ECE 501 Introduction to BME

Slides:



Advertisements
Similar presentations
By Dr.Ahmed Mostafa Assist. Prof. of anesthesia & I.C.U.
Advertisements

The electrocardiogram (ECG or EKG)
Cardiovascular System Block The Electrocardiogram (ECG)
Aims Introduction to the heart.
ECG Rhythm Interpretation
Chapter 6. Biopotential Amplifiers Michael R. Neuman
Technical Seminar Presentation-2004 Presented by : DEVANSHU PRAVEEN SAHOO (EI ) NATIONAL INSTITUTE OF SCIENCE & TECHNOLOGY Presented By Devanshu.
Example Problem You are measuring the EEG of a patient and accidently choose two different types of electrodes for EEG lead. One of them has a source impedance.
ECG Dr. Mohammed Shaat Modified by : Dr. Amal Al Maqadma.
ECE 501 Introduction to BME ECE 501 Dr. Hang. Part IV Bioinstrumentation Electrocardiogram ECE 501 Dr. Hang.
Cardiovascular Block Electrocardiogram (ECG)
Electrocardiography Prof. K. Sivapalan ECG 2 Principle of Electrocardiogram. Trunk as volume conductor. Positively charged and negatively charged.
Co-ordination of the Cardiac Cycle Aims Describe how heart action is coordinated with reference to the sinoatrial node (SAN), the atrioventricular node.
How does the heart function? Control of the heart.
ECG Dr Mahvash Khan MBBS, MPhil. The ECG is a record of the overall spread of electrical activity through the heart.
Electrocardiography Dr. Shafali Singh
EE 5340/7340, SMU Electrical Engineering Department, © Electrocardiogram (ECG) n Generated in the heart n amplitude range: mV n frequency.
Figure 6.1 Rough sketch of the dipole field of the heart when the R wave is maximal The dipole consists of the points of equal positive and negative charge.
Lecture Objectives Describe sinus arrhythmias Describe the main pathophysiological causes of cardiac arrhythmias Explain the mechanism of cardiac block.
Normal electrocardiogram
1 ECG Signal Make-up. 2 Biopotential Requirements n High Input Impedance –Minimum loading of the signal being measured –Minimum 10M  n Input circuit.
Dr.Bayat,MD Assistant professor of cardiology Echocardiologist.
Co-ordination of the Cardiac Cycle SBI3U. The heart is made of cardiac muscle. When the cells receive an electrical impulse they contract - causing a.
Dr. Mona Soliman, MBBS, MSc, PhD Department of Physiology Chair of Cardiovascular Block College of Medicine King Saud University.
ELECTROCARDIOGRAM (ECG) Cardiovascular System (CVS 227) BASIC PRINICPLES Dr.Mohammed Sharique Ahmed Quadri Asst. professor in physiology Al Maarefa College.
Vital Signs Monitor UConn BME 4900 Vital Signs Monitor Purpose As the population ages, many people are required by their doctors to take vital signs.
The Conducting System and EKG Danny Golinskiy, Joel Levy, Emily Brames.
ECG Interpretation Hina Shaikh. What is ECG? Graphical records of electrical current, that is generated by heart Basic equipment: electrodes, wires, amplifier,
Electrocardiographs ECG. OUTLINES : Introducing what is meant by Biopotential signals. what is Electrocardiograph (ECG). The conduction system. The standard.
Ulkumen Rodoplu, MD EuSEM ELECTROCARDIOGRAPHY. Plan A Normal ECG Basic ECG Waveform Initiation of Spread of Electrical Activation in the Hearth The Magnitude.
ECG Basics.
1 Electrocardiography – Normal 6 Faisal I. Mohammed, MD, PhD.
ELECTROCARDIOGRAM (ECG) Cardiovascular System (CVS 227) BASIC PRINICPLES Dr.Mohammed Sharique Ahmed Quadri Asst. professor in physiology Al Maarefa College.
Introduction to the EKG. Electricity of the Heart The contraction of any muscle is associated with electrical changes called depolarizations and can be.
The Heart Continued... Human heartbeats originate from the SA node near the right atrium. Cardiac muscle cells contract, sending a signal to other muscle.
Electrocardiogram (ECG/EKG)
Electrocardiograph. History Italian scientist Carlo Matteucci realizes that electricity is associated with the heart beat Irish scientist.
Electrocardiography (ECG) Electrocardiogram  The tracing made by an electrocardiograph  Electrocardiograph an instrument for recording the changes.
Electrocardiogram Amplifier Design Using Basic Electronic Parts Background Lecture.
Electrocardiogram Amplifier Design Using Basic Electronic Parts
The Heart- PQRST and ECG
Electrocardiogram (ECG/EKG) Allied Health II. Heart Sounds Primarily from blood turbulence caused by closing of the heart valves 4 heart sounds 1 st 2.
22nd April 2009 ECG Recording and Basic Interpretation.
1 Topics to be addressed: Blood Anatomy of Blood Vessels Anatomy of the Heart The Conduction System The Cardiac Cycle Cardiodynamics Blood Flow and its.
Physiologic signals Lecture (2).
Biomedical Instruments Design Biopotential Amplifiers
Biomedical Instrumentation
ECG 1.BIPOLAR LEADS I II III 2.UNIPOLAR LIMBS LEADS AVR AVL AVF
Electrocardiography A recording of the electrical activity of the heart over time Gold standard for diagnosis of cardiac arrhythmias Helps detect electrolyte.
Electrocardiography – Normal 5
Objective 12 Electrocardiograms
The Heart and Circulation
Cardiovascular System Block Cardiac Arrhythmias (Physiology)
ECG ECG is a test that displays the electrical activity of heart muscles over a period of time. Detected with the help of electrodes attached to the outer.
Introduction to the E.C.G.
Fast Response Action Potential of Contractile Cardiac Muscle Cell
ECG Rhythm Interpretation
ECG 1.BIPOLAR LEADS I II III 2.UNIPOLAR LIMB LEADS aVR aVL aVF 3.UNIPOLAR CHEST LEADS C1………..C6 4.RECORDING OF THE ECG.
Electrocardiography – Normal 6
Figure 6.1 Rough sketch of the dipole field of the heart when the R wave is maximal The dipole consists of the points of equal positive and negative charge.
I will understand how to read an Electrocardiogram (EKG)
Electrocardiogram (ECG)
Biomedical Instruments Design Biopotential Amplifiers 1.
Biopotential amplifiers
ECG Basics.
Electrocardiogram (ECG) NOTES
ECG – electrocardiography
(VII.) Electrocardiography
Presentation transcript:

ECE 501 Introduction to BME Dr. Hang

Part IV Bioinstrumentation Electrocardiogram ECE 501 Dr. Hang

Introduction Physiology Medicine Instrumentation ECE 501 Dr. Hang

Medicine The Electrical Cycle ECE 501 Dr. Hang

The Electrocardiogram Medicine The Electrocardiogram The electrocardiogram (ECG) is a standardized way to measure and display the electrical activity of the heart. • Physicians can diagnose problems with the heart by analyzing its ECG and comparing it to the ECG of a healthy heart. ECE 501 Dr. Hang

Medicine ECG Waves ECE 501 Dr. Hang

Medicine ECG Intervals ECE 501 Dr. Hang

Medicine A-V Blocks Complete Hear Block The cells in A-V node are dead and A-V node does not conduct action potential at all Atria and ventricles beat independently Ventricles are driven by abnormal pacemakers ECE 501 Dr. Hang

Medicine A-V Blocks First-degree Hear Block The A-V node is diseased The action potentials from atria can still reach ventricles It is greatly delayed ECE 501 Dr. Hang

Medicine Heart Flutter Paroxysmal Tachycardia An abnormal pacemaker in ventricle may discharge at a rapid rate Ventricles depolarize irregularly at a high rate ECE 501 Dr. Hang

Medicine Heart Flutter Atrial Flutter An abnormal pacemaker in atria may discharge at a rapid rate Atria depolarize irregularly at a high rate ECE 501 Dr. Hang

Premature V- Contraction Medicine Premature V- Contraction An accidental pacemaker within the ventricle causes an extra beat (extrasystole) ECE 501 Dr. Hang

Medicine Fibrillation Atrial Fibrillation Atria twitch irregularly ECE 501 Dr. Hang

Medicine Fibrillation Ventricular Fibrillation Ventricles twitch irregularly ECE 501 Dr. Hang

Medicine Ischemia Deficient in blood supply due to coronary occlusion Early stage: ST elevation Late stage: ST elevation + TQ depression ECE 501 Dr. Hang

Instrumentation The Dipole Model The electrical field generated by this dipole represents the electric activity of the heart at a specific instant Cardiac vector M: dipole moment, having a magnitude of proportional to the amount of charge multiplied by the separation of the two charges The magnitude and direction of M vary over time ECE 501 Dr. Hang

Instrumentation The Dipole Model a1,a2: Lead vectors va1= M.a1=|M|cosθ: ECG signal on lead a1 ECE 501 Dr. Hang

Instrumentation The ECG Model Electrodes are placed on points A and B RAB: Resistance between A and B RT1, RT2: Thoracic medium resistances ΦAB: ECG voltage signal on lead AB ECE 501 Dr. Hang

Instrumentation ECG leads –Limb leads ECE 501 Dr. Hang

ECG leads – Augmented Limb leads Instrumentation ECG leads – Augmented Limb leads a: aVR b: aVL c: aVF d: Lead vectors ECE 501 Dr. Hang

Instrumentation ECG leads – Chest leads Wilson’s Central Terminal ECE 501 Dr. Hang

Instrumentation ECG leads – Chest leads Lead vector: Between the electrode and Wilson’s Central Terminal Electrode positions Lead vectors ECE 501 Dr. Hang

Instrumentation 12 - lead ECG Simultaneous measurement from Three limb leads Three augmented limb leads Six chest leads Process all data from 12 leads to obtain ECG waveform ECE 501 Dr. Hang

Instrumentation Block Diagram of an Electrocardiograph Dr. Hang ECE 501 Dr. Hang

Instrumentation Protection Circuit Prevent high voltages from damage the electrocardiograph Voltage-limiting devices Current-voltage characteristics of Voltage-limiting devices ECE 501 Dr. Hang

Instrumentation Protection Circuit Parallel silicon-diode voltage limiting circuit Low voltage breakdown Vb: 600mV ECE 501 Dr. Hang

Instrumentation Protection Circuit Back-to-back silicon Zener-diode voltage limiting circuit Moderate voltage breakdown Vb: 2 to 20 V ECE 501 Dr. Hang

Instrumentation Protection Circuit Gas-discharge tube (neon light) voltage limiting circuit High voltage breakdown Vb: 50 to 90 V ECE 501 Dr. Hang

Instrumentation Lead Selector Determine which electrodes are necessary for a particular lead and to connect them to the remainder of the circuit Controlled by the operator or by the microcomputer of the electrocardiograph ECE 501 Dr. Hang

Instrumentation Calibration A 1-mV calibration signal is momentarily introduced into the electrocardiograph for each channel that is recorded ECE 501 Dr. Hang

Instrumentation Preamplifier Differential amplifier with high common-mode-rejection ratio (CMRR) Analog filters ECE 501 Dr. Hang

Instrumentation Preamplifier ECE 501 Dr. Hang

Instrumentation Isolation Circuit Contains a barrier to the passage of current from the power line Transformer isolation amplifier Optical isolator Capacitively coupled isolation ampifier ECE 501 Dr. Hang

Transformer isolation amplifier Instrumentation Isolation Circuit Transformer isolation amplifier ECE 501 Dr. Hang

Instrumentation Isolation Circuit Optical Isolator ECE 501 Dr. Hang

Instrumentation Isolation Circuit Capacitively coupled isolation amplifier ECE 501 Dr. Hang

Driven right leg circuit Instrumentation Driven right leg circuit Provides a reference point on the patient that normally is at ground potential The connection is made to an electrode on the patient’s right leg ECE 501 Dr. Hang

Driven right leg circuit Instrumentation Driven right leg circuit ECE 501 Dr. Hang

Instrumentation Driver amplifier Amplify the ECG to a level at which it can appropriately record the signal on the recorder Carry out the bandpass filtering ECE 501 Dr. Hang

Instrumentation Memory System Store digital electrocardiograms from different leads Store patient information ECE 501 Dr. Hang

Instrumentation Microcomputer Control the overall operation Perform a preliminary analysis to determine heart rate recognize some types of arrhythmia calculate intervals ECE 501 Dr. Hang

Instrumentation Recorder-printer Provide a hard copy of the recorded ECG signal Print out patient information, results of the automatic analysis Analog oscillograph-type recorder, thermal recorder, electrostatic recorder ECE 501 Dr. Hang

Instrumentation Inference in ECG Effect of voltage transient from other devices, such as cardiac defibrillation ECE 501 Dr. Hang

60 Hz power-line inference Instrumentation Inference in ECG 60 Hz power-line inference ECE 501 Dr. Hang

Electromyographic(EMG) inference Instrumentation Inference in ECG Electromyographic(EMG) inference ECE 501 Dr. Hang

Instrumentation Inference in ECG Electromagnetic inference Twisting wires to minimize inference ECE 501 Dr. Hang

A mechanism of electric-field pickup from the powerline Instrumentation Inference in ECG A mechanism of electric-field pickup from the powerline ECE 501 Dr. Hang

Another mechanism of electric-field pickup from the powerline Instrumentation Inference in ECG Another mechanism of electric-field pickup from the powerline ECE 501 Dr. Hang