XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 MOS Cooling Results and Current Response Both MOS Cameras.

Slides:



Advertisements
Similar presentations
EPIC Calibration Meeting, Mallorca Low energy calibration with the Vela SNR K. Dennerl, 2009 March 31 Low energy calibration with the Vela SNR Mallorca,
Advertisements

XMM EPIC MOS Steve Sembay Mallorca 01/04/09 A Phenomenological approach to the MOS detector response. Steve Sembay Phenomenological.
WORK PLAN FOR THE EPIC CALIBRATION TEAM (AND BEYOND)
EPIC Calibration Meeting, K. Dennerl VILSPA/Spain, 2003 Mar Temporal variation of the energy response determined with the internal calibration source.
Astrometry and EPIC flux cross-calibration in 2XMMp/2XMM Silvia Mateos R. Saxton, S. Sembay and the XMM-Newton Survey Science Centre.
AGN FEEDBACK IN TWO INTERESTING GROUPS F. Gastaldello (IASF-MI, UCI) D. Buote (UCI), P. Humphrey (UCI), W. Mathews (UCSC), F. Brighenti (U. Bologna), P.
Clusters as calibration tools S. Molendi (IASF-MI)
Comparison of XMM-Newton EPIC, Chandra ACIS-S3, ASCA SIS and GIS, and ROSAT PSPC results for G , 1ES , and MS A “Man on the street”
XMM EPIC MOS Andy Read EPIC Calibration Meeting Mallorca, Spain 01/02/05 What can we do with Pattern 31s? - Central pixel (red) has.
Chandra X-Ray Observatory CXC Paul Plucinsky EPIC Cal April ACIS Calibration: Recent Updates, Planned Updates, & Future Issues 2.E0102: Joint.
Bruno Altieri EPIC Calibration meeting, Feb. 1-3, 2005 European Space Astronomy Centre Page 1 XMM-Newton ESAC Status of XMM-Newton cross-calibration with.
EPIC flux comparison from 2XMM sources S. Mateos, R. Saxton, S. Sembay & A. Read.
XMM 1 EPIC MOS Monitoring CAL/OPS Meeting MPE
XMM EPIC MOS Steve Sembay ESAC 04/10/05 Implementation of the MOS spatial/temporal DRM: Does it work ? Zeta Puppis: Checking the line.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton EPIC The XMM-Newton cross-calibration.
XMM-Newton 1 Michael Smith, ESAC PN Monitoring Palermo, April 2007.
European Space Astronomy Centre Villafranca del Castillo, Madrid, Spain Andy Pollock XMM-Newton SOC RGS calibration Calibration of the XMM-Newton RGS Andy.
XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Saclay 24/09/03 Epoch dependant MOS QE and RMF and Cross-calibration with the PN.
Experience with combined Fe55/Vela SNR calibration - MOS Andrea Tiengo INAF, IASF-Milano In collaboration with: Mariachiara Rossetti (IASF-Milano) Martin.
Marcus Kirsch, XMM-Newton European Space Astronomy Center Page 1 EPIC ESAC Marcus G. F. Kirsch Mallorca, Summary status of the EPIC calibration.
Evaluation of the EPIC flux cross-calibration from 2XMM sources R. Saxton, S. Mateos, A. Read, S. Sembay Mateos et al., 2009, A&A, arXiv
XMM EPIC MOS Andy Read EPIC Calibration Meeting Leicester University 30/09/04 Addendum to 1ES0102 Study - Search for similar low-energy.
EPIC Calibration Meeting, MPE K. Dennerl, 2006 May 04 EPIC pn at low energies: LW mode Properties of EPIC pn at low energies: LW mode.
6 November 2007Cal-Ops Meeting Mallorca Recent Results from EPIC MOS Life Test Facility Tony Abbey and Craig Brown University of Leicester Space Research.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton MOS Timing Mode calibration Marcus.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton EPIC Calibration of the EPIC MOS Timing.
Energy Calibration using SNRs: preliminary results Andrea Tiengo INAF, IASF-Milano Mallorca, 6 November 2007 EPIC Calibration Meeting.
XMM EPIC MOS Steve Sembay MPE 04/05/06 The Effective Area of EPIC A comparison of MOS1, MOS2 and PN derived fluxes of 17 on-axis AGN.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton EPIC Results from the EPIC- Calibration-Workshop.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton XMM Newton cross calibration M.G.F.
The EPIC-pn spectral response Frank Haberl EPIC Cal meeting, Tübingen, February 2003 Status of the pn spectral response in SAS5.4.1 Latest developments.
NRCO #71: preliminary analysis of MOS data Andrea Tiengo INAF, IASF-Milano Mallorca, 8 April 2008 EPIC Calibration Meeting.
Mallorca EPIC Cal Meeting - February 2005 Darren Baskill: XMM project: MOS monitoring.
Marcus Kirsch Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton EPIC Cross calibration PKS
EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 CTI.. update for EPIC pn EPIC pn: Update on the CTI, the absolute energy scale and the energy.
MPE EPIC Cal Meeting – May 2006 Darren Baskill: XMM project: MOS gain monitoring MOS.
EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 Suppression of the low-energy noise Suppression of the low-energy noise: an improved approach.
XMM EPIC MOS Steve Sembay Mallorca 26/10/06 MOS spectral redistribution function has evolved temporarily and spatially whilst in orbit.
Chandra X-Ray Observatory CXC Paul Plucinsky February A Comparison of Spectral Fit Results of E0102 from the Chandra and XMM CCDs Paul Plucinsky.
The EPIC-PN response matrix - update Frank Haberl EPIC Cal meeting, VILSPA, March 2004 Known problems in PN6.6 / SAS6.0.
XMM EPIC MOS Andy Read EPIC CAL/OPS Meeting MPE, Germany 04-05/05/06 XRT Point Spread Function - Shifts in PSF position with energy.
Calibration Meeting Tubingen Feb 2003 MOS Cameras pre and post cooling comparison Tony Abbey, Leicester University Comparison between two closed.
XMM 1 EPIC MOS Monitoring CAL/OPS Meeting Mallorca
Chandra X-Ray Observatory CXC Paul Plucinsky EPIC Cal October Comparison of RGS Model for 1E to the HETG data 2.BI CCD CTI Correction.
Frank Haberl EPIC cal meeting, Palermo, April 2007 EPIC-pn effective area and response Effective area above 4 keV Gain/CTI correction TI mode Re-distribution.
Martin Stuhlinger X-Ray Universe 2005, Sep.26-30, 2005 European Space Astronomy Centre Page 1 XMM-Newton ESAC Status of XMM-Newton cross-calibration with.
Column rejection in the MOS  The core of soft sources is removed by the standard emchain.  Due to very large fluctuations in count rate from one column.
XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Ringberg 2-5/April/2002 Broad-Band MOS v PN Cross-Calibration Study •8 Observations.
XMM 1 EPIC MOS Monitoring. XMM-Newton 2 Martin Stuhlinger, ESAC MOS Bad Pixel Monitoring.
XMM EPIC MOS Steve Sembay Mallorca 02/02/05 Temporal and spatial variability of the MOS RMF 1.Introduction and brief history of the.
Marcus Kirsch, XMM-Newton European Space Astronomy Center Page 1 EPIC ESAC Long term EPIC in-orbit calibration plan Marcus G. F. Kirsch.
RGS/MOS-cooling Results February 03, 2003 L. Metcalfe 1 Science Operations & Data Systems Division Research & Scientific Support Department Page 1 XMM-Newton.
XMM EPIC MOS Steve Sembay Mallorca 06/11/07 Report on three NRCO calibration observations On axis observation of absorbed Seyfert.
Chandra X-Ray Observatory CXC Paul Plucinsky EPIC Cal Update on ACIS Calibration 1) Update to the ACIS contamination model 2) Temperature-dependent.
EPIC Calibration Meeting, Madrid Correcting spatial inhomogeneities of the absolute energy scale K. Dennerl, 2010 March 23 Correcting spatial inhomogeneities.
RXJ a soft X-ray excess in a low luminosity accreting pulsar La Palombara & Mereghetti astro-ph/
NASSP Masters 5003F - Computational Astronomy Lecture 19 EPIC background Event lists and selection The RGA Calibration quantities Exposure calculations.
The IACHEC Matteo Guainazzi XMM-Newton Science Operations Centre ESAC-ESA.
A new look at AGN X-ray spectra - the imprint of massive and energetic outflows Ken Pounds University of Leicester Prague August 2006.
Chandra X-Ray Observatory CXC Paul Plucinsky IACHEC March Thermal SNR Working Group: Summary Report Paul Plucinsky on behalf of the IACHEC Thermal.
C. Y. Hui & W. Becker X-Ray Studies of the Central Compact Objects in Puppis-A & RX J Max Planck Institute for Extraterrestrial Physics, Giessenbachstrasse.
Marcus Kirsch, XMM-Newton European Space Astronomy Center Page 1 EPIC ESAC Status of the EPIC calibration Presentation for XMM Newton UG 18 May 2006 Marcus.
Experience with combined Fe55 / Tycho SNR calibration
The unusual X-ray spectrum of MCG
Multi-wavelength behaviour of III Zw 2
EPIC Calibration & Operations Meeting
Energy Response of the X-ray Imaging Spectrometer (XIS) on Suzaku
XMM-NEWTON EPIC CONTAMINATION MONITORING.
Update of EPIC pn long-term monitoring
Long-term trends in EPIC pn energy response
Presentation transcript:

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 MOS Cooling Results and Current Response Both MOS Cameras cooled to -120C from -100C Rev

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Parallel CTI at Mn and Al for Central CCDs, MOS1 and MOS2 Plots by Bruno Altieri

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Parallel CTI at Al, MOS1Serial CTI at Al, MOS1 Plots by Bruno Altieri

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Parallel CTI at Al, MOS2Serial CTI at Al, MOS2 Plots by Bruno Altieri

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Measured Resolution at Al, MOS1, CCD1

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Measured Resolution at Al, MOS2, CCD1

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Measured Resolution at Mn-Kalpha, MOS1/2, CCD1

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Isolated Neutron Star RX J used for contamination monitoring No evidence for contamination … similar for MOS2

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Vela pre- and post-cooling. No evidence for contamination

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Gain stability test on SNR 1E Ne IX 247: (0.002) 375: (0.002) 552: (0.003) Mg XI 247: (0.002) 375: (0.002) 552: (0.004)

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Low Energy Epoch Dependent Response Variation

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Fit to MOS1/MOS2 spectra of Mrk 421 (Rev 165), using SAS 5.3

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Same as previous, but with data and model extrapolated down to 100 eV

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Fit to MOS1/MOS2 spectra of PKS (Rev 450) using SAS 5.3

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Same as previous, but with data and model extrapolated down to 100 eV Low energy excess indicating response change ?

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Contribution to flux at 150 eV versus incident energy for each spectrum

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Significant change in spectrum between Rev 78 and 533/534

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Significant change in spectrum between Rev 78 and 533/534

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Comparison of Current Response fitted to Rev 78 data and Test Response Fitted to Rev 533 and Rev 534 N H = 1.39(0.5)E20 cm -2 BB = 82.4(1.5) eV

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Current and Test Redistribution at incident E = 250, 300, 350 and 400 eV

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Comparison of spectra extracted from Radii, 0-45, 5-45 and arcsec

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 How does Test Response work with other sources ? Not very well ! Two observations from Rev 450 and 552 Power-Law fit with N H = 1.36E20 cm -2

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Comparison with Current Response

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Quality of fit across main edges from 1 to 3 keV 5-10% Residual at 2.5 keV

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 N132D … Evidence for flux/response change below 500 eV Plot by Marcus Kirsch

XMM EPIC MOS Steve Sembay EPIC Calibration Meeting Tuebingen 03/02/03-06/02/03 Epoch/Flux dependency exhibited by 3C 273