A Segmented Chirped-Pulse Fourier Transform Millimeter Wave Spectrometer (260-290 GHz) with Real-time Signal Averaging Capability Brent J. Harris, Amanda.

Slides:



Advertisements
Similar presentations
Third order nonlinear optics 1. Two Photon pumping 2.. Third harmonic generation 3. Doppler free spectroscopy 4. Lambda structures.
Advertisements

Techniques for High-Bandwidth (> 30 GHz) Chirped-Pulse Millimeter/Submillimeter Spectroscopy Justin L. Neill, Amanda L. Steber, Brent J. Harris, Brooks.
CHIRPED-PULSE TERAHERTZ SPECTROSCOPY FOR BROADBAND TRACE GAS SENSING
Development of a Reduced-Cost CP-FTMW Spectrometer Using Direct Digital Synthesis Ian Finneran Daniel Holland Brandon Carroll Geoffrey Blake California.
Gas Analysis by Fourier Transform Millimeter Wave Spectroscopy Brent J. Harris, Amanda L. Steber, Kevin K. Lehmann, and Brooks H. Pate Department of Chemistry.
Jason J. Pajski, Matt Logan, Brian C. Dian 1, Gordon G. Brown, Kevin O. Douglass, Richard D. Suenram and Brooks H. Pate Department of Chemistry, University.
The Search is Over: Design and Applications of a Chirped Pulse Fourier Transform Microwave (CP- FTMW) Spectrometer for Ground State Rotational Spectroscopy.
EE 230: Optical Fiber Communication Lecture 7 From the movie Warriors of the Net Optical Amplifiers-the Basics.
Single Shot Combined Time Frequency Four Wave Mixing Andrey Shalit, Yuri Paskover and Yehiam Prior Department of Chemical Physics Weizmann Institute of.
Large Area, High Speed Photo-detectors Readout Jean-Francois Genat + On behalf and with the help of Herve Grabas +, Samuel Meehan +, Eric Oberla +, Fukun.
Microwave Spectroscopy I
Lecture 3 INFRARED SPECTROMETRY
Chapter 5 Signals and Noise  Signal carries information about the analyte that is of interest to us.  Noise is made up of extraneous information that.
Digital Communication Techniques
WH04 NUMERICAL AND EXPERIMENTAL ASPECTS OF DATA ACQUISITION AND PROCESSING IN APPLICATION TO TEMPERATURE RESOLVED 3-D SUB-MILLIMETER SPECTROSCOPY FOR ASTROPHYSICS.
Waveguide Chirped-Pulse FTMW Spectroscopy Steven T. Shipman, 1 Justin L. Neill, 1 Brett Kroncke, 1 Brooks H. Pate, 1 and P. Groner 2 1 University of Virginia.
DELIVERING MICROWAVE SPECTROSCOPY TO THE MASSES: A DESIGN OF A LOW-COST MICROWAVE SPECTROMETER OPERATING IN THE GHZ FREQUENCY RANGE Amanda L. Steber.
Spectrum Analyzer Basics Copyright 2000 Agenda Overview: What is spectrum analysis? What measurements do we make? Theory of Operation: Spectrum analyzer.
Ni.com Data Analysis: Time and Frequency Domain. ni.com Typical Data Acquisition System.
June 21st OSU Mol. Spect. Symp.TB02 8:42 a.m Observation of Infrared Free Induction Decay and Optical Nutation Signals from Nitrous Oxide Using.
An Acoustic Demonstration Model for CW and Pulsed Spectroscopy Experiments Torben Starck, Heinrich Mäder Institut für Physikalische Chemie Christian-Albrechts-Universität.
Sebastian Böser Acoustic sensor and transmitter development Amanda/IceCube Collaboration Meeting Berkeley March 2005.
Chirped Pulse Fourier Transform Microwave Spectroscopy of SnCl Garry S. Grubbs II and Stephen A. Cooke Department of Chemistry, University of North Texas,
Resident Categorical Course
IR/THz Double Resonance Spectroscopy in the Pressure Broadened Regime: A Path Towards Atmospheric Gas Sensing Sree H. Srikantaiah Dane J. Phillips Frank.
Spectroscopy with comb-referenced diode lasers
OSU 06/19/08 Ultrabroadband Rotational Spectroscopy: Novel Applications of a Shape Sensitive Detector BRIAN C. DIAN Purdue University Department of Chemistry.
Two-Dimensional Chirped-Pulse Fourier Transform Microwave Spectroscopy Amanda Shirar June 22, th OSU International Symposium on Molecular Spectroscopy.
Chirped-Pulse Fourier Transform mm-Wave Spectroscopy from GHz Brent J. Harris, Amanda L. Steber, Justin L. Neill *, Brooks H. Pate University of.
Kelly Hotopp June 22, 2010 Purdue University.  Demonstration of 2D CP-FTMW spectroscopy ◦ Non-Selective Excitation ◦ Selective Excitation  2D CP-FTMW.
Tunable Mid-IR Frequency Comb for Molecular Spectroscopy
Spectral taxonomy: A semi-automated combination of chirped- pulse and cavity Fourier transform microwave spectroscopy Kyle N. Crabtree, Marie A. Martin-Drumel,
ULTRAVIOLET - CHIRPED PULSE FOURIER TRANSFORM MICROWAVE (UV-CPFTMW) DOUBLE-RESONANCE SPECTROSCOPY Brian C. Dian, Kevin O. Douglass, Gordon G. Brown, Jason.
Spectral Simplification Methods Development Using Waveguide Chirped-Pulse Fourier Transform Microwave Spectroscopy Erin Kent, Steven Shipman New College.
OSU 06/18/08 Ultrabroadband Rotational Spectroscopy: Novel Applications of a Shape Sensitive Detector BRIAN C. DIAN Purdue University Department of Chemistry.
Novel Applications of a Shape Sensitive Detector 2: Double Resonance Amanda Shirar Purdue University Molecular Spectroscopy Symposium June 19, 2008.
Australian Astronomy MNRF Development of Monolithic Microwave Integrated Circuits (MMIC) ATCA Broadband Backend (CABB)
SPECTRUM ANALYZER 9 kHz GHz
Haifeng Huang and Kevin K. Lehmann
Figure 8.3 gives the basic layout of a continuous wave NMR spectrometer. These intruments were the original type of instrument and have largely.
Amanda L. Steber, Brent J. Harris, Justin L. Neill, Kevin K. Lehmann, Brooks H. Pate Department of Chemistry, University of Virginia, McCormick Rd., P.O.
P.W. Gorham et al.. TEST BEAM A SLAC Time relative to beam entry Antenna V/V rms Time relative to beam entry Antenna V/V rms close to shower maximumshower.
Generation of Spurious Signals in Nonlinear Frequency Conversion Tyler Brewer, Russell Barbour, Zeb Barber.
Primarily, Chapter 3 Claridge Also, Chapter 2 Claridge,
Tze-Wei Liu Y-C Hsu & Wang-Yau Cheng
Time-resolved Fourier transform infrared emission spectra of HNC/HCN K. Kawaguchi & A. Fujimoto Okayama University.
Fast Sweeping Direct Absorption (sub)Millimeter Spectroscopy Based on Chirped Pulse Technology Brian Hays 1, Steve Shipman 2, Susanna Widicus Weaver 1.
1 Dual Etalon Frequency Comb Spectrometer David W. Chandler and Kevin E. Strecker Sandia National Laboratories – Biological and Energy Sciences Division.
Infrared--Microwave Double Resonance Spectroscopy of Ar-DF (v = 0,1,2) Justin L. Neill, Gordon G. Brown, and Brooks H. Pate University of Virginia Department.
FAST SCAN SUBMILLIMETER SPECTROSCOPIC TECHNIQUE (FASSST). IVAN R. MEDVEDEV, BRENDA P. WINNEWISSER, MANFRED WINNEWISSER, FRANK C. DE LUCIA, DOUGLAS T. PETKIE,
An Experimental Approach to the Prediction of Complete Millimeter and Submillimeter Spectra at Astrophysical Temperatures Ivan Medvedev and Frank C. De.
Fast Sweeping Double Resonance Microwave - (sub)Millimeter Spectrometer Based on Chirped Pulse Technology Brian Hays 1, Susanna Widicus Weaver 1, Steve.
Direct Observation of Rydberg–Rydberg Transitions in Calcium Atoms K. Kuyanov-Prozument, A.P. Colombo, Y. Zhou, G.B. Park, V.S. Petrović, and R.W. Field.
Beam Diagnostics Seminar, Nov.05, 2009 Das Tune-Meßverfahren für das neue POSI am SIS-18 U. Rauch GSI - Strahldiagnose.
Digital Control System for Microwave Spectroscopy Data Collection Amanda Olmut Dr. Stephen Kukolich, Principle Investigator Dr. Adam Daly, Project Lead.
SEEING IS BELIEVING: An 11 GHz molecular beam rotational spectrum (7.5 – 18.5 GHz) with 100 kHz resolution in 15  s measurement time Brian C. Dian, Kevin.
Primarily, Chapter 3 Claridge Also, Chapter 2 Claridge,
Max Planck Institute for the Structure and Dynamics of Matter
Texas Instruments Introduction to Direct RF Sampling
AMANDA L. STEBER, MARIYAM FATIMA, CRISTÓBAL PÉREZ, and MELANIE SCHNELL
Atmospheric Remote Sensing Via Infrared-Submillimeter Double Resonance
(Instrument part) Thanundon Kongnok M
Multiplexed saturation spectroscopy with electro-optic frequency combs
Optical Coherence Tomography
Two-Photon Absorption Spectroscopy of Rubidium
Frequency Band Performance Comparisons for Room-Temperature Chirped Pulse Millimeter Wave Spectroscopy Justin L. Neill, Brent J. Harris, Robin L. Pulliam,
Principle of Mode Locking
CHIRALITY DETERMINATION FROM PULSED-JET FOURIER TRANSFORM
Tunable Slow Light in Cesium Vapor
Beam Current Monitoring with ICT and BPM Electronics
Presentation transcript:

A Segmented Chirped-Pulse Fourier Transform Millimeter Wave Spectrometer ( GHz) with Real-time Signal Averaging Capability Brent J. Harris, Amanda L. Steber, and Brooks H.Pate Department of Chemistry University of Virginia

Segmented CP-FT: Single Spectrum Acquired in 100  s 100 µs, 35 GHz (260 – 295 GHz) Simple Benchtop Prototype 65 cm sample cell

Advantages for Chirped Pulse mm-wave Spectroscopy A. Steber, B. Harris, J. Neill, B. Pate, J. Mol. Spec 280 (2012) )Rapid spectrum acquisition (35 GHz spectrum acquired in 2  s) 2)Coherent measurement methods for spectrum analysis 3)Transition frequency accuracy 4)Simple spectrum signal processing (no baseline effects) Challenges for Chirped Pulse mm-wave Spectroscopy 1)Real-time signal averaging to achieve high sensitivity 2)Reduction of spurious signals 3)Spectral resolution and Fourier transform signal processing

Instrument Design Principles: The Chirped Pulse 1)The Chirped Pulsed Separates the Pulse Duration from the Pulse Bandwidth High-Resolution Spectroscopy: Line width is much smaller than frequency range Line width: set by the signal dephasing (Doppler, Collisional Relaxation) Optimal Excitation Pulse Duration: ~ 1 / (line width) Transform Limited Pulse Duration: ~ 1 / (frequency range) 2) In the Weak Pulse Limit the Signal is proportional to (Bandwidth) -1/2 Signals for transform limited pulses scale with (Bandwidth) -1 : Spectral energy density reduced by both increased bandwidth and shorter pulse duration.

Instrument Design Principles: Segmenting Segmenting the Measurement Frequency Range Relative Segment Bandwidth: (1/N) Rel.Signal in Segment: (2) ½ Reduction in Averages: ½ Rel. Measurement Time: 2 * ½ = 1 Full Bandwidth CP-FT Rel. Signal in Segment: (N) ½ Reduction in Averages: (1/N) Rel. Measurement Time: N * (1/N) = 1 Things that are segment independent (all things being equal): 1)The time required to reach a target dynamic range (signal-to-noise ratio) 2)The number of points that make up the spectrum For smaller segments, lower digitizer rates can be used (Transform limited pulses: increasing the segment bandwidth increases the measurement time!)

Instrument Design Principles: Segmenting Segmenting the Measurement Frequency Range Things that are segment dependent (all things being equal): 1)The time required for one spectrum scan (N times longer) 2)The total number of data points processed (Reduced by N) Molecules perform some of the averaging Things that are not equal: 1)Digitizer cost and real-time signal averaging performance (favors more segments) Still need bandwidth to use power efficiently and minimizing stitching 2) Proliferation of spurious signals (favors more segments) Exception: Single Segment has best mm-wave practical spur performance

GHz CP-FT Design for Real Time Averaging Number of Spectrum Points: 35 GHz frequency range, ~ 2  s dephasing time Minimum Points: (70 Gs/s) x (2  s) = 140,000 This number of data points can be accommodated in FPGA accumulator 512K points maximum, 8-bit digitizer, 32-bit data size Maximum real-time accumulations: 2 (32-8) = 2 24 = 16M Minimizing the Spurious Signals:Strongly Dependent on AWG purity for LO generation LO Impurity Spurs:LO-to-IF Conversion (self mixing): Easily subtracted Spectrum Images: Improved AWG performance on N*30 MHz This fixes the segment bandwidth to 30MHz x 24 = 720 MHz: 50 segments, 100  s scan time Digitizer/Receiver Spurs: Second Order IM: work in 2 nd Nyquist zone (720 – 1440 MHz) Third Order (Two-tone IM) Signal/Clock Mixing in Digitizer

Segmented Chirped-Pulse Fourier Transform Spectroscopy Separate AWG Channels Generate Chirp Segments (Blue) and Local Oscillator (LO) Frequency (Red) with Phase Reproducibility Output Power: mW GHz GHz 260– 295 GHz 220 – 325 GHz WR 3.4 Low IF (720 – 1440 MHz) 4GS Digitizer

High Speed Segmented CP-FT Measurement Digitized at an IF of 720 – 1440 MHz at 4GS/s Real-Time signal averaging in 32 bit FPGA : sample memory (131 µs, at 4GS/s) 2 µs segments: µs chirped pulse µs decay time -720 MHz bandwidth (per segment) -50 segments (36 GHz, 100 µs )

Spectrum Relative Intensity Performance Frequency dependent spectrometer response: -Source power -Receiver mixer loss -IF amplifier gain -Propagation loss System response correction: -Routine response curve generated with series of stored single frequency waveforms -Correction incorporated into signal processing About 10-15% accuracy

Intensity Reproducibility Intensity variation < 0.1% Suitable for monitoring time evolution of the broadband spectrum

Comparison to Literature mm-wave Benchmark S. Fortman, I. Medvedev, C. Neese, F. De Lucia, ApJ, 725 (2010) 1682 CH 3 CH 2 CN 0.5 mTorr, 6 m path length, 24 s equivalent measurement time Speed: CP-FT is 10,000 times faster Intensity Accuracy: 15% vs. 1% (but 0.1% precision) Line width: 2.5 broader in CP-FT (magnitude FT + windowing)

High Dynamic Range Mode (HDR) HDR sequence: µs chirped pulse µs decay time -24 MHz bandwidth/seg -30 segments for each LO frequency (720mHz) -31 st segment for electrical background subtraction -50 LO frequencies (1500 total segments) 62  s for 720 MHz Two 720 MHz segments can be accommodated in the FPGA Noise Floor Achieved in 1s: 1  V (60,000:1 dynamic range)

Broadband Double Resonance Spectroscopy Acrolein Measurement Protocol: 1) Single Frequency Pump Pulse 2)720 MHz Chirp Observe ~60% signal modulation Sequential and V- Type Level Schemes Show Different Modulation Behavior

Conclusions 1)Real-time Signal Averaging is Achieved using Segmented Chirped Pulse Fourier Transform Spectroscopy 2)High Speed Mode Significantly Reduces Measurement Times Over Absorption Spectroscopy Methods 3)A High Dynamic Range Mode Provides Spurious Signal Reduction and Achieves about 100,000:1 (Largely) Spur Free Performance 4)Capabilities for Broadband Double Resonance Spectroscopy Can Translate Reduced Spectrum Acquisition Times to Reduced Spectrum Analysis Times Acknowledgments Brent Harris is supported by an NSF Graduate Fellowship University of Virginia Equipment Trust Fund NSF I-Corps