Parts of the Anesthetic Machine

Slides:



Advertisements
Similar presentations
Getting Started with the EELD601 After you remove your EELD601 from the box, check to make sure you have all accessories shown below. Click here for next.
Advertisements

1 of Emission Control Theory Support Automotive – Engine Performance Topics covered in this presentation: Types of Emissions Emission Control Devices.
ANESTHETIC MACHINES.
Safety Requirements of the Anesthesia Workstation
Anesthesia Machine Presented by Gil Soto C.R.N.A.
INTERMEDIATE PRESSURE SYSTEM PRESENTED BY: Dr CHITTRA MODERATED BY: Dr JYOTI PATHANIA.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 1 of 4.
Use of the Inhalant Anesthetic Machines
Harkness career center
1 FUEL SYSTEM SERVICE ENGINE CLEAN FUEL TOOL MODEL RX 777.
Oxygen Administration R/C MERT Medic Assistant Lesson V.
Anesthesia Apparatus Checkout Recommendations
Oxy – Fuel Torch Ag Mechanics.
Program: Respiratory Therapy Course: Introduction to Respiratory Sciences Lesson: Gas Physics Handle Gas Cylinders With Care.
Oxyacetylene Torches.
L. P. Gas Changing Tanks. Introduction L.P. gases belong to a family of chemical compounds known as hydrocarbons. This means they are made up of hydrogen.
Check Valves and Manufacturer-Recommended Leak Test
Trigger/Limit/Cycle/Baseline
OXYFUEL WELDING. INTRODUCTION Oxyfuel welding is a welding process where the heat for fusion is supplied by a torch using oxygen and a fuel gas. Several.
Oxy-fuel Welding and Cutting By: John Templeton
SAFETY AND PROCEEDURES
HVACR312 - Refrigeration Metering Devices.
Terms and Definitions Base Metal: Metal to be welded Fusion Welding: process that joins metals by heating them to a melting point and allowing them to.
Scavenging System  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected to a.
Nitric Oxide Set-up for Patients on Ventilator
1 © 2013 Cengage Learning. All Rights Reserved. May not be copied, scanned, or duplicated, in whole or in part, except for use as permitted in a license.
Anesthetic Equipment.
MEDICAL GASES Manufacture, Storage, Transport + Delivery
Aircraft Systems Oxygen Systems A-4N/TA-4. Aircraft Systems Oxygen System LESSON OBJECTIVES To give the TP information on the Oxygen System to include.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
PART 3: Breathing Circuit
The Anesthesia Gas Machine
AccuMax Multi-Point Injection Mechanics
Anesthesia Machine VAPORIZERS. Vaporizers Convert liquid anesthetic into a volatile inhalation agent Based on laws of physics You must memorize the chemical.
A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state.
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics PhD(physiology) Mahatma gandhi medical college and research institute,
Inhaled anesthetic Delivery Systems
Scavenging System A&A Pg 356  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected.
Series Operational Training. 2 Water Tank Water Pump Multiple Discharges 2000 Series Line Strainer Foam Tank Electronic Direct Injection Proportioner.
+ Surgical Procedures 7.01 Implement techniques to prepare and monitor patients for surgery.
Non-Rebreathing System  Used for patients weighing < 7 kg  Attach the clear, plastic hose to the fresh gas inlet (this tube comes from the vaporizer.
CENTRAL GAS SYSTEMS Without reference, identify principles relating to central gas systems as they pertain to facility and equipment interface with at.
Introductory Lecture Series: The Anesthesia Machine
Placing Vapor Distribution Systems and Appliances into Operation MODULE 8 System Tests.
Gasoline fuel supply systems Function: store, transport, Pressurize and filter gasoline going to carburetor or fuel injection system.
EVAPORATIVE EMISSION CONTROL SYSTEMS
Gas Welding.
Without reference, identify principles about Anesthesia Units with at least 70 percent accuracy.
ANAESTHESIA BREATHING CIRCUITS
PART 3: Breathing Circuit
Anesthetic Problems and Emergencies A&A pg Why Do Problems Arise?  Human error  Equipment error  Adverse effects  Patient factors  Anesthetic.
VNTE Prep Anesthetic Review.
Prepared by Dr. Irene Roco
Parts of the Anesthetic Machine A&A Pages
Copyright © 2010 Delmar, Cengage Learning. ALL RIGHTS RESERVED. Unit 23 Using Gas Welding Equipment.
PART 3: Breathing Circuit
STEP 1 Locate the transmission cooler line connection on the vehicle. Identify the correct adapters to be used by referring to the vehicle application.
Introductory Lecture Series: The Anesthesia Machine
BREATHING SYSTEMS.
Veterinary Anesthesia
PART 3: Breathing System Parts
Anesthesia Equipment Session 3
Non-Rebreathing System
Parts of the Anesthetic Machine Parts 1 & 2: A&A pages
Refrigeration Basics By: Mohamed Iqbal Pallipurath.
Anesthetic Gas Machines
Non-Rebreathing System
Sophie Young 21st February 2007
By: JAGDEEP SANGWAN Refrigeration Basics 101.
Presentation transcript:

Parts of the Anesthetic Machine

Purpose of the Anesthetic Machine 2 A liquid inhalant anesthetic is vaporized into a carrier gas, and delivered to the patient via a breathing circuit. Oxygen is delivered at a controlled rate (L/min) Isoflurane is converted to a gas and given at a specific percentage Gases exhaled from the patient are scavenged OR can be re-circulated once CO2 has been removed *Can also be used to deliver just oxygen during recovery or in an emergency*

Components of the Anesthetic Machine 1. Compressed gas supply Gas tanks, their pressure gauges, pressure reducing valves, flow meter 2. Anesthetic vaporizer Precision, out-of-circuit vaporizer is ideal (what we use) Located near the fresh gas outlet 3. Breathing circuit Unidirectional valves, breathing tubes, reservoir bag, pop-off valve, CO2 canister, pressure manometer, negative pressure relief valve 4. Scavenging System Disposes of excess and waste anesthetic gases

PART 1: Compressed Gas Supply

Gas Cylinders Most commonly used carrier gas is oxygen It comes compressed in a metal cylinder and held under pressure- up to 2200 psi Available in various sizes “E” tanks hold 660 L of oxygen and are attached to the anesthesia machine “H” tanks hold 6900 L of oxygen and stand separate from the machine Tanks are delivered and picked up by the oxygen supply company as needed

Gas Cylinders Tanks are color coded for safety and recognition Oxygen tanks: green, white (Canada) Nitrous oxide tanks: blue Carbon dioxide tanks: gray E tanks/cylinders, attach to the anesthetic machine via a yoke. H tanks/cylinders can be attached via hoses or pipes. Gas lines may be used to bring the gas into the surgical area from another room.

Gas Cylinders- Safety In addition to color-coded tanks, the yokes are gas- specific. The valve on the oxygen tank matches the connecter on the yoke specific to oxygen An oxygen yoke has two pins below the outlet port The yoke is connected to a pressure reducing valve to control the amount of gas released. *May or may not be able to see this component*

Gas Cylinders- Safety Combustible gases- avoid flames and sparks! If dropped, has the potential to act as a torpedo Never leave a cylinder unattended! VIDEO

Tank Pressure Gauge Display of the pressure of oxygen currently in the tank. (usually attached to yoke) Only displays when the tank is open/on To open tank: use key Only open once tank has been connected to a yoke! Gauge reads zero when tank is empty AND when tank is turned off/closed What about gas left in the machine? *Turn on O2 and check this gauge BEFORE every anesthetic procedure

Tank Pressure Gauge (psi) Refill line is at 500 psi (red zone) Replace w/ new tank when the pressure is 100 psi Some clinics will require tanks to be changed much earlier. Gauge gives you the amount of O2 in psi but you set your flow rate in liters/minute…how can you tell how much time you have left until empty?

Calculating Oxygen Volume in Liters Total volume of O2 left in your tank (liters) can be calculated by multiplying the pressure (psi) by: 0.3 for E tanks and 3 for H tanks Example: Gauge reads 1100 psi on E tank

Nitrous Oxide A full E cylinder contains 760 psi. Nitrous oxide is present in liquid AND gas forms inside the tank. When the tank is turned on, liquid evaporates into a gas as other gas leaves the tank. The pressure of the tank doesn’t change (because of the constant replacement of the gas) until all liquid has been volatilized. *The gauge will not drop until almost empty* Anesthetist should change the tank as soon as 500 psi is reached.

Pressure Reducing Valve aka Pressure Regulator Regulates the pressure of the gas leaving the tank to flow through the lines, into the flow meter of the anesthesia machine. Reduces the pressure of oxygen that leaves the tank at 2200 psi to a safer 40-50 psi. Allows a constant flow of gas into the machine, despite pressure changes within the tank Can be measured with a line pressure gauge In cases of E tanks and gas lines, the line pressure is preset at 50 psi, so there is usually no gauge seen *Know what line we are measuring the pressure in*

Flow Meter Allows the flow rate of oxygen traveling through the machine to be adjusted by the anesthetist. Reduces the psi AGAIN, from 50 down to 15 psi Oxygen does not reach the patient unless this is turned on (neither does the anesthetic gas). Measured in liters per minute by a ball Ball rises in height, proportional to gas

Calculating How Much Time is Left of O2 We already figured out the # of liters of oxygen in the tank: Divide that amount by the flow rate you’re using: 330 L of O2 / 1 L/min = If you have a higher flow rate: 330 L of O2 / 2 L/min =

Oxygen Flush Valve Button that when depressed, rapidly delivers pure oxygen at a high flow rate (35-75 L/min) Bypasses vaporizer (no anesthetic delivered) O2 goes directly from pressure reducing valve to breathing system  lungs Bypasses flow meter- no way to measure flow! Must be used in short bursts Not to exceed 2 cmH2O on pressure manometer

Oxygen Flush Valve Seems dangerous… When would we want to use this? 1. To dilute anesthetic in lines in emergency 2. To dilute anesthetic in lines on recovery 3. To fill reservoir bag (to give a breath)

PART 2: Anesthetic Vaporizer Next stop as oxygen travels from the flow meter Now that the gas is in the machine, it’s job is to mix with the liquid anesthetic and be delivered to the patient.

Vaporizer Converts the liquid anesthetic agent (usually isofluorane or sevofluorane) into a vapor. Adds controlled amount of these vapors to the carrier gas  “fresh gas” The mixture of the anesthetic gas and carrier gas is known as fresh gas. The vaporizer must be on to deliver any inhalant anesthetics to the patient. Must press down on safety lock to turn dial The flow meter must be on to supply the oxygen; otherwise nothing is delivered.

Vaporizer Amount of anesthetic liquid left in vaporizer is visible in the indicator window. Check BEFORE your surgery to see if it needs refilling! Refill by unscrewing cap of fill port If for some reason the vaporizer is tipped over (usually the whole machine), turn the vaporizer off and run oxygen only through the machine for 15 minutes to flush it out. Use the correct anesthetic with the correct vaporizer! (Iso vs. Sevo)

Vaporizer The vaporizer has a inlet port and an outlet port Optional: Common Gas Outlet aka Fresh Gas Outlet Note: When switching between rebreathing and non- rebreathing systems, this is where you will connect your breathing tubes

Precision vs. Non Precision Non-precision vaporizers are simple, cheaper, and are typically used for anesthetics with low vapor pressure such as methoxyfluorane Non-precision vaporizers are located within the breathing circuit (VIC) Gas flows from the flow meter into the breathing tubes, which contain the vaporizer *They are no longer commonly used; we will focus on precision vaporizers

Non-precision Vaporizer

Precision vs. Non Precision Precision vaporizers deliver a precise, controlled amount of anesthetic to the patient Expressed as a % which is chosen based on an anesthetic’s MAC and the patient’s requirements Commonly used anesthetics can reach concentrations as high as 30% + if they are not controlled (vapor pressure) Precision vaporizers are located outside of the breathing circuit (VOC) Newer models compensate for the factors that determine the concentration of anesthetic delivered (other than what you set it to). Ex. Temperature, respiratory rate, back pressure

Precision Vaporizer Compensation Made to deliver a precise concentration of anesthetic; non-precision vaporizers do not Factors that are compensated for: Temperature changes: Volatile anesthetics vaporize more rapidly at high temperatures If not compensated for: hot room = higher anesthetic output Carrier gas flow rates and respiratory rate effect amount of anesthetic delivered  If not compensated: high L/min or tachypnea = higher anesthetic output

Precision Vaporizer Compensation Factors that are compensated for: Back pressure: Squeezing the reservoir bag while bagging your patient exerts pressure on the vaporizer outlet port If not compensated for: providing a breath = providing a higher anesthetic concentration