BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods

Slides:



Advertisements
Similar presentations
Technical Aspects: the machine the image
Advertisements

Advanced Biomedical Imaging Lecture 3
Introduction to x-rays
Formation Characteristics
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
IMAGE FORMATION  Introduction  The Invisible and Visible Image  Image Characteristics.
Digital Radiography.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Digital Radiographic Imaging 101
X-ray radiography Seminar The "geometrical" unsharpness of X-ray images is limited by finite dimension of the source, i.e. the window on.
Chapter 15 Radiographic Intensifying Screens And Cassettes
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
Computed Tomography RAD309
Main detector types Multi Pixel Photon Counter (MPPC) and Charge Coupled Devices (CCDs) How does it work? 1. Photon hits a pixel producing electron hole.
Radiation Safety level 5 Frits Pleiter 02/07/2015radiation safety - level 51.
Seeram Chapter 5: Data Acquisition in CT
Techniques for detecting X-rays and gamma-rays Pair production Creation of elementary particle and its antiparticle from a photon. Occurs only if enough.
Unit 1 Physics Detailed Study 3.6 Chapter 15: Medical Physics.
Digital Technology 14.2 Data capture; Digital imaging using charge-coupled devices (CCDs)
Direct Digital Radiography or Direct Capture Radiography Bushong Ch. 27.
X-Ray Production & Emission
Detectors The energy loss of particles in matter can be used detect and identify those particles. There are different types of “detectors”: - Gas-filled.
Most of the images recorded during conventional radiography are obtained with film/screen combination image receptors. Which in lessens the patient dose.
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
Instruments for Radiation Detection and Measurement Lab # 3 (1)
ADC PRE-PATIENT COLLIMATION POST-PATIENT COLLIMATION.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
X-Ray Medical Imaging Physics – IB Objectives
Unit III Creating the Image. Unit III Creating the Image.
Chapter 6: Digital Radiographic Imaging
1 Scintillators  One of the most widely used particle detection techniques Ionization -> Excitation -> Photons -> Electronic conversion -> Amplification.
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
DMI 261 Radiation Biology AndProtection. Unit 2 RADIATION Electromagnetic Spectrum –The frequency range of electromagnetic radiation and the photon wavelength.
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
Lecture (2). 2 1.Explain how to produce x-rays and discuss its properties 2.List the basic components of the radiographic machine and identify primary.
Image Receptor Systems
DIGITAL PROJECTION RADIOGRAPHY COMPUTED RADIOGRAPHY
Image Quality Radiographic Resolution.
Computed Radiography By Prof. Stelmark. Presently, an acceleration in the conversion from screen-film radiography (analog) to digital radiography (DR)
Instrumentation Review. Direct and Indirect Ionization Direct - Charge particles that strip away electrons from atoms Indirect - uncharged that have to.
Image Capture. X-Ray X-Ray Generation Film Can be exposed directly to x-ray, but sensitivity is very low Nitrate – 1910s to 1930s Acetate – 1920s to.
Nuclear Medicine: Planar Imaging and the Gamma Camera Katrina Cockburn Nuclear Medicine Physicist.
Seeing the Subatomic Stephen Miller Saturday Morning Physics October 11, 2003.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
Determining Radiation Intensity
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Analog Imaging III By Professor Stelmark.
FILM CASSETTES & INTENSIFYING SCREENS WEEK 9
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Ideal Detector Fast Cheap Rugged Responds to all wavelengths of light Can distinguish different wavelengths Sensitive Low LOD.
Chapter V Radiation Detectors.
DIGITAL IMAGING.
Image Receptor Unsharpness By Professor Stelmark.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
DIGITAL RADIOGRAPHY.
Direct Digital Radiography or Direct Capture Radiography
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
Medical Physics.
AJADI ADETOLA DEPARTMENT OF VETERINARY MEDICINE AND SURGERY.
(Chapter 7) CHEM–E5225 Electron microscopy
Computed Tomography Data Acquisition
PAN-2013: Radiation detectors
כיצד נרכשת התמונה בסרט הרנטגני?
Computed Radiography Feb.
When this happens... We see this... 12/6/2018 Steve Wotton.
COMPUTED RADIOGRAPHY -Dr.SHEFALI MESHRAM
Computed Tomography (C.T)
Presentation transcript:

BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 2

Today Anti-scatter devices X-ray screen-film systems Other methods of X-ray detection

X-ray System Converts X-rays to light and records Produces X-rays from electrical energy Tailors X-ray spectrum Source Filter Restrictor (Collimator) Subject Anti-scatter Detector Determines size and shape of beam Selectively removes scattered photons

Total linear attenuation coefficient At X-ray energies, most photons that interact in the patient are Compton-scattered. Soft Tissue

X-ray Scatter Tube Object Grid Detector Scattered radiation comes into detector from all directions. Result is a relatively uniform background “fog” that reduces dynamic range of the detector available to image true signal. Would like some way to reduce scattered radiation without blocking much direct radiation.

Anti-scatter Strategies Collimation of the beam at the front end Air gaps Grids Scanning Slits

Air Gap Moving the patient away from the detector reduces the scatter reaching the detector. Square-law Solid angle What price do we pay for this?

Anti-scatter Grids Construct a device to collimate photons after they leave the patient. Thin lead strips must be precisely aligned. Performance depends on the grid ratio What is the price paid for a high grid ratio? Typical grid ratios: 5:1 to 16:1 (lower for mammography)

Anti-scatter Grids A stationary grid will leave line artifacts in the image. A Potter-Bucky diaphragm is a movable grid that basically blurs the grid lines during exposure. The grid also blocks some primary radiation in the system.

Anti-scatter Grids Tradeoff between scatter penetrating the grid and primary radiation detected

Anti-scatter Grids Thickness of strips determines the likelihood of penetration. Less scatter penetration = less primary radiation Low-angle scatter may still get through. Multiply-scattered photons may get through.

Anti-scatter Grids Additional exposure is needed to maintain same detector exposure level when using grid. Grid Conversion Factor Typically 3 < GCF < 8 May avoid grid for small body parts or low energies.

Scanning Slits Moving source and collimator Move source, collimator, and slit together. Only takes one part of image at a time. Very high scatter reduction Slow Moving slit Stationary detector

X-ray Detectors Film-Screen Image Intensifiers Panel Detectors

Film-Screen Detectors Roentgen’s first X-rays exposed a photographic plate directly. But photographic film has very low stopping power (a couple of percent). To expose the film to its full dynamic range (contrast) would require high dose and most would be wasted. Augment this with an intensifying screen that converts X-ray photons to visible light.

Screen-film System Reflective Layer Double emulsion film sandwiched between pair of intensifying screens Phosphor particles (high Z) covert X-ray into light photons Screen enhances contrast but lowers resolution Engineering tradeoff: Phosphor thickness Base Film Phosphor Coating

Screen-film System Reflective layer reflects light back into the film Base for mechanical support Phosphor layer material choice: More fluorescent than phosphorescent High linear attenuation coefficient = stopping power Conversion efficiency: total light energy per unit incident X-ray energy (usually 5 – 20%) Energy dependent Base Film Phosphor Coating

Film Very similar to photographic film; must be developed to fix the image Two components: Base: Plastic sheet, dimensionally stable (size and shape do not change under environmental and processing conditions) Emulsion: Crystals of silver bromide suspended in gelatin substance; on one side (single-emulsion), or both sides (double-emulsion) of base. Image is formed in the silver bromide crystals.

Screen-film System Exposure to get a standard level of film density The screen-film combination usually has a speed quoted More sensitive (= fewer X-ray photons to result in a given image density) pairs have higher speed At a particular energy! Speed Sensitivity (mR) 1200 0.1 800 0.16 400 0.32 200 0.64 100 1.28 50 2.56 25 5.0 12 10.0 From Sprawls

Radiographic Cassette Ensures firm and uniform contact between intensifying screens and film sandwiched in between Optical mirrors located outside screens to direct light towards film, maximize light conversion efficiency Contains ID card and loaded only one way into X-ray machine Image source: The Essential Physics of Medical Imaging

Film Density Density describes the overall blackness of the radiograph Image source: http://www.nurseslearning.com/courses/fice/fde0030/Imaging_terms.htm

Screen-film System Thicker screens result in higher sensitivity but increase image blur

Other Detection Schemes Detection is a result of radiation interaction with matter. Radiation interaction results in emission of by products, e.g. electrons, electromagnetic radiation, that can be sensed by instrumentation and recorded by data acquisition systems Gas-Filled Detectors Scintillation Detectors Flat-panel detectors PSP plates Solid State Detectors

Gas Filled Detectors Radiation ionizes the gas. Charges freed by ionization produce a current.

Gas Filled Detectors Radiation interacts with gas and ionizes its atoms Freed electrons interact with gas and ionize more atoms - amplification Ionization chamber: No amplification Proportional counter: Amplification up to 106 times Geiger-Muller counter: Very strong avalanche Spatial sensitivity is lacking – Not used for imaging

Scintillation Detectors Interaction of X-rays with some materials (CsI, cadmium zinc telluride - CZT) produces ‘scintillation’ or “flash of light”. scintillator Not capable of handling high photon flux. Visible light X-ray Electrical pulse photomulitplier The pulse can tell you about the energy of the incident photon.

Photomultiplier Tube Photomultiplier tube (PMT) converts light into electric current by photoelectric effect

Flat-panel Detectors Scintillator Light coupling Light-sensitive digital detector (CCD array) Varian Medical Systems

Photostimulable Phosphor Plates PSP plates X-rays excite electrons which are trapped in the material lattice. Later, the plate is scanned by a laser in a “plate reader” which frees the electrons locally and digitizes the image. The plate can be reused. Plugs in to the film-screen cassette slot.

Solid State Detectors They are compact semiconductors. Electrical conductivity of semiconductor is sensitive to impurities. The depletion layer is sensitive to radiation and electric current flow through, thus the measured current is a measure of radiation.

X-ray Image Detection Screen-film: Still in use PSP Plates: Displacing screen-film in many applications Flat-panel: Increasing use but expensive Solid state: Still in development for X-ray Scintillation detectors: Not fast enough for X-ray imaging, but still important research tools. SPECT imaging Gas counter: Not useful for imaging but used for active measurement of patient exposure.