Studies of optimization methods for dose delivery with a beam scanning system Alexei Trofimov, Thomas Bortfeld Northeast Proton Therapy Center MGH, Boston.

Slides:



Advertisements
Similar presentations
Dept of Radiation Oncology MGH
Advertisements

Optimisation in proton scanning beams
Harvard Medical School Massachusetts General Hospital.
HST.187: Physics of Radiation Oncology #9. Radiation therapy: optimization in the presence of uncertainty Alexei Trofimov, PhD Jan.
Ionization Chamber Array for External Beam Radiotherapy
Skeletons and gamma ray radiosurgery The Mathematics of Shapes.
Energy deposition and neutron background studies for a low energy proton therapy facility Roxana Rata*, Roger Barlow* * International Institute for Accelerator.
FREDONE, A PROJECT ON DOSIMETRY OF HIGH ENERGY NEUTRONS Valérie De Smet (1), Isabelle Gerardy (2), Fréderic Stichelbaut (3), (1) IRISIB, Département nucléaire,
Experience, Expertise and a Commitment to Excellence™
PAUL SCHERRER INSTITUT Future directions and current challenges of proton therapy Tony Lomax, Oxford, 2008 Future directions and current challenges of.
Pascal Storchi Daniel den Hoed Cancer Center
D.G.Lewis Department of Medical Physics Velindre Cancer Centre Whitchurch, Cardiff RT-GRID: Grid Computing for Radiotherapy.
Pair Spectrometer Design Optimization Pair Spectrometer Design Optimization A. Somov, Jefferson Lab GlueX Collaboration Meeting September
TREATMENT PLANNING PHOTONS & ELECTRONS Karen P. Doppke 3/20/2007.
Targeting Issues For Proton Treatments Of The Prostate SJ Rosenthal Ph.D., JA Wolfgang Ph.D., Sashi Kollipara Department of Radiation Oncology Massachusetts.
Beam profile vs time Analyzed Vx vs Vy distributions vs time for –Run 72 (13:26:02 – 28:25) –Run 73 (13:36:10 – 13:37:30) –Run 74 (43:12 – 44:21) Binned.
Interactions of charged particles with the patient I.The depth-dose distribution - How the Bragg Peak comes about - (Thomas Bortfeld) II.The lateral dose.
tomos = slice, graphein = to write
Introduction to Longitudinal Phase Space Tomography Duncan Scott.
Design and Scheduling of Proton Therapy Treatment Centers Stuart Price, University of Maryland Bruce Golden, University of Maryland Edward Wasil, American.
Optimization of Gamma Knife Radiosurgery Michael Ferris University of Wisconsin, Computer Sciences David Shepard University of Maryland School of Medicine.
Data Mining to Aid Beam Angle Selection for IMRT Stuart Price-University of Maryland Bruce Golden- University of Maryland Edward Wasil- American University.
MRI Guided Radiation Therapy: Brachytherapy
Patient Plan Results: Table 3 shows the ratio of the Pinnacle TPS calculation to the DPM recalculation for the mean dose from selected regions of interest.
Kelly Younge, Ph.DKelly Younge, Ph.D Don Roberts, Benedick Fraass, Daniel McShan, and Martha Matuszak University of Michigan, Department of Radiation Oncology,University.
 1  Outline  stages and topics in simulation  generation of random variates.
Applications of Geant4 in Proton Radiotherapy at the University of Texas M.D. Anderson Cancer Center Jerimy C. Polf Assistant Professor Department of Radiation.
112/03/2004 Vision-Realistic Rendering: Simulation of the Scanned Foveal Image from Wavefront Data of Human Subjects, Brian A. Barsky, 2004 Gazihan Alankus.
Surface dose prediction and verification for IMRT plans using line dose profiles † Ronald E. Berg, † Michael S. Gossman and ‡ Stephen J. Klash † Erlanger.
May 31, th PTCOG in Catania, Italy1 Treatment Planning for Broad-Beam 3D Irradiation Heavy-Ion Radiotherapy N. Kanematsu, M. Endo, and T. Kanai,
1 Dosimetric Optimization of Interstitial Gene Therapy Delivery Systems The Gynecologic Disease Program 1 Georgetown University Med. Ctr. 2 Walter Reed.
Medical considerations to guide mixed integer formulations of IMRT planning problems from MD to NP to CR Mark Langer, MD Radiation Oncology Indiana University.
1 Complex Images k’k’ k”k” k0k0 -k0-k0 branch cut   k 0 pole C1C1 C0C0 from the Sommerfeld identity, the complex exponentials must be a function.
1 Radiotherapy, hadrontherapy and treatment planning systems. Faiza Bourhaleb INFN-Torino University Med 1er-Morocco  Radiotherapy  Optimization techniques.
PrimEx collaboration meeting Energy calibration of the Hall B bremsstrahlung tagging system using magnetic pair spectrometer S. Stepanyan (JLAB)
Physics of carbon ions and principles of beam scanning G. Kraft Biophysik, GSI, Darmstadt, Germany PTCOG43 Educational Satellite Meeting: Principles of.
Quality of model and Error Analysis in Variational Data Assimilation François-Xavier LE DIMET Victor SHUTYAEV Université Joseph Fourier+INRIA Projet IDOPT,
Improvement of the Monte Carlo Simulation Efficiency of a Proton Therapy Treatment Head Based on Proton Tracking Analysis and Geometry Simplifications.
Somvilai Mayurasakorn, MD. Division of Therapeutic Radiology and Oncology, Faculty of Medicine, Chiang Mai University Somvilai Mayurasakorn, MD. Division.
Cherrill Spencer, SLAC. MDI Workshop Jan '05 1 Impact of Crossing Angle Value on Magnets near the IP Overview of several unusual quadrupole designs that.
X = Y. Direct variation X 1 X = Y 1 Y 2.
FrontierScience G.P.1 MATRIX an innovative Pixel Ionization Chamber for Online Monitoring of Hadrontherapy Treatments Giuseppe Pittà.
Flair development for the MC TPS Wioletta Kozłowska CERN / Medical University of Vienna.
Development of a superconducting rotating-gantry for carbon therapy
 Multidisciplinary Effort › Surgery › Radiation › Systemic Rx (chemo, “drugs”)
Li-Lens length options for pion yield (OptiM tracking) V.Nagaslaev.
Development of elements of 3D planning program for radiotherapy Graphical editor options  automated enclose of contour  correction of intersections 
E. Mezzenga 1, E. Cagni 1, A. Botti 1, M. Orlandi 1, W.D. Renner 2, M. Iori 1 1. Medical Physics Unit, ASMN-IRCCS of Reggio Emilia, Italy 2. MathResolution.
April 17, Dejan TrbojevicFFAG07 -Non-Scaling FFAG gantries1 Non-Scaling FFAG Gantries Introduction: Motives: The most challenging problem in the carbon/proton.
beam delivery with SC magnets
Medical Dosimetry Presented by (Add Hospital name) in conjunction with The American Association of Medical Dosimetrists.
INTERCOMPARISON P3. Dose distribution of a proton beam
Press release Release of ISOgray Proton Treatment Planning System
A treatment planning code for hadrontherapy,ANCOD++.
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
M. D. Anderson Cancer Center Houston, TX
Karla Leach, BS, CMD Texas Center for Proton Therapy-Irving, TX
Implementation of Object Spot Avoidance in Proton Pencil Beam Treatment on Whole Breast with Implant Metal Injector Peng Wang, PhD, DABR, Karla Leach,
PET-utilized proton range verification using 13N signals
L. A. den Otter. , R. M. Anakotta. , M. Dieters. , C. T. Muijs. , S
Image filtering Images by Pawan Sinha.
Spatial interpolation
Potential clinical impact of laser-accelerated beams in cancer therapy
MCMC Inference over Latent Diffeomorphisms
Simulation of Emittance
Figure 3 Craniospinal irradiation proton therapy for medulloblastoma
Computed Tomography.
Emittance Studies at Extraction of the PS Booster: Emittance Calculation S. Albright, F. Antoniou, F. Asvesta, H. Bartosik, G.P. Di Giovanni, M. Fraser,
Variations table problems
Presentation transcript:

Studies of optimization methods for dose delivery with a beam scanning system Alexei Trofimov, Thomas Bortfeld Northeast Proton Therapy Center MGH, Boston

Alexei Trofimov XXXVI PTCOG Beam scanning at the NPTC First tests have been conducted in collaboration with IBA last week Use the IBA scanning system for delivery Inverse treatment planning with KonRad (DKFZ)

Alexei Trofimov XXXVI PTCOG Treatment planning and delivery For each layer w/in target, treatment planning system generates a discrete beam weight map for regularly spaced pencil beam spots Scanning within a layer is continuous Fluence variation along the path is achieved by simultaneously varying the beam current and scanning speed

Alexei Trofimov XXXVI PTCOG Example: plan for a NPTC patient (medulloblastoma, 3D plan with 2cm-FWHM beam) boost sp.cord target hypoth. cochlea

Alexei Trofimov XXXVI PTCOG Example: plan for a NPTC patient (RPO field, spot spacing  =  = 8.5 mm) beam weight map dose distribution at B.p. range

Alexei Trofimov XXXVI PTCOG Converting a discrete spectrum into a continuous one  vector approximation Triangular approximation

Alexei Trofimov XXXVI PTCOG Difference between planned and delivered doses Along a scanning path element, delivered dose has pseudo-gaussian profile, different from the planned gaussian spot

Alexei Trofimov XXXVI PTCOG Calculated dose difference (  vector approximation)

Alexei Trofimov XXXVI PTCOG Difference between the planned and delivered doses The discrepancy is maximal in the regions of sharp dose gradient (rim of the target, boost) Size of the discrepancy depends on TPS spot spacing ( , range of variation in the weight map, scanning path. Generally, smaller for finer  /  values

Alexei Trofimov XXXVI PTCOG Spot weight optimization Planned dose (conv. of TPS weight map with a gaussian) D TPS = W TPS  g(  ) I teration #i: Delivered dose (convolution with a pseudo-gaussian) D i = W i  [ g(  )  f(  ) ]; f= or Optimized beam weight map for spots at (x,y) : W i+1 (x,y) = W i (x,y) * [ D TPS (x,y) / D i (x,y) ]

Alexei Trofimov XXXVI PTCOG Results of the optimization start1 iteration 10 iterations 100 iterations

Alexei Trofimov XXXVI PTCOG Results of the optimization f(i)   =  (x,y) [ ( D i - D TPS  ) 2 / D TPS ]

Alexei Trofimov XXXVI PTCOG Optimization for a quasi-continuous path: W 0 = W TPS  f(  ); f = or Iteration # i: Delivered dose: D i = W i  g(  ) Optimized beam weight for a quasi-continuous set of points (x,y) along the scanning path : W i+1 (x,y) = W i (x,y) * [ D TPS (x,y) / D i (x,y) ]

Alexei Trofimov XXXVI PTCOG Results of the optimization For a quasi-continuous weight variation along the path 20 iterations start

Alexei Trofimov XXXVI PTCOG Optimization results for one scanning line

Alexei Trofimov XXXVI PTCOG Another example: PA field  -vector approximation optimized (100 iterations)

Alexei Trofimov XXXVI PTCOG Another example: LPO field  -vector approximation optimized (100 iterations)

Alexei Trofimov XXXVI PTCOG Another example: spacing  1.5 *   -vector approximation optimized (100 iterations)

Alexei Trofimov XXXVI PTCOG Results of the optimization Spot spacing (  ) Maximal dose difference (%) Before optimization Optimized on targetpenumbra 0.5 *  < *  *  *  *  *  1048

Alexei Trofimov XXXVI PTCOG Summary Simulation shows that a good dose conformity can be achieved by optimizing the TPS beam weight maps discrepancy reduced 3-fold on the target, 2-fold in the penumbra (from 1-6%) no need to use finer grid Plan to verify the results with the beam