Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QA Maria Sj ö lin Department of Oncology,

Slides:



Advertisements
Similar presentations
QA for the Radiotherapy
Advertisements

F. Foppiano, M.G. Pia, M. Piergentili Medical Linac IEEE NSS, October 2004, Rome, Italy
Backscatter for EPID dosimetry
RapidArc in Bergen Britt Nygaard, Harald Valen and Ellen Wasbø
Pencil-Beam Redefinition Algorithm Robert Boyd, Ph.D.
Ionization Chamber Array for External Beam Radiotherapy
RapidArc plan verification using ArcCHECK™
The Tomotherapy Experience at Advocate Good Samaritan Hospital
RAD309 Patient Dose.
Maria Skłodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warsaw, Poland 8th ECMP, Athens, Dosimetry audits in radiotherapy.
Background:  IMRT has become the choice of treatment for disease sites that require critical structure sparing such as head and neck cancer.  It has.
The TrueBeam System ™ Clinic Name Presenter’s name Clinic location
Real-time tumor tracking with preprogrammed dynamic MLC motion and adaptive dose-rate regulation B.Y Yi, S. Han-Oh, F. Lerma, B. Berman, C. Yu.
Estimating 3D Respiratory Motion from Orbiting Views Rongping Zeng, Jeffrey A. Fessler, James M. Balter The University of Michigan Oct Funding provided.
Lotte Verbunt Investigation of leaf positioning accuracy of two types of Siemens MLCs making use of an EPID.
MONTE CARLO RADIATION DOSE SIMULATIONS AND DOSIMETRY COMPARISON OF THE MODEL 6711 AND I BRACHYTHERAPY SOURCES Mark J. Rivard Department of Radiation.
Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen,
Introduction Modern radiation therapies such as intensity-modulated radiation therapy (IMRT) and volume modulated arc therapy (VMAT) demand from dose calculation.
Quality Control Rad T 110.
Dose Distribution and Scatter Analysis
What Happens in the days before treatment ?
Patient Specific Quality Assurance for IMRT and VMAT based on Radiochromic Film Dosimetry with EBT2/EBT3 Film Sou-Tung Chiu-Tsao, PhD Quality MediPhys.
TWO FIELD BREAST PLAN VS. OPTIMIZED CONFORMAL BREAST PLAN: COMPARISON OF PLAN PARAMETERS Authors: Borko Basarić, Ozren Čudić, Milan Teodorović, Borislava.
Evaluation of New Pre-Treatment In-Air Patient Specific QA Software for TomoTherapy Treatments Lydia L. Handsfield¹, Quan Chen¹, Kai Ding¹, Wendel Renner²,
Quality Control in Radiation Therapy, A New Concept: Dosimetry Check
Challenges for TPS Chunhua Men Elekta Software, Treatment Planning System BIRS Workshop Banff, Canada 3/12/2011.
Quality Assurance for a modern treatment planning system
Introduction to Dosimetry Check
AUTHORS (ALL): Huang, Xiaoyan 1, 2 ; Kuan, K M 2 ; Xiao, G L 2 ; Tsao, S Y 3, 2 ; Qiu, X B 2 ; Ng, K 2. INSTITUTIONS (ALL): 1. Radiation Oncology, Sun.
Kelly Younge, Ph.DKelly Younge, Ph.D Don Roberts, Benedick Fraass, Daniel McShan, and Martha Matuszak University of Michigan, Department of Radiation Oncology,University.
Applications of Geant4 in Proton Radiotherapy at the University of Texas M.D. Anderson Cancer Center Jerimy C. Polf Assistant Professor Department of Radiation.
MONTE CARLO BASED ADAPTIVE EPID DOSE KERNEL ACCOUNTING FOR DIFFERENT FIELD SIZE RESPONSES OF IMAGERS S. Wang, J. Gardner, J. Gordon W. Li, L. Clews, P.
Surface dose prediction and verification for IMRT plans using line dose profiles † Ronald E. Berg, † Michael S. Gossman and ‡ Stephen J. Klash † Erlanger.
IMRT QA Plan Site 5%/3mm3%/3mm2%/2mm 0% noise1% noise2% noise0% noise1% noise2% noise0% noise1% noise2% noise HN
CT physics and instrumentation
Application of a 2-D ionization chamber array for dose verification of dynamic IMRT with a micro-MLC Fujio ARAKI, PhD 1, S. TAJIRI 2, H. TOMINAGA 2, K.
Institute for Advanced Radiation Oncology
Integrating the Health Care Enterprise- Radiation Oncology Use Case: In Vivo Patient Dosimetry Editor: Juan Carlos Celi - IBA Reviewer: Zheng Chang – Duke.
6. Clinical implementation and SBRT quality assurance
Use of the IC Profiler detector array for comprehensive machine QA ESTRO QA & Dosimetry Satellite Symposium Steve Morgan, Medical Physics Dept,
Dosimetric discrepancies due to positional errors in MLC movement during stereotactic lung VMAT A. Abbas, T. Karan, S. Kim, D. J. Moseley, M. M. Taremi,
1 A Comprehensive Study on the Heterogeneity Dose Calculation Accuracy in IMRT using an Anthropomorphic Thorax Phantom S Davidson 1, R Popple 2, G Ibbott.
Araki F. Ikegami T. and Ishidoya T.
Alexis Dimitriadis Prof Karen Kirkby Prof Andrew Nisbet Dr Catharine Clark Dr Anthony Palmer.
K. Pulliam, MS 1,2., D Followill, PhD 2., L Court, PhD 2., L Dong, PhD 3., M Gillin, PhD 2., K Prado, PhD 3., S Kry, PhD 2 1 The University of Texas Graduate.
Introduction The Radiological Physics Center (RPC) anthropomorphic quality assurance (QA) phantom program is one tool the RPC uses to remotely audit institutions.
Part No...., Module No....Lesson No
F. Foppiano, M.G. Pia, M. Piergentili
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
P. Rodrigues, A. Trindade, L.Peralta, J. Varela GEANT4 Medical Applications at LIP GEANT4 Workshop, September – 4 October LIP – Lisbon.
NCS Report 24 VMAT QA NCS Report 24
Commissioning of a commercial treatment planning system for IMAT and Dose Painting treatment delivery. G. Pittomvils 1,,L. Paelinck 1, F. Crop 2, W. De.
The Effects of Small Field Dosimetry on the Biological Models Used In Evaluating IMRT Dose Distributions Gene Cardarelli,PhD, MPH.
Rapid Arc Treatment Verification: post evaluation on Delta-4 and proposal of a new verification protocol G. Pittomvils 1,,L. Paelinck 1, T. Boterberg 1,
MCS overview in radiation therapy
E. Mezzenga 1, E. Cagni 1, A. Botti 1, M. Orlandi 1, W.D. Renner 2, M. Iori 1 1. Medical Physics Unit, ASMN-IRCCS of Reggio Emilia, Italy 2. MathResolution.
Medical Dosimetry Presented by (Add Hospital name) in conjunction with The American Association of Medical Dosimetrists.
Treatment Plan Validation of the Xoft AXXENT™ X-ray Source
Hsiao-Ju Fu Yuk-Wah Tsang Chih-Chia Chang
Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale
Linac Commissioning Overview
OpenGate Technical Meeting May 11th 2017 Clermont-Ferrand
Template Matching Can Accurately Track Tumor Evaluation of Dose Calculation of RayStation Planning System in Heterogeneous Media Huijun Xu, Byongyong Yi,
'Monte Carlo modelling of a novel transmission detector: comparison of simulated and measured VMAT beams' Authors: D. Johnson1, S.J. Weston1, V.P. Cosgrove1,
A. Nisbet 1,2, A. Dimitriadis 1,2,3, A.L. Palmer 1,4, C.H. Clark 2,3
Multidisciplinary Engineering Senior Design Project P06441 See Through Fog Imaging Preliminary Design Review 05/19/06 Project Sponsor: Dr. Rao Team Members:
Ch 10. A System of Dosimetric Calculations
Relationship between dosimetric leaf gap and dose calculation errors for high definition multi-leaf collimators in radiotherapy  Jinkoo Kim, James S.
Mohammad Hussein Royal Surrey County Hospital, UK
GHG meeting at ESTRO36 May, 2017
Presentation transcript:

Implementing RapidArc into clinical routine: A comprehensive program from machine QA to TPS validation and patient QA Maria Sj ö lin Department of Oncology, Division of Radiophysics, Copenhagen University hospital, 2730 Herlev

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Specific tests for RA implementation I.Machine QA II.TPS QA III.Performance assessment of the RA optimization algorithm (PRO8.9) (artificial cases) IV.Patient-specific QA (clinical and artificial cases) i.Phantom QA: GafChromic EBT film/ Delta4/Seven29+Octavius/ ii.Portal dosimetry Med Phys;38(9): , 2011.

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Machine QA (I.) Test of the delivery beyond its clinically used limits (at the machine) A number of tests were developed to focus on angular accuracy Tests were performed with: GafChromic EBT film +solid water +EPSON scanner +VERISOFT FILMSCAN (v2.1) StarCheck + BeamAdjust +inclinometer (+MATLAB) Seven29 + Verisoft + inclinometer (+MATLAB)

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Examples of machine QA tests Static /Dynamic MLC Twinkle  2 deg narrow angular sectors, 1 mm wide static/dynamic central gap → minimal gantry speed  38 deg dose-less sectors → maximum gantry speed  Different MUs were delivered to check dose delivery during acceleration/deceleration for comparison: Increasing the amount of MU → the gantry will slow down!  The test can also be performed with constant gantry speed Introduced errors  The dose rays were shifted 1, 2, and 3 deg → simulate a slow response  The rays were broadening by 1,2, and 3 deg

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Reproduced correct delivery Reference image +introduced error of 3 deg clockwise delay Reference image +2 deg smoothening effect around the actual ctrlpt

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark TPS QA (II.) For RapidArc, the calculation of the dynamic source in the TPS is approximated by summing multiple static gantry, dynamic MLC fields using the AAA algorithm (Anisotropic Analytical Algorithm).

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark The accuracy of the dose calculation is challenged by:  The use of multiple small and off axis openings in relatively large collimator openings…  Leaf-pairs with nearly closed MLC tips commonly… (DLS parameter)  Highly asynchronous leaf movement… (Tongue and groove) An example of a beams-eye-view sequence for four control points, with a gantry spacing of 2 degrees.

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Static field measurements i. DLS validation ii.Small and/or off-axis effective opening in large collimator iii.Tongue and groove

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Examples of line profiles -3 mm leaf-gap  Ionization chamber measurements agree within 2 % with the calculated absolute dose in the center of field opening (4x4 cm)  Satisfactory agreement between dose-peaks under leaf tips  Using a high resolution (0.3 mm) fluence improves the results

Maria Sjölin/ RapidArc Workshop 2012, Aarhus, Denmark Thanks for listening!