Measurement and Detection of Ionizing Radiation

Slides:



Advertisements
Similar presentations
General Characteristics of Gas Detectors
Advertisements

Radiation Safety Program Annual Refresher Training
Radiation Detection ionization chambers (dosimeters, pulse chambers, particle track chambers) scintillation detectors semiconductor detectors photographic.
Lab B4: The Creation and Annihilation of Antimatter SFSU Physics 490 Spring 2004 Professor Roger Bland.
Liquid Scintillation Counter And Contamination Monitoring Training Presented by: Ali Shoushtarian Office of Risk Management October 2009.
GM COUNTER.
Geiger-Muller detector and Ionization chamber
Dispersive property of a G-M tube HV - + In the proportional region a G-M tube has dispersive properties tube voltage.
Radiation Safety Course Radiation Monitoring Heath de la Giroday Dispensing Chemist Radiation Safety Officer.
Radiation Detection Systems
Geiger- Muller counter
Radiation Detectors / Particle Detectors
Scintillators. When radiation interacts with certain types of materials, it produces flashes of light (scintillation) Materials that respond this way.
Molecular Fluorescence Spectroscopy
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Chapter 18 The Nucleus: A Chemist’s View All nuclides with 84 or more protons are unstable with respect to radio active decay. Light nuclides are stable.
Introduction into LS theory and practice. Agenda  Energy deposition  Overview of the LSC process  Theory of operation  Quenching  Machine  Uses.
Computed Tomography RAD309
Nuclear Physics: Radiation, Radioactivity & its Applications.
Lesson 17 Detectors. Introduction When radiation interacts with matter, result is the production of energetic electrons. (Neutrons lead to secondary processes.
Instruments for Radiation Detection and Measurement
. 2 RCRS provides: Pick-up and delivery of meters Required 9-month calibration Preventive maintenance and repairs Shipping for extensive repairs RCRS.
Detectors The energy loss of particles in matter can be used detect and identify those particles. There are different types of “detectors”: - Gas-filled.
Lecture 11  Production of Positron Emitters, Continued  The Positron Tomograph.
Principles of Radiation Detection
Chapter 6 Principles of Radiation Detection. Measurement of Radiation X-rays and electrons produced by radiation therapy treatment machines are measured.
Instruments for Radiation Detection and Measurement Lab # 3 (1)
Radiation & Radioactivity
Type of Material: Instrumentation Review
Atomic Emission Spectroscopy
Photon detection Visible or near-visible wavelengths
1 Alpha Emissions (How a Smoke Alarm Works). 2 Radioactive Emissions (Radiation) Penetrating Power SymbolEquivalentDescriptionType He Stopped by thick.
Who is this?. Marie Curie Born 1867 Poland 1893 degree in physics, 1894 degree in mathematics Denied a place in Krakow University because she was.
1 Scintillators  One of the most widely used particle detection techniques Ionization -> Excitation -> Photons -> Electronic conversion -> Amplification.
Chapter 30: Nuclear Physics and Radioactivity. Radioactivity Radioactivity is the discentigration of an unstable nuclei. when the nuclei decays the nucleus.
FISICA AMBIENTALE 1 Radioattività: misure 1 Lezioni Marie Curie.
HPT001.xxx Rev. x Page 1 of xx TP-1 TVAN Technical Training Health Physics (RADCON) Initial Training Program ACADs (08-006) Covered Keywords Detectors,
SCINTILLATION COUNTER. PRINCIPLE When light radiations strike fluorescent material it produces flashes of light called scintillations. These are detected.
Half-life It's impossible to predict when a specific atom is going to decay, but you can predict the number of atoms that will decay in a certain time.
Instrumentation Review. Direct and Indirect Ionization Direct - Charge particles that strip away electrons from atoms Indirect - uncharged that have to.
Ionization Detectors Basic operation
Seeing the Subatomic Stephen Miller Saturday Morning Physics October 11, 2003.
3/2003 Rev 1 II.3.4 & 13-15a – slide 1 of 31 Part IIQuantities and Measurements Module 3Principles of Radiation Detection and Measurement Session 4,Liquid.
Introduction to nuclear medicine technology NMT 231 Aya Ahmed Saeed.
II. DETECTORS AND HOW THEY WORK
PHYSICS 225, 2 ND YEAR LAB NUCLEAR RADIATION DETECTORS G.F. West Thurs, Jan. 19.
19.3 Detection of radioactivity
Scintillation Counter
Detection of Radiation Contents: Geiger tubes Photo-Multiplier tubes Cloud Chambers Solid State devices.
HPT001.xxx Rev. x Page 1 of xx TP-1 TVAN Technical Training Health Physics (RADCON) Initial Training Program HPT001.xxx Rev. x Page 1 of xx TP-1 TVAN Technical.
Chapter V Radiation Detectors.
Geiger-Mueller Counters Darwin L. Boyd Kent State University School of Technology.
Nuclear Medicine Instrumentation 242 NMT 1 Dr. Abdo Mansour Assistant Professor of radiology
Radiation detectors Ion chamber 2. Geiger Muller counter (GM).
PCS352 LAB #1 Characteristics of a Geiger Counter
Sensors and Detectors - 2
Sensors and Detectors - 2
Interactions of Radiation With Matter
PAN-2013: Radiation detectors
Ionization detectors ∆
Scintillation Counter
ECE699 – 004 Sensor Device Technology
Interaction of Radiation with Matter
1 Principles of Radiation Detection Operational Radiological Safety Course U.S. Army Chemical Biological Radiological & Nuclear School Edwin R. Bradley.
Radiation Protection Service University of Glasgow Lecture 2: Detecting Radiation Detecting Ionising Radiation James Gray University RPA.
Radioactivity B. Sc. -III Feb Dr. Wagh G. S. M. Sc. M. Phil. Ph
Geiger-Mueller Tubes By: Daniel Hancock.
Detecting Ionising Radiation
Atomic Physics Radiation Contiuned.
Presentation transcript:

Measurement and Detection of Ionizing Radiation Methods of detection Electronical instruments

Ionizing radiation is invisible Many methods are available for detection and measurement, including Ionization detectors Scintillation detectors Biological methods Thermo luminescence Chemical methods – free radicals produced Measurement of heat- energy dissipated

Ionization All detecting methods are based on the interaction of the radiation with matter. If a radiation does not interact with any matter, we have no method of the detecting it. Since ionization is an important process for radioactivity, most detectors exploit the signals generated due to ions and electrons . on pairs in a gas produced by ionizing radiation do not recombine until the energies of electrons have dissipated. In a gas, ions and electrons move freely

Ionization Devices contain a gas that can be ionized A voltage is applied to the gas Specific instrumentation and types of measurement depend on amount of voltage applied to the gas. Three types of instruments: Ion chambers Proportional counters Geiger-Mueller counters

Log of electrical signal vs. voltage

The key components of an ionization chamber are shown here The key components of an ionization chamber are shown here. It consists of a detector chamber, a voltage supplier (battery), an ampere meter, and a load resister . Ionizing radiation enters the detector chamber and ionizes the mixture of gas in it. The electrons drift towards the positive electrode and ions move towards the negative electrode. Thus, ampere meter detects a current. The number of ion pairs is proportional to the number of ionizing particles entering the detector chamber. Thus, the current is proportional to the intensity of ionizing radiation. The light electrons drift 100,000 times faster than the heavy ions. The motion of electrons is mostly responsible for the current. Ionization Chamber

Radiation ionizes the gas Radiation ionizes the gas. Ions move toward electrodes, creating current. http://www.science.uwaterloo.ca/~cchieh/cact/nuctek/interactdetector.html

Ion chamber continued Voltage is high enough that ions reach the electrodes, produce current. Proportional to energy: the more energy, the more current. Generally requires some amplification of the signal. Example of use: pocket dosimeters http://www.ludlums.com/images/dosimeter.jpg

Proportional Counters At some hundreds volts, the improvement in sensitivity is more than collecting all the When voltages applied to electrodes of ionization chambers increase, the sensitivities increase. Electrons and ions on the electrode. The currents corresponding to multiples of ions and electrons produced by radioactivity. To distinct them from simple ionization chambers, these detectors are called proportional counters. In proportional counters, the high voltage applied to the electrodes created a strong electric field, which accelerate electrons. The electrons, after having acquired the energy, ionize other molecules. Production of secondary ion pairs initiates an avalanche of ionization by every primary electron generated by radiation. Such a process is called gas multiplication. The gas multiplication makes the detection much more sensitive. Yet, the current is still proportional to the number of primary ion pairs. When voltages applied to proportional counters get still higher, sparks jump (arcs) between the two electrodes along the tracks of ionizing particles. These detectors are called spark chambers, which give internal amplification factors up to 1,000,000 times while still giving an initial signal proportional to the number of primary ion pairs.

Proportional counters -+ -+-+ -+-+-+-+ -+-+-+-+ -+-+-+-+ -+-+-+-+-+-+-+-+ -+-+-+-+-+-+-+-+ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Each ionization electron is accelerated by the voltage so that it ionizes more of the gas. The higher the energy of the radiation event, the greater the avalanche, the higher the current Each ionization event detected separately.

Geiger Mueller counters http://www.pchemlabs.com/images/eberline-rm20-geiger-counter-a.JPG

How Geiger counters work Voltage is high enough that every radiation event triggers a complete avalanche of ionized gas Does not discriminate among different energy levels Each event is registered A quenching agent stops the reaction, resets gas for next event Slow response time (comparatively) but simpler circuitry. Good for simple, sturdy, instruments Best for gamma; low efficiency for alpha, beta.

More Geiger details Higher voltage leads to constant avalanches; instrument “pegs”. Improved efficiency with pancake probe: collects more radiation due to geometry.

Proper use of Geiger counters as “survey meters” http://www.orau.gov/reacts/measure.htm Units of radioactivity and radiation Radiation detection instruments and methods First check battery and check source Enclosed radioactive material of known amount Check level of background radiation Survey area in question Move survey instrument slowly Keep constant distance from object being surveyed; do not make contact.

Solid scintillation counters Crystal-based Radiation hits crystal which releases visible photons Photons amplified by photomultiplier tube, converts to electrical signal Zinc sulfide Good detection of alpha particles, rapid response time Sodium iodide w/ thallium Good for detection of gamma

http://www. fnrf. science. cmu. ac http://www.fnrf.science.cmu.ac.th/theory/radiation/Radiation%20and%20Radioactivity_files/image018.gif

Liquid Scintillation counters Workhorse in biology labs for many years Very useful for beta emitters, some alpha Modern equipment: Computer driven http://www.gmi-inc.com/Genlab/Wallac%201414%20LS.jpg

Basic principles Radioactive sample is mixed with organic solvents (cocktail) Toluene replaced with biodegradable solvents Detergents allow up to 5% aqueous samples Radiation hits solvent, energy is absorbed by solvent; Energy passed to one or more fluors Fluor emits visible light which is detected By fluorescence Amplified by photomultiplier, converted to electrical signal.

Coincidence circuitry Photomultipliers very sensitive Inside of instrument completely dark Tubes give off “thermal electrons” Result would be very high background counts Coincidence circuitry compares results from 2 photomultipliers Event not detected by both: thermal electron Ignored Event detected by both is affect of beta particle Counted.

Counts and energy discrimination As radiation travels through solvent, it gives up energy The more energy it has, the more fluor molecules get excited and release photons Thus, the higher the energy, the brighter the flash The higher the electrical pulse sent from the PMs Instruments can be electronically adjusted Discriminators set for different “pulse height” Able to count betas from H-3 vs. C-14 vs. P-32

Beta energy spectra cpm Pulse height

Summary of capabilities Pulse height From brightness of flash; the more energetic the radiation, the brighter the flash. Discriminators (“gain”) in the instrument can be set so you determine what energy you want counted. Number of pulses Corresponds to how many flashes, that is how many radiation events (decays): the amount of radioactivity.

Difficulties with LSC Static electricity: causes spurious high counts, esp. when humidity is low; don’t wipe outside of vials! Chemiluminescence: chemical reactions in sample, from overhead lights, glass. Suspiciously high counts can be redone; chemi-induced high counts subside over time. Quench Anything that interferes with counting efficiency. Measured: counts per minute (cpm) Desired: decompositions per minute (dpm)

Counting efficiency Because samples are usually dispersed in clear containers, geometry is favorable for energy transfer in all directions and good light emission Not all decay events will get registered, however, because no system is 100% efficient We seek to know the # of decompositions per minute (dpm) but measure the counts per minute (cpm). Using standards helps determine efficiency.

Effect of Quench

All about quench Chemical quench Acids, bases, high salt, any chemical that interferes with transfer of energy from the solvent to the fluor. Result: fewer activated fluor molecules, less intense flash, interpreted as a lower energy event. Color quench Colored material absorbs visible light from fluor Less intense flash, appears as lower energy event

About quench -2 Self absorption If particulate matter not well suspended, energy not absorbed by fluor, not detected as well. Both lowering of cpm and forcing into lower energy range.

Counting statistics Radioactive decay is a random event To be sure results are reliable, a minimum number of decay events must be recorded. Reliability depends on total number of counts! Example Statistical significance is the same in these two cases; 10 minute count yielding 500 cpm 1 minute count yielding 5000 cpm. Both have total of 5000 counts Instruments have settings for stopping count when a certain statistical threshold is reached.