ATOMIC ABSORPTION AND ATOMIC FLUORESCENCE SPECTROMETRY Chap 9 Source Modulation Interferences in Atomic Absorption Interferences in Atomic Absorption Spectral.

Slides:



Advertisements
Similar presentations
Atomic Absorption.
Advertisements

Furnace aa. Today is Guy Fawkes Day L’vov platform furnace Sample is placed on platform Temp of platform rises more slowly than that of walls.
Atomic Absorption Spectroscopy (AAS)I
Atomic Absorption & Emission Spectroscopy
1.1 Atomic Absorption Spectrometry (AAS) determination of elements not compounds needs radiation source high temperature for atomization Atomization a.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
ATOMIC ABSORPTION SPECTROMETER
1 Atomic Absorption Spectroscopy Atomic Emission Spectroscopy Lecture 18.
Atomic Absorption and Atomic Fluorescence Spectrometry Wang-yingte Department of Chemistry
1 A TOMIC SPECTROMETRIC METHODS PART 9. 2 Interferences : Four classes of interferences: 1-Spectral interferences 2-Chemical interferences 3- Refractory.
Metal Analysis by Flame and Plasma Atomic Spectroscopy Flame A. Atomization 1. Types of Atomization Processes a.) Nebulizers b. Electrothermal atomization.
AAS and FES (Ch 9, 7th e, WMDS)
AA and Atomic Fluorescence Spectroscopy Chapter 9
Atomic Spectroscopy Atomic Spectroscopic Methods Covered in Ch 313: Optical Atomic Spectrometry (Ch 8-10) Atomic X-ray Spectrometry (Ch 12) Atomic Mass.
INTRODUCTION TO OPTICAL METHODS
Atomic Absorption Spectroscopy AAS Comparatively easy to use Low maintenance Low consumables Good for measuring one element at a time. Comparatively easy.
Analytical Chemistry –Atomic absorption Spectroscopy
1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission.
Molecular Fluorescence Spectroscopy
Atomic Emission Spectroscopy
ATOMIC SPECTROMETRY 1. Flames 2. Electrothermal Atomizers 3. Plasmas.
ContrAA ® - A Revolution in Spectroscopy Basics and Applications of High-Resolution Continuum Source AAS.
announcements and reminders
Lecture 2 M.Sc.. AA Spectrometer Components Lamp and FlameDetector Fuel Oxidant Nebulizer Double-Click picture for VIDEO.
Atomic Absorption Spectroscopy (AAS)
Fluorescence. Topics Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-Induced Fluorescence.
Atomic Absorption Spectroscopy (AAS)
Double-Beam AAS Single-Beam I0 Double-Beam Isample
Monroe L. Weber-Shirk S chool of Civil and Environmental Engineering Atomic-Absorption (AA) Spectroscopy 
Atomic Absorption Spectroscopy Elemental analysis Dissolved metals Source –Hallow Cathode Lamp –Cathode of specific element Sample –Acetylene-air flame,
Atomic Absorption Spectroscopy
ATOMIC ABSORPTION SPECTROSCOPY
BC ILN Atomic Absorption Spectroscopy (AAS) 1 Thompson Rivers University.
Spectroscopy (continued) Last time we discussed what spectroscopy was, and how we could use the interaction of light with atoms and molecules to measure.
Atomic Emission Spectrometry
Atomic Absorption Spectrophotometer
Atomic Absorption Spectroscopy
AAS Atomic Absorption Spectrophotometry. AAS – Widely in clinical laboratories to measure elements such as aluminum, calcium, copper, lead, lithium, magnesium,
1 Atomic Absorption Spectroscopy Lecture Emission in Flames There can be significant amounts of emission produced in flames due to presence of flame.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 20
Atomic spectroscopy Elemental composition Atoms have a number of excited energy levels accessible by visible-UV optical methods ä Must have atoms (break.
Chapter 15 Molecular Luminescence Spectrometry Three types of Luminescence methods are: (i) molecular fluorescence (ii) phosphorescence (iii) chemiluminescence.
Beers Law for a Single Component Sample I0I0 A = Absorbance = - log 10 I I / I 0 b = Optical path length c = Solution Concentration (M/L) ε = Molar Absorptivity.
Atomic Fluorescence Spectroscopy. Background l First significant research by Wineforder and Vickers in 1964 as an analytical technique l Used for element.
Chapter 9 Atomic Absorption and Atomic Fluorescence Spectrometry.
Chapter 9 Atomic Absorption and Atomic Fluorescence Spectrometry
Plasma A Plasma consists of a collection of free-moving electrons and ions and is very hot. Energy must be continually applied to sustain the plasma.
What is Atomic Spectroscopy? – Sec Atomic Absorption Spectroscopy (AAS) versus Atomic Emission Spectroscopy (AES)
Atomic-absorption spectroscopy Lab3 Atomic-absorption spectroscopy.
Atomic Absorption Spectroscopy.  Qualitative test for metals in solution  Can also be used for coloured lights  This emission of light is called an.
Chem. 133 – 4/12 Lecture. Announcements I Strike – No Strike now planned Exam 2: – Average was 77 (range 61 to 93) Last HW Set (Set 3 – see handout) Lab.
1 Chapter 8 Atomic Absorption Spectroscopy ( AAS ) Yang Yi College of Science, BUCT.
Chapter 28 Atomic Spectroscopy.
Chem. 133 – 4/4 Lecture.
ATOMIC ABSORPTION AND ATOMIC FLUORESCENCE SPECTROMETRY
Flame Emission Spectrometry
Chem. 133 – 3/14 Lecture.
Flame Photometer.
Atomic Absorption and Atomic Fluorescence Spectrometry
Spectroscopy Chapter 7.
Elemental composition
Atomic Absorption and Emission
Atomic Absorption and Atomic Fluorescence Spectrometry
Atomic Absorption Spectroscopy
Пламена Спектроскопија
Atomic Absorption Spectroscopy. Atomic absorption spectroscopy is based on the same principle as the flame test used in qualitative analysis.
Fluorescence.
Spectroscopy Uses emission and absorption of light by electrons moving between ground and excited state configuration, hence electronic configuration.
FLAME SPECTROSCOPY The concentration of an element in a solution is determined by measuring the absorption, emission or fluorescence of electromagnetic.
Inductively Coupled Plasma
Presentation transcript:

ATOMIC ABSORPTION AND ATOMIC FLUORESCENCE SPECTROMETRY Chap 9 Source Modulation Interferences in Atomic Absorption Interferences in Atomic Absorption Spectral Spectral Chemical Chemical Atomic Fluorescence Spectroscopy Atomic Fluorescence Spectroscopy

AA Source Modulation Need to eliminate emission from analyte atoms Need to eliminate emission from analyte atoms Source beam is chopped Source beam is chopped Chopper placed here

Beam chopper for subtracting flame subtracting flame background emission (a) Lamp and flame emission reach detector (b) Only flame emission reaches detector (c) Resulting signal

ATOMIC ABSORPTION AND ATOMIC FLUORESCENCE SPECTROMETRY Chap 9 Source Modulation Source Modulation Interferences in Atomic Absorption Spectral Chemical Atomic Fluorescence Spectroscopy Atomic Fluorescence Spectroscopy

Spectral Interferences 1) Undesired signals overlapping analyte signal e.g., V at nm near Al at nm ∴ use Al at nm 2) combustion products  broadband absorption subtract signal from a blank subtract signal from a blank 3) matrix interferences (most serious) use higher T use higher T use radiation buffer use radiation buffer add excess of known interferent add excess of known interferent

Other methods of correcting for matrix effects Continuum Source (D 2 ) Correction Method Continuum Source (D 2 ) Correction Method D 2 lamp provides continuum D 2 lamp provides continuum UV abs. is subtracted from analyte signal UV abs. is subtracted from analyte signal Fig Fig. 9-14

Fig 9-14Continuum-source Fig 9-14 Continuum-source background correction

Other methods of correcting for matrix effects Continuum Source (D 2 ) Correction Method Continuum Source (D 2 ) Correction Method D 2 lamp provides continuum D 2 lamp provides continuum UV abs. is subtracted from analyte signal UV abs. is subtracted from analyte signal Fig Fig Zeeman Background Correction Zeeman Background Correction based on splitting of absorption lines based on splitting of absorption lines by a magnetic field by a magnetic field lines absorb different polarizations lines absorb different polarizations

Fig 9-14 Background correction by Zeeman effect B field splits atomic absorption lines (Zeeman effect) Line absorption differs with polarization of source

Chemical interferences (very common) 1) Certain components of the sample decrease the extent of atomization the extent of atomization e.g., SO 4 2- and PO 4 3- hinder atomization of Ca 2+ Add releasing agent: Sr +, La 3+, etc.Add releasing agent: Sr +, La 3+, etc. Add protecting agent: EDTA, hydroxyquinolineAdd protecting agent: EDTA, hydroxyquinoline 2)Ionization interference Occurs when O 2 or N 2 O is oxidantOccurs when O 2 or N 2 O is oxidant Analyte ionizes and releases electrons at high TAnalyte ionizes and releases electrons at high T M ⇌ M + + e -

Table 9-2 Degree of Ionization with temperature Add ionization supressor: K, Rb, Cs saltsAdd ionization supressor: K, Rb, Cs salts Provides high concentration of electrons to flameProvides high concentration of electrons to flame M ⇌ M + + e - shifted to the leftM ⇌ M + + e - shifted to the left

Fig 9-17 Effect of K as ionization suppressor for Ca

Table 9-3 Detection limits (ppb) for selected elements

ATOMIC ABSORPTION AND ATOMIC FLUORESCENCE SPECTROMETRY Chap 9 Source Modulation Source Modulation Interferences in Atomic Absorption Interferences in Atomic Absorption Spectral Spectral Chemical Chemical Atomic Fluorescence Spectroscopy

Absorption and fluorescence by atoms in a flame

Set-up for a Luminescence Experiment If source is at fixed λ and monochromator is scanned an emission spectrum results. If source is scanned, and monochromator is at a fixed λ, an excitation spectrum results. Fig. 7-1 (b) 90°