Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd
Functions of keys Start / Reset Key Escape Key Background light Key Run / Pause Key Setup Key Left / Right Keys Enter Key Delete Key Up / Down Keys
Boot up meter… start/reset Press the start/reset key for at least 2 seconds to boot up the meter. And enter to the main menu. Main menu
Calibration… leftright enter Press left and right keys to move the cursor to ‘Cali’ and press enter to go to interface of Calibration. Calibration Window MIC Serial No.
No. Calibration… Put on the calibrator and turn it on. Notes: keep the meter and calibrator connected tightly and solidly. Calibration Window left right enter Press left and right to move the curser to ‘CALI’, and press enter to start calibrating.
MIC Serial Calibration… In processing of calibration, the upright corner will shows the # 1-9. (Which means the calibration takes 9 seconds totally.) MIC Serial No. When the digit becomes 9, means the calibration is completed 。 MIC Serial No. MIC Serial Notes: the LC value should be around 93.8 ±0.2dB. Otherwise, redo the calibration. enter After finishing calibration, move cursor to ‘USE ’, press enter to save the setup.
Over 3dB difference … 1) The screen inside the microphone is broken. In this case, the instrument should be sent back to the manufactory to get repaired. 2) Check whether or not the calibrator work normally. MIC Serial No. When the Sensiti and Lpx have over 3dB difference , the new calibration setting won’t be saved. Then please follow the steps to resolve the problem.
3). We recommend to replace the microphone when you found the Lpx is smaller than sensiti too much. Though, when you are in emergency case, you can try manually adjust the sensitive of microphone. First of all, record down the Lpx value. Over 3dB difference (Cont.) … MO DI Reset enterleft/right up/down Press Reset to reset the meter first, then come back to the calibration interface again. Move the cursor to ‘MODI’, press enter. Then use left/right keys to move cursor to ‘Sensiti’ and use up/down keys to adjust the sensitivity to equal the Lpx value that you have recorded down earlier.. MIC Serial No.