1 M.G. Pia et al. The application of GEANT4 simulation code for brachytherapy treatment Maria Grazia Pia INFN Genova, Italy and CERN/IT

Slides:



Advertisements
Similar presentations
A selection of Geant4 medical physics applications
Advertisements

The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
Maria Grazia Pia, INFN Genova Test & Analysis Project Maria Grazia Pia, INFN Genova on behalf of the T&A team
1 COMPARISON BETWEEN PLATO ISODOSE DISTRIBUTION OF A 192 IR SOURCE AND THOSE SIMULATED WITH GEANT4 TOOLKIT F. Foppiano 1, S. Agostinelli 1, S. Garelli.
Precision validation of Geant4 electromagnetic physics Katsuya Amako, Susanna Guatelli, Vladimir Ivanchenko, Michel Maire, Barbara Mascialino, Koichi Murakami,
F. Foppiano, M.G. Pia, M. Piergentili Medical Linac IEEE NSS, October 2004, Rome, Italy
Maria Grazia Pia, INFN Genova Atomic Relaxation Models A. Mantero, B. Mascialino, Maria Grazia Pia INFN Genova, Italy P. Nieminen ESA/ESTEC
Maria Grazia Pia, INFN Genova 1 Part V The lesson learned Summary and conclusions.
Low Energy Electromagnetic Physics
Geant4-Genova Group Validation of Susanna Guatelli, Alfonso Mantero, Barbara Mascialino, Maria Grazia Pia, Valentina Zampichelli INFN Genova, Italy IEEE.
A general purpose dosimetric system for brachytherapy
Maria Grazia Pia, INFN Genova PhysicsLists in Geant4 Advanced Examples M.G.
20 February 2002Geant4 Users' Workhsop, SLAC1 Low-Energy Electromagnetic Processes in P. Nieminen (ESA-ESTEC)
14 User Documents and Examples II SLAC Geant4 Tutorial 17 May 2007 Dennis Wright Geant4 V8.3.
Maria Grazia Pia, INFN Genova CERN, 26 July 2004 Background of the Project.
Maria Grazia Pia, INFN Genova Low Energy Electromagnetic Physics Maria Grazia Pia INFN Genova
Maria Grazia Pia, INFN Genova – CHEP 2001 ow Energy Electromagnetic Physics ow Energy Electromagnetic Physics S. Chauvie,G. Depaola, F. Longo, V. Ivanchenko,
Geant4-INFN (Genova-LNS) Team Validation of Geant4 electromagnetic and hadronic models against proton data Validation of Geant4 electromagnetic and hadronic.
The brachytherapy advanced example Susanna Guatelli (CERN/INFN)
Maria Grazia Pia Systematic validation of Geant4 electromagnetic and hadronic models against proton data Systematic validation of Geant4 electromagnetic.
Budker Inst. of Physics IHEP Protvino MEPHI Moscow Pittsburg University.
Particle Physics Software aids Medicine Accurate geometry and material modeling in the fight against cancer Precise physics models for radiation interactions.
Workshop on Physics on Nuclei at Extremes, Tokyo Institute of Technology, Institute for Nuclear Research and Nuclear Energy Bulgarian Academy.
M.G. Pia et al. Brachytherapy at IST Results from an atypical Comparison Project Stefano Agostinelli 1,2, Franca Foppiano 1, Stefania Garelli 1, Matteo.
G. Bartesaghi, 11° ICATPP, Como, 5-9 October 2009 MONTE CARLO SIMULATIONS ON NEUTRON TRANSPORT AND ABSORBED DOSE IN TISSUE-EQUIVALENT PHANTOMS EXPOSED.
F. Foppiano, B. Mascialino, M. G. Pia, M. Piergentili Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy Monte Carlo 2005 Topical.
Maria Grazia Pia, INFN Genova Low Energy Electromagnetic Physics Maria Grazia Pia INFN Genova on behalf of the Low Energy Electromagnetic.
IEEE Nuclear Science Symposium and Medical Imaging Conference Short Course The Geant4 Simulation Toolkit Sunanda Banerjee (Saha Inst. Nucl. Phys., Kolkata,
IEEE Nuclear Science Symposium and Medical Imaging Conference Short Course The Geant4 Simulation Toolkit Sunanda Banerjee (Saha Inst. Nucl. Phys., Kolkata,
OOAD… LowE Electrons From HEP computing to medical research and vice versa Bidirectional From HEP computing to medical research and vice versa Bidirectional.
Maria Grazia Pia Simulation for LHC Radiation Background Optimisation of monitoring detectors and experimental validation Simulation for LHC Radiation.
Riccardo Capra 1, Stéphane Chauvie 2, Ziad Francis 3, Sebastien Incerti 4, Barbara Mascialino 1, Gerard Montarou 3, Philippe Moretto 4, Petteri Nieminen.
A General Purpose Brachytherapy Software Simulation + Analysis (isodose calculation) 2/10/2002 Geant4 Workshop CERN Susanna Guatelli Univ. and INFN Genova.
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Susanna Guatelli & Barbara Mascialino G.A.P. Cirrone (INFN LNS), G. Cuttone (INFN LNS), S. Donadio (INFN,Genova), S. Guatelli (INFN Genova), M. Maire (LAPP),
Technological Transfer from HEP to Medical Physics How precise Brachytherapy MonteCarlo simulations can be applied in Clinics Reality Problem: How to achieve.
IEEE Nuclear Science Symposium and Medical Imaging Conference Short Course The Geant4 Simulation Toolkit Sunanda Banerjee (Saha Inst. Nucl. Phys., Kolkata,
Precision Validation of Geant4 Electromagnetic Physics Geant4 DNA Project Meeting 26 July 2004, CERN Michela.
Precision analysis of Geant4 condensed transport effects on energy deposition in detectors M. Batič 1,2, G. Hoff 1,3, M. G. Pia 1 1 INFN Sezione di Genova,
Geant4 Training 2006 Short Course Katsuya Amako (KEK) Gabriele Cosmo (CERN) Susanna Guatelli (INFN Genova) Aatos Heikkinen (Helsinki Institute of Physics)
F. Foppiano, M.G. Pia, M. Piergentili
N 0 primary photons generated N d primary photons detected Determination of the photon mass attenuation coefficients Check on ParentID( ) Energy value.
1 Neutron Effective Dose calculation behind Concrete Shielding of Charge Particle Accelerators with Energy up to 100 MeV V. E Aleinikov, L. G. Beskrovnaja,
ESTRO, Geneva 2003 The importance of nuclear interactions for dose calculations in proton therapy M.Soukup 1, M.Fippel 2, F. Nuesslin 2 1 Department of.
Geant4 Training 2004 Short Course Katsuya Amako (KEK) Gabriele Cosmo (CERN) Giuseppe Daquino (CERN) Susanna Guatelli (INFN Genova) Aatos Heikkinen (Helsinki.
Maria Grazia Pia, INFN Genova and CERN1 Geant4 highlights of relevance for medical physics applications Maria Grazia Pia INFN Genova and CERN.
Luciano Pandola, INFN Gran Sasso Luciano Pandola INFN Gran Sasso Genova, July 18 th, 2005 Geant4 and the underground physics community.
AIDA Workshop, 3/6/2002S. Guatelli Overview of AIDA in Geant4 bio-medical applications A collection of contributions from various user groups AIDA Workshop.
Implementation of a New Monte Carlo Simulation Tool for the Development of a Proton Therapy Beam Line and Verification of the related Dose Distributions.
Validation of GEANT4 versus EGSnrc Yann PERROT LPC, CNRS/IN2P3
A Short Course on Geant4 Simulation Toolkit Introduction
MCS overview in radiation therapy
F. Foppiano3, S. Guatelli2, J. Moscicki1, M.G. Pia2 CERN1 INFN Genova2
International Workshop on radiosensitization
Geant4 REMSIM application
F. Foppiano, S. Guatelli, B. Mascialino, M. G. Pia, M. Piergentili
Gamma Ray Satellites Simulations with Geant4
Introductory Course PTB, Braunschweig, June 2009
Hadronic physics validation of Geant4
Short Course Siena, 5-6 October 2006
The Hadrontherapy Geant4 advanced example
Introductory Course ORNL, May 2008
Geant4 at IST Applications in Brachytherapy
Short Course IEEE NSS/MIC 2003 Katsuya Amako (KEK) Makoto Asai (SLAC)
Precision validation of Geant4 electromagnetic physics
G. A. P. Cirrone1, G. Cuttone1, F. Di Rosa1, S. Guatelli1, A
Radiation damage at cellular and DNA level
AN ORIGINAL MODEL FOR THE SIMULATION OF LOW ENERGY ANTIPROTONS
The Geant4 Hadrontherapy Advanced Example
Presentation transcript:

1 M.G. Pia et al. The application of GEANT4 simulation code for brachytherapy treatment Maria Grazia Pia INFN Genova, Italy and CERN/IT F. Foppiano 1, S. Agostinelli 1,2, S. Garelli 1, G. Paoli 1, P. Nieminen 3 1 National Cancer Institute and 2 Physics Dept. of Genova, 3 European Space Agency IX th International Conference on Calorimetry in High Energy Physics Annecy, 10 October 2000

2 M.G. Pia et al. Outline /Introduction to brachytherapy /Application of Monte Carlo simulation to brachytherapy /The GEANT4 toolkit and its extension targeted to medical physics /Results /Simulation of the attenuation coefficients for various materials /Simulation of a brachytherapy radioactive source / Description of 192 Ir source geometry used in real treatments / Simulation of the anisotropy function in water / Simulation of isodoses in water /Conclusions and future goals

3 M.G. Pia et al. ] In HDR endocavitary brachytherapy: a radioactive source, for example 192 Ir, is used the source moves along catheters inserted in natural cavities of the body, e.g. vagina or bronchi; this allows the deposition of the therapeutic tumor dose right where it is needed the source track is programmed by an after-loading unit What is brachytherapy? ]Brachytherapy is a medical therapy used for cancer treatments ]Radioactive sources are used to deposit therapeutic doses near tumors while preserving surrounding healthy tissues

4 M.G. Pia et al. Brachytherapy treatment set-up After-loading unit Catheter along which source moves A naso-pharynx endocavitary treatment

5 M.G. Pia et al. Brachytherapy treatment planning (1) A typical vaginal treatment plan: source moves along a single catheter

6 M.G. Pia et al. Brachytherapy treatment planning (2) A typical intra-uterine treatment plan: source moves along 3 catheters

7 M.G. Pia et al. Monte Carlo for brachytherapy Monte Carlo simulation topics for brachytherapy: oDose calculation n Computation of dose deposition kernels for treatment planning dose calculation algorithms based on convolution/superposition methods n Separation of primary, first scatter and multiple scatter components for complex dose deposition models / Computation of other model-dependent parameters, e.g. anisotropy function / Accurate computation of dose deposition in high gradient regions (i. e. near sources) oVerification of experimental calibration procedures

8 M.G. Pia et al.GEANT4 is an Object Oriented Toolkit for the simulation of the passage of particles through matter The transparency of physics Advanced functionalities in geometry, physics, visualisation etc. Extensibility to satisfy new user requirements thanks to the OO technology Adopts standards wherever available (de jure or de facto) Use of evaluated data libraries Quality Assurance based on sound software engineering Subject to independent validation by a large user community worldwide User support organization by a large international Collaboration of experts Geant4 provides various features relevant for medical applications

9 M.G. Pia et al. GEANT4 Low Energy Electromagnetic Physics Further extensions are in progress Relevant for medical, space science, astrophysics etc. applications Geant4 Low Energy Electromagnetic package extends the coverage of physics interactions 250 eV down to 250 eV for electrons and photons based on the LLNL data libraries shell effects ~ 1 keV down to ~ 1 keV for hadrons and ions Bethe-Block above 2 MeV Ziegler and ICRU parameterisations (with material dependence) free electron gas model quantal harmonic oscillator model charge dependence (Barkas effect)

1010 M.G. Pia et al. Simulation of  Simulated water  (attentuation coefficient) versus NIST data with Geant4 Standard electromagnetic package and Low Energy extension

11 M.G. Pia et al. Description of  -Selectron 192 Ir source ]GEANT4 allows complete flexible description of the real geometry ] 192 Ir energy spectrum l currently described as monochromatic at 356 keV l will soon be described by the new GEANT4 RadioactiveDecay class

1212 M.G. Pia et al. Simulation of dose deposition in water ]The simulated source is placed in a 30 cm water box ]The dose deposition is investigated in the longitudinal plane ]Plane is partitioned in 1 million 1mm 3 voxels ]A minimum of 10 millions photons are generated on the 4  solid angle Longitudinal plane partitioned in cells  -Selectron 192 Ir source

1313 M.G. Pia et al. Investigated quantities: anisotropy ]The dose deposition is not isotropic due to source geometry and auto- absorption, encapsulation and shielding effects ]Anisotropy can be described by a simple angular function which can be computed by re-sampling our simulated voxels grid calculations

1414 M.G. Pia et al. Investigated quantities: isodoses ]The simulated dose deposition data can also be used to derive isodoses

1515 M.G. Pia et al. Conclusions and future goals ]Monte Carlo simulation is useful in brachytherapy both to obtain model-dependent parameters and to verify experimental data ]GEANT4 offers reliable particle-matter Monte Carlo simulation in a flexible modern object-oriented toolkit ]We have used GEANT4 to simulate  coefficients and a commercial brachytherapy source with full dose deposition /More realistic description of 192 Ir source energy spectrum with the new GEANT4 RadioactiveDecay class /Simulation of shielded brachytherapy applicators