RECTIFICATION
X-RAY CIRCUIT
Whenever the art of medicine is loved, there is also love for humanity. Hippocrates
RECTIFICATION PROCESS BY WHICH ALTERNATING CURRENT IS CHANGED TO PULSATING DIRECT CURRENT
ALTERNATING CURRENT 1 PHASE A/C ( UNRECTIFIED)- FREQUENCY 60 Hz
60 POSITIVE PULSES/SEC 60 NEGATIVE PULSES/SEC
60 POSITIVE PULSES 60 NEGATIVE PULSES 120 PULSES 60 FULL CYCLES
BATTERY – CURRENT OUTPUT STEADY D/C
D/C HALF WAVE RECTIFIED D/C HALF-WAVE – 60 POSITIVE PULSES/SEC NEGATIVE PULSES ARE SUPPRESSED 1 PULSE/CYCLE
D/C FULL WAVE RECTIFIED D/C HALF-WAVE – 120 POSITIVE PULSES/SEC 2 PULSES/CYCLE
D/C -3 PHASE 6 PULSE CURRENT D/C 3 PHASE 6 PULSE – 6 PULSES/CYCLE
D/C - 3 PHASE 12 PULSE D/C 3 PHASE 12 PULSE – 12 PULSES/CYCLE
DIODES OR VALVE TUBES HELP RECTIFY A/C CURRENT DIODE ELECTRONS VALVE TUBE
X-RAY TUBE FORWARD BIAS
REVERSE BIAS + - NO X-RAY
FORWARD vs REVERSE
RECTIFICATION TYPES ► HALF-WAVE ► FULL WAVE ► THREE PHASE SIX PULSE ► THREE PHASE TWELVE PULSE ► HIGH FREQUENCY SINGLE PHASE
RECTIFICATION NO RECTIFICATION HALF-WAVE FULL-WAVE
3 PHASE RECTIFICATION
HIGH FREQUENCY RECTIFICATION
HALF-WAVE RECTIFICATION
FULL-WAVE RECTIFICATION
3 PHASE 6 PULSE
3 PHASE 12 PULSE
HIGH FREQUENCY
VOLTAGE RIPPLE ► HALF WAVE % ► FULL WAVE % ► 3 PHASE 6 PULSE % ► 3 PHASE 12 PULSE % ► HIGH FREQUENCY <1%
RIPPLE 100%
RIPPLE-3 PHASE 13%
RIPPLE-HIGH FREQUENCY <1%
Home-made x-ray apparatus
CHECKING FAILURE OF RECTIFICATION SYSTEM ► SPINNING TOP TEST – 1 PHASE EQUIPMENT ► SYNCHRONOUS SPINNING TOP TEST- 3 PHASE WITH LONGER EXPOSURE TIMES ► OSCILLOSCOPE- 3 PHASE EQUIPMENT WITH SHORT EXPOSURE TIMES
SPINNING TOP TEST
RESULT OF SPINNING TOP TEST- 1 PHASE - DASHES
HALF- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 60 PULSES
FULL- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 120 PULSES
RESULT OF SYNCHRONOUS SPINNING TOP TEST-3PHASE ARC
3 PHASE RECTIFICATION- LONGER EXPOSURE TIME DEG. OF ARC = 360°/ sec x TIME (sec)
UTILIZING OSCILLOSCOPE
UTILIZING OSCILLOSCOPE RECTIFIED SYSTEM