PH - MEASUREMENT pH : - It is a convenient measure of acidity / alkalinity of a aqueous solution at a specific temperature. It is measured on a continuous.

Slides:



Advertisements
Similar presentations
Analyze-Immediately Parameters pH Analyze Immediately Within 15 minutes of collection: pH Residual Chlorine Temperature Chlorine Dioxide Dissolved Oxygen.
Advertisements

SVCC Respiratory Care Programs Analysis of ABG Samples.
PH measurements. Measuring the pH of Solutions The activity of a single ion cannot be measured Only measure our ‘best approximation’ to the hydrogen ion.
Summary of Potentiometry:
Lecture 6a Cyclic Voltammetry.
ISE Ion Selective Electrodes Prepared By Michigan Department of Environmental Quality Operator Training and Certification Unit.
Electrodes and Potentiometry
Section 11 Potentiometry. Potentiometric Electrodes Potentiometric electrodes measure: Activity not concentration Concepts to review: Activity and affect.
14-6 Cells as chemical probes
Potentiometry. pH is a Unit of Measurement  pH = Power of Hydrogen (H + )  Defined as the Negative Logarithm of Hydrogen Ion Activity  pH = log (1/H.
The world leader in serving science Technical pH Seminar.
Chem. 133 – 3/5 Lecture. Announcements Lab –Set 2 Period 2 Labs Set to Finish 3/10 –3/12 will be make up day (for both Period 1 and Period 2 labs) –Set.
Chapter 15 Solutions. Chapter 15 Table of Contents Copyright © Cengage Learning. All rights reserved Solubility 15.2 Solution Composition: An Introduction.
1 Objective:The objective of this experiment is to determine the mass of acetylsalicylic acid (ASA) in an Aspirin tablet. Chemicals Potassium hydrogen.
Chapter 3 – Volumetric Analysis Week 1, Lesson 1.
ANALYTICAL CHEMISTRY CHEM 3811 CHAPTER 15 DR. AUGUSTINE OFORI AGYEMAN Assistant professor of chemistry Department of natural sciences Clayton state university.
Selection of measuring instruments
Titremetric analysis Dr. Mohammad Khanfar. Concept of Titremetric analysis In general, we utilize certain property of a substance to be analyzed in order.
Gravimetric Analysis l A Gravimetric analysis is based upon the measurement of the weight of a substance that has a KNOWN composition AND IS chemically.
Chapter 14 Principles of Neutralization Titrations
This teaching material has been made freely available by the KEMRI-Wellcome Trust (Kilifi, Kenya). You can freely download,
Classification of Instruments :
IC Controls Quality Water Solutions for pH R1.0 © 2004 IC CONTROLS pH / ORP Conductivity Dissolved Oxygen Chlorine Standards An Overview.
Chemometrics Method comparison
Acids and Bases Biotechnology I. Life Chemistry  Based on water  Cells contain 80-90% water  Proper pH essential to ALL living systems Plants cannot.
Standardisation of potassium permanganate solution Ex 5
Solutions Solubility -the amount of solute that can be dissolved to form a solution. Solvent – the substance in a solution present in the greatest amount.
Chapter 15 Table of Contents
Chemistry 232 Applications of Aqueous Equilbria. The Brønsted Definitions  Brønsted Acid  proton donor  Brønsted Base  proton acceptor  Conjugate.
A Seminar on Potentiometric and Diazotization Titrations
Analytical chemistry MLAB 243 Level 4 Lecture time: every WED 8 -10
CHEMISTRY. Composition of Matter Matter - Everything in universe is composed of matter Matter is anything that occupies space or has mass Mass – quantity.
CHEMISTRY. Composition of Matter Matter - Everything in universe is composed of matter Matter is anything that occupies space or has mass Mass – quantity.
THE MEASUREMENT OF pH. I GENERAL PRINCIPLES Convenient way to express hydrogen ion concentration, or acidity pH = - log[ H +
Chemistry I Unit 11: Acids and Bases Text Questions from Wilbraham, et. al Addison Wesley Chemistry by Michael S. Matta, Dennis D. Staley, A. Wilbraham,
Chapter 15 Preview Lesson Starter Objectives
Solutions.
CHAPTER 11 ELEMENTS OF ELECTROCHEMISTRY Introduction to Analytical Chemistry.
1 Gravimetric Analysis. 2 Gravimetric analysis is the quantitative determination of analyte concentration through a process of precipitation of the analyte,
Solutions used in medical laboratory. O Solution is a homogeneous mixture of two or more substances. O Solute is the dissolved substance, whereas solvent.
Introduction The Equipment The Terms The Process Calculations
1 Titrations (Review) In a titration a solution of accurately known concentration is added gradually added to another solution of unknown concentration.
Preparation of BUFFER. Buffers - Buffers are the solutions which resist changes in pH when small - amounts of acid or alkali is added to them. - A buffer.
PH meter. What is pH? pH is a measure of the activity of the hydrogen ion in a solution. The pH value states the degree of the hydrogen ion activity in.
1 Chapter 2 Steps in a chemical analysis Plan of analysis Before doing any quantitative analysis, the following questions should be answered: 1-
Arrhenius Acids and Bases Acid: Acid: A substance that produces H 3 O + ions in aqueous solution. Base: Base: A substance that produces OH - ions in aqueous.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
Petra Spitzer, Ralf Eberhardt, Janine Giera PTB, Germany Barbara Werner, Anett Czysch ZMK, Germany Traceable calibration procedures in chemistry Example:
PH THEORY What is it pH?What is it pH?/How is pH Measured?/What Equipment is Required to Measure pH?/How is a pH Measurement Device Calibrated?/Why is.
Ch. 21 Potentiometry 1. General principles
PH meter Bushra mubarak. pH pH is defined as the negative logarithm of the hydrogen ion activity aH. Unit of measure which describes the degree of acidity.
Chapter 14 Principles of Neutralization Titrations.
Stoichiometry: Quantitative Information About Chemical Reactions Chapter 4.
POTENTIALS AND THERMODYNAMICS OF CELLS (1) POTENTIALS AND THERMODYNAMICS OF CELLS (1)
Chapter 5: Errors in Chemical Analysis. Errors are caused by faulty calibrations or standardizations or by random variations and uncertainties in results.
Chapter 1: The Nature of Analytical Chemistry
Analytical biochemistry laboratory (BIOC 343) L. Nouf Aalshareef KAU- Biochemistry department Second semester 2013.
Hydronium Ions and Hydroxide Ions Self-Ionization of Water In the self-ionization of water, two water molecules produce a hydronium ion and a hydroxide.
Chapter 13 Titrations in Analytical Chemistry. Titration methods are based on determining the quantity of a reagent of known concentration that is required.
Fundamentals of Analytical Chemistry Eighth Edition.
PH - MEASUREMENT pH: - It is a convenient measure of acidity / alkalinity of a aqueous solution at a specific temperature. It is measured on a continuous.
Practical Analytical Chemistry (1)
PH & Buffers.
Ion Selective Electrode
Conductivity Lecture.
Titration Volumetric analysis
Chapter 3 – Volumetric Analysis
Instrumental Methods II
Intellical Red-ROD technology
EXP. NO. 6 Acid Base Titration
Presentation transcript:

pH - MEASUREMENT pH : - It is a convenient measure of acidity / alkalinity of a aqueous solution at a specific temperature. It is measured on a continuous scale from 0 to 14 In a chemical laboratory, pH is commonly measured using an electronic pH meter with a scale 0.01 pH meter

pH - MEASUREMENT WHY IS pH MEASURED? To test a sample against a legal requirement To test a chemical against a specification As a part of analytical method Process control in chemical industry Environmental monitoring of waste and effluents Monitoring and controlling biochemical reactions, many of which only take place in a particular and sometimes narrow pH range.

pH - MEASUREMENT PRINCIPLES: - The pH value of a given solution is a measure of the activity of the hydrogen ion (H + ) in that solution The scale is logarithmic In a solution of pH 5, the hydrogen ion activity is 10 times higher than it is in a solution of pH 6 In aqueous solutions, the actual species is the hydronium ion H 3 +, rather than the hydrogen ion H + pH = log10 1/a H + = -log10aH + Activity of a species is a measure of its effective or available concentration rather than its actual or total concentration. aH + = γ x cH + γ = activity of coefficient

pH - MEASUREMENT PRINCIPLES: - In dilute solution (< mol) of simple univalent electrolyte γ approximate to unity and activity and concentration are approximately equal In other type of aqueous solution, there is no easily determined relationship between activity and concentration From the theoretical, thermodynamic view point, the activity of single ion species such as H+ is an exact quantity

pH - MEASUREMENT PRINCIPLES: - pH value to be regarded as convenient and comparitative measure of acidity pH is usually determined by electrochemical measurement, in which the potential of a pH electrode immersed in the test solution is measured. The pH electrode respond quantitatively and specifically to hydrogen ions even in the presence of other positive ions

pH - MEASUREMENT ELECTROMETRIC DETERMINATION OF pH The pH measurement is possible because emf of certain chemical cell varies with the hydrogen ion concentration of the solution on the cell. pH electrode/test solution to be measured// reference electrode symbol//signifies the presence of a liquid junction between test solution and reference electrode If other variables in the cells are controlled, emf of the cell can be correlated with pH Pt, H 2 (p) \H+ (a) \ reference electrode. - left hand electrode is the hydrogen electrode - right hand electrode whose potential is not affected by pH

pH - MEASUREMENT ELECTROMETRIC DETERMINATION OF pH E cell = E H2, H + + E ref E cell - E ref = E H2, H + E cell - E ref = RT/nF x log10 aH + R = gas constant (8.314 joule k-1mol-1) T = Absolute temperature (Kelvin) F = A electrochemical constant (96487 coulomb mol-1) N = 1 for univalent species such as H + E cell - E ref = pH

pH - MEASUREMENT ELECTROMETRIC DETERMINATION OF pH Theoretical slope of a pH electrode is such that change in pH of 1 unit results in a change in the potential (E) of the electrode of 59.1 mV at 25°C Mathematically Delta E / Delta pH = 59.1 mV The value of E Ref include the following - The standard potential of the pH electrode (i.e. the potential when a H + = 1) - The potential of the reference electrode - The liquid junction potential - The asymmetry potential

pH - MEASUREMENT ELECTROMETRIC DETERMINATION OF pH The value of E ref cannot be determined in routine measurements mainly due to the indeterminate - Nature of the liquid junction potential - Asymmetry potential - Value of asymmetry potential tends to drift as the condition of the electrodes changes - The equation not used in a basic way to determine pH of a solution - The response of the glass electrode is calibrated using standard aqueous buffer solutions with known reference pH values. Reading of pH meter adjusted so that they correspond - to the reference values.

pH - MEASUREMENT ELECTROMETRIC DETERMINATION OF pH The pH of a test solution may then be measured pH = -log a H + pH measurement reflects the activity rather the concentration of hydrogen ion

pH - MEASUREMENT BUFFERS:- The resistance of a solution to change in Hydrogen ion concentration upon the addition of small amounts of acid or alkali is termed as buffer action and a solution which posses this property is called buffer solution. The reference pH values for primary pH standards are established by high accuracy potentiometric measurements Using specially designed electrochemical cells and platinum- hydrogen gas electrode rather than a glass electrode, as the H + - sensing electrode Avoids error in potential measurement arising from liquid junction potential and the asymmetry potential

pH - MEASUREMENT BUFFERS:- Enables the pH of the standard solution to be determined from the Nernest equation. Such specialized measurements done at NIST NIST supplies a number of high purity salts as standard reference materials for pH Each of which has a certificate Detailed instructions of preparation and use of solution The pH values certified to 3 decimal places with typical uncertainties of ± pH unit at a range temperature Enables pH meters to be calibrated in a manner that is traceable to a services of internationally recognized standards

pH - MEASUREMENT HIGH PURITY SALTS USED AS PRIMARY pH STANDARD Note: The uncertainty of the tabulated pH values are estimated to be ± 0.01 High purity saltsConcentration/g/LpH at stated temperature 15°C20°C25°C30°C Potassium tetraoxolate Potassium hydrogen phthalate Potassium di-hydrogen phosphate Di sodium hydrogen phosphate Sodium tetra borate decahydrate Sodium hydrogen carbonate Sodium carbonate

pH - MEASUREMENT BUFFERS:- Buffer solutions to be prepared using salts of the highest purity available Certain of the salts should be dried before use Potassium hydrogen phthalate- Dry at 110°C for one hour Potassium di-hydrogen phosphate - Dry at 110°C for one hour Disodium hydrogen phosphate - Dry at 110°C for one hour Sodium carbonate – Dry at 270°C for one hour Carbon dioxide-free distilled water to be used to prepare the solutions Important for those buffer solution with a pH > 6

pH - MEASUREMENT BUFFERS:- Prepared solutions to be stored in a well stoppered Pyrex or polythene bottles Normally to be replaced after 2 to 4 weeks and sooner if mould or sediment is observed

pH - MEASUREMENT GLASS ELECTRODE A – Glass bulb B – Tube filled with 0.1 mol HCl C – Silver-Silver chloride D – Saturated KCl solution saturated with AgCl E – Silver-Silver chloride electrode

pH - MEASUREMENT GLASS ELECTRODE Potential is developed in an aqueous solution is proportional to hydrogen activity or pH of the solution Hydrogen ion in the solution forms a dynamic equilibrium with hydrogen ions that are “bound” to the membrane surface in ion exchange type process, thereby establishing a potential across the membrane Magnitude of potential is proportional to the pH of the solution

pH - MEASUREMENT CARE OF pH ELECTRODES Manufacturer’s instructions to be consulted for specific guidance on particular electrodes and situation For reliable pH measurement, pH electrodes are properly stored and maintained Response of the pH electrode to hydrogen ion activity depends acutely on such matters as: - Cleanliness and condition of the glass membrane - Conditions of the reference electrode liquid junction - Conditions of reference electrode and its filling solution - Cleanliness of the electrode body

pH - MEASUREMENT CAR E OF pH ELECTRODES On exposure to air and allowing to dry out, glass membrane becomes dehydrated and liquid junction may also deteriorate pH electrode should be stored in an appropriate storage solution, when not in use Mixture of different salts of similar composition to the reference electrode filling solution, a buffer to provide suitable pH(e.g.4 to7) and mild cleansing agent Electrode to be immersed to a sufficient depth to cover both the glass membrane and the liquid junction A pH electrode should not be stored in water Periodically the level of the filling solution in reference electrode should be checked and topped-up as necessary

pH - MEASUREMENT CARE OF pH ELECTRODES When not in use, the filling hole of the reference electrode should be covered with the plug provided for this If pH electrode has been left in air and dried out, it should be re- hydrated by immersion in the storage solution for at least 12 hours A pH electrode should be regularly examined for salt crystal build up and membrane and liquid junction deposits For additional cleaning soak the pH electrode in either 0.1 molar HCl or 0.1 molar Nitric acid

pH - MEASUREMENT SELECTION OF pH ELECTRODE: For pH measurements in the range 1 to 10, general purpose glass electrodes suitable When the concentration of alkali metal ions become large these ions seems to permeate the membrane, resulting in a significant errors in the measured pH This is referred to as ‘alkaline error’ or a negative error in pH is observed New glasses rich in lithium have been introduced and gives reliable results At very low pH values the activity of water is reduced due to salting out type effect arising from very high concentration of Hydrogen ions Nature of glass membrane is altered and values are higher than true values. This is referred as acidic errors

pH - MEASUREMENT ACCURACY Electrodes available that are capable of measuring pH to typically 0.01, 0.02, 0.02 and 0.1 pH unit Possible to record pH to pH unit, only done to study small differences or changes in pH

pH - MEASUREMENT COMPOSITE LIQUIDS Composite liquid samples such colloids, suspensions sludge, slurries, emulsions etc presents number of difficulties for pH measurements The liquid junction easily becomes clogged and contaminated, leading to poor and inaccurate response Electrodes with modified liquid junctions (e.g. sleeve or double junctions) may over come the problem Attention to correct depth of immersion of the electrode in the liquid and electrode cleaning after use are important

pH - MEASUREMENT CALIBRATION Essential to calibrate the response of a pH electrode by using standard buffer solutions of known pH At least two buffer solution should be used pH of the test sample is bracketed by two buffers The chosen buffers should not be more than 3 pH unit or no less than 1 pH unit apart To be free from contamination, sediment and mould Each laboratory to determine the shelf life of their buffer solutions based on practical experience, frequency of use, storage conditions and the importance of the pH measurements being made.

pH - MEASUREMENT CALIBRATION When a buffer solutions requirement for use, a suitable quantity should be transferred to clean dry beaker and close buffer solution bottle quickly, used buffer solution should not be returned to the stock bottle and pH electrode should not be immersed directly into stock bottle Ensure that buffer solutions and test solutions whose pH values are to be measured are equilibrated to ambient temperature

pH - MEASUREMENT USE OF BUFFER SOLUTION:- First buffer solution - Immerse the pH electrode in the first buffer solution ensuring that the glass membrane and the liquid junction are covered by the solution - The depth of immersion should be such that the level of the filling liquid in the reference compartment is about 25 mm above the level of the solution being measured - This will ensure a uniform flow of filling solution through the liquid junction and a stable liquid junction potential - Solution should be stirred using a magnetic stirrer at a moderate speed taking care not to contact and damage the glass membrane

pH - MEASUREMENT USE OF BUFFER SOLUTION:- - When the reading has stabilized to within about ± 0.02 pH unit, adjust the reading using ‘set-buffer’ control so that the displayed reading is equal to the reference pH value of the buffer solution - Discard the buffer solution and re-fill the beaker with a fresh portion of the same buffer - Observe the pH reading, if it is not within about ±0.02 pH unit of the reference value, re-adjust the ‘set-buffer’ control so that the displayed reading is equal to the reference pH value of the buffer. - Repeat this procedure until two successive readings agree to within about ±0.02 pH unit - Record details of buffer solution and relevant results obtained

pH - MEASUREMENT SECOND BUFFER SOLUTION - Rinse the beaker with water and discard rinsing - Dry the interior of the beaker with a tissue and add the required amount of the second buffer solution - Rinse the electrode with water and then the buffer solution and dry by touching with a tissue - Immerse the electrode in the second buffer solution, allow the reading to stabilize to within about ±0.02 pH unit and note the pH reading - If the electrode is giving a Nernstian response, the displayed pH reading will not differ significantly (i.e. by more than about ±0.04) from the reference value of 2nd buffer - Adjust the displayed reading if necessary to equal the reference pH values of the second buffer

pH - MEASUREMENT SECOND BUFFER SOLUTION - Make this adjustment using slope control - The adjust required should not be unduly large, typically no more than ±0.3 pH unit - If the difference between the displayed readings and reference value for the second buffer solution exceeds +0.3 unit, it indicates electrode response could be markedly non- Nernstanian - That is electrode slope value is significantly different from the theoretical value of 59.1mV per ph unit - Manufacturer of electrodes often specify the slope value for an electrode in good conditions

pH - MEASUREMENT SECOND BUFFER SOLUTION - Specified slope values are typically in the range of 90 – 105% - Discard buffer solution, replace it with fresh portion of the second buffer, immerse the electrode and observe the displayed pH value. It should be within ±0.02 pH unit of the reference value

pH - MEASUREMENT THE COMPOSITION OF THE AQUEOUS SOLUTION: - Aqueous solution with a relatively high ionic strength is to be measured, a special pH electrode with a modified reference electrode, to counter act effects due to liquid junction potential Water or aqueous solution of very low ionic strength are to be measured, response of the glass electrode is slow, noisy, subject to drift and inaccurate These samples have low buffering capacity and absorption of carbon-dioxide from the atmosphere can cause unsteady reading Use of additive to increase ionic strength of the sample Sample contains components that react with silver, the use of a pH electrode with a silver/silver chloride reference electrode is prohibited.

pH - MEASUREMENT THE COMPOSITION OF THE AQUEOUS SOLUTION: - Reaction products deposited in the liquid junction causing a slow response or no response Sample causing such problems include waste waters and lea hates containing sulfur compounds These samples have low buffering capacity and absorption of carbon-dioxide from the atmosphere can cause unsteady reading Use of additive to increase ionic strength of the sample Sample contains components that react with silver, the use of a pH electrode with a silver/silver chloride reference electrode is prohibited.

pH - MEASUREMENT MEASUREMENT OF TEST SOLUTION - Rinse the beaker with water and discard rinsing - Rinse it with test solution and add required amount of the test solution whose pH is to be determined - Rinse the electrode with water and then with test solution and dry by touching with a tissue - Immerse the electrode in the test solution and observe the pH reading - Discard the test solution and replace it with fresh portion of the test solution and re-measure the pH - For a well buffered test solution duplicate pH result should normally agree within about ± 0.02 pH unit - For slightly buffered test solution, replicate results may only agree within about ±0.1 pH unit

pH - MEASUREMENT MEASUREMENT OF TEST SOLUTION - Record pH values obtained - It lengthy series of measurements is made, re-measure the buffer solutions at regular intervals e.g. every 10 test solutions

pH - MEASUREMENT UNCERTAINTY OF PH MEASUREMENTS Using a glass electrode Below are some possible source of uncertainty in pH measurements along with the estimate of their magnitude Source of uncertaintyStandard Uncertainty pH value of the standard buffer solutions±0.01 The repeatability of pH measurements±0.01 Drift in the electrode response between the calibration ±0.01 The liquid junction potential±0.015 Temperature (where known to ±1°C)0.005

pH - MEASUREMENT THE STANDARD UNCERTAINTIES IN THE TABLE ARE IN THE FORM OF STANDARD DEVIATION Estimates of the contribution that each of these sources of uncertainty makes to the overall error of a pH measurement are some what subjective Vary with the precise conditions of the measurement and equipment used Using above data, the total standard uncertainty in a pH measurement is estimated to be about ±0.025 pH unit When temperature difference is ± 5°C, the total standard uncertainty is about ±0.035 pH units For very dilute or concentrated test solutions higher uncertainty may apply

pH - MEASUREMENT THE ANALYST MUST KNOW: The meaning of pH and the pH scale Acid/base reactions, the relationship or pH to acidity and basicity, the Nernest equation The different techniques for measuring pH (indicators, pH papers and pH meters) How to operate a pH meter – read the appropriate instruction manuals How to follow written instructions and obtain satisfactory results – standard test procedures How to check that the equipment is working correctly and what to do if there is a problem

pH - MEASUREMENT THE ANALYST MUST KNOW: The effect of temperature on pH - Information issued with buffer solutions states the temperature at which the pH should be determined - Samples should be allowed to equilibrate to the same temperature as the buffer solutions before measurement - Some pH meters have a thermometer on the probe, these will correct for temperature differences The importance of a homogeneous solution That the solutions used for calibration must be relevant to the pH range of the sample

pH - MEASUREMENT PRACTICAL TECHNIQUE : ESSENTIAL KNOWLEDGE SHOULD INCLUDE: Knowing how to stir solutions without damaging probe Knowing when to decide when a reading is stable Knowing how to achieve acceptable precision when measuring the pH of samples Knowing how to clean and maintain a pH electrode - General cleaning - Wash with de-ionised water before and after each measurement - Removal of deposits - Storage requirements, electrodes should be stored in a solution appropriate to the filling medium

pH - MEASUREMENT ANALYSTS SHOULD BE AWARE OF: The limitations of the equipment in terms of accuracy Interferences: - Potential problems measuring solutions with very high or very low pH values - Measurement of pH in non-aqueous ·solutions - Matrix effects of the samples under test, e.g., samples with high solids or high organic solvent content Troubleshooting - Drifting readings - stability of the reading should be obtained - within 2 minutes - Slow response - Erratic readings

pH - MEASUREMENT ANALYSTS SHOULD BE AWARE OF: Recording results and logbook entries - accurate transcription How to set up control charts The level of precision possible to achieve with the available instrument

pH - MEASUREMENT CALIBRATION : Buffer standards - Know how they work - Know different types of buffer and how to use them - Know how to make up buffer solutions for calibration - Know how to store buffer standard solutions correctly - Know that it is necessary to use more than one buffer standard (one either side of the expected pH range of samples to be tested)

pH - MEASUREMENT CALIBRATION : Calibrate the instrument according to the operation manual or an STP - Check electrode slope is within manufacturers' tolerance - Know when to change the electrode, e.g., when the slope reading is not within the tolerance limits - Know how often the pH meter needs calibrating - Check that the pH meter has maintained its performance in the event of a power cut

pH - MEASUREMENT OBSERVATION SIGNS WHICH INDICATE THAT ALL IS NOT WELL Messy bench area where the analyst has been working Liquid spilled on the work area Tops left off reagent bottles - bottles not returned to the shelf Solutions labeled inadequately, etc. - also labeled correctly but unreadable due to damaged label Tops left off buffer solutions or solutions inappropriately stored Evidence of incorrect care of pH electrode - build up of deposits on surface etc Dirty glassware left lying about rather than being cleaned and put away

pH - MEASUREMENT OBSERVATION SIGNS WHICH INDICATE THAT ALL IS NOT WELL Lack of notes in work book relating to calibration of equipment or temperature of measurement if appropriate crossings out in workbook, indicating that the analyst was having trouble deciding when a stable reading had been reached Only one value recorded - no evidence that a stable reading had been achieved.

pH - MEASUREMENT PRACTICAL TESTS Check that all buffer solutions are labeled with the following data: - The analysts initials - Date of preparation - Identity and review date Store all buffer solutions in an appropriate manner