Arizona’s First University. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis.

Slides:



Advertisements
Similar presentations
Orthogonal and Least-Squares Based Coordinate Transforms for Optical Alignment Verification in Radiosurgery Ernesto Gomez PhD, Yasha Karant PhD, Veysi.
Advertisements

DR.RAVINDRA SRIVASTAVA, CONSULTANT NEUROSURGEON VIMHANS HOSPITAL.NEW DELHI.
In the past few years the usage of conformal and IMRT treatments has been increasing rapidly. These treatments employ the use of tighter margins around.
The Tomotherapy Experience at Advocate Good Samaritan Hospital
Copyright ©2008 Accuray, Incorporated. All rights reserved E CyberKnife ® Robotic Radiosurgery System Radiosurgery System Comparisons.
Seeram Chapter 13: Single Slice Spiral - Helical CT
The TrueBeam System ™ Clinic Name Presenter’s name Clinic location
Targeting Issues For Proton Treatments Of The Prostate SJ Rosenthal Ph.D., JA Wolfgang Ph.D., Sashi Kollipara Department of Radiation Oncology Massachusetts.
1 ADC Calculations Lars Ewell Radiation Oncology University of Arizona Medical Center 2/8/08 1.
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID.
An algorithm for metal streaking artifact reduction in cone beam CT M. Bazalova 1,2, G. Landry 1, L. Beaulieu 3, F. Verhaegen 1,4 1-McGill University,
tomos = slice, graphein = to write
Chapter 2 Stewart C. Bushong
Innovation/Impact: By designing a simulated human shaped (anthropomorphic) plastic phantom with targets, organs at risk (OAR) and heterogeneities, the.
Design and Scheduling of Proton Therapy Treatment Centers Stuart Price, University of Maryland Bruce Golden, University of Maryland Edward Wasil, American.
COMPUTED TOMOGRAPHY I – RAD 365 CT - Scan
Radiotherapy Treatment Planning
Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator Andrew J Reilly Imaging Physicist Oncology Physics Edinburgh Cancer.
بسم الله الرحمن الرحيم و قل رب زدنى علماً ﴿و قل رب زدنى علماً﴾ صدق الله العظيم.
What Happens in the days before treatment ?
 Quality & Safety Considerationsin Stereotactic Radiosurgery and Stereotactic Body Radiation Therapy.
Simulating Potential Layouts for a Proton Therapy Treatment Center Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil-
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
Department of Radiation Oncology Henry Ford Health System
Safety Consideration Software limits on robot controller Limit switches on the robot “wrist” to prevent excess rotations Limit switches on the vertical.
The days before Treatment
Navigation in Robotic Radiosurgery Achim Schweikard.
Steven Marsh, James Eagle, Juergen Meyer and Adrian Clark
Kelly Younge, Ph.DKelly Younge, Ph.D Don Roberts, Benedick Fraass, Daniel McShan, and Martha Matuszak University of Michigan, Department of Radiation Oncology,University.
1 Overview of Gamma Knife ® Surgery Dr. Sandra Vermeulen, M.D. Swedish Cancer Institute Northwest Hospital Gamma Knife Center Seattle, Washington Learn.
Evaluation of the Performance of the Fast Scanning Platform of an OCT System Malcolm Heard 1, Miguel Herrera 1, Geoffrey Ibbott 1 1 Department of Radiation.
Software Development For Correction of Gradient- Nonlinearity Distortions in MR Images T.S. Lee, K.E. Schubert Computer Science CSUSB R.W. Schulte Radiation.
Comparison of Clinical Parameters for Proton Therapy in the United States Paige Summers, MS.
MRI SIMULATION FOR CONFORMAL RADIATION THERAPY OF PROSTATE CANCER Pasquier D 1-3, Palos G 3, Castelain B 1, Lartigau E 1,2, Rousseau J 2,3 1 Department.
HAMBURG BEAM PIPE POSSIBLE LAYOUT AND MANUFACTURING JANUARY 2012.
PRINCIPLES OF CT Dr Mohamed El Safwany, MD. Intended learning outcome The student should learn at the end of this lecture principles of CT.
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Principles and Practice of Radiation Therapy
Physics of carbon ions and principles of beam scanning G. Kraft Biophysik, GSI, Darmstadt, Germany PTCOG43 Educational Satellite Meeting: Principles of.
Part No...., Module No....Lesson No
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Quality Assurance.
Introduction to Radiation Therapy
Measurements of DB supporting systems Fiducialisation Sylvain GRIFFET, 21/03/2011 Measurements performed in February and early March /13 EDMS n°
Purpose N-isopropylacrylamide (NIPAM) polymer gel dosimeters were employed to verify the dose distribution of clinical intensity modulated radiation therapy.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 26 Three-Dimensional Digital Imaging.
Carmel McDerby Clatterbridge Centre for Oncology, Merseyside,UK
A 2D/3D correspondence building method for reconstruction of a 3D bone surface model Longwei Fang
A virtual simulator in an multivendor radiotherapy department, an overview of the commissioning process Geert Pittomvils, Lobke Mommaerts, Frederic Crop,
Rapid Arc Treatment Verification: post evaluation on Delta-4 and proposal of a new verification protocol G. Pittomvils 1,,L. Paelinck 1, T. Boterberg 1,
Cone Beam CT at the Ghent University Hospital: first clinical results and evaluation of the selected workflow. G. Pittomvils 1,,M. Coghe 1, A. Impens 1,
A Clinical Audit in Stereotactic Radiotherapy Lucy Richley Bristol Haematology and Oncology Centre 8 th June 2007.
Treatment Chart Record of patients radiation therapy history. Must contain: History and diagnosis Rationale for treatment Treatment plan Consent Documentation.
Medical Dosimetry Presented by (Add Hospital name) in conjunction with The American Association of Medical Dosimetrists.
Dr. Malhar Patel DNB (Radiation Oncology)
You Zhang, Jeffrey Meyer, Joubin Nasehi Tehrani, Jing Wang
Linac Commissioning Overview
An Investigation of the Effect of Window and Level Controls on the accuracy of the ExacTrac Patient Repositioning System Dan Goldbaum and Russell Hamilton.
Left Posterior Superior Right Anterior Inferior
Orthogonal and Least-Squares Based Coordinate Transforms for Optical Alignment Verification in Radiosurgery Ernesto Gomez PhD, Yasha Karant PhD, Veysi.
Using the CRW Frame for the Stereotactic Biopsy of Cerebral Lesions
Reducing Treatment Time and MUs by using Dynamic Conformal Arc Therapy for SBRT Breath-Hold Patients Timothy Miller, Sebastian Nunez Albermann, Besil Raju,
Insert tables Insert figure
Doc.Ing. Josef Novotný,CSc
LEKSELL GAMMA KNIFE PERFEXION
L. A. den Otter. , R. M. Anakotta. , M. Dieters. , C. T. Muijs. , S
Novalis Frameless Radiosurgery
C-13 O-arm Scatter Cloud.
GHG meeting at ESTRO36 May, 2017
Presentation transcript:

Arizona’s First University. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis

Physicists (University of Arizona) RJ Hamilton CJ Watchman J Gordon NJ Stanley Radiation Oncologist (University of Arizona) B Stea Neurosurgeon (Center for Neurosciences, Tucson, AZ) A Sanan Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis

Accuracy of Frame-Based Novalis SRS for TGN Rahimian et al (J. Neurosurg S3 101:351-55, 2004) –Image Fusion 0.4 mm –MR 0.2 mm –Gantry 0.3 mm –Couch 0.6 mm –Overall 0.9 mm (0.6 mm w/o couch) Target position 0.2 mm Laser 0.2 mm Phantom Radiographic Test 0.7 ± 0.1 mm

Frame-Based SRS for TGN Human Issues –Pins fix frame to skull –Uncomfortable during treatment delivery too –Patient remains in department for several hours waiting for treatment –Difficult to abort procedure (machine problems) –Scheduling: patient, neurosurgeon, radiation oncologist, physicist, treatment time, etc…

Frame-Based SRS for TGN Technical Issues –Frame-based systems require that the position of a patient in the frame is unchanged from application of the pins through treatment completion –The stereotactic frame (not the patient) is aligned with the machine isocenter for treatment –Differential torque on the head ring deforms it –Head ring is bolted to the table assembly, patient may be displaced relative to it.

Image-Based SRS for TGN Human Issues –Plastic mask (BrainLab U-PLAST mask fixation system) helps patient remain still –Scheduling: specialists (neurosurgeon, radiation oncologist) may work at their convenience, treatment time set for patient

Image-Based SRS for TGN Technical Issues –Positioning accuracy Depends on the accuracy of the registration of images between the planning system (DRRs) and imaging system (ExacTrac) Depends on the calibration of the imaging system with the mechanical isocenter –Variations of the position of the patient in the mask (stereotactic coordinate space) are not important provided that the imaging system can correct for it (ExacTrac X-Ray 6D and Robotics) –Once positioned for treatment, the patient must remain fixed.

Feasibility of Image-Based SRS for TGN Demonstrate System Accuracy –Anthropomorphic Phantom Experiment Demonstrate that variations of the patient position in the mask are not important –Anthropomorphic Phantom Experiment Confirm for each patient that deviations of their head within the mask may be corrected by the ExacTrac X-ray 6D and robotics –Perform planning CT scan plus two addition CT scans. The final scan is performed 1 hour prior to the scheduled treatment time. Confirm that the patient does not move during treatment –Repeat the ExacTrac X-ray positioning 3-4 times (every 3 rd couch position) during treatment delivery

Demonstrate System Accuracy Anthropomorphic Phantom –0.25” BB embedded in head –BrainLab U-PLAST mask constructed –Axial CT scan (35 cm FOV 1.5mm slice) with BrainLab Head & Neck Localizer –BrainScan Plan –ExacTrac 6D X-Ray and Robotics

Anthropomorphic Phantom CT

Anthropomorphic Phantom Plan

Accuracy of Isocenter Determination Two planners contoured the BB in the planning system and then placed the isocenter at the center. Repeated the procedure 10 times each.

Pixel Size Pixel Size = 0.5 mm

Pixel Size Pixel Size = 0.5 mm

Pixel Size Pixel Size = 0.5 mm

Pixel Size Pixel Size = 0.5 mm 0.5 mm

ExacTrac Calibration

Anthropomorphic Phantom Treatment Head Holder

Anthropomorphic Phantom Treatment

Select Isocenter

Anthropomorphic Phantom Treatment Automatic Preposition

Anthropomorphic Phantom Treatment

Automatic Preposition Film Verification Difference = ( 0.3, -0.2 ) = 0.4 mm 10 mm cone BB shadow

Anthropomorphic Phantom Treatment

Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm 10 mm cone BB shadow

Phantom Displacement Experiment Phantom moved by exactly 2.0 cm in all three directions as determined by a ruler Perform X-ray Alignment with ExacTrac Shifts Found by ExacTrac –AP mm –Lateral20.05 mm –Vertical20.40 mm

Phantom Moved: Not in Mask !!! Head Rotated

Phantom Moved: Not in Mask !!! No Head Holder  Head Lower and Rotated Head Holder Removed

Phantom Moved: Not in Mask !!! Position using standard procedures with array

Phantom Moved: Not in Mask !!!

Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm 10 mm cone BB shadow

Phantom Moved: Not in Mask !!! Rotation too large for robotics

Phantom Moved: Not in Mask !!!

Uncorrected Rotations Isocenter position is still fine Uncorrected rotation changes geometry of beams relative to skull and therefore changes the calculated doses

Patient Repeat CT Scans Original CT scan includes mask fabrication Second CT scan scheduled between date of original scan and treatment day Third CT scan obtained one hour prior to scheduled treatment time

Patient Repeat CT Scans Results The mean observed isocenter displacements from repeat CT localizations in the mask –AP 0.9 mm –Lateral 0.6 mm –Vertical 1.7 mm

Patient Motion During Treatment ExacTrac localization was preformed during treatment every three arcs and at the end of the procedure to evaluate the extent of patient movement in the mask during treatment.

Patient Motion During Treatment: Results The mean displacements from isocenter –AP 0.3 mm –Lateral 0.5mm –Vertical 0.3 mm The mean rotations were all less than 0.4 degrees.

Dosimetric Effects of Isocenter Misplacement Five patients treated to 90 Gy at the isocenter with a 4 mm cone Isocenters of original plans were displaced towards the brainstem by 0.5, 1.0, 1.5, and 2.0 mm (worst case scenarios). Maximum brainstem and CN5 doses computed

Isocenter Misplacement Results: CN5

Isocenter Misplacement Results: Brainstem

Conclusions Accuracy is better than image pixel size Mask reproducibility allows accurate positioning with ExacTrac 6D and Robotics Patients move minimally during treatment. Monitoring the position adds little to the overall treatment time and insures accuracy Patients tolerate the procedure well and have not experienced complications

Thank You!