1 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Laurens Pluimers Supervisors: Dr.ir. W.M. van Spengen Prof.dr.ir. A. van Keulen
2 Challenge the future Micrometer(µm) Nanometer(nm) Picometer(pm) Millimeter(mm) Meter(m ) Kilometer(km ) Scaling
3 Challenge the future Microscopes Hair: µm DNA: nm Atoms: pm Optical microscope Resolution: 200nm Resolution: 100pm Source: andrew.cmu.edu Atomic force microscope (AFM)
4 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability
5 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability
6 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability
7 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Outline Introduction Atomic Force Microscope (AFM)
8 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Outline Introduction Atomic Force Microscope (AFM) Probe calibration
9 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Outline Introduction Atomic Force Microscope (AFM) Probe calibration Electrostatic pull-in instability
10 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Outline Introduction Atomic Force Microscope (AFM) Probe calibration Electrostatic pull-in instability Results of feasibility study
11 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Outline Introduction Atomic Force Microscope (AFM) Probe calibration Electrostatic pull-in instability Results of feasibility study Conclusions & Recommendations
12 Challenge the future Atomic Force Microscope Working principle Quadrant detector Laser Cantilever beam(probe) Sample Source:
13 Challenge the future Atomic Force Microscope Working principle Source:
14 Challenge the future Atomic Force Microscope Setup: Optical beam deflection system
15 Challenge the future Atomic Force Microscope AFM probe 20μm Source:
16 Challenge the future Atomic Force Microscope Images Topography image of metallic nanoparticles deposited on graphite Source:
17 Challenge the future Recap What is an Atomic Force Microscope (AFM)? “Feeling” the sample surface with probe Optical beam deflection system Resolution ~100pm √
18 Challenge the future Atomic Force Microscope Modes of operation Imaging Topography scan Force measurements Material properties
19 Challenge the future Atomic Force Microscope Mode of operation: Force measurements Measurement tip / sample interaction forces: Atomic bonding Van der Waals forces Magnetic forces Chemical bonding Probe Sample h Source:
20 Challenge the future Atomic Force Microscope Interaction forces Material A Material B Quadrant detector Laser Probe F int
21 Challenge the future Atomic Force Microscope Interaction forces x y “Force” image Material A Material B
22 Challenge the future Atomic Force Microscope Probe calibration k F int x Hooke’s law F int =k ·x Probe Laser Quadrant detector k=spring constant
23 Challenge the future Probe calibration Added mass M x Hooke’s law k
24 Challenge the future Probe calibration Euler-Bernoulli beam theory t L b Cantilever base
25 Challenge the future Probe calibration Other calibration methods MethodAccuracyDisadvantages Added mass15-25%Destructive, slow Euler-Bernoulli beam theory 20-40%Inaccurate, slow Nano-Force Balance0.4%External equipment, expensive Thermal tune20%Only compliant beams
26 Challenge the future Recap Why do you need to calibrate the probe? To determine the exact interaction forces between tip and sample Bonding forces Material properties Disadvantages other methods Need for new method √
27 Challenge the future Probe calibration New calibration method Based on probe’s Electrostatic Pull-in Instability (EPI) Inventor: Prof.dr.ir. F. van Keulen Improvements: Wide range of cantilever beams (k= 0.1 – 50 N/m) Non-destructive Integrated system in AFM Fast and easy to use
28 Challenge the future Probe calibration New calibration method Based on probe’s Electrostatic Pull-in Instability (EPI) EPI Probe calibration using EPI Experimental setup
29 Challenge the future Electrostatic Pull-in Instability V u=d 0 u Probe Counter electrode DC voltage source Pull-in point
30 Challenge the future Electrostatic Pull-in Instability Top view cantilever beam
31 Challenge the future Non-linear behaviour of the cantilever beam Elastic restoring forces are linear Electrostatic forces are quadratic Main advantage: well defined instability point(pull-in) measurement Electrostatic Pull-in Instability
32 Challenge the future Probe calibration Electrostatic pull-in instability L b d0d0
33 Challenge the future Probe calibration EPI: differential gap method V p1 V V V p2 Δd Δd
34 Challenge the future EPI probe calibration Experimental setup Variables: Differential gap ( Δd ) Pull-in voltage (V pi ) Length (L) Width (b) Accuracy: % Model Source: AFM system
35 Challenge the future EPI probe calibration Experimental setup XYZ stage Variables: Differential gap ( Δd ) XYZ stage Source:
36 Challenge the future EPI probe calibration Experimental setup Variables: Differential gap ( Δd ) Pull-in voltage (V pi ) Source: XYZ stage Counter electrode XYZ stage
37 Challenge the future EPI probe calibration Experimental setup Variables: Differential gap ( Δd ) Pull-in voltage (V pi ) Source: Counter electrode XYZ stage
38 Challenge the future EPI probe calibration Experimental setup Variables: Differential gap ( Δd ) Pull-in voltage (V pi ) Length (L) Width (b) Source: Counter electrode XYZ stage Aspheric lens
39 Challenge the future EPI probe calibration Calibration mode Source: Variable: Pull-in voltage (V pi ) Source:
40 Challenge the future EPI probe calibration Width scan x Source: Variable: Width (b) Source:
41 Challenge the future EPI probe calibration Length scan y Source: Variable: Length (L) Source:
42 Challenge the future EPI probe calibration Experimental setup Source:
43 Challenge the future Probe calibration Experimental setup Optical path Laser Aspheric lens Quadrant detector
44 Challenge the future Probe calibration Experimental setup
45 Challenge the future Probe calibration Experimental setup
46 Challenge the future Probe calibration Experimental setup
47 Challenge the future Probe calibration Experimental setup
48 Challenge the future Probe calibration Experimental setup
49 Challenge the future Results Performance check: Differential gap ( Δd ) Pull-in voltage (V pi ) Length (L) Width (w) Calibration test probe
50 Challenge the future Results Width scan Width Position stage [µm] QD output [V] Width scan EPI
51 Challenge the future Results Length scan Length Position stage [µm] QD output [V] Length scan EPI
52 Challenge the future Results Length/Width scan Width [µm]Length[µm] EPI50.59 ± ±0.40 Bruker WL50.71 ± ±0.3 Error [µm] 0.12 ± ±0.5 Error [%]
53 Challenge the future Results Calibration test probe ProbeSpring constant k [N/m]Δk [%] NanoWorldEPI 1 (compliant) (stiff) Requirement: Accuracy %
54 Challenge the future Conclusions Performance check: EPI method can be implemented as integrated system Calibration test probe: EPI calibration method is able to determine the spring constant of AFM probes Accuracy system not within requirements
55 Challenge the future Recomendations Increase accuracy by improving model Include fringing field effects Tapered end beam My model Reality
56 Challenge the future Recommendations Increase accuracy by improving model Include fringing field effects Tapered end
57 Challenge the future Recommendations Increase accuracy by improving model Include fringing field effects Tapered end Cantilever beam
58 Challenge the future Recommendations Increase accuracy by improving model Include fringing field effects Tapered end New model in progress
59 Challenge the future Feasibility study for AFM probe calibration using the probe’s electrostatic pull-in instability Questions?
60 Challenge the future Extra sheet Width scan Width Position stage [µm] QD output [V] Width scan EPI
61 Challenge the future Extra sheet Width scan
62 Challenge the future Laser + Lens Quadrant detector Laser beam Width cantilever beam Extra sheet Width scan
63 Challenge the future Extra sheet Extended model