QUALITY ASSURANCE & ACR ACCREDITATION. West Physics Consulting is a proud supporter of MTMI’s Diagnostic Imaging Programs. Thank you West Physics!

Slides:



Advertisements
Similar presentations
Receiver Bandwidth affects signal-to-noise ratio (SNR)
Advertisements

BOLD Imaging at 7T Mark Elliott CfN Symposium 4/9/2008.
Principles of the MRI Signal Contrast Mechanisms MR Image Formation John VanMeter, Ph.D. Center for Functional and Molecular Imaging Georgetown University.
Magnetic Resonance Imaging
Imaging Sequences part I
Introduction to MRI Magnetic Resonance Imaging
Proton Spin In absence of a magnetic field, protons spin at random
MRI Phillip W Patton, Ph.D..
Richard Wise FMRI Director +44(0)
M R I Pulse Sequences Jerry Allison Ph.D..
PHYSICS OF MAGNETIC RESONANCE
Six-week project Lauren Villemaire MBP 3970Z Department of Medical Biophysics University of Western Ontario.
Magnetic Resonance Imaging - MRI
MR Sequences and Techniques
Statistical Parametric Mapping
Clinical Evaluation of Fast T2-Corrected MR Spectroscopy Compared to Multi-Point 3D Dixon for Hepatic Lipid and Iron Quantification Puneet Sharma 1, Xiaodong.
DIAGNOSTIC ROLE OF STATIC AND DYNAMIC CONTRAST ENHANCED MAGNETIC RESONANCE IMAGING IN THE EVALUATION OF SOFT TISSUE TUMOURS Abstract No. IRIA
IMAGE QUALITY.
Parameters and Trade-offs
Topics spatial encoding - part 2. Slice Selection  z y x 0 imaging plane    z gradient.
ACR Ultrasound PM / Standards
Terry M. Button, Ph.D. Principals of Magnetic Resonance Image Formation.
Chapter 10 Lecture Mark D. Herbst, M.D., Ph.D..
FMRI: Biological Basis and Experiment Design Lecture 12: Signal-to-Noise Ratio Things that determine signal strength –voxel size –RF coil Things that determine.
Tissue Contrast intrinsic factors –relative quantity of protons tissue proton density –relaxation properties of tissues T1 & T2 relaxation secondary factors.
Medical Imaging Systems: MRI Image Formation
Magnetic Resonance Imaging 4
Principles of MRI Physics and Engineering
Imaging Sequences part II
The comparison of MRI imaging In 3.0 T & 1.5T By a.r.shoaie.
Magnetic Resonance Imaging
Seminar October, 2008 j. brnjas-kraljević. Imaging (MRI)  tomography technique  tomography technique – the volume image is built up by images of thin.
Medical Imaging Systems: MRI Image Formation
HOW TO OPTIMIZE MRI OF EXTREMITIES ?
B1 inhomogeneities lead to significant variations in image intensity and tissue contrast across the brain, particularly in spin echo sequences. While intensity.
RT 4912 Review (A) Rex T. Christensen MHA RT (R) (MR) (CT) (ARRT) CIIP.
Contrast Mechanism and Pulse Sequences Allen W. Song Brain Imaging and Analysis Center Duke University.
ACRIN CV Committee Imaging Procedures Scott Akers, MD, PhD, Radiologist Walter Witschey, PhD, Physicist Dena Flamini, RT (R) (MR), (M) Imaging Analyst,
Contrast Mechanisms in MRI Introduction to Cardiovascular Engineering Michael Jay Schillaci, PhD Managing Director, Physicist Tuesday, September 16 th,
Contrast Mechanism and Pulse Sequences
Chapter 10 Lecture Mark D. Herbst, M.D., Ph.D.. Image Detail and Noise Two image characteristics that reduce the visibility of anatomy/pathology –Blurring.
MRI Physics Dr Mohamed El Safwany, MD.. MRI Magnetic Resonance Imaging Magnetic Resonance Imaging.
NEW SEQUENCES LAVA Liver Acquisition with Volume Acceleration.
Protons (hydrogen nuclei act like little magnets) MRI Collective Magnetic Moment of Protons (M 0 ) Each pixel is a glass of protons B 0 = 3T (not to scale)
Basic MRI Chapter 1 Lecture. Introduction MRI uses radio waves and a magnetic field to make images MRI uses radio waves and a magnetic field to make images.
Quality Assurance.
V.G.Wimalasena Principal School of Radiography
2002 symposium1 Craig Goodrich 2002 MIRL Symposium.
MRI Physics: Spatial Encoding Anna Beaumont FRCR Part I Physics.
MRI: Contrast Mechanisms and Pulse Sequences
Magnetic Resonance Learning Objectives
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
Jalal jalal shokouhi – MD General secretary of Iranian society of radiology President of Iranian radiologists cooperative 17th -18th December 2010, Shahid.
Real time shimming (RTS) for compensation of respiratory induced field changes P van Gelderen, JA de Zwart, P Starewicz, RS Hinks, JH Duyn Introduction.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
SWIFT Dual Breast Imaging Sequence and Coil with Interleaved Adiabatic Fat Suppression high temporal and spatial resolution #3369 Computer 8 breast imaging.
Chapter 5 Mark D. Herbst, M.D., Ph.D.. The MR Imaging Process Two major functions –Acquisition of RF signals –Reconstruction of images.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
FMRI data acquisition.
Impact of Obesity on Medical Imaging and Image-Guided Intervention
Sunday Case of the Day Physics
بسم الله الرحمن الرحيم.
Monday Case of the Day Physics
Magnetic Resonance Imaging: Physical Principles
Bioengineering 280A Principles of Biomedical Imaging Fall Quarter 2005 MRI Lecture 5 Thomas Liu, BE280A, UCSD, Fall 2005.
MRI Pulse Sequences: IR, EPI, PC, 2D and 3D
MRI ARTEFACTS PG Student : Dr Aditi Shah PG Guide: Dr Rajendra Chavan
Basics of fMRI and fMRI experiment design
The echo time (TE) The echo time (TE) refers to the time between the application of the radiofrequency excitation pulse and the peak of the signal induced.
Presentation transcript:

QUALITY ASSURANCE & ACR ACCREDITATION

West Physics Consulting is a proud supporter of MTMI’s Diagnostic Imaging Programs. Thank you West Physics!

WHY ACR ? Images property of Yu Liu PhD

WHY ACR ? Images property of Yu Liu PhD

IMAGE QUALITY Diagnostic Accuracy Signal to Noise (SNR) Contrast to Noise (CNR) Spatial Resolution Absence of Artifacts

SNR- Is the relationship between MR signal and background noise. Signal- Is the processing of induced voltage (Faraday’s Law of Induction) in a receiver coil after excitation. The voltage is dependent on the strength of the magnet (NMV) used for induction. Signal is not the only factor contributing to the quality of the image. Noise is refer as unavoidable background signals By evaluating SNR we can assess the overall quality of the MR Signal Images property of e-MRI.org

CONTRAST to NOISE RATIO (CNR) CNR: CNR: Also refers as Contrast Resolution,is the ability to distinguish between differences in intensity in an image. CNR is influence by the SNR and spatial resolution in the MR image.

Matrix (In-Plane) CONTRAST to NOISE (CNR) CNR : Is the ability to distinguished between two adjacent structures or tissue 192 x x 320

Slice Thickness CONTRAST to NOISE (CNR) CNR : Is the ability to distinguished between two adjacent structures or tissue 5mm 1.5mm

IMAGE QUALITY SNR/Spatial Resolution & Absence of Artifacts Scanner Performance

ACR (Q/A) Quaterly Preventive Maintenance Weekly Technologist Phantom – Phantom Scans Annual Testing by a Physicist – Phantom Scans – Coil Testing Re- Accreditation by the ACR – Every three years The MRI Accreditation Program evaluates the qualifications of personnel, the quality control program, MR safety policies and image quality specific to MRI.

ACR MRI ACCREDITATION Get the Team Together Get the Team Together – Radiologist (Medical Director MRI) – Chief Technologist (Clinical Imaging) – MR Physicist (Scanner Testing) Buy ACR Phantom Buy ACR Phantom (JM Specialty Part – San Diego,CA) Weekly Technologist Phantom Weekly Technologist Phantom (3-6 months ACR recommends) Scanner Testing by a Physicist Scanner Testing by a Physicist (Annual/Initial) Apply online for ACR Accreditation Apply online for ACR Accreditation (Select Module desire) – Survey Agreement (Site) – Six Modules (Brain,Spine,Msk,Body,Vascular,Cardiac ) – 45 days to complete your submission and return it to ACR – Could go back 6 months to provide clinical images. Getting Started:

ACR PHANTOM Images property of Chen Lin PhD Level Phantom Correctly

ACR PHANTOM ACR WEEKLY PROTOCOL : Sag Scout Sag Scout Ax T1 SE Ax T1 SE TR 500 TR 500 TE 20 TE 20 FOV 25 FOV 25 Matrix 256 x 256 Matrix 256 x 256 Slices 11 Slices 11 Slice Thickness 5mm Slice Thickness 5mm Gap 5mm Gap 5mm

ACR PHANTOM CENTER FREQUENCY: Set to the resonance frequency of water protons Set to the resonance frequency of water protons Depends on Main Magnetic Field (1.5T vs. 3T) Depends on Main Magnetic Field (1.5T vs. 3T) Affects fat saturations Affects fat saturations Severity off resonance artifacts Severity off resonance artifacts ACR recommended criteria < +/- 1.5ppm different between weekly test. Possible Causes of Failures : Phantom not Isocenter Phantom not Isocenter Magnet Drift Magnet Drift

ACR PHANTOM TRANSMITTER GAIN: RF power required to achived certain flip angles RF power required to achived certain flip angles Depends on RF Coils,Phantom,and Pulse Sequence Depends on RF Coils,Phantom,and Pulse Sequence Should remain constant Should remain constant Severity off resonance artifacts Severity off resonance artifacts. Possible Causes of Failures : Defective RF Transmitter Coil Defective RF Transmitter Coil RF Amplifier RF Amplifier

ACR PHANTOM ACR Sequence Positioning:

ACR PHANTOM Images property of Chen Lin PhD Geometric Accuracy: Related to image distortion 190 mm 148 mm

ACR PHANTOM. Possible Causes of Failures Phantom not level appropriate Phantom not level appropriate Gradient Non Linearity Gradient Non Linearity Eddy Currents Eddy Currents Geometric Accuracy: Related to image distortion ACR recommended criteria +/- 2mm in all directions

ACR PHANTOM SPATIAL RESOLUTION: Ability to detect differences between small structures Ability to detect differences between small structures Depends on Pixel/Voxel size (FOV/Matrix) Depends on Pixel/Voxel size (FOV/Matrix)

ACR PHANTOM. Possible Causes of Failures Poor SNR Poor SNR Bad Receiver Coil Bad Receiver Coil Eddy Currents Eddy Currents Gradient Calibration Gradient Calibration Spatial Resolution: ACR recommended criteria <= 1.0mm 1.1 mm 1.0mm 0.9mm

ACR PHANTOM Low Contrast Detectability (LCD): Ability to detect objects with subtle differences in signal intensity Ability to detect objects with subtle differences in signal intensity Possible Causes of Failures Poor SNR Poor SNR Inaccurate slice positioning Inaccurate slice positioning Artifacts Artifacts

ACR PHANTOM Additional Testing: Image uniformity Image uniformity Slice Thickness Slice Thickness Ghosting/Artifacts Ghosting/Artifacts

ACR PHANTOM Artifacts: Zipper/RF Leak

ACR ACCREDITATION Clinical Modules: Images property of Chen Lin PhD

ACR ACCREDITATION Clinical Images Evaluations : Pulse Sequence Pulse Sequence Image Contrast Image Contrast Anatomical Coverage Anatomical Coverage Imaging Planes Imaging Planes Spatial Resolution (Pixel & Voxel size) Spatial Resolution (Pixel & Voxel size) Artifacts Artifacts ID Exam ID Exam

ACR ACCREDITATION Tips & Suggestions All clinical images must be done in patients (No Volunteer). All clinical images must be done in patients (No Volunteer). Include all identification required (Patient & Parameters). Include all identification required (Patient & Parameters). DICOM images with a viewer. DICOM images with a viewer. Except Cardiac and Phantom DICOM only Except Cardiac and Phantom DICOM only Record True Matrix,not Reconstructed Matrix (Pixel size). Record True Matrix,not Reconstructed Matrix (Pixel size). Check for Artifacts prior to submission. Check for Artifacts prior to submission. Present best cases,no motion on images. Present best cases,no motion on images. Submit all required sequences,include localizers. Submit all required sequences,include localizers.

ACR ACCREDITATION Be Aware : Spine Axial GRE T2*/TSE at 3T,CSF must be bright(no flow artifact) Spine Axial GRE T2*/TSE at 3T,CSF must be bright(no flow artifact) MRCP include entire Biliary Tree Anatomy (Best when using 3D rather 2D MRCP) MRCP include entire Biliary Tree Anatomy (Best when using 3D rather 2D MRCP) Liver MRI,use flouro technique to acquired true arterial phase. Liver MRI,use flouro technique to acquired true arterial phase. CE-MRA,no venous contamination (Neck Carotid) CE-MRA,no venous contamination (Neck Carotid) Cardiac MRI,Viability exam myocardium fully saturated (Scar visible) Cardiac MRI,Viability exam myocardium fully saturated (Scar visible) Considered using Dixon for In/Out Phase at 3T Considered using Dixon for In/Out Phase at 3T Forms fill them out correctly(take your time) Forms fill them out correctly(take your time)

ACR ACCREDITATION Breast Accreditation: Breast Accreditation: Facilities performing breast MRI must have the capacity to perform mammographic correlation, directed breast ultrasound, and MRI-guided intervention, or create a referral arrangement with a cooperating facility that could provide these services. Radiologists – Requirement of Breast MRI read in the last 36 months Facilities must use the Breast Imaging Reporting and Data System (BI- RADS®)7 final assessment codes and terminology for reporting and tracking outcomes. Technologists – Requirement of 50 MRI Breast examination prior 24 months.

ACR ACCREDITATION Breast Accreditation: Facility must submit 1 case with a known, enhancing, biopsy- proven carcinoma clearly visible in the breast parenchyma. No Masectomy cases Pulse Sequences T2W/Bright Fluid T1W (Pre,Arterial,Delay) - Bilateral Slice Thickness < 3mm(No Gap),In Plane Resolution/Pixel Size < 1.0

ACR ACCREDITATION Summary: ACR Q/A Phantom needs to be done weekly Check scanner image quality Yearly Evaluation (PhD) Compares year to year scanner performance ACR Accreditation Done every three years,includes clinical images and phantoms. Why ACR ? Is all about image quality!!!!!! Is all about image quality!!!!!!