Basics of Magnetic Resonance Imaging

Slides:



Advertisements
Similar presentations
Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Advertisements

Fund BioImag : Echo formation and spatial encoding 1.What makes the magnetic resonance signal spatially dependent ? 2.How is the position of.
Ari Borthakur, PhD Associate Director, Center for Magnetic Resonance & Optical Imaging Department of Radiology Perelman school of Medicine, University.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
BE 581 Lecture 3- Intro to MRI.
PHYSICS OF MAGNETIC RESONANCE
Nuclei With Spin Align in Magnetic Fields HoHo anti-parallel parallel Alignment Energy  E = h  H o Efficiency factor- nucleus ConstantsStrength of magnet.
MR TRACKING METHODS Dr. Dan Gamliel, Dept. of Medical Physics,
Magnetic Resonance Imaging
Magnetic Resonance Imaging
Principles of MRI: Image Formation
Basic Principles MRI related to Neuroimaging Xiaoping Hu Department of Biomedical Engineering Emory University/Georgia Tech
Magnetic Resonance Imaging
NMR Spectroscopy Part I. Origin of NMR. Nuclei in Magnetic Field Nucleus rotate about an axis -- spin Nucleus bears a charge, its spin gives rise to a.
Lecture 2 1 H Nuclear Magnetic Resonance. Gas Chromatograph of Molecular Hydrogen at –100 °C Thermoconductivity Detector 12.
MRI. Magnetic Resonance 1.Principle first observed in Used for spectroscopy and imaging 3.Imaging techniques are a form of tomography, where slices.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
Relaxation Exponential time constants T1 T2 T2*
FMRI: Biological Basis and Experiment Design Lecture 5: non-BOLD MRI Equilibrium and excitation Relaxation rates Image contrast –TE –TR.
Imagen de resonancia magnética Magnetic resonance imaging, G.A. WRIGHT IEEE SIGNAL PROCESSING MAGAZINE pp:56-66.
Nuclear Magnetic Resonance Spectrometry Chap 19
Nuclear Magnetic Resonance Spectrometry Chap 19. Absorption in CW Experiments Energy of precessing particle E = -μ z B o = -μ B o cos θ When an RF photon.
Nuclear Magnetic Resonance (NMR) Magnetic Resonance Imaging (MRI)
Spectroscopy 3: Magnetic Resonance CHAPTER 15. Pulse Techniques in NMR The “new technique” Rather than search for and detect each individual resonance,
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Resonance condition. Pulse A coil of wire placed around the X axis will provide a magnetic field along the X axis when a direct current is passed through.
G Practical MRI 1 Magnetization, Relaxation, Bloch Equation, Basic Radiofrequency (RF) Pulse Shapes.
Radiofrequency Pulse Shapes and Functions
Physics of Magnetic Resonance Chapter 12
Principles of MRI Physics and Engineering
Medical Physics Physics 421 Course Description: Medical Physics is a course with two main parts:  Physics of the body  Physics of Diagnostic and Therapeutic.
Principles of Magnetic Resonance
ELEG 479 Lecture #12 Magnetic Resonance (MR) Imaging
MRI Physics: Just the Basics
Determination of Spin-Lattice Relaxation Time using 13C NMR
Prof. Dr. Moustafa. M. Mohamed Dr. Eng. Safa Ahmed El-Askary
BE 581 Intro to MRI.
CT “Computer tomography”. Contrast mechanisms in X-ray imaging: X-ray absorption X-ray absorption mechanisms: 1. Photoelectric effect 2. Compton scatter.
Magnetic Material Engineering. Chapter 6: Applications in Medical and Biology Magnetic Material Engineering.
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Chapter 4 Mark D. Herbst, M.D., Ph.D.. Magnetization MRI depends on magnetization of tissues Temporary magnetization occurs when the hydrogen nuclei are.
Chapter 6 Lecture Spin Echo Imaging Methods Mark D. Herbst, MD, PhD Two Main Types of MR methods –Spin Echo –uses RF pulse to get an echo –Gradient Echo.
Basic Concept of MRI Chun Yuan. Magnetic Moment Magnetic dipole and magnetic moment Nuclei with an odd number of protons or neutrons have a net magnetic.
Basic of Magnetic Resonance Imaging Seong-Gi Kim Paul C. Lauterbur Chair in Imaging Research Professor of Radiology, Neurobiology and Bioengineering University.
fMRI Methods Lecture2 – MRI Physics
NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY Basics of …….. NMR phenomenonNMR phenomenon Chemical shiftChemical shift Spin-spin splittingSpin-spin splitting.
Variable Rate Selective Excitation Radio Frequency Pulse in Magnetic Resonance Imaging Stephen Stoyan.
Nuclear Magnetic Resonance I Magnetization properties Generation and detection of signals.
Magnetic Resonance Imaging – Basic Principles –
MRI. Vector Review x y z Vector Review (2) The Dot Product The Cross Product (a scalar) (a vector) (a scalar)
BMI2 SS08 – Class 6 “functional MRI” Slide 1 Biomedical Imaging 2 Class 6 – Magnetic Resonance Imaging (MRI) Functional MRI (fMRI): Magnetic Resonance.
BME1450 Intro to MRI February 2002 The Basics The Details – Physics The Details – Imaging.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
V.G.Wimalasena Principal School of Radiography
MR Image Formation FMRI Graduate Course (NBIO 381, PSY 362)
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
RT 4912 Review (C) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Magnetic Resonance Imaging Glenn Pierce, King’s College London, Department of Physics Introduction Edward Purcell and Felix Bloch were both awarded the.
Spinning Nucleus Produces Magnetic Moment
Magnetic Resonance Learning Objectives
Charged particle. Moving charge = current Associated magnetic field - B.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Lecture 1: Magnetic Resonance
Physics of Magnetic Resonance Imaging
Medical Physics Physics 421 Course Description:
Basic MRI I Chapter 3 Notes.
FT NMR WORKSHOP/===/ S.A.I.F./===/ NEHU/==/ Shillong
How MRI Works By Wesley Eastridge, adapted from and with illustrations from The Basics of MRI by Joseph P. Hornak, Ph.D.
Introduction to MR Robert Frost
Introduction to MR Robert Frost
Presentation transcript:

Basics of Magnetic Resonance Imaging

Angular Momentum Orbital Angular Momentum Principles of Medical Imaging – Shung, Smith and Tsui

Angular Momentum Spin Angular Momentum Spin is an intrinsic property of the nucleons (protons and neutrons) in a nucleus HOWEVER – The name doesn’t mean that spin results from the nucleons rotating about an axis!!! http://svs.gsfc.nasa.gov/vis/a000000/a001300/a001319/

Spin Angular Momentum Spin is quantized – it can only take certain values Here I is the total spin quantum number of the nucleus. The proton has I = ½. Lz is the angular momentum due to that spin.

Spin Angular Momentum To get the total spin of a nucleus we add up (separately) the spins of the protons and neutrons 15N has spin ½ . We do pairwise addition: 7 protons 8 neutrons Only nuclei with an odd number of protons or neutrons will be visible to MRI Note: 14N has spin 1.

Alignment of Spins in a Magnetic Field, B0 The spin angular momentum yields a magnetic moment Principles of Medical Imaging – Shung, Smith and Tsui

Energy Levels of Spins and B0 Principles of Medical Imaging – Shung, Smith and Tsui

Energy Levels and Spin E1 = B0 E2 = -B0 E = E1 - E2 = 2 B0 = (h/2) B0 for spin-1/2 particles B0 is the main magnetic field

Energy Level Population and Field Strength 2,000,000 1,000,002 999,999 999,995 1,000,005 999,987 1,000,014 B0 = 0 n = 0 B0 = 0.5T n = 3 B0 = 1.5T n = 10 B0 = 4.0T n = 27 Spins are distributed according to the Boltzmann distribution

Larmor Frequency Principles of Medical Imaging – Shung, Smith and Tsui

Excitation Energy and Frames of Reference B0 B1 Beff B0 = main magnetic field B1 = applied field (pulse) Beff = vector sum B0+B1 z z   y y lab frame rotating frame x x

Net Magnetization, M, and the Rotating Frame While B1  0, M precesses around B1 z z M M y y B1 x x B0  0 B1 = 0 B0  0 B1  0

Net Magnetization, M, and the Rotating Frame We turn B1 “on” by a applying radiofrequency (RF) to the sample at the Larmor frequency. This is a resonant absorption of energy. If we leave B1 on just long enough for M to rotate into the x - y plane, then we have applied a “90 pulse”. In this case, Nupper = Nlower. If we leave B1 on just long enough for M to rotate along the -z axis, then we have applied a “180 pulse” (inversion). In this case, Nupper = Nlower + M.

Free Induction Decay What is the effect of applying a 90  pulse? /2 RF time

Free Induction Decay The effect of a 90 pulse is to rotate M into the x - y (transverse) plane. If we place a detection coil (a loop of wire) perpendicular to the transverse place, we will detect an induced current in the loop as M precesses by (in the lab frame). Minitial y x z Mfinal B1 Principles of Medical Imaging – Shung, Smith and Tsui

Signal Processing of Free Induction Decay Principles of Medical Imaging – Shung, Smith and Tsui We can characterize the signal by its: Amplitude Phase Frequency Fourier Transform

Signal Processing of Free Induction Decay We see that after a 90 pulse, we get a cosinusoidal signal. To quantitatively describe the signal we calculate its Fourier transform. (think: Larmor frequency) Principles of Medical Imaging – Shung, Smith and Tsui

Fourier Transform of Time Domain Data http://www.med.harvard.edu/JPNM/physics/didactics/improc/intro/fourier3.html

Will it be a useful image? Image Contrast (I) We can detect the signal from water molecules in the body. Can we make an image? Will it be a useful image?

Relaxation Processes Fortunately for us, the signal we get from water molecules in the body depends on their local environment. Spins can interact by exchanging or losing energy (or both). As in all spectroscopy methods, we put energy into the system and we then detect the emitted energy to learn about the composition of the sample. We then use some variables to characterize the emission of energy which (indirectly) tell us about the environment of the spins. Image Contrast!

Relaxation Processes (T2) 1. Spin-spin relaxation time (T2): when spins interact With each others magnetic field, they can exchange energy (perform a spin flip). They can lose phase coherence, however. Only affects Mxy. Signal without T2 interaction between spins T2* Signal including T2 interactions between spins T2*??? http://irm-francophone.com/htm/signal.htm

T2* and T2 T2 is an intrinsic property of the sample. This is what we are interested in to use for contrast generation. T2* is the time constant of the decay of the free induction decay. It is related to the intrinsic T2 in the following way: depends upon the local magnetic field sample only

T2* and T2 Not a random random process process Inhomogeneity term - dephasing due to magnet (B0) imperfections depends upon position Susceptibility term - dephasing due to the interaction of different sample regions with B0 (depends upon position)

Relaxation Processes (T2) /2 pulse  (delay)

Effect of Spin Coherence on Signal x y z x y z T2* http://irm-francophone.com/htm/signal.htm

Irreversible versus Reversible http://www.cchem.berkeley.edu/demolab/images/HahnEchoSpinRes.htm Start Reverse

Hahn Spin Echo Pulse Sequence http://www.esr.ethz.ch/intro/spinecho.html http://www.chem.queensu.ca/FACILITIES/NMR/nmr/webcourse/t2.htm

Hahn Spin Echo and T2 We can calculate T2 by changing the echo spacing, , and recording the signal at 2. http://spiff.rit.edu/classes/phys273/exponential/exponential.html Signal Echo spacing , 

Spin-Lattice Relaxation, T1 To look at the behavior of the longitudinal component of M (Mz), we start by putting M along the -z axis and then read it out with a 90 pulse.

Spin-Lattice Relaxation, T1 Energy levels and Inversion Equilibrium After Inversion Net Magnetization: = =

Relaxation Processes (T1)  pulse long TI short TI /2 pulse /2 pulse

Image Contrast and T1 In (a) the TI is chosen to null the signal from curve [ii], while the TI in (b) nulls out [i] http://www.fmrib.ox.ac.uk/~stuart/thesis/chapter_2/section2_4.html

Now Can We Make A Useful Image?

Magnetic Resonance Imaging

Magnetic Resonance Image Formation What do we need? Ability to image thin slices 2. No projections - image slices with arbitrary orientation 3. Way to control the spatial resolution 4. Way to carry over the spectroscopic contrast mechanisms to imaging.

Magnetic Resonance Image Formation How can we spatially encode this signal? Principles of Medical Imaging – Shung, Smith and Tsui

Magnetic Resonance Image Formation - Slice Selection Magnetic field gradients: B = B(position) precessional frequency is now a function of position

Magnetic Resonance Image Formation - Slice Selection  = frequency bandwidth z = slice thickness MRI: Basic Principles and Applications - Brown, Semelka

Magnetic Resonance Image Formation - Slice Selection How do we excite only a slice of spins? Fourier Pairs! sinc rect FT time frequency

Magnetic Resonance Image Formation - Phase Encoding So far we have a slab of tissue whose spins are excited. The next step is to place a grid over the slab and define pixels.

Magnetic Resonance Image Formation - Phase Encoding eff < 0 Gy eff > 0 Center of magnet

Magnetic Resonance Image Formation - Phase Encoding Apply gradient for a finite duration   = (y) (the phase of M over each region of the sample depends upon it’s position) This is because the gradient makes each spin precess with an angular frequency that depends on it’s position. For the duration of the gradient, t, spins move faster or slower than 0 depending upon where they are. After the gradient is turned off, all spins again precess at 0. The phase accumulated during time, t, is:

Magnetic Resonance Image Formation - Phase Encoding x’ y’ z’ M after a 90 pulse Turn on the phase encoding gradient x’ y’ z’ x’ y’ z’ x’ y’ z’ Gy = 0 Gy < 0 Gy > 0

Magnetic Resonance Image Formation Frequency Encoding (Readout) Now we need to encode the x-direction... How do we spatially encode the frequency of the signal? Can we turn on another gradient? When? And for how long?

Magnetic Resonance Image Formation Frequency Encoding (Readout) We apply a gradient while the signal is being acquired as the spin-echo is being formed.  /2 RF Signal Gx time Therefore, the precessional frequency is a function of position

Magnetic Resonance Image Formation Spin-echo Pulse Sequence ?(3) ? (1) ?(2) http://spl.harvard.edu:8000/pages/papers/zientara/fast/fastimaging.html

(1) Phase Encoding: Each acquisition is separately encoded with a different phase. The sum total of the N acquisitions is called the ‘k-space’ data.

The Effect of Frequency Encoding on the Signal (dephasing gradient) (2) MRI: Basic Principles and Applications - Brown, Semelka

Slice-select refocusing gradient (3) z y x The effect of having the gradient on during the time that the magnetization is moving from the z-axis to the y-axis is to curve the path. In general, one needs to apply a slice refocussing gradient of opposite magnitude after the RF pulse so that the spins are in phase at the end of the pulse. The area of the negative gradient must be one half the area of the slice selection gradient pulse.

Structure of MRI Data: k-space http://www.indyrad.iupui.edu/public/lectures/mri/iu_lectures/mri_homepage.htm

Structure of MRI Data: k-space What is k-space? The time-domain signal that we collect from each spatially encoded spin echo gets put in a matrix. This data is called k-space data and is a space of spatial frequencies in an image. To get from spatial frequencies to image space we perform a 2-D Fourier Transform of the k-space data. General intensity level is represented by low spatial frequencies; detail is represented by high spatial freq’s.

low spatial frequencies high spatial frequencies all frequencies http://www.indyrad.iupui.edu/public/lectures/mri/iu_lectures/mri_homepage.htm

Structure of MRI Data: k-space FT k-space data image data http://www.jsdi.or.jp/~fumipon/mri/K-space.htm