Theory & Appl. Light Microscopy Phase Contrast Optics.

Slides:



Advertisements
Similar presentations
24.6 Diffraction Huygen’s principle requires that the waves spread out after they pass through slits This spreading out of light from its initial line.
Advertisements

Adjusting a Microscope 1Center components on optic axis 2Focus objective 3Focus condenser 4Adjust illumination lamp voltage (intensity) iris diaphragm.
Chapter 4 Notes Part II: Microscopy (refer to pg.60-61)
Biology 177: Principles of Modern Microscopy Lecture 09: Polarization and DIC.
IPC Friedrich-Schiller-Universität Jena Contrast modes in light microscopy: Bright field.
The Wave Nature of Light
Microscopy Outline 1.Resolution and Simple Optical Microscope 2.Contrast enhancement: Dark field, Fluorescence (Chelsea & Peter), Phase Contrast, DIC 3.Newer.
Chapter 3: Observing Microorganisms Through a Microscope
ERT107 MICROBIOLOGY FOR BIOPROCESS ENGINEERING Pn Syazni Zainul Kamal PPK Bioprocess.
POWERPOINT PRESENTATION ON POLARISED MICROSCOPE
02 Dissecting Microscope. A B Carrying a Microscope.
Diffraction of Light Waves
Chapter 24 Wave Optics.
Chapter 24 Wave Optics.
Microscopy Boot Camp /08/25 Nikitchenko Maxim Baktash Babadi.
Circular Dichroism Part I. Introduction.
P. Moghe, 125:583 1 Microscopy Techniques for Biomaterials and Cell Based Interfaces Professor Prabhas V. Moghe October 26, :583 Fall 2006.
USE AND CARE OF THE MICROSCOPE LECTURE 1. MICROSCOPY u Light Microscopy: any microscope that uses visible light to observe specimens u Compound Light.
Newton’s Rings Another method for observing interference in light waves is to place a planoconvex lens on top of a flat glass surface, as in Figure 24.8a.
Chapter 24 Wave Optics.
Microscopy Techniques for Biomaterial Characterization: A Primer Prabhas V. Moghe Lecture 3 September 21, 1999 RU CBE 533 or BME 553; NJIT BME 698.
Chapter 25:Optical Instruments Cameras Homework assignment : Read Chap.25, Sample exercises : 4,21,24,41,43  Principle of a camera ss’ D Intensity of.
Optical microscopy Optics Lenses and image formation Depth of field Numerical aperture Resolution Instrument Specimen preparation Contrast Examples Optics.
Diffraction vs. Interference
3: Interference, Diffraction and Polarization
Chapter 37 Wave Optics. Wave optics is a study concerned with phenomena that cannot be adequately explained by geometric (ray) optics.  Sometimes called.
Microscope.
Microscopy.
Nano-Electronics S. Mohajerzadeh University of Tehran.
The wave nature of light Interference Diffraction Polarization
Identification of minerals with the petrographic microscope
Microscopy. Scale Lenses and the Bending of Light light is refracted (bent) when passing from one medium to another refractive index –a measure of how.
Honors Microbiology: Chapter 3 Microscopy and Staining
FNI 2B OM 1 Optical Microscopes. FNI 2B OM2 Outline Justification History Components of the Optical Microscope Theory of operation  Basic Microscope.
Chapter 36 In Chapter 35, we saw how light beams passing through different slits can interfere with each other and how a beam after passing through a single.
Agenda How to make the specimen visible – Definition of Contrast
Chapter 24 Wave Optics. General Physics Review – waves T=1/f period, frequency T=1/f period, frequency v = f velocity, wavelength v = f velocity, wavelength.
Chapter 36 Diffraction In Chapter 35, we saw how light beams passing through different slits can interfere with each other and how a beam after passing.
Basic Microscopy – An Overview – October 2005 Protistology Course MBL, Woods Hole, MA.
Optical path length opl = path length x η Δopl = Δ path length x Δη δ (phase shift) = 2π x Δopl / λ 0o0o 90 o 270 o 360 o 180 o How far out of phase have.
Advanced Biology Visualizing Cells. The Human Eye  Resolution – The minimum distance two points can be apart and still be distinguished as two separate.
Chapter 24 Wave Optics. The particle nature of light was the basis for ray (geometric) optics The wave nature of light is needed to explain various phenomena.
Interference in Thin Films, final
Chapter 6 Companion site for Light and Video Microscopy Author: Wayne.
Molecular Cell Biology Light Microscopy in Cell Biology Cooper Modified from a 2010 lecture by Richard McIntosh, University of Colorado.
Contrast Microscopy Koehler Illumination Dark Field Phase Contrast Differential Interference Contrast Hoffman Modulation.
Lecture Nine: Interference of Light Waves: I
Lecture 24 Interference of Light.
Chapter 38 Diffraction Patterns and Polarization.
Announcements HW set 10 due this week; covers Ch (skip 24.8) and Office hours: Prof. Kumar’s Tea and Cookies 5-6 pm today My office hours.
The Microscope and Forensic Identification. Magnification of Images A microscope is an optical instrument that uses a lens or a combination of lenses.
Chapter 24 The Wave Nature of Light
Brightfield Contrasting Techniques Kurt Thorn Nikon Imaging Center University of California, San Francisco USA.
Phase contrast and DIC Gerd A. Guenther 2011 Honorable Mention Nikon Small World photomicrograhy competition Image of a freshwater ciliate. Ryan McGorty,
Chapter 24 Wave Optics. Young’s Double Slit Experiment Thomas Young first demonstrated interference in light waves from two sources in Light is.
THE MICROSCOPE. Antony van Leeuwenhoek ( ) Inventor of the first microscope.
Brightfield Contrasting Techniques Kurt Thorn NIC.
Microscopy.
Microscopy Group 2 Cabatit, Mendoza, Ramos, Rodriguez, Tan.
By c.Keerthana.  First described by Dutch physicist frits Zernike in  It is a type of light microscopy.  It is a contrast enhancing optical technique.
Advanced Higher Physics Interference
Chapter 25 Wave Optics.
Microscope.
Two-Beam Interference
Chapter 36 In Chapter 35, we saw how light beams passing through different slits can interfere with each other and how a beam after passing through a single.
LIGHT MICROSCOPY basic
LIGHT MICROSCOPY variations
140MIC: Microbiology Lecture-6 Microscopes.
Theoretical Background Challenges and Significance
Presentation transcript:

Theory & Appl. Light Microscopy Phase Contrast Optics

Theory & Appl. Light Microscopy Abbé Theory Designed optics for amplitude objects Absorb light without change in phase of light waves Based on assumption of no difference in index of refraction between specimen and background

Theory & Appl. Light Microscopy

Criterion for Resolution Lens must capture undiffracted light plus at least first order of diffracted rays Combine these in image plane by interference But — most biological specimens (esp. living) are not amplitude objects Phase Objects

Theory & Appl. Light Microscopy Phase Objects Do not absorb light Difference in index of refraction between specimen and background

Theory & Appl. Light Microscopy

Example: Cell Object 1.25  m thick, i.r. = 1.35; i.r. water = 1.30 (0.05 difference) Difference in path length for light = 1.25 (0.05) =  m 62.5/500 nm = 1/8 wavelength /8 =  /4 radians = 45° This is difference in phase of wave passing through cell against wave passing next to cell

Theory & Appl. Light Microscopy

Phase Differences Our eyes cannot see this Eyes set for amplitude differences, so cell is essentially transparent But — information is present in light beams from specimen and in image How do we see this?

Theory & Appl. Light Microscopy Frits Zernike (1888–1966) Dutch physicist Developed vector notation for theory of light propagation through phase objects Invented phase contrast optics in 1930; not manufactured until 1941 by Zeiss

Theory & Appl. Light Microscopy P S  Zernike Phase Vector Diagram For propagation of light through phase object Length of P = amplitude specimen/amplitude medium =  transmission ratio S = incindent wave P = particle wave  P = phase shift of ray through specimen (S = U, undiffracted (0- order) ray

Theory & Appl. Light Microscopy P  U D D =  of all diffracted orders of light from specimen U = undiffracted light P = resulting specimen light, produced by interference between U and D in image formation Calculate P by vector addition U + D = P By the law of sines

Theory & Appl. Light Microscopy

Brightfield Optics Shifts all vectors in phase equally, and may change all amplitudes equally: U + D = P U = P No amplitude image Information in P is present in , not in amplitude — eye cannot see this

Theory & Appl. Light Microscopy

Phase Contrast Imaging Basic principle: –Shift phases (  s) and/or amplitudes of U and D differentially –This can produce a change in amplitude of P (length of vector)

U'U' D P  U D P'P' U'U' D'D'D'D' In specimen In microscopeAt image plane U = P U'  P' Amplitude!

Theory & Appl. Light Microscopy Phase Contrast Optics Physically separates U and D light and subjects one or the other to phase shift and/or amplitude shift In theory, any shift of U and D are possible In practice, a shift of  90° ( /4) is appropriate for most biological specimens

Theory & Appl. Light Microscopy

Optical Arrangements Several possible, but major design challenge to keep U and D rays separate and handled differently In practice, use a hollow cone of light to illuminate specimen –Phase Annulus below condenser –Phase plate at back focal plane of objective Only 0 order rays from annulus pass through plate

Theory & Appl. Light Microscopy

Phase Plate Rings in phase plate can include –Attenuating layer (absorption but no phase shift), or –Phase-shifting layer (no absorption, phase shift only), or –Any combination of the two

Theory & Appl. Light Microscopy Positive/Negative Phase Positive Phase Specimen dark against light background (usual now) Negative Phase Specimen bright against dark background (looks like darkfield optics)

P  U D U = P P'P' U'U' D'D' U' > P' Positive Phase Retard D relative to U (move D vector clockwise)

P  U D U = P P'P' U'U' D'D' U' < P' Negative Phase Advance D relative to U (move D vector counterclockwise)

Theory & Appl. Light Microscopy

Example Systems Anoptral Phase Contrast Change amplitude of U (soot on ring), no phase shifts for either U or D rays. Bright image — negative phase Popular among algae workers in Great Britain in 50s–60s

P  U D U = P P'P' U'U' D'D' U' < P' Anoptral Phase Produces delicate image against brown background No phase shifts on ring

Theory & Appl. Light Microscopy

Example Systems Zernike Phase Contrast Differential changes in amplitude and phase of U and D rays. All combinations possible: –Amplitude absorption with no phase shift (metal coating) –Phase shift wavefront with no absorption (silica coating)

From: Rose & Pomerat (1960) J. Biophys. Biochem. Cytol. 8:423.

Theory & Appl. Light Microscopy Use/Limitation of Phaseco Use for qualitative, not quantitative evaluation of specimens Reasons: –Intensity differences in image not uniquely related to index of refraction differences of specimen –Phase halo — optical artifact Cannot completely separate U and D rays in optics

Theory & Appl. Light Microscopy Intensity Differences Two points may have same image intensity, but have different  values (different i.r.s) I.e., if IP/IU of  at 240° identical to ratio at 320°, then how distinguish different i.r.?

Theory & Appl. Light Microscopy Phase Halo Serious artifact, most prominent at boundaries of sharp differences in i.r. Exceeds ability of optics to produce an accurate image So identification of exact boundary of specimen from image is very difficult

Theory & Appl. Light Microscopy

Reducing Phase Halo Modification of design of phase plate Apodized Phase Contrast Addition of neutral density filters to phase plate to suppress halo Optical Process

Theory & Appl. Light Microscopy

Reducing Phase Halo Modification of specimen and medium Worst halo comes from abrupt i.r. difference between specimen (cell) and medium it is in Match i.r. of medium to i.r. of specimen to reduce halo Barer & Joseph (1957) Symp. Soc. Exp. Biol. 10:160–184. Use of non-osmotic solutes to increase medium index of refraction

Theory & Appl. Light Microscopy Interference Microscopy Like phaseco in that imaging produces amplitude differences from phase differences in specimen Quantitative Techniques Qualitative Techniques

Theory & Appl. Light Microscopy Optical Path Difference Specimen vs. medium  ' = (  s -  m )t  ' = optical path length t = physical thickness Can measure  ', then calculate  s = (  '/t) +  m

Theory & Appl. Light Microscopy Dry Mass Calculations Derived from  ' Need to determine , the refractive increment (difficult) (For most biological specimens,  = 1.8 x i.r./gm solute/100 ml)

Theory & Appl. Light Microscopy C (dry weight concentration) = (  specimen -  water )/  = (  s – 1.33)/1.8 x = gm/100 ml = gm solids x 100/(area x thickness)  ' =  C t Mass of solids per cell = (  ' x area)/100  = (  ' x area)/0.18

Theory & Appl. Light Microscopy Double Beam Interference Phaseco — image formed from interference between 0 order and diffracted orders from specimen Double Beam Interference — image arises from interference between light from specimen and from a reference beam that does not pass through specimen (No phase halos from incomplete separation of U and D rays)

Vector Diagrams U'U' U R R = reference beam = U = P = A 0 U' =  2 A 0  1.4 A 0 P' P R Interference between P and R produces P'  1.8 A 0

Theory & Appl. Light Microscopy Image –Specimen bright against background –Ratio of intensities (1.8/1.4) 2  1.6 Can vary amplitude and phase of R vector to produce negative contrast as well

Theory & Appl. Light Microscopy Coherent Optics For this to work, the specimen and reference beams must be coherent to one another (Not needed for phaseco: U and D emerge from same point in specimen and are automatically coherent) Light from source must be split into 2 beams and reunite these in image

Theory & Appl. Light Microscopy Mach-Zender Double Microscope Classical form Difficult to construct Difficult to set up optics Difficult to interpret images Beam splitter system must have twin matched objectives and condensers (and add appropriate compensators)

Theory & Appl. Light Microscopy Image contains interference fringes in a gradient across field: /2, 3 /2, 5 /2, 7 /2, etc. Displacement of fringe is related to difference in optical path through the specimen:  ' Measure physical thickness of specimen and calculate C and dry weight

Theory & Appl. Light Microscopy Not Commonly Used Mach-Zender expensive and specialized More commonly used systems: split beam interference optics Single condenser and objective used Reference and Specimen beams present in same system Double Beam Interference Optics

Theory & Appl. Light Microscopy Jamin-Lebedeff Microscope Special attachments applied to condenser and objective, as well as polarizer and analyzer system About 2/3 of field has useable image (rest has ghost image) Rotation of analyzer allows quantification of image information Angle information produces  ' Then measure vertical thickness of specimen to calculate dry weight

Theory & Appl. Light Microscopy

Problems with Designs Image deteriorates with higher magnification objectives (40x max) Optical path differences in different scopes Contrast is lost with open aperture Condenser and Objective must be specially modified and are not useable for other optics

Theory & Appl. Light Microscopy Common Biological Use Nomarski Differential Interference Contrast (DIC) Qualitative, not quantitative use Nomarski 1952 patent (Allen, et al. (1969) Zeit. fur Wiss. Mikros. 69:193) DIC sensitive to d  /ds, so shows refractive gradients or interfaces

Theory & Appl. Light Microscopy Georges (Jerzy) Nomarski (1919–1997) Polish-born, lived in France after World War II Physicist, many inventions Developed modification of interference microscopes now known as differential interference contrast (DIC) optics

Theory & Appl. Light Microscopy Robert Day Allen (1927–1986) Pioneered practical applications of Nomarski’s system

Theory & Appl. Light Microscopy DIC Complicated optical arrangement involving polarizer, analyzer, double wollaston prisms. Polarizer produces light; lower wollaston prism separates that into 2 component beams polarized at right angles to one another

Theory & Appl. Light Microscopy Lower wollaston also modified to separate two beams in space Each beam is R for the other Displacement of beams is set for each objective’s resolution: –100x, NA 1.25 — 0.2  m –40x, NA 0.65 — 0.55  m –16x, NA 0.32 — 1.32  m Upper wollaston recombines 2 beams into same path, but is adjustable Usually displace from precise recombination

Theory & Appl. Light Microscopy Nomarski Image Result is extinction (shadow) on one side of specimen and reinforcement (bright) on the other Shear of image False relief 3D image Consider wavefront diagrams

Theory & Appl. Light Microscopy

Shear in Image Degree of shear is set by wollaston combination Bias of shear adjustable by shifting upper wollaston position to retard one beam more or less relative to other Cannot be used for quantitative measurements of dry mass But extremely useful for observing living cells

Theory & Appl. Light Microscopy

Comparison of Nomarski and Phase Contrast Optics Phase Contrast Cheaper Easier to set up Uses less than full aperture of objective Phase Halo — surrounds specimen and other changes in i.r. Nomarski More expensive Fussy alignment Uses full aperture — closet to theoretical limit Shadow Effect — contrast greatest at shear direction maximum

Phase Contrast Insensitive to birefringence in specimen or slides Extremely large depth of field — sensitive to artifacts far out of plane of specimen Doesn’t work well with stained specimens Nomarski Optics disrupted by birefriengence Extremely shallow depth of field — useful for optical sectioning of specimen Works well with stained specimens; optics can be adjusted to enhance contrast

Theory & Appl. Light Microscopy