1 INTRODUCTION TO CONTINUOUS GLUCOSE MONITORS H. Peter Chase, MD Vicky Gage, RN, CDE Laurel Messer, RN, CDE Susie Owen, RN, CDE Sally Sullivan, RN, CDE.

Slides:



Advertisements
Similar presentations
Making The Most of Continuous Glucose Monitoring Gary Scheiner MS, CDE
Advertisements

Making The Most of Continuous Glucose Monitoring Gary Scheiner MS, CDE
Do you require any assistance? Do you experience any symptoms? Calit2 Summer Research Program Minimize Hypoglycemic Episodes Utilizing Remote Assistance.
Diabetes and Technology: Insulin Delivery and Monitoring
Advanced Pumping. Objectives: Identify situations to utilize temporary basal rate in pump therapy patients. Identify examples of when to use combination.
Newest Trends in Diabetes Management in Schools
1-800-DIABETES DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to Know DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to.
Diabetes & Insulin Pens Prepared by: Alison Deux, 4 th year pharmacy student.
Monitoring diabetes Diabetes Outreach (March 2011)
INTRODUCTION TO CONTINUOUS GLUCOSE MONITORS
1-800-DIABETES DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to Know DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to.
WHAT’S NEW IN DIABETES Lisa Still, RN,CDE Diabetes Specialty Nurse Arkansas Children’s Hospital November 2011.
Continuous Glucose Monitoring. Diabetes Management Evolution Insulin Delivery Glucose Monitoring 2000 First CGM system 2006 Paradigm REAL- Time, combining.
Insulin Pump What to tell your patient!! Prakash Abraham Isla Fairley.
Continuous Glucose Monitoring
Putting Pump Policies Into Practice- Case Study Conference Call Elizabeth Blair, ANP-BC,CDE Joyce Lekarcyk, RN, CDE.
PRACTICAL ASPECTS OF CONTINUOUS GLUCOSE MONITORING (CGM) Barb Tallant, RN, MA Laurel Messer, RN, BSN, CCRP.
Diabetes Control in Youth: The American Experience Georgeanna J. Klingensmith, MD Keystone Colorado July 2008.
Suhyla Alam (Eastern Virginia Medical School), Amy West, Maura Downey, Jane EB Reusch, Kristen Nadeau University of Colorado Denver and Children’s Hospital.
Accuracy of the FreeStyle Navigator ™ Continuous Glucose Monitor Diabetes Research in Children Network Larry Fox, 1 Roy Beck, 2 Stuart Weinzimer, 3 Katrina.
Management Tools and CGM Kathryn Moe, RN CDE Medtronic Diabetes.
Pumps and Sensors In the School Setting. 2 Agenda Comparison of available insulin pumps Continuous glucose monitors Pump therapy & CGM in the school setting.
Tools and Technology for Diabetes Amy Hess Fischl, MS RD LDN BC-ADM CDE.
Practical Aspects of Continuous Glucose Monitoring 2008 Rosanna Fiallo-Scharer, MD Laurel Messer, RN, BSN, CDE Barbara Davis Center for Childhood Diabetes.
4-06 CHANGE IS GOOD: THE BASAL BOLUS INSULIN CONCEPT Management of Hyperglycemia in the Adult Hospitalized Patient TEAM MEMBERS: Physicians: Maryann Emanuele,
Diabetes Technology Update
Mealtime Glycemic Excursions in Pediatric Subjects with Type 1 Diabetes: Results of the Diabetes Research in Children (DirecNet) Accuracy Study Study Group.
Abstract Background: As part of a study to evaluate the accuracy of the GlucoWatch ® G2 TM Biographer and the Continuous Glucose Monitoring System (CGMS.
Diabetes Control and Complications Trial (DCCT) Results indicate that most youth with T1DM should be treated intensively in order to reduce the risk of.
Background: DirecNet Diabetes Research in Children Network NIH funded collaborative study group 5 clinical centers, central laboratory, coordinating center,
Making the Most of Continuous Glucose Monitoring Gary Scheiner MS, CDE Owner & Clinical Director Integrated Diabetes Services LLC Wynnewood, PA AADE 2014.
Diabetes Research in Children Network Pilot Study of the Navigator TM Continuous Glucose Monitoring System in Children with Type 1 Diabetes: Safety, Tolerability,
Diabetes Care Tasks at School: What Key Personnel Need to Know Diabetes Care Tasks at School: What Key Personnel Need to Know BLOOD GLUCOSE MONITORING.
Use of The FreeStyle Navigator TM Continuous Glucose Monitoring System in Children on Glargine- based Multiple Daily Injection Therapy Stuart Weinzimer.
Use of the FreeStyle Navigator ™ Continuous Glucose Monitoring System in Children with Type 1 Diabetes Diabetes Research in Children Network L. A. Fox,
A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children.
Healthcare Across Borders - September 2003 Head-To-Head Comparison Of The Two Currently Available Continuous Monitors North County Endocrine 700 West El.
Evaluation of Factors Affecting CGMS Calibration Bruce Buckingham, 1 Craig Kollman, 2 Roy W Beck, 2 Andrea Kalajian, 2 Rosanna Fiallo-Scharer, 3 Michael.
Accuracy Study of the Medtronic Minimed Continuous Glucose Monitoring System (CGMS) and GlucoWatch® G2TM Biographer (GW2B) in Children with Type 1 Diabetes.
INSULIN PUMPS Shelby Polk DNP, FNP-BC, CDE. 2 MANAGEMENT OF DIABETES IN SCHOOLS Exercise Legal Rights Health & Learning Nutrition Insulin Administration.
In order to get the most out of your diabetes clinic consultation, we would be grateful if you could complete this brief set of questions, each time you.
Clarification of Pump Orders from Barbara Davis Center Diabetes Resource Nurse Winter/Spring 2013.
Diabetes Update: Michael Gottschalk, M.D, Ph.D.
DirecNet Study of the Accuracy of the Navigator Continuous Glucose Monitoring System in Children and Adolescents with Type 1 Diabetes Darrell Wilson 1,
The Super Bolus And The Projected BG Alert New Insulin Pump Ideas To Improve Glucose Levels, Avoid Hypoglycemia And Speed Correction Of Hyperglycemia John.
Managing Diabetes All About Blood Glucose. Topics What affects blood glucose levels? What blood glucose tests are important for people with diabetes?
DIABETES IN SCHOOL DR. SARAH LAWRENCE PEDIATRIC ENDOCRINOLOGIST CHILDREN’S HOSPITAL OF EASTERN ONTARIO ASSOCIATE PROFESSOR OF PEDIATRICS UNIVERSITY OF.
January 2013 Standards of Care for Diabetes Management in the School Setting – Colorado January 2013.
Global Continuous Glucose Monitoring Market to Rise at 13.60% CAGR till 2019, Propelled by Rise in Diabetic Population
Monitoring and Management Jennifer Danielson, PharmD, MBA, CDE Clinical Assistant Professor University of Washington School of Pharmacy.
Dr Zaranyika MBChB(Hons) UZ, MPH, FCP SA Department of Medicine UZ-CHS
 History of advancement in diabetes treatment  New technology  Features  Pros and Cons  Safety in school  Questions.
Blood Glucose Monitoring Liz Gregory / Sue Russell October 2013.
GLOBAL CONTINUOUS GLUCOSE MONITORING (CGM) MARKET - Opportunity Analysis on the basis of Product, End Use, Geography, Industry Forecast 2025
The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is designed with convenience, performance, and simplicity in mind. You are already using the.
Making the Most of Continuous Glucose Monitoring
1-800-DIABETES DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to Know DIABETES CARE TASKS AT SCHOOL: What Key Personnel Need to.
Consultant/ Clinical Senior Lecturer
Battery life of transmitter is a minimum of 6 months.
Diabetes Education.
Diabetes Care Tasks at School: What Key Personnel Need To Know
Components of a CGM record over a 3
LBL Rev 01.
Study Objective & Methods
Pharmacist Involvement in Continuous Glucose Monitoring
Introduction to Continuous Glucose Monitoring
User-based design strategy
Meal control achieved in the five meals of the study (the first dinner was handled in open-loop mode, and all other meals were handled by closed-loop).
Insulin in Type 2 Diabetes
Practical Implementation and Optimization of A Closed Loop System
Presentation transcript:

1 INTRODUCTION TO CONTINUOUS GLUCOSE MONITORS H. Peter Chase, MD Vicky Gage, RN, CDE Laurel Messer, RN, CDE Susie Owen, RN, CDE Sally Sullivan, RN, CDE Barbara Davis Center for Childhood Diabetes April 2010

2 Agenda Dr. Chase will present a general CGM overview Nurses will explain device features and comparisons between sensors You will have an opportunity to handle each device and ask questions Barbara Davis Center for Childhood Diabetes April 2010

3 What is a CGM? (Continuous Glucose Monitor) A device that provides “real-time” glucose readings and data about trends in glucose levels Reads the glucose levels under the skin every 1- 5 minutes (10-15 minute delay) Provides alarms for high and low glucose levels and trend information The 3rd era in diabetes management Barbara Davis Center for Childhood Diabetes April 2010

4 Who Should Use a CGM? 1)The person and the family must both want a CGM 2)A youth must be willing to wear the sensor (and carry the receiver) 3)Using good diabetes care (4 BGs/day) 4)Good support system 5)Adequate body “real estate” 6)Cost of CGM (RNs to elaborate) Understanding Pumps and CGMs, p.100 Barbara Davis Center for Childhood Diabetes April 2010

5 WHY Use CGM? A.Prevention of low blood sugars (alarms) B.Prevention of high blood sugars (ketones) C.Minimize wide glucose fluctuations D.Behavior Modification E.Prevention of Complications (?) Barbara Davis Center for Childhood Diabetes April 2010

6 How common are glucose levels <60mg/dl during the night in children with T1D? –French (i) and Australian (ii) data showed approximately 50% of children with low BG (<60mg/dl) during the night (on NPH bid) –DirecNet data (one night in hospital with blood sugars every 30 min.) A) : 39 of 91 (43%) low BG (44% of children on insulin pumps/56% on NPH) B) 2004: 14 of 50 (28%) with low BG (all on insulin pumps or Lantus) (i) Beregszaszi M, et al. J Pediatr. 131, 27, 1997 (ii) Porter PA, et al. J. Pediatr. 13, 366, 1997 Barbara Davis Center for Childhood Diabetes April 2010

7 WHY Use CGM? A.Prevention of low blood sugars (alarms) B.Prevention of high blood sugars (ketones) C.Minimize wide glucose fluctuations D.Behavior Modification E.Prevention of Complications (?) Barbara Davis Center for Childhood Diabetes April 2010

8 “Snapshot of BG levels” Barbara Davis Center for Childhood Diabetes April 2010

9 Continuous Glucose Monitoring Barbara Davis Center for Childhood Diabetes April 2010

10 Hyperglycemia is common, especially after meals 0% 10% 20% 30% 40% 50% < > 300 Breakfast Lunch Dinner Boland et al, Diabetes Care 24:1858, 2001 Barbara Davis Center for Childhood Diabetes April 2010

11 WHY Use CGM? A.Prevention of low blood sugars (alarms) B.Prevention of high blood sugars (ketones) C.Minimize wide glucose fluctuations D.Behavior Modification E.Prevention of Complications? Barbara Davis Center for Childhood Diabetes April 2010

12 Three Parts to CGMs: A.Sensor B.Transmitter C.Receiver/Monitor Understanding Pumps and CGMs, p.103 Barbara Davis Center for Childhood Diabetes April 2010

13 A) Sensor (p.103) Barbara Davis Center for Childhood Diabetes April 2010

14 B) Transmitter (p.103) Barbara Davis Center for Childhood Diabetes April 2010

15 C) Receiver or Monitor (p.103) Barbara Davis Center for Childhood Diabetes April 2010

16 What does “Calibration” mean and why do I need to do it? Calibration is a process that gives a fingerstick BG value to the CGM system so the values will align with each other Number of Calibrations vary by device Best times to calibrate are when the BG values are stable: before meals and before bed Do not calibrate when arrows are present Barbara Davis Center for Childhood Diabetes April 2010

17 What type of data will we get? Two types of CGM data: Real Time data: seen on CGM Retrospective data: download to a computer Barbara Davis Center for Childhood Diabetes April 2010

18 Real Time Data Three types of Real Time Data: A) Trend graphs B) Alarms C) Trend Arrows Understanding Pumps and CGM: pages Barbara Davis Center for Childhood Diabetes April 2010

Trend graphs – Knowing a glucose level is 240 mg/dl may not be as important as knowing the “trend.” TREND GRAPHS Understanding Pumps and CGM: p.103 Barbara Davis Center for Childhood Diabetes April 2010 Real Time data

ALARMS ALARMS Barbara Davis Center for Childhood Diabetes April 2010 Warn patients of current or projected high and low blood sugar Real Time data

21 Alarms continued Alarms continued Threshold alarm: Warning when glucose is above or below a set value --all devices have this Projected Alarms: 10,20 or 30 minute warning of Impending hypo- or hyperglycemia Real-Time Revel System, Guardian Real-Time and the Navigator have these Real Time data Barbara Davis Center for Childhood Diabetes April 2010

TREND ARROWS Arrows that indicate the rate and direction of change Glucose going down -1 to -2 (mg/dL)/min Glucose going up 1 to 2 (mg/dL)/min Glucose falling quickly >-2 (mg/dL)/min Fairly stable glucose -1 to 1 (mg/dL)/min Glucose rising quickly >2 (mg/dL)/min Barbara Davis Center for Childhood Diabetes April 2010 Real Time data

23 Retrospective Data Three types of Retrospective data (viewed on computer download) A)Trend graphs B)Pie charts C)Data tables Understanding Pumps and CGMs, Chapter 17 Barbara Davis Center for Childhood Diabetes April 2010

24 Reports that show one or more days of CGM data– also called sensor overlay A) Trend Graphs Retrospective data Barbara Davis Center for Childhood Diabetes August 2009 Barbara Davis Center for Childhood Diabetes April 2010

25 Case Study: Using Trend Graphs Three Trend Graphs showing change over time: –Teenager with T1D for 9.5 years –Started Navigator: Sept –Starting HbA1c: 7.1% –Most recent HbA1c: 6.0% –Current number of low BGs per week (<60 mg/dL or <3.3 mmol/L): 1/week Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data

26 BASELINE GLUCOSE Trend Graph #1 Prior to CGM Use BASELINE GLUCOSE Trend Graph #1 Prior to CGM Use Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data

27 Glucose Trend Graph #2 After three months of CGM use Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data

28 Most recent CGM report GLUCOSE Trend Graph #3 Most recent CGM report Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data

29 B) PIE CHARTS Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data Show percentage of time glucose is above, below and in target range

30 C) Data Tables Barbara Davis Center for Childhood Diabetes April 2010 Retrospective data Show statistical information about different periods of the day

31 USING CGM RESULTS: (To make insulin adjustments) Important not to get overwhelmed by data ** Make One Change At A Time Look for patterns 2 out of 3 days A behavior modification device  Missed boluses, snacking, low BGs on CGM Good initial communication with HCP Barbara Davis Center for Childhood Diabetes April 2010

32 Interpreting CGM data Barbara Davis Center for Childhood Diabetes April 2010

33 Interpreting CGM data Barbara Davis Center for Childhood Diabetes April 2010

Realistic Expectations of CGM You will still need to test your blood sugar levels Using CGM does not make Diabetes Management a “Breeze” You will still experience low and high blood sugar values Sensor values will not always “MATCH” the blood sugar values The most significant improvements will be seen with consistent CGM wear Barbara Davis Center for Childhood Diabetes April 2010

You will still need to test your blood sugar levels for: Realistic Expectations of CGM Barbara Davis Center for Childhood Diabetes April 2010 Calibrations Insulin dosing Treating high and low blood sugar levels Questioning the accuracy of the CGM values Times when you do not feel right

36 Realistic Expectations of CGM Using CGM does not make Diabetes Management a “Breeze” Initially, people are overwhelmed from all the data Subjects will follow an algorithm early on in the use of CGM Family weekly downloads of CGM data are important CGM works as a behavior modification device Barbara Davis Center for Childhood Diabetes April 2010

Realistic Expectations of CGM Barbara Davis Center for Childhood Diabetes April 2010 You will still experience low and high blood sugar levels There will ALWAYS be some lows and highs Alarms may NOT be sensitive to slowly falling blood sugar values The time spent in hypoglycemia and hyperglycemia can be reduced CGM is an additional tool to help make decisions CGM values are not very helpful in treating lows

Realistic Expectations of CGM Barbara Davis Center for Childhood Diabetes April 2010 Sensor values will not always “MATCH” the blood sugar values CGM values are about 10 minutes behind blood sugar values Values are less accurate early after insertion Values are furthest off when blood sugars are rapidly rising or falling Calibrations are important determinants of accuracy of the CGM

The most significant improvements will be seen with consistent CGM wear ( >6 days/week) Realistic Expectations of CGM Barbara Davis Center for Childhood Diabetes April 2010 Good initial education helps patients and families to learn what to expect Regular use of Real-time and Retrospective data is necessary for optimal benefits Follow up with Diabetes Team is important to continue to make adjustments

40 Questions? Next: Device Information from the nurses Hands on with CGM devices Barbara Davis Center for Childhood Diabetes April 2010

41 CGMs available April 2010 Barbara Davis Center for Childhood Diabetes April 2010

42 Paradigm Real Time systems (Paradigm pump and Revel pump) A: Insulin pump and CGM receiver B: Infusion set C: Sensor D: Transmitter Barbara Davis Center for Childhood Diabetes April 2010

43 Paradigm Real-Time system (Paradigm pump and Revel pump) Pump is not controlled by CGM readings No extra receiver to carry Has high and low alarms Revel has predictive alarms Sensors last 3 days (6 days) Calibrations every 12 hours Great online download Barbara Davis Center for Childhood Diabetes April 2010

44 Guardian Real Time For people not using a pump Has high and low alarms Has predictive alarms Sensor lasts 3 days (6 days) Calibrations every 12 hours Why not buy a pump? Barbara Davis Center for Childhood Diabetes April 2010

45 Navigator CGM Barbara Davis Center for Childhood Diabetes April 2010

46 Navigator CGM Has built in Freestyle BG meter Larger transmitter Has high and low alarms Has predictive alarms Sensors last 5 days Calibrations at 10, 12, 24 and 72 hours Well studied and highly accurate Barbara Davis Center for Childhood Diabetes April 2010

47 DexCom 7 Plus Barbara Davis Center for Childhood Diabetes April 2010

48 DexCom 7 Plus Most simple system to use Smallest transmitter Has high and low alarms Has rate of change alarms Sensor lasts 7 days Calibrations every 12 hours Basic download software Barbara Davis Center for Childhood Diabetes April 2010

49 MiniMed Guardian MiniMed Paradigm 522/722 Freestyle Navigator Dexcom Seven Plus Sensor life 3 days 5 days 7 days Initial calibration period 2 hours 10 hours 2 hours Number of calibrations 2-4 per day 4 in 5 days 2-4 per day Trend arrows YesYesYes High/low alarms YesYesYes Predictive high/low alarms Yes No-Paradigm Real-time Yes-Revel pump Yes Rate of change alarms Cost ~$1300 system $35/sensor $1000 system + 722/522 insulin pump $35/sensor ~$1000 system ~$35-50/sensor ~$ system ~$60/sensor Barbara Davis Center for Childhood Diabetes April 2010

50 CGM Reimbursement Family will fill out an Authorization to release insurance information Physician will write prescription (Letter of Medical Necessity) BDC will provide any other paperwork requested by the insurance company Call Mireya at BDC to get started phone: , fax: Barbara Davis Center for Childhood Diabetes April 2010

51 Websites Paradigm REAL-Time, Paradigm REAL-time Revel and Guardian REAL-time: www.minimed.com Navigator: Dexcom: www.dexcom.com Barbara Davis Center for Childhood Diabetes April 2010

52 CGM Start and Follow up Medtronic Device Training— Contact Mireya at to schedule Navigator Training— Trainer will contact you when your device ships Dexcom Training— Contact Michelle Perrot at Follow up CGM with Stephanie Kassels, FNP– call Dahlia at to schedule Barbara Davis Center for Childhood Diabetes April 2010