BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods

Slides:



Advertisements
Similar presentations
X-RAY INTERACTION WITH MATTER
Advertisements

Fund BioImag : Interaction of ionizing radiation with matter 1.What is the basis of contrast for x-ray imaging ? 2.By which mechanisms does ionizing.
X Rays Medical Physics Notes.
X-rays : Their Production Their Interaction with Matter
X-Ray & γ-Ray Interactions with Matter
Gamma-Ray Spectra _ + The photomultiplier records the (UV) light emitted during electronic recombination in the scintillator. Therefore, the spectrum collected.
X-Ray Interaction with Matter & Human Biology
Interactions of Radiation With Matter
2. The Particle-like Properties Of Electromagnetic Radiation
Physics of Radiotherapy Lecture II: Interaction of Ionizing Radiation With Matter.
Medical Imaging X-Rays I.
Radiation Interaction Q&A
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
INTERACTION OF IONISING RADIATION WITH MATTER
Interactions with Matter
Particle Interactions
Interaction of radiation with matter - 3
Physics of Radiography Interaction with matter. By the end of the first part of the session you should be able to: 1.Understand what can happen as x-ray.
Interaction of Gamma-Rays - General Considerations uncharged transfer of energy creation of fast electrons.
1 Photon Interactions  When a photon beam enters matter, it undergoes an interaction at random and is removed from the beam.
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
Assumptions: 1)Matter is composed of discrete particles (i.e. electrons, nucleus) 2)Distance between particles >> particle size 3)X-ray photons are small.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
For spRs sitting FRCR Part I Examinations Interaction of Radiation with Matter Amanda Barry, Ph.D.
DMI 261 Radiation Biology AndProtection. Unit 2 RADIATION Electromagnetic Spectrum –The frequency range of electromagnetic radiation and the photon wavelength.
Radiation Interactions
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Interactions of Radiation With Matter
Resident Physics Lectures
1 Patient Interactions 2010FINAL. 2 1.______________ 2.______________ 3.______________ 4.______________ 5.______________.
Lecture 1.3: Interaction of Radiation with Matter
Patient Interactions Photoelectric Classic Coherent Scatter
1 Dosimetry part 1: X-ray interactions with matter G.Haddadi, PhD, MSc Associated prof. of Medical Physics Fassa University of Medical Sciences.
Photon Tissue Interactions
Resident Physics Lectures Christensen, Chapter 4 Basic Interactions Between X-Rays and Matter George David Associate Professor Medical College of Georgia.
Medical Imaging Radiation I. Naked to the Bone: Medical Imaging in the Twentieth Century (Paperback)by Bettyann Kevles Bettyann Kevles E=mc2: A Biography.
Interactions of radiation with Matter
© Jimoid.com 2005 Ionising Radiation There are two types of radiation; ionising and non-ionising. Radiation Ionising Non-ionising Indirectly ionising (neutral.
Medical Image Analysis Interaction of Electromagnetic Radiation with Matter in Medical Imaging Figures come from the textbook: Medical Image Analysis,
X-rays discovered in 1895 by Wilhem Roentgen Roentgenology – Branch of medicine dealing with x-ray or gamma rays in diagnosis and treatment. Radiology.
Gamma ray interaction with matter A) Primary interactions 1) Coherent scattering (Rayleigh scattering) 2) Incoherent scattering (Compton scattering) 3)
By Dr: Nahla Nagy Assistant Professor Radiological Science Interactions of X-Rays with matter.
Chapter 5 Interactions of Ionizing Radiation. Ionization The process by which a neutral atom acquires a positive or a negative charge Directly ionizing.
Interactions of EM Radiation with Matter
Appendix G1: Chapter 12: X-Ray Interaction with Matter
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Gamma and X ray interactions
RAD 354 Chapt. 9 Interaction with Matter Five Interactions to know – Coherent (classical, unmodified, Thompson) – Compton – Photo Electric – Pair Production.
Interaction of Radiation with Matter
Interaction of x-ray photons (and gamma ray photons) with matter.
Alhanouf Alshedi Basic Interactions of Radiation with Matter 2 ed Lecture.
Barium & Iodine Imaging Physics
Interactions of Ionizing Radiation
Interaction of Radiation with Matter
Chapter 9 Concepts you already know!!! INTENSITY = ROENTGENS AKA EXPOSURE l mAs and mR proportional? l kVp and mRproportional? l Distance and mRproportional?
Chapter 2 Radiation Interactions with Matter East China Institute of Technology School of Nuclear Engineering and Technology LIU Yi-Bao Wang Ling.
Chp 2 : Interaction of radiation with matter Radiation Contrast
2.4.2 interaction of x-rays with matter
Chapter 5 Interactions of Ionizing Radiation
Interactions of Radiation With Matter
RAD 254 Chapter 10 X-ray interaction with matter
X-ray Interactions with Matter
Resident Physics Lectures (Year 1)
Concepts you already know!!!
Resident Physics Lectures (year 1)
Resident Physics Lectures (Year 1)
Lecturer Radiological Science
INTERACTION OF X-RAYS WITH MATTER
Computed Tomography (C.T)
Presentation transcript:

BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods Radiation Physics Part 2

Today EM Radiation interactions with matter Rayleigh scatter Photoelectric effect Compton scatter Pair production Attenuation of EM Radiation

X-ray interactions with matter Possible outcomes for an x-ray or gamma ray when traveling through matter are: No effect (a) (it travels along the original path with original energy) Absorption (b) (it is absorbed in the matter) Scatter (c) (It changes its direction of travel and/or energy) (a) (b) (c) Imaging detector

EM Radiation Four possible interactions with matter Rayleigh scatter Photoelectric effect Compton scatter Pair production Each has its own physical properties

huincident = huscattered ; lincident = lscattered Rayleigh Scattering Elastic (coherent, classical) scattering - scatter with direction change but no energy loss. huincident = huscattered ; lincident = lscattered Characteristics an interaction of the x-ray with the entire atom via a refractory mechanism (wavelength ~ atomic radius) occurs for longer wavelengths (low x-ray energies in the mammography range - 15 - 30keV) reduces imaging quality in x-ray imaging. E(keV) P(%) ~30 <12 >70 <5 P: probability

Compton scattering It is also referred as inelastic or non-classical scattering an interaction of incident x-ray with outer shell electrons; almost atomic (proton) number (Z) independent; proportional to re; weakly energy dependent until energy is high (1MeV); a disturbance to x-ray image quality.

Compton Scatter Equation Planck’s constant Scatter angle Electron rest mass wavelengths

Compton Scattering energy and angular dependence hu0 husc Ee- q f Energy and momentum conservation q

f = 00 & q = 1800 Compton scattering (Max energy transfer to electron) hu0 husc Ee- q f

q = 900 Compton scattering hu0 husc Ee- q f

Compton scatter angular distribution hu0 husc Ee- q f The lower the incident x-ray energy, the more isotropic (angular distribution) the (compton) scattered x-ray is. (Isotropic - each point of compton interaction becomes an x-ray source - bad for imaging)

Photoelectric Effect (PE) An total absorption event (not a scatter): all or none; An interaction of incident x-ray with inner shell electron; Highly dependent on atomic (proton) number (Z) - Z3; Highly energy dependent - E-3;

Photoelectric effect Incident electron deposits all of its energy into a shell electron Must overcome electron binding energy The shell electron (photoelectron) escapes Ionization event Energetic electron interacts with tissue (short range) The photon must start with energy greater than the electron binding energy.

Photoelectric Effect “K-edges” Photoelectric attenuation as a function of energy is discontinuous at the shell binding energies. We make use of this property in several ways. From Sprawls, The Physical Basis of Medical Imaging

Photoelectric effect energy and Z dependence High Z dependence (Z3) of PE is the basic reason why diagnostic x-ray is suited to medical imaging: PE completely removes the incident x-ray (no scatter or partial absorption) PE differentiates different atomic compositions in tissue (bone and fat); PE is behind the use of high Z image contrast materials; PE decreases drastically (E-3) with increasing x-ray energy (not good for imaging but good for therapy)

X-ray imaging contrast ZBa=56; r =3.5g/cm3 Zwater, effective=7.25; r =1g/cm3 (ZBa/Zwater)3=512

Pair Production occurs under strong electric field of nucleus (=>Z dependent); only occurs if the incident x-ray energy is > 1.022MeV (enough to create the rest masses of an e- and a e+); completely removes the incident x-ray (no scatter or partial absorption) increases with increasing x-ray energy generates an electron and a positron

Positron annihilation 2nd part of fig 3-12, redbook Positron annihilation gamma rays are ~1800 apart; each gamma ray has 511keV; principle of PET scanner (a measure of PET agent distribution in patient).

EM Radiation Interactions Rayleigh scatter: only significant at low energies Compton scatter: dominates at imaging energies Photoelectric effect: significant at imaging energies Pair production: cannot occur at imaging energies

Total linear attenuation coefficient Soft Tissue

X-ray attenuation incident N Absorbed & scattered n N-n Dx transmitted Linear attenuation coefficient m (cm-1): m= 0.10 cm-1 ==> 10% per cm.

Total attenuation coefficients of water and lead Characteristic x-ray peak (K-shell) Table of interaction vs. x-ray energy (Khan?)

Relative importance of each x-ray interaction in water

m/r (cm2/g) = m (cm-1) /r (g/cm3) Nx = N0 e- (m/r) r x Mass attenuation coefficient (m/r) linear attenuation coefficient (m) normalized to the density (r) of the irradiated material. Therefore it is independent of density. m/r (cm2/g) = m (cm-1) /r (g/cm3) Nx = N0 e- (m/r) r x

Exercise xap xat Given: CT x-ray energy is120kVp. The patient thickness is 20 cm and 25 cm in the AP and lateral direction, respectively. If the x-rays attenuate 20% per cm of tissue, what is the x-ray intensity difference as measured by the detector in the AP and lateral positions? xap patient xat Solution: m = 0.20 cm-1 X-ray detector

Half Value Layer (HVL) 100 50 = 100/2 (a) (b) (c) HVL or m is a function of BOTH the x-ray (E) and the matter (Z, r). T1/2 is a function of radionucleus only. 50 = 100/2

Narrow-beam and broad-beam HVL which HVL is larger, broad-beam or narrow-beam?

Beam Hardening: HVL (x) Polyenergetic x-ray beam Beam Hardening: HVL (x) Beam hardening - mean energy of a poly energetic x-ray beam increases as it travels through the irradiated material. Reason - lower energy x-rays have larger m than higher energy x-rays and thus attenuate quicker. HVL(x1) < HVL (x2) when x1 < x2 (for polyenergetic x-ray beams)

Review of interactions Particles (electrons) EM Collisional Ionization, heat Radiative X-rays (characteristic, Bremsstrahlung) Rayleigh scatter Change direction, no energy loss Compton scatter Change direction, energy loss Photoelectric Absorption, photoelectron released Pair production Absorption, electron-positron pair released