based energy flux measurements for calibration of the Advanced Thermal and Land Application Sensor (ATLAS) Eric Harmsen, Associate Professor Dept.
Heat Transfer Simulation for the Welding of the T-Section.
OEAS 604: Introduction to Physical Oceanography Surface heat balance and flux Chapters 2,3 – Knauss Chapter 5 – Talley et al. 1.
A STUDENT’S EXPERIMENT WITH MULTIPLE REFLECTIONS AND REFRACTIONS ON A GLASS PLATE AND VALIDATION OF THE FRESNEL’S EQUATIONS N. Mahmudi 1, S. Rendevski.
Scintillometer and eddy covariance comparisons over sagebrush and invasive cheatgrass communities in Idaho Scintillometer and eddy covariance comparisons.
Estimation of Potential Evapotranspiration from Merged CERES and MODIS Observations Anand Inamdar & A. French Arid Land Agricultural Research Center (ALARC/ARS/USDA)
Ground-based energy flux measurements for calibration of the Advanced Thermal and Land Application Sensor (ATLAS) 1 Eric W. Harmsen and Richard Díaz Román,
METR 415/715 Wednesday February Imaginary part of index of refraction When n i is non-zero, absorption of an EM wave occurs as it passes through.
Evaporation What is evaporation? How is evaporation measured?
RESERVA ECOLÓGICA DO IBGE RECOR SITE BRASÍLIA. IMAGEM DO SATÉLITE IKONOS, 2001 RECOR CAMPO SUJO CERRADO.
Evaporation What is evaporation? How is evaporation measured? How is evaporation estimated? Reading for today: Applied Hydrology Sections 3.5 and 3.6 Reading.
The “Monitor to measure the Integral TRAnsmittance” (MITRA)
The “Monitor to measure the Integral TRAnsmittance” (MITRA)
SPLIT-WINDOW TECHNIQUE
Miscellaneous Measurements
Jun Zhang NOAA/AOML/HRD William Drennan Univ. of Miami/ RSMAS
EVAPORATION.
Eddy covariance (EC) is the most direct method for estimating fluxes of energy, heat, and mass. The fundamental theory of EC relies on the conservation.
The Cabauw Experimental Site for Atmospheric Research (CESAR): New developments Fred C. Bosveld (KNMI) Content CESAR and its research themes Long term.
Absolute calibration of sky radiances, colour indices and O4 DSCDs obtained from MAX-DOAS measurements T. Wagner1, S. Beirle1, S. Dörner1, M. Penning de.
The Nocturnal Surface Energy Budget (SEB) at Cabauw Fred C
The summer 2017 seen by LMDZOR and two French instrumented sites
Oceanography By Brenda Owens
Spatial Variation in Corn and Soybean Fields in Central Iowa
Guo Jianxia1,2 Bian Lingen1 Gao Zhiqiu1
The upstream effect on back pressure ratio
Atmospheric phase correction for ALMA
Operational Oceanography
Atmospheric reanalysis at ECMWF
Presentation transcript:

Instrumentation in Environmental Physics -- Is Factory Calibration Reliable? Instrumentation in Environmental Physics -- Is Factory Calibration Reliable? EPSCoR Project Supported by NSF Wenguang Zhao & Richard G. Allen October 24, 2011, Xi’an, China

AuthorsAuthors University of Idaho – University of Idaho – Wenguang Zhao and Richard G. Allen Idaho State University – Idaho State University – Matt Germino Boise State University – Boise State University – Sridhar V. University of Idaho – University of Idaho – Wenguang Zhao and Richard G. Allen Idaho State University – Idaho State University – Matt Germino Boise State University – Boise State University – Sridhar V.

BackgroundBackground 1. Most people believe the factory calibration. 1. Most people believe the factory calibration. 2. Representation of H and ET measured by traditional methods (EC and BR etc) is limited, especially for heterogeneous fields. 2. Representation of H and ET measured by traditional methods (EC and BR etc) is limited, especially for heterogeneous fields. 3. Large aperture scintillometry (LAS) is an alternative method to estimate H from a relatively large footprint (source) area. 3. Large aperture scintillometry (LAS) is an alternative method to estimate H from a relatively large footprint (source) area. 1. Most people believe the factory calibration. 1. Most people believe the factory calibration. 2. Representation of H and ET measured by traditional methods (EC and BR etc) is limited, especially for heterogeneous fields. 2. Representation of H and ET measured by traditional methods (EC and BR etc) is limited, especially for heterogeneous fields. 3. Large aperture scintillometry (LAS) is an alternative method to estimate H from a relatively large footprint (source) area. 3. Large aperture scintillometry (LAS) is an alternative method to estimate H from a relatively large footprint (source) area.

QuestionsQuestions 1. How can we accurately calculate H from the LAS measurement, structure function constant of refractive index fluctuations for the wavelength used by the LAS (C n 2 )? 1. How can we accurately calculate H from the LAS measurement, structure function constant of refractive index fluctuations for the wavelength used by the LAS (C n 2 )? 2. How does the H calculated by the LAS measurement compare to the traditional EC measurement result? 2. How does the H calculated by the LAS measurement compare to the traditional EC measurement result? 1. How can we accurately calculate H from the LAS measurement, structure function constant of refractive index fluctuations for the wavelength used by the LAS (C n 2 )? 1. How can we accurately calculate H from the LAS measurement, structure function constant of refractive index fluctuations for the wavelength used by the LAS (C n 2 )? 2. How does the H calculated by the LAS measurement compare to the traditional EC measurement result? 2. How does the H calculated by the LAS measurement compare to the traditional EC measurement result?

4-way net radiometers

(A)(B) Intercomparison of 5 NR01 4-way net radiometers in 2009

Measured atmospheric long wave radiation by using factory calibration coefficients of each NR01’a Measured atmospheric long wave radiation by using the modified calibration coefficients of ourselves Intercomparison in summer 2009 Intercomparison

Intercomparison of 10 4-way net radiometers (8 NR01s & 2 CNR1s) in 2010

Used factory calibration Coef. Measured atmospheric long wave radiation by using factory calibration coefficients of each NR01’a Measured atmospheric long wave radiation by using the modified calibration coefficients of ourselves Intercomparison in summer 2010 Intercomparison Used self calibration Coef.

Intercomparison of 8 4-way net radiometers (6 NR01s, a CNR1 & a CNR4) in 2011

Intercomparison in summer 2011 Intercomparison Used factory calibration coefficient Used self calibration coeficient (from day+night data) Used self calibration coefficient (from night data) + about 2.3% (from 2.0% to 2.5%) short wave radiation correction

3-D Sonic anemometers

3 sets of EC systems (RM Young D sonic + LI-7500)

-3.54 mm mm mm

Comparison of the 3 EC systems

BLS 9000 Boundary Layer Scintillometers

Transmitter Receiver Intercomparison of 3 receivers and their SPU with the same transmitter Intercomparison of 3 receivers and their SPU with the same transmitter

T1T1T1T1 T2T2T2T2 R1R1R1R1 R2R2R2R2 Intercomparison of 2 independent sets of BLS 9000 scintillometers with the opposite light transmiting directions

SN:T-E-0112 measured a lower sensible heat flux (H) than SN: T-E-0115 SN:T-E-0112 measured a lower sensible heat flux (H) than SN: T-E-0114

REBS HFT-3.1 Heat Flow Transducer

Multiple Depths of Plates (~4 and 8 cm) Soil Temperature profile and 6cm)

LAS: Transmitter A & Receiver B

ConclusionsConclusions Good agreement was obtained between H measured by CSAT3 and RM Young D sonic anemometers 2. H measured by 2. H measured by Scintec BLS900 compared well with both EC systems (CSAT3 and RM Young). Thank you! Good agreement was obtained between H measured by CSAT3 and RM Young D sonic anemometers 2. H measured by 2. H measured by Scintec BLS900 compared well with both EC systems (CSAT3 and RM Young). Thank you!

Questions?Questions? Do I need to STOP here?