Basic quality control tests in medical radiography

Slides:



Advertisements
Similar presentations
PROCESSOR QA.
Advertisements

Darkroom Film processor
Technique Guidance Systems
X-ray films are generally mad of an emulsion-gelatin containing radiation sensitive silver halide. with a flexible, blue tented base X-ray films are sensitive.
 The panoramic unit is often used to cut down on the amount of radiation the patient is exposed too pared to a CMX. This technique is also utilized to.
Lecture (4). Step by Step Guide to a Typical Radiographic Examination Request Form Check form is fully completed and signed. Check pregnancy question.
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
Quality Control in Diagnostic Radiology
Veterinary Radiation Safety 4 Students Radiation Safety MA Law requires that persons who perform diagnostic radiology procedures on animal patients be.
Knee & Patella Radiography
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
Easi-Sterilise Standard Operating Procedures Section 2 Packaging and Wrapping Procedures.
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
MAMMO QC – covered in week 8
Chapter 18 Foot Radiography
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
CVT 102.  Time  Distance  Shielding.
Technical Artifacts and Errors
Essentials of Dental Radiography for Dental Assistants and Hygienists CHAPTER Copyright ©2012 by Pearson Education, Inc. All rights reserved. Essentials.
Radiography Introduction. Objectives To describe  Properties of x-rays  Production of x-rays  Formation of radiographic image  Components of an x-ray.
TIMERS.
Chapter 11 The X-ray Darkroom.
Lecture 4 Beam restricting Devices, their Use and Maintenance
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
Darkroom Fog and Viewbox Illumination Check
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
The X-ray Imaging System
Conventional and Computed Tomography
THE X-RAY UNIT THE BASICS WHAT FOUR COMPONENTS ARE IN TYPICAL RADIOGRAPHIC ROOM? 1. X-RAY TUBE 2. OPERATING CONSOLE 3. HIGH VOLTAGE GENERATOR 4. X-RAY.
Quality Control Rad T 110.
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.
Control of Scatter Radiation
Radiography Testing 1 Radiography Testing (RT), or industrial radiography is a nondestructive testing (NDT) method of inspecting materials for hidden flaws.
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
CHAPTER 3 EQUIPMENT OPERATION AND QUALITY CONTROL
Yr 11 Photography. Topics – Part 1 Equipment and chemicals Safety Techniques –Processing film –Creating proof sheet.
Resident Physics Lectures
Equipment Operation and Maintenance. Grid-controlled tube Designed to turn on and off rapidly, providing short precise exposures Use in fluoro and in.
Shoulder 1. Intended Learning Outcomes The student should be able to recognize clinical radiographic technical principles of the shoulder.
RADIOLOGIC IMAGING EQUIPMENT 1. THE X-RAY ROOM 2.
Week 1 Review Don’t forget: You can copy- paste this slide into other presentations, and move or resize the poll.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. Chapter 23 Extraoral Imaging.
Radiographic Equipment
Designing for Radiation Protection. Design Standards for Radiation Protection – Leakage Radiation.
Week 2: Radiographic Equipment
Alhanouf Alshedi X-ray film basic structure 2 ed Lecture.
Quality Assurance in the Dental Office
Part No...., Module No....Lesson No
Analog Imaging III By Professor Stelmark.
Part No...., Module No....Lesson No
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
DARK ROOM TECHNIQUES-PROCESSING
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. 1 Extraoral Imaging.
Week 2: Radiographic Equipment
Quality Assurance in Dental Radiography
Stacy Kopso, M.Ed.,RT(R)(M). Quality Control (QC)-A program that specifically addresses the safe and reliable operation of equipment Required by the Joint.
Radiography Dentalelle Tutoring.
THE X-RAY UNIT THE BASICS.
X-Ray Generators and Ancillary Radiographic Equipment
THE X-RAY UNIT THE BASICS.
Factors Effecting the Production
Quality Control Testing of Screen Speed
Artifacts and Errors.
Presentation transcript:

Basic quality control tests in medical radiography Finals Topics

-DARKROOM QUALITY CONTROL -QUALITY CONTROL FOR X-RAY EQUIPMENT AND ACCESSORIES

DARKROOM QUALITY CONTROL Films Determination of the Optimum Developing Time Check for the consistency of the generator output Check for mAs linearity Check for Timer Accuracy

QUALITY CONTROL FOR X-RAY EQUIPMENT AND ACCESSORIES Collimator Check Check for Alignment of the X-ray Tube Column and the X-ray Table Test for Grids Test for Bucky Movement Test for Lead Rubber Aprons and Gloves

FILMS Checking for: DARKROOM QC - Expiry date * “FIFO” method should be observed - Storage conditions * temperature at about 20-25 C * away from sources of heat and where chemical vapors may leak and fog the film * boxes of films should not be stored on the floor; should be stacked vertically to avoid pressure marks

DARKROOM QC DETERMINATION OF THE OPTIMUM DEVELOPING TIME Objective - To determine the developing time that will give the film’s maximum density without an increase in base + fog. Procedure - Expose the loaded cassette in strips (6 strips) using the lowest kVp and mAs setting at 100cm SID. The exposure factors may be adjusted until we are able to produce a density of 1.0 on the middle strip. The last strip is not exposed to radiation.

PRODUCTION OF SENSITOMETRIC STRIPS X-ray film Lead sheet 5 exposures 4 exposures 3 exposures 2 exposures 1 exposure No exposure Cut into a number of strips

Procedure DETERMINATION OF THE OPTIMUM DEVELOPING TIME DARKROOM QC - Inside the darkroom, cut the film into strips and process them using different developing times, e.g. 30s, 60s, 90s, 120s, 150s & 180s.

Evaluation DETERMINATION OF THE OPTIMUM DEVELOPING TIME DARKROOM QC DETERMINATION OF THE OPTIMUM DEVELOPING TIME Evaluation - Examine the film strips, look at the middle step and determine which strip exhibit the maximum density at an acceptable basic fog. The time used to develop this strip is considered the optimum developing time. 1 (30s) 2 (60s) 3 (90s) 4 (120s) 5 (150s) 6 (180s)

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR THE CONSISTENCY OF THE GENERATOR OUTPUT Objective - To determine whether the generator produces the same density every time the same exposure factors are selected. Procedure - Position the water phantom on the loaded cassette and expose the film with set factors. Process the film the usual way.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR THE CONSISTENCY OF THE GENERATOR OUTPUT 10cm H2O 10cm H2O 10cm H2O 10cm H2O Exposure 1 Exposure 2 Exposure 1 Exposure 2 A. Consistent generator output B. Inconsistent generator output

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR mAs LINEARITY Objective - To determine any differences in photographic density when different mA settings are used for equal mAs value. Procedure - Position the water phantom on the loaded cassette and expose the film with set factors. Process the film the usual way.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES Good Linearity CHECK FOR mAs LINEARITY 10cm H2O 10cm H2O 10cm H2O We are going to check their receprocity Exposure 1 50mA & 0.2s Exposure 2 100mA & 0.1s Exposure 3 200mA & 0.05s

QC FOR X-RAY EQUIPMENT AND ACCESSORIES Poor Linearity CHECK FOR mAs LINEARITY 10cm H2O 10cm H2O 10cm H2O Exposure 1 50mA & 0.2s Exposure 2 100mA & 0.1s Exposure 3 200mA & 0.05s

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR TIMER ACCURACY Objective - To determine the accuracy of the x-ray timer. Procedure - Position the spinning top on the loaded cassette and expose the film using the clinically used exposure times. Process the film the usual way.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR TIMER ACCURACY

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR TIMER ACCURACY For mains supply with 60Hz frequency: One-pulse (self rectified) unit Set time (sec) No. of dots 0.05 3 0.10 6 0.50 30 1.00 60 Two-pulse (full wave rectified) unit 0.05 6 0.10 12 0.50 60 1.00 120

QC FOR X-RAY EQUIPMENT AND ACCESSORIES COLLIMATOR CHECK Objective - To check if the radiation field is the same as the area illuminated by the light from the collimator. Procedure - Center the loaded cassette in the useful beam at a commonly used SID. Collimate the field to approximately half the film’s size and place a metal marker at the center of the illuminated area and mark the inside edges with lead wires. Make an exposure using appropriate technique factors. Make a second exposure with the collimator fully opened. Process the film the usual way.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES COLLIMATOR CHECK 1st exposure 2nd exposure

QC FOR X-RAY EQUIPMENT AND ACCESSORIES COLLIMATOR CHECK Radiation field Light field The misalignment should not exceed 2% of the SID

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR THE ALIGNMENT OF THE X-RAY TUBE COLUMN AND THE X-RAY TABLE Check for table alignment with the x-ray beam

QC FOR X-RAY EQUIPMENT AND ACCESSORIES CHECK FOR THE ALIGNMENT OF THE X-RAY TUBE COLUMN AND THE X-RAY TABLE Check whether the column is vertical - Using a plumb bob, hold the string’s end against the tube column making sure that the string hangs unobstructed. If the column is parallel to the string, the column is vertical. Check whether the table top is horizontal - Place a bowl of water on top of the table at different locations and measure the depth of water around the edge of the bowl. If the table is horizontal, the depth of the water will not vary across the bowl. Check whether the horizontal arm is still horizontal - Measure the distance between the cross-arm and the table top in front of the cross-arm and from the back. The arm is horizontal if the distance is the same from both ends.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES TEST FOR GRIDS Objective - To check whether the grid gives a uniform density over its area. Procedure - Place the loaded cassette on the cassette tray or under the loose stationary grid. Place metal markers on the cassette to indicate the position of the grid on the film. Expose the film using the recommended SID and process the film the usual way.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES TEST FOR BUCKY MOVEMENT Objective - To check whether the grid bucky is moving throughout the whole exposure. Procedure - Place the loaded cassette on the cassette tray. Place metal markers on the cassette to indicate the position of the grid on the film. Expose the film in the bucky mode and process the film the usual way. Check for presence of grid lines on the radiograph.

QC FOR X-RAY EQUIPMENT AND ACCESSORIES LEAD RUBBER APRONS AND GLOVES Care -clean with soap and water, weekly or as necessary -should not be stored near a heat source -aprons should be hung up using hangers; should not be fold -gloves should be stored flat or upright on a custom holder Checking for damage -to surface covering, seams and fasteners of aprons (if any) -check for flexibility, feeling for cracks

QC FOR X-RAY EQUIPMENT AND ACCESSORIES LEAD RUBBER APRONS AND GLOVES Test to detect cracks on aprons and gloves Method -place the apron or gloves to be tested on the table top -use fluoroscopy system to examine the whole article *if radiography is to be used, placed cassettes under test articles and make separate exposures, ensuring that the whole article is examined Evaluation -Look for cracks or other abnormal variations of density in the image