MRI Physics 3: Hardware Douglas C. Noll Biomedical Engineering University of Michigan.

Slides:



Advertisements
Similar presentations
MRI Phillip W Patton, Ph.D..
Advertisements

1 of 24 CCS&BS MR Safety SeminarCenter for Complex Systems and Brain Sciences Magnetic Resonance Safety Seminar: An Introduction to the Potential Hazards.
The importance of MRI, a few numbers  MRI units worldwide in 2003  75 millions scans per year performed  Constant need for over 1000 MRI technologists.
Magnetic Resonance Imaging
Noll An Overview of Echo Planar Imaging (EPI) Douglas C. Noll, Ph.D. Depts. of Biomedical Engineering and Radiology University of Michigan, Ann Arbor.
MRI Safety Jeopardy* ForcesProjectilesPotpourriHeatingScreening Because when you think MRI safety, think * Jeopardy is a trademark of Sony.
Methods: Scanner: GE 7T Excite (bore 60cm diam x 335cm length) and GE 3T LX (bore 60cm diam x 262cm length). The gradient performance of both scanners.
ELEG 479 Lecture #9 Magnetic Resonance (MR) Imaging
FMRI: Biological Basis and Experiment Design Lecture 18: Physical practicalities Digression: analysis ICE9: Example for WA8 Safety limits –dB/dt –SAR –Acoustic.
The NMR spectrometer Magnet Probe Coils Transmitters
EPI – Echo Planar Imaging Joakim Rydell
Introduction to Medical Imaging
FMRI: Biological Basis and Experiment Design Lecture 12: Signal-to-Noise Ratio Things that determine signal strength –voxel size –RF coil Things that determine.
Magnetic Resonance Imaging Basic principles of MRI This lecture was taken from “Simply Physics” Click here to link to this site.
Magnetic Resonance Imaging Mary Holleboom ENGR 302 May 7, 2002.
Magnetic Resonance Imaging
MRI SAFETY FOR NON-MRI PERSONNEL.
MAGENETIC RESONANCE IMAGINING T2, T5, T7.
MRI: an Introduction By Mohammad Ali Ahmadi Pajouh Amirkabir University of Technology Biomedical Eng. Dep.
Magnetic Resonance Imaging (MRI) By Isabelle! and sofia and ronaldo.
Magnetic Resonance Imaging
RF Coils Used In MRI.
Chapter 5 Lecture 10 Spring Nonlinear Elements 1. A nonlinear resistance 2. A nonlinear reactance 3. A time varying element in you circuit or system.
Chapter 15: MRI Safety Mark D. Herbst, MD, PhD. Two main sources of danger Magnetic field RF Energy Also –Gradients –Acoustic noise.
Principles of MRI Physics and Engineering
Principles of Magnetic Resonance
Magnetic Resonance Imaging
CT “Computer tomography”. Contrast mechanisms in X-ray imaging: X-ray absorption X-ray absorption mechanisms: 1. Photoelectric effect 2. Compton scatter.
2012 spring fMRI: theory & practice
Medical Imaging Systems: MRI Image Formation
Stimulus Presentation and Image Acquisition Valentina Petre.
Hearing Impairment Hair cells are responsible for translating mechanical information into neural information. Thus, with damaged hair cells, the auditory.
Brain Research Methods Maddie Coates. Direct Brain Stimulation Direct brain stimulation is when a device is sends a weak electrical current to disrupt.
MRI DRAFT Principles of MRI ► Sectional Anatomy ► Provides anatomic and physiologic info ► Non- invasive ► No radiation  Magnetic fields and.
Rad T 265 MRI Lecture. No Magnetic Field = No Net Magnetization Protons align with a magnetic field…
Pei-Ann Lin and PJ Velez December 13, 2011
Review of Ultrasonic Imaging
Human Subjects in fMRI Research Credits: Robert Savoy, Ph.D. Franz Schmitt, Ph.D.
1 Introduction to Magnetic Resonance Imaging درس مقدمه ای بر فیزیک پزشکی Sahand University of Technology Faculty of Electrical Engineering M. Shamsi.
Magnet Safety.
G Practical MRI 1 Gradients.
Brain Research Methods.
Magnetic Resonance Imaging
Functional Brain Signal Processing: EEG & fMRI Lesson 11 Kaushik Majumdar Indian Statistical Institute Bangalore Center M.Tech.
Lesson Plan MRI Scan Experience
Magnetic Resonance Imaging (MRI). The Components: A magnet which produces a very powerful uniform magnetic field. A magnet which produces a very powerful.
Principles of MRI Physics and Engineering Allen W. Song Brain Imaging and Analysis Center Duke University.
 This depends on a property of nuclei called spin.  Gyroscope: Principle: As long as its disc remains spinning rapidly the direction of the spin axis.
MAGNETIC RESONANCE IMAGING by PRADEEP V.EPAKAYAL. Mem.no L.
Medical Imaging. X-ray Radiography 2-D projection image created from difference in x-ray absorption rates between tissues 2-D projection image created.
Medical Resonance Imaging
Medical Imaging.
Methodist Hospital for Surgery
Lesson Plan MRI Scan Experience
The strength of the flux density at the Earth’s surface ranges from less than 30μT in an area including most of South America and South Africa to over.
Jody Culham Brain and Mind Institute Department of Psychology
INTRODUCTION to RADIOLOGY
Medical Imaging Imagining Modalities.
Magnetic field safety October 2018.
Basics of MRI By Dr. Mohamed Samieh.
Human Subjects in fMRI Research
Lesson Plan MRI Scan Experience
Lesson Plan MRI Scan Experience
Magnetic Resonance Imaging [MRI]
Lesson Plan MRI Scan Experience
When a person lies in a scanner, the hydrogen nuclei (i. e When a person lies in a scanner, the hydrogen nuclei (i.e., protons) found in abundance.
Basics of MRI and MRI safety
Hoglund Brain Imaging Center
(4)ELECTRONIC SUPPORT SYSTEM
Superconducting Magnets
Presentation transcript:

MRI Physics 3: Hardware Douglas C. Noll Biomedical Engineering University of Michigan

Noll MRI Hardware Recall the three magnetic fieldsRecall the three magnetic fields –B 0 –B 1 –G x, G y, G z Also, we need pulse control and data acquisition systemsAlso, we need pulse control and data acquisition systems

Noll Major MRI Components

Noll RF Screen Room Encloses the MRI deviceEncloses the MRI device Shields MRI scanner from electromagnetic noise from computers, radio stations, etc.Shields MRI scanner from electromagnetic noise from computers, radio stations, etc. –MRI signals are weak compared to noise sources –Any wires going in/out must be filtered and grounded so they don’t introduce noise These rooms can also incorporate magnetic shielding to contain the magnetic field.These rooms can also incorporate magnetic shielding to contain the magnetic field.

Noll The Main Field B 0 range: 0.2 T to 7 TB 0 range: 0.2 T to 7 T Higher fields:Higher fields: –Higher SNR –Slightly longer T1’s, –Poorer RF homogeneity –Higher power deposition –Higher cost Typically superconductingTypically superconducting

Noll The Main Field Safety issues:Safety issues: –Attraction of ferrous objects –Attraction of metallic implants (e.g. cochlear implants, neurostimulators, pacemakers, and poorly designed/manufactured stents, screws, pins, aneurysm clips, etc.) –Attraction of foreign metal objects (metal in eyes, shapnel, ingested ferrous objects). –Affects magnetic switches in pacemakers –Some reports of dizziness, light flashes, unusual tastes, etc. at very high magnetic fields. –The FDA has classified 7 T and under as a “non- significant risk.”

Noll RF Fields RF coils must generate a rotating (or oscillating) magnetic fieldRF coils must generate a rotating (or oscillating) magnetic field –Ideally uniform application of B 1 RF coils also detect precessing magnetizationRF coils also detect precessing magnetization –Ideally sensitive mainly to tissue of interest –Uniformity is of secondary importance Three main types:Three main types: –Volume Coils –Surface Coils –Arrays Coils (which are basically arrays of surface coils)

Noll RF Coils Array Coil Volume Coils

Noll RF Coils Typical CoilsTypical Coils –Head Coil (volume birdcage coil) –Body Coil (volume, mostly transmit only) –Surface Coils (e.g. occipital coils for visual system studies, mostly receive only) Improved SNR vs. volume coilsImproved SNR vs. volume coils –Head Array (array of surface coils, receive only) The idea here is that one can receive the benefits of surface coils, but over a larger volume.The idea here is that one can receive the benefits of surface coils, but over a larger volume. Also allows use of parallel MRI technology.Also allows use of parallel MRI technology.

Noll RF Fields Safety issues:Safety issues: –RF heating of the body (FDA limits our power deposition) –Focal RF heating around metals (glasses, jewelry, implants, some tattoos, etc.) can lead to burns –Focal heating near implanted devices e.g. cochlear implants, neurostimulators, pacemakers, and poorly designed/manufactured stents, screws, pins, aneurysm clips, etc.)

Noll Gradient Fields High power amplifiers generate magnetic fields in the same direction as B 0, but with variations along x, y and z.High power amplifiers generate magnetic fields in the same direction as B 0, but with variations along x, y and z.

Noll Gradients x-gradient (G x ) y-gradient (G y ) z-gradient (G z )

Noll Gradient Coil Designs Z Gradient X,Y Gradient

Noll Gradient Fields Safety issues:Safety issues: –Rapid changes of fields can lead to peripheral nerve stimulation (e.g. like a twitching in the back). FDA guidelines state that stimulation cannot be “painful.” –The gradients are also responsible for the substantial acoustic noise. FDA says it can’t exceed 99 dBA with hearing protection in place.

Noll Additional Slides

Noll Typical fMRI Protocol Scout AcquisitionScout Acquisition –Where am I? T1-weighted imagesT1-weighted images –To align structure to fMRI maps T2-weighted imagesT2-weighted images –To rule out pathology fMRI – T2*-weighted, dynamicfMRI – T2*-weighted, dynamic –EPI or spiral High-resolution 3D T1-weighted imagesHigh-resolution 3D T1-weighted images –For segmentation of anatomy