Isotope Ratio Performance of an Axial Time of Flight ICP-MS Stuart Georgitis 1, Lloyd Allen 1, and Janos Fucsko 1, Frank Vanhaecke 2 1 LECO Corporation.

Slides:



Advertisements
Similar presentations
Chemistry 30 – Unit 2 – Solubility – Ch. 16 in Text.
Advertisements

Inductively Coupled Plasma Mass Spectrometry Dr. Lloyd Allen and Dr
Ion Sources Some characteristics of ion sources (especially in high precision work): It should have high efficiency in generating ions of the element of.
Elemental Scientific | | PC 3 Advantages Thermal stabilization of spray chamber improves long-term stability Reduces effect.
Inductively Coupled Plasma Mass Spectrometry or ICP-MS
Atomic Emission Spectroscopy
427 PHC.  Atomic emission spectroscopy (AES) is based upon emission of electromagnetic radiation by atoms.
Atomic Emission - AES M* → M + hn Thermal excitation M → M*
Fluorescence. Topics Sensitivity (nM-pM) –Contrast UV-vis measurements Derivatization Laser-Induced Fluorescence.
Atomic Mass Spectrometry
Homework Chapter 0 - 0, 1, 2, 4 Chapter 1 – 15, 16, 19, 20, 29, 31, 34.
Chem. 133 – 2/17 Lecture. Announcements Lab Work –Electronics Lab Report due 2/19 –Let me know by today if you plan to do a lab practical instead –Half.
Atomic Absorption Spectroscopy Elemental analysis Dissolved metals Source –Hallow Cathode Lamp –Cathode of specific element Sample –Acetylene-air flame,
Atomic Mass & Molar Mass
Chapter 5: Signals and Noise
Mass spectrometry 8/23/12 What are the principles behind MS?
Types of Analyses Single Collector –Uses only one collector or detector –This is the case with a Daly or EM Advantages: –Need only one detector –Peaks.
BALZ S. KAMBER Laurentian University Ultra-high purity ICP-MS.
Signal-to-Noise Optimization Noise Sources Most Commonly Encountered 1. Detector Noise 2. Shot Noise 3. Flicker Noise.
Typical chronometers. LA-ICPMS and SIMS challenges from ~
Atomic Emission Spectroscopy
Materials Characterization Lab Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008.
Solution Chemistry (Chp. 7)
Solution Chemistry (Chp. 7) Chemistry Topics Molar Concentration (mol/L) Dilutions % Concentration (pp. 255 – 263) Solution Process Solution Preparation.
Mass spectrometry Ions are analyzed on the basis of their m/z Chlorine has 2 isotopes, 35 Cl and 37 Cl, in the approximate ratio of 3 :1. Electrons are.
Atomic Emission Spectrometry By: Alexa Kunch
What is analytical chemistry? 5 subdisciplines of chemistry Some definitions In every day life… Analytical chemist versus chemical analyst.
Solution Chemistry (Chp. 7)
NOISE N. Libatique ECE nd Semester
Solution Concentration. Review  A solution is a homogeneous mixture.  The solvent is the major component of the solution.  The solute is the minor.
Define Problem Select Appropriate Methods Obtain and store sample Pre-treat sample Perform required measurements Compare results with standards Apply necessary.
Lecture 2 Measurements Concentration Solution Preparation.
Advances in Determinations of metals by ICPMS, from ultra-small sampling to ultra- trace analysis. AES Department Dr. Otto Herrmann Teresa Switzer May.
The Mass Spectrograph & Average Atomic Mass. Inventor Francis William Aston won the 1922 Nobel Prize in Chemistry for his work in mass spectrometry.
Effect of Sample Introduction on ICP-MS Detection of Nonmetal Analytes in a Simulated HPLC Matrix Adam E. Dill and Jon W. Carnahan Department of Chemistry.
to Optical Atomic Spectroscopy
Measurement of “Difficult” Elements and Isotopes Using a Hexapole ICPMS Zenon Palacz, Simon Meffan-Main. Micromass U.K. Ltd.
FEMTOSECOND LASER FABRICATION OF MICRO/NANO-STRUCTURES FOR CHEMICAL SENSING AND DETECTION Student: Yukun Han MAE Department Faculty Advisors: Dr. Hai-Lung.
1 LASER ABLATION. 2 Laser Ablation ICP-MS CETAC LSX 500 Thermo Finnigan Element
Overall efficiency and speed of analysis – key factors are ions per sample molecule desorbed and ions produced per laser pulse Precision and accuracy of.
Detection Limits N no longer >> N B at low concentration What value of N-N B can be measured with statistical significance? Liebhafsky limit: Element is.
Mass spectrometry (Test) Mass spectrometry (MS) is an analytical technique that measures masses of particles and for determining the elemental composition.
Data Analysis and Presentation Chapter 5- Calibration Methods and Quality Assurance EXCEL – How To Do 1- least squares and linear calibration curve/function.
Sources of noise in instrumental analysis
1 Nuclear Activation Techniques to measure the energy distribution of laser-accelerated protons bunches T.Bonnet, M.Comet, D.Denis-petit, F. Gobet, F.
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Basics of Ion Beam Analysis
An evaluation of the isotope ratio performance of an axial TOF - ICP - mass spectrometer F. Vanhaecke, L. Moens, R. Dams, L. Allen* and S. Georgitis* University.
Atomic-absorption spectroscopy Lab3 Atomic-absorption spectroscopy.
For the past several years, one of the primary focuses of our research group has been the development of novel ion-exchange systems for the purpose of.
Chapter 1: Introduction 1. Type of instrumental methods Radiation Electrical method Thermal properties Others 2. Instruments for analysis Non-electrical.
Chem. 133 – 2/16 Lecture. Announcements Lab today –Will cover last 4 set 2 labs + start on set 2 labs –Lab Report on electronics labs – due 2/23 (I planned.
Beats Array Counter Sensor ab Experimental device: a – scheme, b – external view.
Chapter 29 Mass Spectrometry. 29 A Principles of mass spectrometry In the mass spectrometer, analyte molecules are converted to ions by applying energy.
Different source designs for particle ionization in a pulsed glow discharge Farzad Fani Pakdel November 16, 2006.
Chem. 133 – 2/14 Lecture.
Solution Concentration
M. Fernández, E. Conde, A.I. Barrado Department of Technology
Chapter 5: Signals and Noise
Experiments at LCLS wavelength: 0.62 nm (2 keV)
Types of Fractures by Projectiles
DO NOW Pick up Notes. Get out Solubility handout and Concept Review handout.
Unit 5: Solutions ppm and ppb.
Strategies for Eliminating Interferences in Optical Emission Spectroscopy Best practices to optimize your method and correct for interferences to produce.
Concentrated vs Dilute Solutions
Signals and Noise Signal to Noise Ratio Types of Noise
Elements in Eliminated Jacobian = 133,515
Elements in Eliminated Jacobian = 133,515
Пламена Спектроскопија
Unit 5: Solutions ppm and ppb.
Presentation transcript:

Isotope Ratio Performance of an Axial Time of Flight ICP-MS Stuart Georgitis 1, Lloyd Allen 1, and Janos Fucsko 1, Frank Vanhaecke 2 1 LECO Corporation 2 University of Ghent

Introduction Nature of noise in ICP-MS measurement Sequential and simultaneous detection: fundamental differences in signal ratios Axial TOF ICP-MS: Is it really better for isotope ratio determinations Isotope ratios of transient and steady state signals with liquid and solid sampling methods Characterization of TOF ICP-MS performance Limitations of measurements

Isotope Ratio Fundamentals Sources of Noise in ICP-MS –Flicker Noise: Non Fundamental, Caused by Sample Introduction system and ICP.   s –Shot Noise: Fundamental, Due to the Random Arrival Rate of Particles (photons, electrons, ions) at a detector.   s 1/2

Isotope Ratios 50 ng/mL Ag 30 min. period Each point –5 repetitions –10 s integration/repetition Relative Standard Deviation (%) – 107 Ag: 0.18% – 109 Ag: 0.17% – 107 Ag/ 109 Ag: 0.02%

Isotope Ratios Do you need simultaneous techniques to measure? How is signal to noise ratio improved? Examples for solution and for solid material sampling.

Isotope Ratio Fundamentals Flicker Noise can be minimized or eliminated by ratio pairing. Flicker noise elimination is most effectively done using simultaneous acquisition. Should Flicker noise be eliminated, shot noise should be the dominant remaining source of noise.

Isotope Ratio Fundamentals The theoretical shot noise limit can be calculated: RSD = (  /s) at the Shot Noise Limit  = s 1/2 RSD = s -1/2 RSD 2 A/B = RSD 2 A + RSD 2 B or RSD 2 A/B = s B -1 + s B -1

Solid Sample Isotope Ratios NIST 610 Glass

Solid Sample Isotope Ratios 206 Pb/ 207 Pb in NIST Glass Conc (ppm) RSD Signal RSD Ratio % 0.8% % 0.2% % 0.09% 10 second integration n = 10

Transient Signal Isotope Ratio Precision (1)

Transient Signal Isotope Ratio Precision Time (s) Signal (mV) Ag107 Ag109

Transient Signal Isotope Ratio Precision* Ratio5 ng (%RSD) 50 ng (%RSD) Ag (107/109) Ba (138/137) Cu (63/65) Pb (208/207) Pb (208/206) Pb (206/207) Zn (64/66) *10  l Injection n = 5

Isotope Ratio Precision Ratio Precision = 0.34% Ratio Time (min) Pb-208 = 1.3 % Pb-206 = 1.3 % Peak Area

Isotope Ratio Precision (%RSD) 50  g/L208/206208/207206/20763/ % 0.11% 0.09% 0.10% 500  g/L208/206208/207206/20763/ % 0.05% 0.02%0.05% 30 Second Integration Time n=10

Isotope Ratio Limitations Simultaneous Techniques Even at the Shot noise limit, practical limitations arise –In order to obtain a %RSD of 0.01 on a 1:1 Ratio, 200 Million counts must be accumulate –In order to obtain a %RSD of on a 1:1 Ratio, 20 Billion counts must be accumulated –Ultimately, detector saturation limits the overall count rate which can be tolerated and integration for infinite time (2000 s/rep) is not possible

Silver Isotope Ratios %RSD vs Concentration 0.06% RSD, 100 ppb n = Ag/ 109 Ag Figure6

Lead Isotope Ratios %RSD vs Concentration Figure 7

1  g/L Steady State Solution Nebulization, %RSD vs Integration Time 207 Pb/ 206 Pb Figure 9

1  g/L Steady State Solution Nebulization, %RSD vs Integration Time 107 Ag/ 109 Ag Figure Ag/ 109 Ag RSD = 0.29%

Conclusions Fast simultaneous detection provides better element and isotope ratios. Precision of signal ratios are primarily controlled by counting statistics if practical (<2000 sec) integration time is used. The improved performance helps applications: –isotope ratio analysis from small or heterogeneous samples, using steady state or transient signals –isotope dilution analysis –internal standardization even for fast changing transient signals: speciation, chromatography, laser ablation