Electrical Surgical Unit

Slides:



Advertisements
Similar presentations
Effect of Electricity on Living Tissue
Advertisements

BIOPOTENTIAL AMPLIFIERS
Electronic Devices Ninth Edition Floyd Chapter 11.
Surgical Diathermy Machine
Kenny Low Dec 2009 Urology registrar Thursday teaching
EE 4BD4 Lecture 25 Electrosurgery Biomedical Device Technology: Principles and Design, Charles C. Thomas Publisher 2007 Anthony K. Chan 1.
Shortwave and Microwave Diathermy part(2). Induction Electrodes (Drum Electrode) One Or More Monopolar Coils Rigidly Fixed In A Housing Unit May Use More.
Chapter 8 Inverters AC Power • Inverters • Power Conditioning Units • Inverter Features and Specifications.
POWER SUPPILES LECTURE 20.
Pacemaker & its Classification
Diathermy Interactions with Implanted Leads and Implanted Systems with Leads Marian Kroen Office of Surveillance and Biometrics Issues Management Staff.
CHAPTER 13 TRANSMITTERS AND RECEIVERS. Frequency Modulation (FM) Receiver.
28. Alternating Current Circuits
Chapter 22 Alternating-Current Circuits and Machines.
NFPA ELECTRIC VEHICLE SAFETY FOR EMERGENCY RESPONDERS Module II : Basic Electrical Concepts And Hazards Module II : Basic Electrical Concepts And Hazards.
Electrical, Electronic and Digital Principles (EEDP)
General Licensing Class G7A – G7C Practical Circuits Your organization and dates here.
What is a Watt? Everything you ever wanted to know about Electrosurgery. Fahad bamehriz, MD.
Power Electronics and Drives (Version ) Dr. Zainal Salam, UTM-JB 1 Chapter 3 DC to DC CONVERTER (CHOPPER) General Buck converter Boost converter.
Physiological Effects of Electricity on Human Body by Mohd Yusof Baharuddin.
Alternating Current Circuits
What is electrosurgery?
Diathermy.
Lasers for eye treatment The laser was invented in 1960, and in 1961 this laser (Ruby) was used by eye doctors. It is natural that the eye was chosen to.
UNIT-1 Rectifiers & Power Supplies. Rectifier A rectifier is an electrical device that converts alternating current (AC), which periodically reverses.
Without reference, identify principles about Electrosurgical Units with at least 70 percent accuracy.
HAZARD OF MEDICAL INSTRUMENT by Mohd Yusof Baharuddin.
Applied Circuit Analysis Chapter 3 - Power and Energy Copyright © 2013 The McGraw-Hill Companies, Inc. Permission required for reproduction or display.
Electrical Principles 1 1 G5 - ELECTRICAL PRINCIPLES [3 exam questions - 3 groups] G5A - Reactance; inductance; capacitance; impedance; impedance matching.
PSST 10819_11/2008 Arrhythmia in Practice Today ™ Content provided by Boston Scientific September 2008 Lessons in Lesions.
HAZARD OF MEDICAL INSTRUMENT II by Mohd Yusof Baharuddin.
DEVICES TO PROTECT AGAINST ELECTRICAL HAZARDS
BASICS OF ELECTROSURGERY
Electrosurgery & Gyn Surgery: Get the Point Across
Computer controlled I NSTANT R ESPONSE ™ system provides safer and more precise electrosurgery. TM.
Cauterization Catheter – An Advancement in Conductive Biomaterials and Medicine C. Blyth 1, C. Fernandez 1, S. Hittinger 1, C. Jones 1, B. McGee 1, B.
ELECTRICAL SAFETY OUTLINE  The basic effect of electricity on human body  Ventricular Fibrillation  Electric shock  Electric Shock – Treatment  Important.
Electro surg i cal unit  PURPOSE ESU are used for surgical cutting and for controlling bleeding by causing coagulation “ hemostasis ” at the surgical.
Chapter 4 Bipolar Junction Transistors
Dr.Moallemy Radiofrequency Lesioning. Dr.Moallemy  Radiofrequency (RF) current is used in pain medicine to make discrete therapeutic lesions in various.
Electro surgery in Gynaecology
Cauterization Catheter – An Advancement in Conductive Biomaterials and Medicine C. Blyth 1, C. Fernandez 1, S. Hittinger 1, C. Jones 1, B. McGee 1 Advisors:
This presentation may contain some forward looking statements, particularly regarding operational prospects in 2009 and beyond, which involve a number.
RDT Hospital, Dept of Gen Surgery
Electrosurgical Unit.
Ultrasound 10cm 2 Ultrasound (US) Therapeutic ultrasound –Uses acoustic energy, delivered at high frequencies for therapeutic purposes Electrical current.
In practice, it is often required to make or break an electric al circuit in many operations. In some applications, it is desirable that this make and.
Signal Analyzers. Introduction In the first 14 chapters we discussed measurement techniques in the time domain, that is, measurement of parameters that.
Switch Mode Power Supply(SMPS) BY: Arijit Acharya NETAJI SUBHASH ENGINEERING COLLEGE M.tech(P.S.) Roll No - 1.
Electrosurgery Units ELECTROSURGICAL UNITS (ESUs)
SAFE USE OF DIATHERMY RACHEAL AFOLABI, RN, RM, H. Dip. PERI-OPERATIVE NURSING, H.Dip. HEALTH SYSTEM MANAGEMENT.
The Working Theory of an RC Coupled Amplifier in Electronics.
Chapter 12 Therapeutic Heating Applications of Radio Frequency Energy C-K. Chou.
BME 312 BIOMEDICAL INSTRUMENTATION II LECTURER: ALİ IŞIN LECTURE NOTE 7 Electrosurgical Devices BME 312-BMI II-L7- ALİ IŞIN 2014 FACULTY OF ENGINEERING.
DIATHERMY Diathermy was once one of the most popular modalities used in rehab. Became less popular with development of more modern tech like US and because.
Electro surgical unit.
Electrosurgical/Surgical Diathermy Units
VUSE Medical and Electromechanical Design Lab
A. In monopolar electrosurgery, current from an electrosurgical generator passes from an active electrode (the “Bovie” tip) through the patient to a return.
Hemodialysis Lecture 3.
Maintenance schedule for different equipment and their records in a hospital By Mr. Rahul kumar.
POWER AMPLIFIERS.
SURGICAL DIATHERMY.
General Licensing Class
SURGICAL DIATHERMY.
Iontophoresis Vineela.U 08B21A0538 CSE.
Amplifiers Classes Electronics-II
Diathermy and its safe use
Amplifiers Classes Electronics-II
Fundamental Use of Surgical Energy
Presentation transcript:

Electrical Surgical Unit Dr Fadhl Al-Akwaa fadlwork@gmail.com www.Fadhl-alakwa.weebly.com Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material ESU Importance Radio frequency current has been used for many years to reduce time under anesthesia and improve the clinical result. The ability to simultaneously cut tissue and coagulate blood vessels greatly reduces morbidity and mortality, and enables a large number of procedures that are otherwise not possible, especially in neurosurgery. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material AGENDA 81.1 Theory of Operation 81.2 Monopolar Mode 81.3 Bipolar Mode 81.4 ESU Design 81.5 Active Electrodes 81.6 Dispersive Electrodes 81.7 ESU Hazards 81.8 Recent Developments Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Terminology Cutting Coagulation= fulgurate + desiccation Desiccation OR contact coagulation. تجفيف: process of drying food Fulgurate: (fulgur is Latin for lightning). destroy using electricity (such as a tumor, etc.) Hemostasis: stoppage of bleeding Blended mixed, merged, combined choice of blend waveforms to allow the surgeon to select the degree of hemostasis desired. Bleeding RF frequency 300 kHz and 5 MHz Cautery: process of burning or scarring with a hot iron or other implement الكي Ablating: remove by cutting Please contact Dr Fadhl to use this material

Electrosurgery vs electrocautery Electrosurgery is sometimes erroneously referred to as electrocautery. In cauterization the essential physical mechanism behind the treatment is conduction heat transfer from a hot object placed on the surface to raise the temperature high enough to denature the tissue proteins. Cutting and coagulation by means of electrosurgery is also accomplished by heating tissue to high temperatures, but the essential difference is that the primary mechanism is electrical power dissipation directly in the affected tissues themselves, rather than heat transfer from an external hot object on the tissue surface. Please contact Dr Fadhl to use this material

Electrosurgery vs Diathermy Electrosurgery is also referred to as surgical diathermy, particularly in Europe. diathermy literally means through-heating, such as might be applied in physical medicine for the relief of pain or in hyperthermia therapy for tumor treatment. Electrosurgical devices in operating rooms are designed and built for surgical use only. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Clinical application Electrosurgery is commonly used in dermatological, gynecological, cardiac, plastic, ocular, spine, ENT, Ear Nose & Throat orthopedic, urological, neuro- and general surgical procedures. Please contact Dr Fadhl to use this material

Duty cycle=on time/off time Continuous Interrupted Please contact Dr Fadhl to use this material

Crest Factor The Crest Factor is equal to the peak amplitude of a waveform divided by the RMS value. Electrical engineering — for describing the quality of an AC power waveform http://en.wikipedia.org/wiki/Crest_factor http://www.dliengineering.com/published/TECH-CrestFactor.pdf

Current density (high or low) Please contact Dr Fadhl to use this material

Electrosurgical modes Please contact Dr Fadhl to use this material

Electrosurgical modes Monopolar mode AND Bipolar mode The most noticeable difference between these two modes is the method in which the electric current enters and leaves the tissue. In the monopolar mode, the current flows from a small active electrode into the surgical site, spreads through the body, and returns to a large dispersive electrode on the skin. In the bipolar mode, the current flows only through the tissue held between two forceps electrodes. The monopolar mode is used for both cutting and coagulation. The bipolar mode is used primarily for coagulation. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material History Development of the first commercial electrosurgical device is credited to Dr. William T. Bovie, who worked on it from 1914 to 1927 while employed at Harvard University[8] [9] The first use of an electrosurgical generator in operating room occurred on October 1, 1926. The surgery was performed by Dr. Harvey Cushing. Please contact Dr Fadhl to use this material

Theory of Operation: Temperature What is the general effects of heat on biologic tissue? What happen when tissue temperature increase from normal body temperature to 45°C? neither microscopic nor macroscopic changes, some cytochemical changes do in fact occur. the cells return to their normal function when the temperature returns to normal values. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Theory of Operation What is the general effects of heat on biologic tissue? Above 45°C, Normal cell functions was inhibited and lead cell was death. Further increasing the temperature up to 100°C leads to tissue drying; that is, the aqueous cell contents evaporate. This process is called desiccation. If the temperature is increased beyond 100°C, the solid contents of the tissue reduce to carbon, a process referred to as carbonization. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Bioheat equation Where T and To are the final and initial temperatures (K), σ is the electrical conductivity (S/m), ρ is the tissue density (kg/m3) C is the specific heat of the tissue (Jkg-1k-1) J is the current density (A/m2), and t is the duration of heat applications Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Bioheat equation The bioheat equation is valid for short application times where secondary effects such as heat transfer to surrounding tissues, blood perfusion, and metabolic heat can be neglected. the surgeon has primarily three means of controlling the cutting or coagulation effect during electrosurgery: the contact area between active electrode and tissue, The electrical current density, and the activation time. output power vs. tissue impedance Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Output values for ESUs, Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Waveform Interrupted waveforms, such as exponentially damped sinusoids (obtained from spark gap or other relaxation-type oscillators) are effective for coagulation techniques requiring fulguration, or intense sparking (fulgur is Latin for lightning). Continuous sine waveforms (e.g., those obtained from vacuum tube or transistor oscillators) have proven most effective for cutting. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Cut/Coag Cut/Coag Most wet field electrosurgical systems operate in two modes; Cut where the tissue is vaporized and Coag where the tissue is dried. Cut If the voltage level is high enough then the heat generated can generate a vapour pocket. The vapour pocket typically reaches temperatures of approximately 400 Degrees Centigrade which vaporizes soft tissues. Coag When the system is operating in coag mode the voltage output is usually lower than in cut mode and less power is delivered. This therefore generates less heat and a vapour pocket is not generated. Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Why high frequency? sensitivity to electrical stimulation decreases with increasing frequency in the kHz range and above. Neural and muscle cells are electrically-excitable, i.e. they can be stimulated by electric current. In human patients such stimulation may cause acute pain, muscle spasms, and even cardiac arrest. To minimize the effects of muscle and neural stimulation, electrosurgical equipment typically operates in the radio frequency (RF) range of 100 kHz to 5 MHz. Please contact Dr Fadhl to use this material

high-frequency power amplifiers vacuum tube/ spark gap 1907, Lee De Forrest, power transistors, solid-state electrosurgical generators(Valleylab and EMS) 1970 Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material ESU Design RF Generators for Electrosurgery “high-frequency power amplifiers” control input switches, safety monitor. Microcontroller Power supply Microcontroller Electrodes Active electrode return electrode, Please contact Dr Fadhl to use this material

Block diagram for typical solid state ESU Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Safety Appliances. American National Standard for Electrosurgical Devices the return electrode losing its attachment to the skin>> skin burns at the dispersive, or return, electrode site. Patient return monitor circuit Please contact Dr Fadhl to use this material

Electrosurgery Analyzers QA-ES Electrosurgery Analyzer Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Performance testing Tests include: • Automatic power distribution measurement • Crest factor measurement • RF leak measurement • Return electrode monitor (REM) test Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material Power Output Tests output current (A), power (W), peak-to-peak voltage (V), crest factor values. Please contact Dr Fadhl to use this material

HF Current Leakage Test This test checks whether the active and dispersive leakage currents are within acceptable limits. Please contact Dr Fadhl to use this material

Measuring Leakage Current with HF Isolated Equipment Please contact Dr Fadhl to use this material

Measuring Leakage Current with Grounded HF Equipment Please contact Dr Fadhl to use this material

Please contact Dr Fadhl to use this material REM Alarm Test The REM (Return Electrode Monitoring) alarm test ensures that the ESU sounds an alarm if the resistance between the two neutral electrodes exceeds its specified limit Please contact Dr Fadhl to use this material

Electrosurgery Analyzers Typical measurement parameters: RF current (ma) RF power (watts) Crest Factor (CF) – the measure of Vpeak / VRMS 1.414 to 1000 range Duty cycle (pulsed waveforms) 1% - 100% range Through “non-inductive” test loads from 0Ω to > 5000Ω 5th generation parameters include Frequency Pulsed parameters such as time on, time off, total cycle time, duty cycle