Emma Muir, Sam Muir, Jacob Sandlund, & David Smith Advisor: Dr. José Sánchez Co-Advisor: Dr. James Irwin.

Slides:



Advertisements
Similar presentations
Controller Tests Stephen Kaye Controller Test Motivation Testing the controller before the next generation helps to shake out any remaining.
Advertisements

Signal Encoding Techniques
Chapter3 Pulse-Echo Ultrasound Instrumentation
Foundations of Medical Ultrasonic Imaging
Ultrasound Microscopy and High Frequency Coded Signals Antti Meriläinen, Edward Hæggström.
Fundamentals of Data & Signals (Part II) School of Business Eastern Illinois University © Abdou Illia, Spring 2015 (February18, 2015)
University Of Vaasa Telecommunications Engineering Automation Seminar Signal Generator By Tibebu Sime 13 th December 2011.
Motivation Application driven -- VoD, Information on Demand (WWW), education, telemedicine, videoconference, videophone Storage capacity Large capacity.
5/4/2006BAE Analog to Digital (A/D) Conversion An overview of A/D techniques.
Resident Physics Lectures
1. Output signal alternates between on and off within specified period Controls power received by a device The voltage seen by the load is directly proportional.
Using an FPGA to Control the Protection of National Security and Sailor Lives at Sea Brenda G. Martinez, Undergraduate Student K.L. Butler-Purry, Ph.D.,
CAPSTONE Demonstration Radio TiVo FPGA Thundercats: Hariklia Karagiannis Hasina Jamal Osato Edo-Osagie Brad Mazan Chad Griffith.
Chapter 9 Data Acquisition A/D Conversion Introduction
ECE 501 Introduction to BME
1 Autonomously Controlled Vehicles with Collision Avoidance Mike Gregoire Rob Beauchamp Dan Holcomb Tim Brett.
PWM Audio Amplifiers Zhiming Deng Chinwuba Ezekwe Dimitrios Katsis.
Design of a 1-D Sonic Anemometer MDR Presentation Group Members: Vanessa Dubé, Michael Jao, Chethan Srinivasa, Robert Vice Advisors: Professor Jackson.
Development and Implementation of a FPGA-based digital beamformer Supervisors: Nandita Bhattacharjee Dr. Andrew Paplinski Dr. Andrew Paplinski Dale Harders.
The World Leader in High Performance Signal Processing Solutions Audio ADC/DACs Primer David Hossack.
Despeckle Filtering in Medical Ultrasound Imaging
THE ULTRASOUND IMAGE: GENERATION AND DISPLAY
Oscilloscope Tutorial
Ultrasonic Imaging using Resolution Enhancement Compression and GPU- Accelerated Synthetic Aperture Techniques Presenter: Anthony Podkowa May 2, 2013 Advisor:
Formatting and Baseband Modulation
ORE 654 Applications of Ocean Acoustics Lecture 6a Signal processing
A/D Conversion No. 1  Seattle Pacific University Analog to Digital Conversion Based on Chapter 5 of William Stallings, Data and Computer Communication.
Fundamentals of Digital Communication
DATA ACQUISITION Today’s Topics Define DAQ and DAQ systems Signals (digital and analogue types) Transducers Signal Conditioning - Importance of grounding.
ECE 4371, Fall, 2014 Introduction to Telecommunication Engineering/Telecommunication Laboratory Zhu Han Department of Electrical and Computer Engineering.
Introduction to Adaptive Digital Filters Algorithms
1 Digital Voltage Transducer family DV from 1200 to 4200 V RMS DVL from 50 to 2000 V RMS.
RITEC MEDARS MULTIPLE EMAT DRIVER AND RECEIVER SYSTEM Your Logo Here Drive Phased Arrays of EMATs at high powers.
1 Resolution Enhancement Compression- Synthetic Aperture Focusing Student: Hans Bethe Advisor: Dr. Jose R. Sanchez Bradley University Department of Electrical.
1HSSPG Georgia Tech High Speed Image Acquisition System for Focal-Plane-Arrays Doctoral Dissertation Presentation by Youngjoong Joo School of Electrical.
Comprehensive Ultrasound Research Platform Emma Muir Sam Muir Jacob Sandlund David Smith Advisor: Dr. Sánchez Co-advisor: Dr. Irwin.
COMMUNICATION SYSTEM EEEB453 Chapter 5 (Part IV) DIGITAL TRANSMISSION.
INTERPOLATED HALFTONING, REHALFTONING, AND HALFTONE COMPRESSION Prof. Brian L. Evans Collaboration.
SIGMA-DELTA ADC SD16_A Sigma-Delta ADC Shruthi Sujendra.
Building Three-Dimensional Images Using a Time-Reversal Chaotic Cavity
Comprehensive Ultrasound Research Platform Emma Muir Sam Muir Jacob Sandlund David Smith Advisor: Dr. José Sánchez.
Review of Ultrasonic Imaging
Sarah Gillies Ultrasound Sarah Gillies
ENG3640 Review and Exam Question1 ENG3640 Microcomputer Interfacing Review & Final Exam Structure.
Digital Beamforming. Beamforming Manipulation of transmit and receive apertures. Trade-off performance/cost to achieve: –Steer and focus the transmit.
Ultrasound Simulations using REC and SAFT Presenter: Tony Podkowa November 13, 2012 Advisor: Dr José R. Sánchez Department of Electrical and Computer Engineering.
VOCODERS. Vocoders Speech Coding Systems Implemented in the transmitter for analysis of the voice signal Complex than waveform coders High economy in.
Frank Ludwig, DESY Content : 1 Stability requirements for phase and amplitude for the XFEL 2 Next LLRF system for optimized detector operation 3 Limitations.
Chapter 6: Real-Time Image Formation. display B, M, Doppler image processing Doppler processing digital receive beamformer system control keyboard beamformer.
Microphone Array Project ECE5525 – Speech Processing Robert Villmow 12/11/03.
IMAGE DATA ACQUISITION
Topics discussed in this section:
A 5-Pulse Sequence for Harmonic and Sub-Harmonic Imaging
Ultrasound Computed Tomography 何祚明 陳彥甫 2002/06/12.
1 Resolution Enhancement Compression- Synthetic Aperture Focusing Techniques Student: Hans Bethe Advisor: Dr. Jose R. Sanchez Bradley University Department.
Artifacts Ultrasound Physics George David, M.S.
Bar code scanner Department of Computer Engineering, M.S.P.V.L. Polytechnic College, Pavoorchatram.
PSF What is PSF 1way/2way PSFs PSFs for continious & discrete apertures.
 Ultrasound waves are longitudinal with high frequencies ( ≈ > 20,000 Hz, though medical Ultrasound is between 1 to 15 MHz.)  When an ultrasound reaches.
Electric Air Ukulele Ivan Setiawan (setiawa2) Satyo Iswara (iswara2) ECE 445 Senior Design Spring 2012 Team #32 TA: Jane Tu.
Ultrasound Physics Image Formation ‘97. Real-time Scanning Each pulse generates one line Except for multiple focal zones frame one frame consists of many.
Chapter 10 Digital Signal and Image Processing
Dr. Clincy Professor of CS
Ultrasound Physics Image Formation ‘97.
Ultrasound Physics Image Formation ‘97.
Optical Coherence Tomography
Spatially Varying Frequency Compounding of Ultrasound Images
Direct Digital Synthesis: Applications to Radar
Dr. Clincy Professor of CS
Review of Ultrasonic Imaging
Presentation transcript:

Emma Muir, Sam Muir, Jacob Sandlund, & David Smith Advisor: Dr. José Sánchez Co-Advisor: Dr. James Irwin

2

3 [1] Benign Malignant

Introduction How Ultrasound Works Coded Excitation Objective Motivation Significance Design Comparison 4

5

6 Conventional Ultrasound [2] Coded Excitation Ultrasound [2]

7 Research Platforms Mostly single-element Large multi-element  RASMUS RASMUS [3]

8 Ultrasound Research Platform Prototype  Arbitrary Waveforms o Coded excitation signals  Multi-element o Beamforming  Reduced size and cost Lecroy Oscilloscope

9 Improve…  Ultrasound Techniques  Ultrasound Research Reduce size and cost

10 Medical Applications  Detect and Diagnose Tumors  Noninvasive  Faster Results

11 Previous Designs: Our Design: Digital Device Amplifier Transducer Digital Device Transducer Switching Amplifier D/A

12 Oversample  1-bit  Densities represent voltages

13 Transducer acts as a (BP) filter  Smooths / Averages

14 Example:  0.5 V DC  -1 to 1 V Dynamic range 8-bit Two’s Complement (-128 to 127):  Value = 64 ( ) Sigma Delta Modulation:  Oversample  1-bit

Introduction Functional Description Methods Results and Discussion Conclusion Questions 15

Introduction Functional Description Methods Results and Discussion Conclusion Questions 16

Up to 4 transducer channels Excitations <= 3 μs SNR > 50 dB 17

18

19 Generate Waveform

20 Transmit Waveform

21 Receive Image

22 Create Image

23

24

25

26

27

Time Gain Compensation (TGC)  Attenuation  TGC = Att * Depth * (Probe frequency)  white noise for larger depths 28

29

Envelope Detection  Determines the bounds of the processed signal  Detects width and contains the display information  Absolute value of the Hilbert Transform 30

31

Introduction Functional Description Methods Results and Discussion Conclusion Questions 32

Sigma Delta Modulation PC/FPGA Interface FPGA Data Processing  Pulse Compression  Delay Sum Beamforming 33

34

35 < 10% Mean Squared Error (MSE) 500 M samples/second Accuracy vs. Overloading (Saturation)  Order = 2nd  OSR = 16 o must be a power of 2 o 16*2 = 32 samples per period

36 [4]

37 [4]

38

39 [5]

40

41 Assign waveform to pins  Independent for each pin  (3 μs) * (500 MHz) = 1500 bits/waveform  = 1536 bits/waveform (divisible by 512) Assign delay to pins  Increments of 4ns = (1/250 MHz)  250 MHz = memory clock rate of FPGA

42 Transfer information for 4 pins in < 1 sec  <32 sec for 128 pins  (4 pins) * (1536 bits/waveform) sent within 1 sec  ~6 Kbps Start transmission

UART connection  baud o Fastest FPGA baud rate  Sends as o 1 start bit o 8 data bits o 2 stop bits  (1536/8)*11*4 = 8448 bits  ~73 ms for 4 channels  ~2.3 s for 128 channels 43 Start Stop Data

44

45 Transmit at 500 MHz Output waveforms in parallel  4 individualized waveforms  Length of 3  s per waveform  1536-bits per waveform

46 Yes No

47 Transmit at 500 MHz  Two 250 MHz clock edges (transmits on rising and falling edge)  250 MHz * 2 = 500 MHz XOR

48

Data Processing  Less than 2 minutes Display an image  Depths between 0.25 cm and 30 cm  Dynamic range between 40 dB and 60 dB 49

50

Restore the spatial resolution Match reflected wave to original excitation Use Wiener filter  Optimal solution between a match filter and an inverse filter [6]  Solution determined by o Smoothing Factor (SF) o Predicted signal-to-noise-ratio (SNR) Predict SNR = 50 dB 51

Matched filter  Cross correlation of original coded excitation and received signal  Creates side lobes  Does not amplify noise  Optimal for large noise – small SNR Inverse filter  Inverse of the original coded excitation  No side lobes  Amplifies noise  Optimal for no noise – large SNR 52

53 Wiener Filter Equation = Coded Excitation = Smoothing Factor = SNR of system Noise increases SNR decreases λ/S increases Closer to a Match Filter Noise decreases SNR increases λ/S decreases Closer to an Inverse Filter [1]

54 SNR = 60 dB

55

Sensor Array Transducer Focal Point Narrowest beam Greatest amplitude Beamforming not necessary at this point 20 mm 4 mm mm

57 Amplitude at Point = Σ i=1 S i ( Depth + Delay(S i,Point)) [7] Delay(S,P) = (D SP - D SF )/c [7] S = sensor P = point D SP = distance from sensor to point D SF = distance from sensor to point c = 1540m/s (speed of sound in tissue) 8 Sensors Focal Point Point Depth

Introduction Functional Description Methods Results and Discussion Conclusion Questions 58

59

60

61

62

63

64

% MSE

66

Correlations Filtered Sigma-delta Modulated Linear Chirp Filtered Captured Data Filtered Linear Chirp 99.84%99.33% Filtered Sigma-delta Modulated Linear Chirp % 67

Correlations Filtered Sigma-delta Modulated Linear Chirp Filtered Captured Data Filtered Linear Chirp 99.84%99.33% Filtered Sigma-delta Modulated Linear Chirp % 68

Correlations Filtered Sigma-delta Modulated Linear Chirp Filtered Captured Data Filtered Linear Chirp 99.84%99.33% Filtered Sigma-delta Modulated Linear Chirp % 69

Correlations Filtered Sigma-delta Modulated Linear Chirp Filtered Captured Data Filtered Linear Chirp 99.84%99.33% Filtered Sigma-delta Modulated Linear Chirp % 70

71

72

73 h 1 (n) * c 1 (n) = h 2 (n) * c 2 (n)

74 Field II Software [8] 10 mm separation 46 dB SNR Distance in mm 100 Transducer

75 Without BeamformingWith Beamforming

76 REC Excitation and Pulse Compression Impulse Excitation

77

Introduction Functional Description Methods Results and Discussion Conclusion Questions 78

Valid waveform transmission Portable system Multi-channel Research potential 79 Lecroy Oscilloscope

The authors would like to thank Analog Devices and Texas instruments for their donation of parts. This work is partially supported by a grant from Bradley University ( REC) Dr. Lu Mr. Mattus Mr. Schmidt Andy Fouts 80

[1] J. R. Sanchez et al., "A Novel Coded Excitation Scheme to Improve Spatial and Contrast Resolution of Quantitative Ultrasound Imaging," IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 56, no. 10, pp , October [2] "Clinical Image Library." GE Healthcare-. GE Healthcare. Web. 14 Apr [3] J. A. Jensen et al., “Ultrasound Research Scanner for Real-time Synthetic Aperture Data Acquisition,” IEEE Trans. Ultrason., Ferroelect., Freq. Contr., vol. 52, no. 5, pp. 881–891, [4] R. Schreier and G. C. Temes. Understanding Delta-Sigma Data Converters, John Wiley & Sons, Inc., [5] R. Schreier, The Delta-Sigma Toolbox Version 7.3. Analog Devices, Inc,

[6] T. Misaridis and J. A. Jensen, “Use of Modulated Excitation Signals in Medical Ultrasound Part I: Basic Concepts and Expected Benefits,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 52, no. 2, pp , February [7] Thomeniu, Kai E. "Evolution of Ultrasound Beamformers." IEEE Ultrasonics Symposium (1996): Print. [8] J.A. Jensen. Field: A Program for Simulating Ultrasound Systems, Medical & Biological Engineering & Computing, pp , Volume 34, Supplement 1, Part 1,