Resident Physics Lectures

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

Control of Scatter Radiation
Quality Control in Diagnostic Radiology
X-Ray Beam Restrictors
Radiographic Quality Chapter 5.
Quality Assurance and Digital Radiography
Resident Physics Lectures
POWERPOINT PRESENTATION ON POLARISED MICROSCOPE
Chapter 6 Control of Secondary and Scattered Radiation.
Control of Scatter Radiation Beam Restricting Devices
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Resident Physics Lectures
Chapter 6, Fauber Supplement-Chapter 14, Bushong
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
Basic quality control tests in medical radiography
Lecture 4 Beam restricting Devices, their Use and Maintenance
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
The Grading System For Lab. Lab Break Down Competencies65% Quizzes10% Final Comp25%
Quality Control Rad T 110.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Control of Scatter Radiation
Plain Film Tomography George David Associate Professor Department of Radiology Medical College of Georgia Resident Physics Lectures.
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Dr Mohamed El Safwany, MD.
Resident Physics Lectures
Resident Physics Lectures
Parts of a Compound Light Microscope
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Beam Restricting Devices
Radiographic Equipment
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Resident Physics Lectures Christensen, Chapter 4 Basic Interactions Between X-Rays and Matter, Grid Attenuation and Filtration George David Associate.
Part No...., Module No....Lesson No
Grids George David Associate Professor Department of Radiology
Chapter 2: The Lens. Focal Length is the distance between the center of a lens and the film plane when focused at infinity.
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Resident Physics Lectures Christensen, Chapter 2C Production of X-Rays George David Associate Professor Department of Radiology Medical College of Georgia.
Chapter 18 Quality Control
Week 2: Radiographic Equipment
M Manser WAVES : Optics (t riple science) By the end of this presentation you should be able to: Identify the differences between converging and diverging.
Control of Scatter Radiation Beam Restricting Devices By Prof. Stelmark.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Diagnostic Equipment Quality Control
Lenses Are classified by their Focal Length.
Resident Physics Lectures
Physics Case of the Day - Sunday
Resident Physics Lectures (year 1)
Protection of Patients
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (Year 1)
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (year 1)
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (year 1)
Resident Physics Lectures (Year 1)
Chapter 5, Fauber Supplement-Chapter 11, Bushong
Filters, Collimators and Grids
Fluoroscopy Static Image Recording Systems
Artifacts and Errors.
GRIDS PRESENTER-Dr.Vishwanath patil PG resident
Presentation transcript:

Resident Physics Lectures Christensen, Chapter 7 X-Ray Beam Restrictors George David Associate Professor Department of Radiology Medical College of Georgia

Restrictors Types Aperture Diaphragms Cones and Cylinders Collimators

Aperture Diaphragms Lead sheet with hole in center simple large penumbra (shadow) Lead close to focal spot Focal Spot Penumbra Film

Cones & Cylinders Focal Spot Focal Spot Film Film With cone angle as above, this functions as an aperture Reduced Penumbra

Collimators Advantages Configuration Provides adjustable rectangular field fluoro may also have circular field light beam indicates x-ray field Configuration two sets of shutters longitudinal transverse sets may actually have two pairs of shutters operating together One set; two pairs

Collimators Generally include filter & light Focal Spot Filter Lamp Mirror Filter Focal Spot Shutters (only 1 set shown)

Collimator Features Numerical field size indicators Field size is SID dependent 8 10 12 14 17 40” 8 10 12 14 17

Collimator Features Field center indicator (crosshairs) plastic sheet attached to bottom of collimator two black lines drawn on it light projects field center on patient Shutters Crosshairs Bottom of Collimator

Positive Beam Limitation (PBL or Automatic Collimation) senses film size, orientation, & SID source - image distance limits x-ray field to size of film No longer required PBL does not insure tube & bucky are aligned! Cassette

Collimator Checks X-Ray / Light Field alignment Alignment of x-ray field to bucky Field Size Indicator accuracy PBL accuracy