8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS

Slides:



Advertisements
Similar presentations
Advanced Biomedical Imaging Lecture 3
Advertisements

X-ray tube.
Formation Characteristics
The X-Ray Tube Bushong Ch 7.
X-Ray Interaction with Matter & Human Biology
{ Projection Radiography Chapter 5 (in more detail)
Resident Physics Lectures
X-ray tube and detection of X-rays Lecture 5. Reminder: The rough schematics of an X-ray tube filament cathod target anode photon flux e-e- electron kinetic.
FRCR: Physics Lectures Diagnostic Radiology
ACVR Artifacts Artifacts of Diagnostic Radiology
ENTC 4390 PRODUCTION OF X RAYS.
Radiation Physics II.
Medical Imaging X-Rays I.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods X-ray Instrumentation Part 1.
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Types of Radiation Interactions All or Nothing Many Small There is a finite probability per unit length that the radiation is absorbed. If not, there is.
The Origins of X-Rays. The X-Ray Spectrum The X-Ray Spectrum (Changes in Voltage) The characteristic lines are a result of electrons ejecting orbital.
Factors affecting the X-Ray output
X-Ray Production & Emission
X-Ray Production & Emission
Conventional and Computed Tomography
Reference Reading Chapter 2: pp  X-rays are produced within the dental x-ray machine  The x-ray machine can be divided into 3 study areas.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University.
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
Quality Control Rad T 110.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Topic 29: Remote Sensing 29.1 Production and use of X-rays 29.2 Production and uses of ultrasound 29.3 Use of magnetic resonance as an imaging technique.
Medical Image Analysis Medical Imaging Modalities: X-Ray Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan,
A-LEVEL PHYSICS 15/09/ :49 Medical Physics.
Ch. 2 – Anatomy of the X-ray Machine
HABIS X-RAY PRODUCTION AND EXPOSURE FACTORS X-RAY PRODUCTION AND EXPOSURE FACTORS PREPARED BY PREPARED BY Dr fahad albadr radiology chairman radiology.
Resident Physics Lectures
Image Quality Radiographic Resolution.
Mammography Questions and Answers
Scanning Electron Microscope (SEM)
Chapter 2 The X-ray Beam.
X - RAYS IN DIAGNOSTICS D. Krilov
1 Dosimetry part 1: X-ray interactions with matter G.Haddadi, PhD, MSc Associated prof. of Medical Physics Fassa University of Medical Sciences.
Photon Tissue Interactions
Radiation Quality Chapter 4. X-ray Intensity Intensity: the amount of energy present per unit time per unit area perpendicular to the beam direction at.
Image Quality Q&A. RAPHEX Diagnostic Question 2001 D9: The imaging system which is best for visualizing small high contrast objects is: A. Computed tomography.
X-rays discovered in 1895 by Wilhem Roentgen Roentgenology – Branch of medicine dealing with x-ray or gamma rays in diagnosis and treatment. Radiology.
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Medical Equipment Technology Department 1 Introduction to Biomed. Imaging Systems Lecture No.5-6 Dr. Yousif Mohamed Y. Abdallah.
Interaction of x-ray photons (and gamma ray photons) with matter.
Image Quality and Patient Dose. X-ray tube design - basic principles Electrons generated by thermionic emission from a heated filament (cathode) Accelerating.
RTT 425 Radiation Therapy Physics Radiation Quality, Chapter 4 From Stanton and Stinson: Applied physics for Radiation Oncology.
Tube Exposure Factors Math Technique Contrast and Density
X-Rays Unit 5: Medical Physics. OCR Specification Learning Outcomes: Describe the nature of x-rays Describe how x-rays are produced Describe how x-rays.
X-ray tube.
An early x-ray by Wilhem Rontgen
Optimizing The Image RVT: Chapter 7
3 July July July Conventional X-rays Generator Basic components of an X-ray machine:  Electron source.  Vacuum where electrons were.
Patient is placed between X-ray tube and silver halide film.
Diagnostic Radiology II X-ray Tubes. Anode angle Anode angle defined as the angle of the target surface with respect to the central ray in the x-ray field.
Chapter 4: Diagnostic X-Ray Production
01 X-ray Tube.
Fundamentals of radiography.
X RAYS.
X-RAY PRODUCTION AND EXPOSURE FACTORS
X-ray Production Sharif Qatarneh Medical Physics Division
Part No...., Module No....Lesson No
X-Radiation.
Stacy Kopso, M.Ed., RT(R)(M)
Exposure Factors Chapter 4
Resident Physics Lectures (Year 1)
Resident Physics Lectures (Year 1)
Generation of X-rays Q&A
Computed Tomography (C.T)
Presentation transcript:

8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS 8. Diagnostic Radiology 8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS

X-rays are one of the main diagnostical tools in medicine since its discovery by Wilhelm Roentgen in 1895. Current estimates show that there are approximately 650 medical and dental X-ray examinations per 1000 patients per year. X-rays are produced when high energetic electrons interact with matter. The kinetic energy of the electrons is converted into electromagnetic energy by atomic interactions (see chapter 7.1.)

The X-ray tube provides an environment for X-ray production via bremsstrahlimg and characteristic radiation mechanisms. The classical X-ray tube requires: electron source electron acceleration potential target for X-ray production

The intensity of the electron beam determines the intensity of the X-ray radiation. The electron energy determines the shape of the bremsstrahlungs spectrum, in particular the endpoint of the spectrum. Low energy X-rays are absorbed in the tube material.

The X-ray energy determines also the emission of characteristic lines from the target material.

The major components of the modern X-ray tube are: cathode (electron source) anode (acceleration potential) rotor/stator (target device) glass/metal envelope (vacuum tube)

The figure shows a modern X-ray tube and housing assembly. Typical operation conditions are: Acceleration Voltage: 20 to 150 kV Electron Current: 1 to 5 mA (for continuous operation) Electron Current: 0.1 to 1.0 A (for short exposures)

The cathode consists of: a. a spiral of heated low resistance R tungsten wire (filament) for electron emission. Wire is heated by filament current I = U / R. ( U  10 V, I  3-6 A ) Electrons are released by thermionic emission, the electron current is determined by the temperature which depends on the wire current. The electron current is approximately 5 to 10 times less than the wire current. b. a focusing cup with a negative bias voltage applied to focus the electron distribution.

The anode is the target electrode and is maintained at a positive potential difference Va with respect to the cathode. Electrons are therefore accelerated towards the anode: E = wVa Upon impact, energy loss of electrons takes place by scattering and excitation processes, producing heat, electromagnetic radiation and X-rays.  0.5% of the electron energy is converted into X-rays.

Because of the relatively low X-ray production efficiency, most of the released energy comes in form of heat: heat generation is a major limitation for X-ray machines high melting point material with high X-ray output tungsten (high melting point) good overall radiative emission

molybdenum (high melting points) high emission of characteristic X-rays

The two major anode configurations are: The stationary anode is the classical configuration, tungsten target for X-ray production and copper block as heat sink

The rotating anode is a tungsten disc, large rotating surface area warrants heat distribution, radiative heat loss (thermally decoupled from motor to avoid overheating of the shaft)

The anode angle is defined as the angle of the target surface to the central axis of the X-ray tube. The focal spot size is the anode area that is hit by the electrons. The anode angle q determines the effective focal spot size: effective focal length = focal length • sinq The angle q also determines the X-ray field size coverage. For small angles the X-ray field extension is limited due to absorption and attenuation effects of X-ray photons parallel to the anode surface.

Typical angles are: q = T to 20°. A small angle in close distance is recommended for small spot coverage, a large angle is necessary for large area coverage.

The X-rays pass through a tube window (with low X-ray absorption) perpendicular to the electron beam. Usually the low energy component of the X-ray spectrum does not provide any information because it is completely absorbed in the body tissue of the patient. It does however contribute significantly to the absorbed dose of the patient which excess the acceptable dose limit. These lower energies are therefore filtered out by aluminum or copper absorbers of various thickness.

The minimum thickness d depends on the maximum operating potential of the X-ray tube but is typically d  2.5 mm for Va  100 kV The intensity drops exponentially with the thickness d: with eff as material dependent absorption coefficient.

The absorption coefficient is determined in terms of the Half-Value Layer HVL which is the thickness of a material necessary to reduce the intensity to 50% of its original value. The solution yields:

Graph showing how the intensity of an x-ray beam is reduced by an absorber whose linear absorption coefficient is  = 0.10 cm1.

The spectral distribution of the X-rays can be defined by the appropriate choice of filters. The filter material depends on the energy range of the original X-ray distribution!

The influence of different filter combinations for a 200 kV X-ray spectrum is shown in the figure.

The X-ray beam size is limited by a collimator system, the collimators are lead for complete absorption.

Collimator design allows to optimize the point exposure!

The size of the collimator (object size) determines the geometric "unsharpness" (blurring) of the image. The blurring B in the image is given by: where a is the effective size of the collimator of the X-ray tube and m is the image magnification: The resulting geometric unsharpeness Ug is defined: Additional unsharpeness can be caused by the image receptor (grain size, resolution of the film, etc) and by movement of the object (restless person).

For general radiography purposes the geometric unsharpeness dominates the other components Therefore the unsharpeness will increase with increasing magnification. To keep magnification small (close to m=1) requires the image receptor to be as close as possible to the patient and the focus patient distance to be large. Typical conditions are: a  1mm d1  1 m d2  10 cm

For a close dental X-ray shot the conditions are: a  1mm d1  5 cm d2  1 cm

The radiographic image of the X-ray exposure is determined by the interaction of the X-rays which are transmitted through the patient with a photon detector (film, camera etc.)

Primary X-ray photons have passed through the patient without interaction, they carry useful information. They give a measure for the probability that a photon pass through the patient without interaction which is a function of the body tissue attenuation coefficients. Secondary photons result from interaction inside the patient, they are usually deflected from their original direction and carry therefore only little information. They create background noise which degrades the contrast of the image. Scattered photons are often absorbed in grids between the patient and the image receptor.

The two dimensional image I(x, y) of the three dimensional distribution of the X-ray attenuating body tissue of the patient can be described as a function of the initial photon intensity N of energy E, the energy absorption efficiency of the image receptor (E) (film) and the attenuation coefficients  which have to be considered along the photon path in z-direction. with S(E) as distribution of the scattered secondary X-ray photons. The expression can be simplified to: with R as the ratio of secondary to primary radiation.

As higher the attenuation coefficient, as larger absorption, as lower the final intensity of the image. For bone tissue the attenuation coefficient is considerably larger than for soft body tissue, therefore increased absorption.

The quality of the image can be assessed by a few physical parameters: radiographic contrast noise and dose CONTRAST OF THE IMAGE Consider that you want to image clearly a target tissue of thickness x with an attenuation coefficient 2 inside the body of thickness t with a lower soft body tissue attenuation coefficient 1

The contrast C of the target tissue volume is defined in terms of the image distribution function I1 and I2: I1 gives the energy absorbed outside the target tissue I2 gives the energy absorbed inside the target volume. Approximating for an X-ray energy E:

The expression can be simplified to: The contrast depends mainly on the difference of attenuation coefficients 1 and 2 as well as on the ratio of scattered to primary X-ray photons. As higher the ratio R (the number of scattered photons), as lower the contrast. Therefore it is important to understand and to reduce secondary scattered photon intensity to minimize R.

The number of scattered photons depends on several parameters: X-ray field size; an increase in field size increases R 3.5 Thickness of radiated volume (increase is roughly proportional with thickness due to increase in scattering events) X-ray energy dependence - decrease of scatter with increasing energy

To reduce the number of secondary scattered photons a led grid is typically between object and image receptor. Because scattered photons will not meet the grid at normal incidence, they will be absorbed by the grid stripes.

NOISE AND DOSE Even if the imaging system may have high contrast the noise level may prevent identification of the object. Two major noise components are: statistical fluctuations in the number of X-ray photons fluctuations in the receptor and display system

The first component of the noise is called quantum noise can usually be reduced by increasing the number of photons used to form an image. This however will increase the dose absorbed by the patient which should be minimized. What is the minimum surface dose required on a body of thickness t to see a contrast C for an object of size x over an area A against a background of pure quantum noise? The signal to be detect is: The image noise in an area A results from a statistical Poisson process and can be derived as:

This yields for the signal to noise ratio SNR: An object becomes detectable if the SNR exceeds a threshold value of: At these conditions the number of incident photons N for the patient can be calculated to:

The absorbed dose D for the patient is determined by the number of photons per area N, the mass energy absorption coefficient for tissue (), and the photon energy E: The minimum dose required to visualize a fixed object increases with the fourth power of the object size. For a fixed dose and contrast there is a minimum object size which can be visualized.