Machine-specific QC in Aarhus using EPID Mai-Britt K. Jørgensen, Lone Hoffmann, L.P. Muren, Lars H. Præstegaard, Rune Hansen, Jørgen B.B. Pedersen, Maria F. Jensen Aarhus University Hospital Department of Medical Physics
Rutine QC 2 ▪ IMRTVMAT Machine specificSweeping gap Picket fence (PF) Sweeping gap Picket fence (PF) Dose rate versus gantry speed (DRGS) Dose rate versus MLC speed (DRMLC) Patient specificPDI (Delta4) (PTW729 array) PDI (Delta4)
Picket fence, PF For every MLC and gab… Position: Relative rotation of EPID and collimator Intensity (”width”): 3 ▪
PF: ex. ”Eye test”
PF: ex. Deviation in position, tolerance ±0.3mm 5 ▪
PF: ex. Deviation in intensity, tolerance ±300
PF: ex. ”Eye test version 2”
PF: In total ~30 test on each accelerator, one every month
Dose rate vs. gantry speed, DRGS 9 ▪
DRGS: ex. typical deviation
DRGS: ex. cable issue 11 ▪
DRGS: In total ~30 test on each accelerator, one every month
Dose rate vs. MLC speed, DRMLC 13 ▪
DRMLC: ex. segment 4 DR120, 0.4cm/s 14 ▪
DRMLC: ex. segment 2 DR600, 2.4cm/s
DRMLC: ex. segment 4 DR120, 0.4cm/s segment 1 DR480, 1.6cm/s
DRMLC: In total ~30 test on each accelerator, one every month
Time consumption Measurements:~10 min Analysis (home-made software):~1 min Total:~11 min 18 ▪
Effect of chain adjustments on the DRGS and DRMLC tests Mehtod 3 DRGS and 3 DRMLC test with a loose chain 3 DRGS and 3 DRMLC test with a thight chain Result, comparing loose and thight chain DRGS: Standard deviation was unchanged DRMLC: Std 0.6 (loose) -> 0.2 (tight) 19 ▪
Loose chain: unstable characteristic
Tight chain: Stable characteristic
Discussion Deviations in PF are covered in next, next session Deviations in DRGS and DRMLC Chain adjustments have an effect on DRGS and DRMLC test, but it does not solve everything Cable issues Beam adjustments? For more information on the calculations and tolerances: Med. Phys. 38, 1425 (2011); 22 ▪