Lecture 4 Beam restricting Devices, their Use and Maintenance

Slides:



Advertisements
Similar presentations
Technical Aspects: the machine the image
Advertisements

Advanced Biomedical Imaging Lecture 3
Control of Scatter Radiation
Advanced Biomedical Imaging Lecture 4 Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
X-Ray Beam Restrictors
Radiographic Quality Chapter 5.
Topic 3 Selection of kV High kV scattered radiation is energetic. Angle of rays is same as those of primary beam and is not deflected very much. Follows.
RADIOLOGY. NEXT GENERATION SCIENCE / COMMON CORE STANDARDS ADDRESSED! CCSS.ELA-Literacy.RST Determine the central ideas or conclusions of a text;
LIGHT THIN LENSES Name: ________________ Class: _________________
Dorsal/ Lumbar/ LS Spines
Chapter 6 Control of Secondary and Scattered Radiation.
Control of Scatter Radiation Beam Restricting Devices
Chapter 17 The Grid So far we have discussed how kVp, patient size and collimation impact scatter radiation. As the part size and kVp increase, scatter.
Radiographic Grids II By Professor Stelmark.
The Grid Kyle Thornton DMI 50B.
Analyzing the Image Density. Density Overall blackening of the image.
X-ray Beam Composition and Collimation
DESIGNING FOR RADIATION PROTECTION. TUBE HOUSING  REDUCES LEAKAGE TO LESS THAN 100 mR PER HOUR AT A DISTANCE OF ONE METER FROM HOUSING  One meter is.
Chapter 23 Mirrors and Lenses. Notation for Mirrors and Lenses The object distance is the distance from the object to the mirror or lens Denoted by p.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Radiographic Technique - I
Resident Physics Lectures
Basic quality control tests in medical radiography
Factors affecting the X-Ray output
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
THE X-RAY UNIT THE BASICS WHAT FOUR COMPONENTS ARE IN TYPICAL RADIOGRAPHIC ROOM? 1. X-RAY TUBE 2. OPERATING CONSOLE 3. HIGH VOLTAGE GENERATOR 4. X-RAY.
COLUMNS. COLUMNS Introduction According to ACI Code 2.1, a structural element with a ratio of height-to least lateral dimension exceeding three used.
Quality Control Rad T 110.
Resident Physics Lectures Christensen, Chapter 6Filters George David Associate Professor Department of Radiology Medical College of Georgia.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Control of Scatter Radiation
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography.
Radiation Monitoring and Protection
Spot Filming Cinefluorography Fluoro Room Shielding Radiation Areas.
Dr Mohamed El Safwany, MD.
RAD 354 Chapt. 11 Control of Scatter Break down into: Those that reduce pt. dose and those that are geometrical in nature and those NOT! 3 factors affecting.
Reinforced Concrete Design
IAEA International Atomic Energy Agency Dental Radiology Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 2(2)
INVERSE SQUARE LAW. The picture above demonstrates the typical x-ray tube used to produce a point source of x-rays. Then as radiation exits the tube it.
CT Instrumentation and X-ray system
Preparing Variable kVp Technique Charts By Prof. Stelmark.
Graphic Communication
Beam And Scatter Radiation Restricting Devices DMI 50B Kyle Thornton.
Beam Restricting Devices
Radiographic Equipment
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
Grids George David Associate Professor Department of Radiology
IMAGE RECEPTORS. Follow the Sequence- Film System Tube Tube Patient Patient (bucky) or non-bucky (bucky) or non-bucky Cassette Cassette INTENSIFYING SCREEN.
Microscope Notes. Parts of a microscope 1. Eye piece/ocular- the piece that is looked through, has a power of Body tube- skinny tube attached to.
Revision of terminology and drawing a Ray diagram
Stacy Kopso, M.Ed., R.T.(R)(M).  Xray photon loses energy and changes direction  Responsible for most of the scatter that fogs the image  Related to.
IAEA International Atomic Energy Agency General Radiography Radiation Sources in medicine diagnostic Radiology Day 7 – Lecture 1(1)
Lavin: Chapter 7 CTVT: pp
MICROSCOPE A compound microscope is an expensive precision instrument that requires special care and handling. let’s learn the parts of a compound microscope,
Week 2: Radiographic Equipment
Optimizing The Image RVT: Chapter 7
Exposure Factors or Prime Factors
Control of Scatter Radiation Beam Restricting Devices By Prof. Stelmark.
Pinhole Camera.
RADIATION PROTECTION 04/12/2016.
Resident Physics Lectures (year 1)
Protection of Patients
Factors Effecting the Production
Resident Physics Lectures
Resident Physics Lectures
Resident Physics Lectures (year 1)
Chapter 5, Fauber Supplement-Chapter 11, Bushong
Artifacts and Errors.
Presentation transcript:

Lecture 4 Beam restricting Devices, their Use and Maintenance Topic 1 Importance of beam restricting devices Topic 2 Types of beam restricting devices Topic 3 Quality assurance tests of beam restricting devices Topic 4 Compression of area of interest

Topic 1 Importance of beam- restricting devices

Topic 1 Importance of beam restricting devices These devices restrict the primary x-ray beam to the area of interest. The larger the area covered by primary x-ray beam, the greater the scattered radiation produced. Scattered radiation must be minimized. They increase radiation dose and add to film darkening thus reducing visualization of details.

Topic 1 Importance of beam- restricting devices Beam was not restricted to the chest – the area of interest. Entire abdomen and part of face of baby was exposed unnecessarily.

Topic 1 Importance of beam- restricting devices Unrestricted beam-considerable scatter Restricted beam – less scattered radiation

Topic 2 Types of beam-restricting devices a. Attached to x-ray tube housing Aperture diaphragms - lead sheets with circular, square, or rectangular openings Cones – detachable metal tubes of different lengths, shapes, and sizes of opening Variable aperture collimators – adjustable lead plates or shutters , contain "cross hairs", a light source, and a mirror to project the light Aperture diaphragms are inserted near the tube window and are usually used with a cone and a variable aperture collimator. The cone length and size of opening affect the size of the x ray field. Variable aperture collimators are also called light beam diaphragms.

Beam- restricting devices Removable metal cone Variable aperture collimator

Variable aperture, multiple shutter collimator Diaphragm closer to window reduces off-focus x rays. Multiple shutters Multiple shutters Smaller sized field reduces size of primary beam.

Topic 2 Types of beam - restricting devices b. Not attached to x ray tube housing Lead blockers – used to divide a cassette when more than one exposure is made on a single film; reduces unnecessary exposure of film thus improving image quality Lead blockers limit the chances of film fog due to the unnecessary x ray exposure.

Use of lead blockers Place blocker on couch for lateral projections of thoracic, lumbar, sacrum or coccyx spine. Position blocker next to patient's back to absorb radiation that will not pass through spine. Defective lead rubber aprons could be cut into a range of sizes to serve as blockers.

Topic 3 Quality assurance tests of beam limiting devices Collimation or limiting the x-ray beam to the area of interest requires use of beam restricting devices. It is good practice to collimate such that the resulting image has collimated edges on all four sides of film

Topic 3 Quality assurance tests of beam- restricting devices Bring tube to rest on table top to check that all four edges of square light beam diaphragm touch the table top when using vertical beam. Do several times during the day. Do the same test with cone inserted in front of the x ray tube window. Caution: Cones sometimes become bent when dropped or bumped.

Topic 3 Quality Assurance tests of beam restricting devices Do collimator-beam alignment check at least every six months. Do check of alignment of centre of x ray beam at least every six months.

Collimator-beam alignment test Using 100 cm focus-film distance (FFD), set collimator at 20 cm x 20 cm field size on surface of loaded cassette. Place metal coins or paper clips on top of cassette as shown in next slide. Expose the film at 60 kVp and 4 – 8 mAs. Process the film. Check whether the distance between outer edges of image and of paper clips/coins is within +/- 2% of FFD.

Line diagram showing metal coins and field size Coin to identify right upper corner. Edge of light. Cassette.

Collimator-beam alignment test result

Check of alignment of centre of x-ray beam Place unexposed loaded cassette in centre of bucky tray and centre tube to cassette. Move tube to 100 cm FFD to bucky tray. Reduce longitudinal collimators to a thin slit (e.g. 0.5 cm). Close lateral collimators. Expose using 60 kVp and 4 – 8 mAs. Do not remove cassette.

Check of alignment of centre of x ray beam (cont) Close longitudinal collimators and open lateral collimators to a thin slit. Expose film again. Process film. Bend film in half and check that exposed "cross" is in film centre. Acceptable deviation: 1 cm on either side of centre. Can also be done for non-bucky radiography. Just move tube so 100 cm FFD is to cassette top.

Centre of beam alignment test result

Topic 3 Summary Beam restricting devices may be attached or not attached to tube housing require periodic checks are necessary for radiation protection and contribute to good image quality minimize primary radiation minimize scatter radiation Use a compression band. Use a compression band

Topic 4 Compression of area of interest Reducing patient thickness by compression minimizes production of scattered radiation, thus reducing radiation dose. Device used : compression band.

Topic 4 Compression of area of interest Compression band over abdomen reduces its volume. This reduces scatter radiation.

Lecture 4 Summary Beam-restricting devices are important in radiation protection. These devices should be checked regularly. Regular performance of quality assurance tests is in accordance with ALARA.