Servo-Controlled Blood Vessel Occluder Ahmed El-Gawish, Alan Chen, Hugo Loo, & Imad Mohammad Advisor: Ki Chon
Objectives Renal autoregulation research The device increases or decreases blood pressure by ~20 mmHg with fast response times The response time should be in order of 100 ms.
The Output KI Chon project proposal
Original Design
Modifications Cylinder Linear actuator BiStep Motor Controller
Cylinder
Linear Actuator 50 lb Stepping actuator rated nominal force 1600 steps per inch 5 inch stroke length 4 connecting leads Herbach & Rademan Company (
Motor Connection Diagrams
BiStep Motor Controller Complete Unipolar/Bipolar dual stepper motor controller system Terminal block connectors for its TTL input control as well as its motor output and power connections Jumper for switching between use of TTL-serial and RS232- serial Parallax inc
Driver Chip
Progress Overview LabVIEW VI
Warning! The following slides may be unsuitable, offensive, or possibly even kinda-sorta profoundly disturbing to some viewers. Viewer discretion is advised
Results
Output Red is the Blood Pressure signal, blue is the average over time, response time is approximately 300 ms in this plot
Setbacks Response time Undershoot due to PI Automation Funds not fully exploited
Further Improvements Faster/more accurate motor Optimize cylinder diameter VI improvements (more controls/analysis) Control, control, control
Testing Integration Testing Pressurized Tube PI Calibration Live Subject