Code P008 Date: Clarity User Training Advanced Clarity User Training
Code P008 Date: For advanced users User settings and display options Sequences Multidetector systems Export data Report setup Batch data reprocessing
Code P008 Date: For advanced users Data archivation Column Performance Noise and Drift Special calculations Control modules GLP a CFR21part11
Code P008 Date: User settings are stored in individual users desktop *.dsk files. For this reason they are set in the instrument window and not in the main window User settings
Code P008 Date: Text Label and Lines for legends and arrows Chromatogram/Create Label Moving and Scaling of Chromatograms Math operation s Overlay activation Labels and display options
Code P008 Date: Colored peak area in chromatogram Peaks in calibration can have a color assigned. When these peaks are identified in chromatogram they will be automatically colored
Code P008 Date: Colored peak area in chromatogram Highlight currently selected peaks Simply select a peak in the chromatogram result table. The result table selection overrides the color defined in calibration
Code P008 Date: Colored peak area in chromatogram Highlight peaks selected in Results table (signal color used) Use Peak Color set in Calibration file
Code P008 Date: Context menu invoked by right mouse click in table Table customizing options
Code P008 Date: Common columns in Summary table Independent setting of columns Common - for all compounds Summary – compound specific
Code P008 Date: Summary table options
Code P008 Date: Context menu invoked by right mouse click in Chromatogram window Axis properties setting Graph properties setting Graph display options
Code P008 Date: Gradient display options Setting for individual signals Chromatogram display options
Code P008 Date: Customizing Toolbars Reset All restores original settings
Code P008 Date: Define a serie of samples to be measured Essential for working with autosamplers For controled autosamplers the Vial number and Injection volume are used from this table Allows modifications during a run for not yet processed samples Simply add samples by filling any field in the last empty row. Other fields will be copied from previous. Several samples can be added by copying from a table Allows for import of sample data from text files Allows reprocessing of last measured sample set Sequence table
Code P008 Date: Start, End vial and Injections number. Several vials and/or injections defined on single row File name, automatic %variables for SV, EV, IV Automatic recalibration from sequence Row status indicator Use row, add new rows during sequence run Tooltip with actual file name Individual Post-Run settings for each row Sequence table window
Code P008 Date: Pasive sequence Clarity only expects Start signal, analysis time is set on autosampler Active sequence Clarity sends a Ready signal to autosampler and waits for Start signal AS control (Active sequence) Autosampler injects according to SV, EV, IV and Inj. Volume in sequence table Editation of template method used on selected row Initial value for %n Sequence options
Code P008 Date: Run sequence starts complete sequence (resets the status) Pause sequence Resume sequence continues from last injection Status of the sequence. Stored at Pause, Abort, Stop Can be changed by Edit/Reset Status for selected rows Repeat Injection Skip Vial Stop sequence will stop the sequence run. Repeated Stop will stop the current analysis Sequence control
Code P008 Date: Select the file to import Assign the respective fields to be imported Only Start vial (SV) and file name must be set (marked in bold) Other fields will be filled by default values Sequence import
Code P008 Date: Export data to other programs to prevent typing errors Simple Copy and Paste (Ctrl C – Ctrl V) Export Data settings common for manual export from chromatogram window automatic export from postrun at the end of each run automatic export by batch reprocess of chromatograms or sequence Export of chromatograms in common formats Export data
Code P008 Date: Copy a picture to MS Office (including labels) via Clipboard or store it in *.emf format Copy a selected part of table using Copy (Ctrl C) and Paste (Ctrl V) Export data - simplest
Code P008 Date: Headers: Column headers and chromatogram info Full format: Filename, date and time before each row blank field = actual chromatogram name +.txt extension Path can be used All exported data are in single file Export time/signal data pairs (import to other programs – for example graphics) Export data: manual - from Chromatogram window automatic - according to Setting/Postrun batch - from Analysis/Batch Export data - options
Code P008 Date: Export data - example
Code P008 Date: print Report style from Report Setup dialog will be used do not forget to switch off the Overlay mode export reprocess of chromatograms according to instrument method reprocess whole sequence electronic signature File type Chromatograms Calibration standards Sequence Files Export according to Settings - Export Data Electronic signature for series of chromatograms Batch processing
Code P008 Date: Assignment of detectors to Clarity Instruments in the System Configuration dialog All signals stored in a single chromatogram Result table, Integration table and Acquisition parameters are individual for each chromatogram Common calibration file calibration curves are individual for each signal Summary table for display of results from all signals Multidetector systems
Code P008 Date: Results are displayed individually, some properties like calibration are common for the whole chromatogram Connecting link marks signals from one chromatogram. Click icon to set signal active Detector Delay compensates for interdetector volume, inactive for first signal Multi signal chromatograms
Code P008 Date: Summary table displays results for several chromatograms or signals simultaneously In overlay, several chromatograms can be compared. Maximum number can be set in user options Doubleclick on legend will set signal active Scale, move and math operations are active only in overlay Ctrl + click will hide/show the signal Ctrl + click will lock the tool for repe- titive use. Overlay
Code P008 Date: Report Setup - logo
Code P008 Date: Report Setup – tiled signals
Code P008 Date: To reduce the data size compressed archive *.dgz files should be used *.dgz archives always replace all data (files deleted from project will be during archiving deleted also from the archive). It is advantageous to archive whole projects – should be considered when planning project structure Archiving and backup
Code P008 Date: Archive and Restore for copying or moving projects Selected file types Without compressing = original format and structure compressed *.dgz archives = single file Archive data
Code P008 Date: Separation parameters Values used for Capacity ratio and Efficiency calculation Calculation type Performance tab (hidden in default layout) Advanced tab in the Method Setup (displays template method Performance parameters
Code P008 Date: Actual noise monitoring in the Data Acquisition window Noise monitoring
Code P008 Date: Drift and Noise displayed for selected intervals in Result table header Drift and Noise can be used for in User column calculations Noise and Drift
Code P008 Date: ISTD calculations Normalized % calculations Groups calibration of peak group like single compound example - isomers (xylen) Scale factor recalculation of result to other units Uncalibrated response quantification of unidentified peaks using Uncalibrated response factor Injection volume Injected volume correction for manual injection in GC Calculations
Code P008 Date: ISTD calculations are performed: a) Amount for ISTD is set in both calibration and sample b) both are zero ISTD compound is calibrated ISTD amount in sample is given Only one ISTD compound is allowed in calibration. If the conditions are not met, no results will be calculated ISTD peak can be hidden in the Result table ISTD calculations
Code P008 Date: If Sample amount is non- zero, it will replace the sum in the Amount column and serve as a base for Amount % calculation Total line is calculated as a sum of all groups + sum of peaks not belonging to any group. Groups are calibrated, thus peaks should not be both calibrated and in a group to get correct Amount% values Amount % gives the normalized % if all integrated peaks are calibrated and Sample amount is zero Normalized % calculations
Code P008 Date: st step Adding peaks to a group in the integration table 2 nd step Setting a Group in Calibration table. The compound name and amount is set as for ordinary peaks Groups
Code P008 Date: Scale factor and Units after scaling: Multiplies values in Amount column and changes the units in column header. Method specific parameter, multiplies also the entered Sample Amount value Uncal. Response: Used for quantification of unidentified compounds. For universal response detectors (ELSD, FID, RI) Dilution: Multiplies values in Amount column. Sample specific parameter (entered from Sequence or Single Analysis window). Scale Factor and Uncal.Response
Code P008 Date: Default Injection Volume must be set first in calibration and Injection Volume set in standards before calibrating Supposes linear detector response Can lead to incorrect results, if used improperly Setting Default Injection Volume in calibration Setting Injection Volume for sample and standard Injection Volume
Code P008 Date: Command Calibration/Show History displays data used for selected level Deviation of point from calibration curve in % Residual sum of deviation squares Calibration
Code P008 Date: Calibration Retention (Kovats) indexes LOD and LOQ limits Columns for Retention (Kovats) indexes are hidden in default settings. Use setup columns to disdplay them Amounts lower then specified LOD or LOQ limits are marked in Peak type column Retention indexes, LOD, LOQ
Code P008 Date: For selected instruments the direct control is available. The advantages are The instrument parameters are stored in a method It is not necessary to enter the parameters twice, especially useful for autosamplers The parameters are recorded in the chromatogram Control Modules
Code P008 Date: Control modules enable full control of connected devices like gradient programe and other settings. The used settings are stored within a chromatogram. Control modules (GC)
Code P008 Date: Control modules (LC, AS)
Code P008 Date: Option to Send/Confirm/Do not send method to instruments after method change Send Method options Confirmation dialog Control Modules
Code P008 Date: Control modules overview
Code P008 Date: Other devices can be controlled by A/D board digital outputs: gradient or detector program start detector autozero valve switching fraction collector control Synchronization in Active Sequence Dig. output set in System Configuration, not in the Event table Additional control
Code P008 Date: Set events in the Event Table tab in the Method Setup Named events Digital event input type Output event can trigger run/sequence commands: Stop Abort Skip vial Shutdown Repeat injection Event table
Code P008 Date: Chromatogram history All saved versions are stored in the chromatogram Versions can be opened (including calibration) Audit trails Station, Chromatogram, Sequence, Calibration User access Can be restricted by User Accounts settings GLP a CFR21 part 11
Code P008 Date: Electronic signatures based on User Accounts based on Certificates Validation Installation qualification - standard feature Performance qualification - by optional Validation kit Operation qualification - SST module for statistical evaluation and pre-set limits control GLP a CFR21 part 11