Radiation Sources in medicine diagnostic Radiology

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

X Rays Medical Physics Notes.
Introduction to x-rays
Quality Control in Diagnostic Radiology
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Radiation Sources in medicine diagnostic Radiology
Control of Scatter Radiation Beam Restricting Devices
Resident Physics Lectures
MAMMO QC – covered in week 8
Analyzing the Image Density. Density Overall blackening of the image.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 5 Advanced X ray machines &Mammogram.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
BMME 560 & BME 590I Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Imaging Part 3.
Part No...., Module No....Lesson No
Radiation Exposure and Risk John Hamaty. Evaluation of Radiation Exposure Levels in Cine Cardiac Catheterization Laboratories American Association of.
Robert L. Metzger, Ph.D., C.H.P. Roland Wong, Sc.M., D.A.B.M.P.
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
Factors affecting the X-Ray output
X-Ray Production & Emission
X-Ray Production & Emission
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
Mammography Regulatory Issues Bruce Matkovich Radiation Safety Section Michigan Dept. Of Consumer & Industry Services.
Why Digital Mammography. Why Cogdell Memorial Hospital?  Cogdell Memorial Hospital delivers the highest quality patient care available –Dedicated staff.
INSTRUMENTATION. ELECTRICAL REQUIREMENTS AND EFFICIENCY  Effective voltage  High frequency inverter technology (over 10,000Hz)  Constant potential.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography.
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
Current UK legislation and guidelines for radiation protection of patients and staff Prepared by: Dr D. Mladenova.
FLUROSCOPY. What is Fluoroscopy? Fluoroscopy is a method of using low intensity X-ray beams to continuously visualize the area of interest in "real.
Quality Assurance, Quality Control & Equipment Bushong Ch. 19 & 20
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
IAEA International Atomic Energy Agency Interventional Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(3)
Resident Physics Lectures
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
IAEA International Atomic Energy Agency Dental Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 2(2)
Equipment Design for Radiation Protection
Mammography Questions and Answers
Characteristic Radiation in Tungsten Targets Shel l # of electrons Binding energy L Char M Char N Char O Char P Char Eff X-ray Energy.
Factors affecting CT image RAD
Exposure Factors or Prime Factors
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
THE EFFECT OF NON-IDEAL DETECTORS ON ENERGY WEIGHTED SPECTRA USED IN X-RAY MEDICAL IMAGING George D. Patatoukas 1, Panagiotis F. Liaparinos 1, Anastasios.
Historical Review and Modalities in Diagnostics
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.07: Optimization of protection in Mammography.
Radiation Sources in medicine diagnostic Radiology
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Rad T 275 Mammography. Types of Mammography Diagnostic –Patients with symptoms or risk factors Screening –Asymptomatic patients.
RAD 254 Chapter 19 Mammography
Part No...., Module No....Lesson No
Radiation Sources in medicine diagnostic Radiology
Trends in Patients’ Exposure Doses during Radiographic Examination in Japan Shoichi Suzuki 1, Yasuki Asada 1, Masanao Kobayashi 2, Kenichi Kobayashi 2,
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Mammography. Introduction and History Breast cancer is 2nd only to lung cancer as cause of death in women –Very treatable with early detection! 1st innovation.
Basis of Mammography F. Milano Dept. Clinical Physiopathology University of Florence
IAEA International Atomic Energy Agency Fluoroscopy Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(2)
Date of download: 7/5/2016 From: Radiation-Induced Breast Cancer Incidence and Mortality From Digital Mammography Screening: A Modeling Study Ann Intern.
Exposure Factors or Prime Factors
RAD 354 Chapt. 19 Mammo Also know as soft tissue radiography
Moayyad E Suleiman Prof. Patrick C Brennan Dr. Peter Kench
Chapter 6: Mammography Systems
Mammography Mammography is a specific type of imaging that uses a low-dose x-ray system for examination of the breasts. Common uses of the procedure: Diagnostic.
Optimisation of Patient Protection for Radiography
Part No...., Module No....Lesson No
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
Stacy Kopso, M.Ed., R.T.(R)(M)
Resident Physics Lectures (Year 1)
X Rays Medical Physics Notes.
Resident Physics Lectures (Year 1)
Presentation transcript:

Radiation Sources in medicine diagnostic Radiology Mammography Day 7 – Lecture 2(1)

Objective To become familiar with mammography x-ray systems. To become familiar with specific radiation risks associated with this equipment.

Contents Description and physical characteristics of mammography systems. Equipment malfunction affecting radiation protection.

Mammography Mammography is presently the most reliable method for detecting lesions in the breast. It: requires high standards of image quality and equipment performance because the contrast between normal and pathological areas in the breast is extremely low; is performed on symptomatic (medically referred) patients as well as on asymptomatic women who satisfy selection criteria for approved breast cancer screening programmes. Such programmes are common in many countries.

Specific requirements Mammography shall be carried out using dedicated, special purpose x-ray equipment with: generators capable of relatively low x-ray tube potentials: e.g. 25-30 kV peak; x-ray tubes with a molybdenum or rhodium target (anode) and Mo or Rh filtration. In modern mammography units different anode / filter combinations are available; the use of an anti-scatter grid and automatic exposure control (AEC) system are strongly recommended.

Specific requirements (cont) Radiolucent breast compression device - the application of firm compression to the breast during mammography provides immobilisation, reduces tissue thickness and ensures greater uniformity in thickness. Compression contributes to improved image quality by minimizing blurring and by reducing both the exposure required and the intensity of scattered radiation. A standard breast phantom approximating an average breast (designed to standard specifications) for equipment performance checks and estimation of the mean glandular dose (MGD).

Mammographic Equipment Special (dedicated) equipment for mammography IMAGE RECEPTOR COMPRESSION PLATE X-RAY TUBE ASSEMBLY OPERATOR’S PROTECTIVE SCREEN

Malfunctions affecting radiation protection Basically the same as for general x-ray systems (see previous lectures) but tests performed and measuring instruments used must be adapted to the characteristics of mammography systems, e.g. inaccuracy and inconsistency of the x-ray tube voltage and radiation output; misalignment between the x-ray beam and the image receptor, non-uniformity of the x-ray field; unsatisfactory film storage conditions, image development and viewing conditions improperly calibrated AEC, etc.