Nuclear Cardiology Guidelines Benchmark to compare local practice against
UK Guidelines British Nuclear Cardiology Society British Nuclear medicine Society British Cardiac Society http://www.bnms.org.uk Draft - Rev 3 2002 21 pages
USA (1) - SNM Society of Nuclear Medicine Version 2.0 1999 http://www.snm.org 14 pages
USA (2) - ASNC American Society of Nuclear Cardiology Two part guidelines (1998-2000) http://www.asnc.org 84 close-typed pages
Techniques Covered BNCS SNM ASNC MPI -Planar MPI - SPECT MPI - Gated MUGA First Pass PET = Covered in detail = Mentioned
MPI – SPECT: Aspects Covered BNCS SNM ASNC Clinical Indications Stress Techniques Radiopharmaceuticals Equipment QA Acquisition Processing and Display Artefacts and Pitfalls Report Content/Style
Doses (MBq) BNCS SNM ASNC Thallium 80 < 150 110 Tc-99m (1-day) 250+750 <1400 300+950 (2-day) 2 * 400 2 * 1100 (max) 2 * 800 Boost for obese pt? FDA max
Stress Methods Methods Exercise (treadmill or bike) Pharmacological Adenosine Dipyridamole Dobutamine Contraindications Stop conditions
Acquisition LEHR parallel hole collimator Circular or non-circular orbit 20% symmetric energy window 64 x 64 or 128 x 128 (optional) Zoom can be used Take care heart always in view (BNCS) Pixel size around 6mm (BNCS) 6.4±0.2 mm (ASNC) cf Picker 4.9mm (64*64, 1.8mag) imaging time 20-30 min Continuous or S&S with 32/64 stops
Processing: stages and options
Motion Correction All guidelines state that patient motion should be assessed SNM gives acceptable limits: ±1 pixel in axial direction in a 64x64 study Repeat acquisition or optionally use software methods for minor motion
Filtering - ASNC Not prescriptive Discusses filter types “Conventional” Butterworth/Hamming/Hanning “Restorative” / “Contrast Enhancement” Wiener/Metz For gated, recommend “0.55 or 0.45 Nyquist frequency cutoff” for Butterworth Iterative reconstruction an option if available
Filtering - BNCS “Filtered back projection using Butterworth and Hanning filters is the most common method of reconstruction. Cut-off frequencies of 0.5 and 0.75 respectively are normally chosen, and these should be the same for each patient and should not be altered to compensate for low-count images in order to maintain consistency of appearance. Iterative reconstruction is preferred if attenuation correction has been performed and it can also be used without attenuation correction.”
Attenuation Correction (Transmission sources) Attenuation correction methods “in rapid development”..“difficult to provide ‘cookbook’ guidelines” (ASNC) “The effectiveness of these techniques in routine clinical practice is currently uncertain. They should currently be used only in experienced centres and preferably as part of a formal evaluation of their value” (BNCS)”
Reformatting & Display SA / HLA / VLA slices Stress and rest with frames aligned or interactive manipulation Report from computer display (not hardcopy) A “continuous” colour scale (BNCS) NO guidance on combining adjacent slices
Normalisation of frames Each set of tomograms should be displayed using a window maximum that corresponds to the maximum within the myocardium in each set. Displays that set the maximum of the window to the maximum of each tomogram and those that use the same maximum for stress and rest images should not be used. (BNCS) “Preferred” ..”frame normalization” ..normalized to the brightest pixel in each frame. That method provides optimal image quality of each slice. Drawback..gradations in activity between slices in a series may be lost. (ASNC)
Quantification ASNC guidelines detail semi-quantitative segmental “scoring” systems Full quantification (QGS etc) Mentioned briefly in ASNC and BNCS Always as a supplement to visual analysis May have benefits for inexperienced observers and for serial comparison of studies
Report Contents Patient Details Indications Stress agent / dose / protocol / response /ECG changes Study Quality Findings Ventricular dilatation, Lung Uptake Perfusion defects (severity/extent/“reversibility”) Conclusion Prognosis / recommendations
Sources of error Submaximal stress Inadequate dose for BMI Acquisition Poor positioning/orbit, Wrong energy window Patient motion, Attenuation artefacts Reconstruction and processing Inappropriate filter / reconstruction technique Wrong alignment Image display Inappropriate colour scale or windowing Comparison of inappropriate frames/slices
Conclusion Lots of useful information Not a complete “cookbook” - still lots of scope for differences that may affect final result, especially in processing and display Local audit may have value