Strain Gages Electrical resistance in material changes when the material is deformed R – Resistance ρ – Resistivity l – Length A – Cross-sectional area.

Slides:



Advertisements
Similar presentations
FORCE MEASUREMENT.
Advertisements

EE 4BD4 Lecture 15 Strain ( force) Gauges, pressure sensors and Load Cells 1.
BASIC SENSORS AND PRINCIPLES. ① Strain gages Measurement of extremely small displacement ① Potentiometers Translational and Rotational displacement ②.
Lecture 4: Signal Conditioning
Lecture 2: Pressure Measurements
Variable Capacitance Transducers The Capacitance of a two plate capacitor is given by A – Overlapping Area x – Gap width k – Dielectric constant Permitivity.
Piezoresistive sensors
Measurement Lab 19 Feb 2003 Note: this material may be copyright protected and may only be used for personal use. 0 Force - I Calibration Force Measurement.
Bike3.ppt1 H167 Hands-on Lab LAB 4: Stress and Strain.
FORCE-STRAIN-STRESS measurements.
Instrumentation (AMME2700) 1 Instrumentation Dr. Xiaofeng Wu.
Ken Youssefi Mechanical & Aerospace Engr., SJSU Discontinuity in Cross Section Stepped shafts Discontinuity.
1 Chapter 27 Current and Resistance. 2 Electric Current Electric current is the rate of flow of charge through some region of space The SI unit of current.
ME 411/511Prof. Sailor Strain Measurements Module goals… –Introduce students to operating principles behind strain gauges –Discuss practical issues regarding.
Self-Induction Transducers ~ AC Supply v ref Inductance Measuring Circuit x (Measurand) Ferromagnetic Target Object Coil is activated by the supply and.
Beams Beams: Comparison with trusses, plates t
SENSORS AND TRANSDUCERS
1 ENGR 512 Experimental Methods in Engineering Spring 2011 Dr. Mustafa Arafa Mechanical Engineering Department
Chang Liu Micro Actuators, Sensors, Systems Group
Homework # Lim Myeong-Jun. Index (Axid) Stress δ & ε Curve Cantilever Strain gage Gage factor Problem solution.
Differential Transformer
Classification of Instruments :
1 5. Strain and Pressure Sensors Piezoresistivity Applied stress gives the change in resistance  = F/A  =  x/x  R/R (stress) (strain) In the case of.
Lecture 3: Bridge Circuits
Topic 1 different attributes that characterize sensors ETEC 6405.
EKT314/4 Electronic Instrumentation
CHAPTER 8 Sensors and Transducers.
Engineering H192 - Computer Programming Gateway Engineering Education Coalition Lab 5P. 1Winter Quarter Stress and Strain Lab 5.
Instrumentation Dr. Xiaofeng Wu 1 1.
MEDICAL INSTEUMENTATIO N. 동서의료공학과 정 진 웅.
Instrumentation (AMME2700) 1 Instrumentation Dr. Xiaofeng Wu.
Medical Instrumentation II. 1. (Axial) Stress On the surface, the (average) force per unit area is denoted as σ : “Stress” (cf. pressure) M y x A Medical.
EKT 451/4 SEM 2 Chapter 6 SENSOR & TRANSDUCER SHAIFUL NIZAM MOHYAR
9AEI It is defined as the change of position of a body with respect to a reference It may be Linear motion Rotational motion Displacement.
Dept. of Biomedical Engineering YOUNHO HONG RESISTIVE SENSORS.
ABE425 Engineering Measurement Systems Strain Gages, Bridges and Load Cells Dr. Tony E. Grift Dept. of Agricultural & Biological Engineering University.
Lecture 3: Bridge Circuits
INTRODUCTION Strain gauge measurement is a point strain measurement method developed in the USA c.a independently by E. Simmons and A. Ruge. It.
BASIC MECHANICAL SENSORS AND SENSOR PRINCIPLES. Definitions Transducer: a device that converts one form of energy into another. Sensor: a device that.
Force Measurement Force is a quantity capable of changing the size, shape, or motion of an object. It is a vector quantity and, as such, it has both direction.
Medical Instrumentation Ⅱ 동서의료공학과 이건우 -Sensor-
Types Of Transducers Resistive Position Transducer: The principle of the resistive position transducer is that the physical variable under measurement.
WHEATSTONE BRIDGE OVERVIEW APPLIED APPROACH! Wheatstone Bridge Overview Learning Module.
Signal Conditioning Elements (SCE). 6/13/2016Measurement & Transducers2 1. Voltage dividers Example :Potentiometer circuit.
HALL EFFECT TRANSDUCERS As already explained in- Art page 562, when a conductor is kept perpendicular to the magnetic field and a direct current.
Additional information on Passive Probes (10x) Take note that the scope’s input capacitance, C in must falls within the probe’s compensation range For.
EKT 451 CHAPTER 6 Sensor & Transducers.
1 Transducers EKT 451 CHAPTER 6. 2 Definition of a Transducers   Transducer is any device that converts energy in one form to energy in another.  
حسگرها (دريابه)و مبدل ها Sensors & Transducer. حسگر يا مبدل هر دو از اجزاء وسايل اندازه گيری هستند هر دو کميتی فيزيکی را به شکل قابل استفاده ای تبديل.
MECH 373 Instrumentation and Measurements
Semiconductor Strain Gages
MECH 373 Instrumentation and Measurements
TRANSDUCERS PRESENTATION BY: Dr.Mohammed Abdulrazzaq
MECH 373 Instrumentation and Measurements
Figure 8.1 A simple strain gage.
(2) Bridge Circuits.
MECH 373 Instrumentation and Measurements
5. Strain and Pressure Sensors
Variable Capacitance Transducers
Measurements 17GN1001 Measurement of FORCE and STRAIN
MECH 373 Instrumentation and Measurements
Measurement of Pressure
ME 322: Instrumentation Lecture 7
MECH 373 Instrumentation and Measurements
CHAPTER 8 Sensors and Transducers.
Types of Resistance Strain Gauge
ME 311 Mechanical Measurements Instrumentation II 5/6/03
Prepared by Dr. Mohamed Ahmed Awad
MECH 373 Instrumentation and Measurements
Measurement of Motion.
Presentation transcript:

Strain Gages Electrical resistance in material changes when the material is deformed R – Resistance ρ – Resistivity l – Length A – Cross-sectional area Taking the differential Change in resistance is from change in shape as well as change in resistivity For linear deformations – strain Ss – sensitivity or gage factor (2-6 for metals and 40 – 200 for semiconductor)

Strain gage accelerometer The change in resistance is measured using an electrical circuit Many variables can be measured – displacement, acceleration, pressure, temperature, liquid level, stress, force and torque Some variables (stress, force, torque) can be determined by measuring the strain directly Other variables can be measured by converting the measurand into stress using a front-end device Output vo Direction of Sensitivity (Acceleration) Strain Gage Housing Seismic Mass m Base Mounting Threads Strain Member Cantilever Strain gage accelerometer

Semiconductor (silicon with impurity) Strain gages are manufactured as metallic foil (copper-nickel alloy – constantan) Direction of Sensitivity Foil Grid Single Element Two-Element Rosette Backing Film Solder Tabs (For Leads) Three-Element Rosettes Semiconductor (silicon with impurity) Doped Silicon Crystal (P or N Type) Welded Gold Leads Nickle-Plated Copper Ribbons Phenolic Glass Backing Plate

Potentiometer or Ballast Circuit + v Strain Gage o Output v R ref (Supply) - Rc Ambient temperature changes will introduce error Variations in supply voltage will affect the output Electrical loading effect will be significant Change in voltage due to strain is a very small percentage of the output Question: Show that errors due to ambient temperature changes will cancel if the temperature coefficients of R and Rc are the same

Wheatstone Bridge Circuit Small i + R1 vo R2 RL Load (High) R4 R3 - B - + vref (Constant Voltage) When the bridge is balanced True for any RL

Direct Measurement of Output Voltage Null Balance Method When the stain gage in the bridge deforms, the balance is upset. Balance is restored by changing a variable resistor The amount of change corresponds to the change in stain Time consuming – servo balancing can be used Direct Measurement of Output Voltage Measure the output voltage resulting from the imbalance Determine the calibration constant Bridge sensitivity To compensate for temperature changes, temperature coefficients of adjacent pairs should be the same

The Bridge Constant More than one resistor in the bridge can be active If all four resistors are active, best sensitivity can be obtained R1 and R4 in tension and R2 and R3 in compression gives the largest sensitivity The bridge sensitivity can be expressed as Bridge Constant

Example 4.4 A strain gage load cell (force sensor) consists of four identical strain gages, forming a Wheatstone bridge, that are mounted on a rod that has square cross-section.  One opposite pair of strain gages is mounted axially and the other pair is mounted in the transverse direction, as shown below.  To maximize the bridge sensitivity, the strain gages are connected to the bridge as shown.  Determine the bridge constant k in terms of Poisson’s ratio v of the rod material. Axial Gage 1 2 1 + vo 3 2 Transverse Gage − Cross Section Of Sensing Member 3 4 4 − + vref Transverse strain = (-v) x longitudinal strain

Calibration Constant k – Bridge Constant Ss – Sensitivity or gage factor

Example 4.5 A schematic diagram of a strain gage accelerometer is shown below.  A point mass of weight W is used as the acceleration sensing element, and a light cantilever with rectangular cross-section, mounted inside the accelerometer casing, converts the inertia force of the mass into a strain.  The maximum bending strain at the root of the cantilever is measured using four identical active semiconductor strain gages. Two of the strain gages (A and B) are mounted axially on the top surface of the cantilever, and the remaining two (C and D) are mounted on the bottom surface. In order to maximize the sensitivity of the accelerometer, indicate the manner in which the four strain gages A, B, C, and D should be connected to a Wheatstone bridge circuit.  What is the bridge constant of the resulting circuit? A C Strain Gages A, B + W δvo − C, D D B l b A B + C − h vref D

Obtain an expression relating applied acceleration a (in units of g) to bridge output (bridge balanced at zero acceleration) in terms of the following parameters: W = Mg = weight of the seismic mass at the free end of the cantilever element E = Young’s modulus of the cantilever l = length of the cantilever b = cross-section width of the cantilever h = cross-section height of the cantilever Ss = gage factor (sensitivity) of each strain gage vref = supply voltage to the bridge. If M = 5 gm, E = 5x1010 N/m2, l = 1 cm, b = 1 mm, h = 0.5 mm, Ss = 200, and vref = 20 V, determine the sensitivity of the accelerometer in mV/g. If the yield strength of the cantilever element is 5xl07 N/m2, what is the maximum acceleration that could be measured using the accelerometer? If the ADC which reads the strain signal into a process computer has the range 0 to 10 V, how much amplification (bridge amplifier gain) would be needed at the bridge output so that this maximum acceleration corresponds to the upper limit of the ADC (10 V)? Is the cross-sensitivity (i.e., the sensitivity in the two directions orthogonal to the direction of sensitivity small with this arrangement? Explain. Hint: For a cantilever subjected to force F at the free end, the maximum stress at the root is given by

MEMS Accelerometer Signal Conditioning Mechanical Structure Applications: Airbag Deployment

Data Acquisition Supply frequency ~ 1kHz Bridge Calibration Constant Oscillator Power Supply Amplifier Demodulator And Filter Dynamic Strain Reading Supply frequency ~ 1kHz Output Voltage ~ few micro volts – 1 mV Advantages – Stability (less drift), low power consumption Foil gages - 50Ω – kΩ Power consumption decreases with resistance Resolutions on the order of 1 m/m

Semiconductor Strain Gages Ribbons Single Crystal of Semiconductor Gold Leads Phenolic Glass Backing Plate Gage factor – 40 – 200 Resitivity is higher – reduced power consumption Resistance – 5kΩ Smaller and lighter

Temperature Coefficient of Resistance (10-6/C) Properties of common strain gage material Material Composition Gage Factor (Sensitivity) Temperature Coefficient of Resistance (10-6/C) Constantan 45% Ni, 55% Cu 2.0 15 Isoelastic 36% Ni, 52% Fe, 8% Cr, 4% (Mn, Si, Mo) 3.5 200 Karma 74% Ni, 20% Cr, 3% Fe, 3% Al 2.3 20 Monel 67% Ni, 33% Cu 1.9 2000 Silicon p-type 100 to 170 70 to 700 n-type -140 to –100

Disadvantages of Semiconductor Strain Gages The strain-resistance relationship is nonlinear They are brittle and difficult to mount on curved surfaces. The maximum strain that can be measured is an order of magnitude smaller 0.003 m/m (typically, less than 0.01 m/m) They are more costly They have a much larger temperature sensitivity. Resistance Change Resistance Change P-type N-type 0.4 0.4 0.3  = 1 Microstrain = Strain of 1×10-6 0.3 0.2 0.2 0.1 0.1 −3 −2 −1 1 2 3 ×103  −3 −2 −1 1 2 3 ×103  −0.1 −0.1 Strain Strain −0.2 −0.2 −0.3 −0.3

For semiconductor strain gages S1 – linear sensitivity Positive for p-type gages Negative for n-type gages Magnitude is larger for p-type S2 – nonlinearity Positive for both types Magnitude is smaller for p-type

Linear Approximation Error Quadratic Error Minimize Error = 0 Curve Change in Resistance Linear Approximation −max Error max  Strain Quadratic Error Minimize Error = 0 Maximum Error

Range – change in resistance Percentage nonlinearity error

Temperature Compensation α = Temperature Coefficient of Resistance β = Temperature Coefficient of Gage Factor 3 α Compensation Feasible Temperature coefficients (per °F) 2 (−β) Compensation Not Feasible Compensation Feasible 1 Concentration of Trace Material (Atoms/cc) Resistance change due to temperature Sensitivity change due to temperature

Self Compensation with a Resistor + δvo R R − + − vi Rc − R3 R4 + Compensating Resistor Rc − + vref vi For self compensation the output after the temperature change must be the same vref Possible only for certain ranges