A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state.

Slides:



Advertisements
Similar presentations
INHALATION ANAESTHESIA BREATHING SYSTEMS Prepare and monitor anaesthesia in animals INHALATION ANAESTHESIA BREATHING SYSTEMS.
Advertisements

Associate Professor of Anesthesia
ANESTHETIC MACHINES.
CARBON DIOXIDE ABSORPTION
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 1 of 4.
Presented by: Dr. manoj kumar Moderated by: Dr. Aparna Sharma
Inhaled Anesthetic Delivery Systems
Paediatric breathing systems Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab.DCA, Dip. Software statistics PhD (physio) Mahatma Gandhi medical.
Check Valves and Manufacturer-Recommended Leak Test
Scavenging System  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected to a.
Anesthetic Equipment.
1 Oxygen Administration. 2 Introduction  Oxygen is a colorless, odorless, tasteless gas that is essential for the body to function properly and to survive.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 4 of 4.
Inhaled Anesthetic Delivery Systems
 The annular space. The clearance between the head of the float and the flow tube is known as the annular space.
Technology Topics 2005 CA1 - CA40 focus BWH Anesthesia Clinical Conference August 31, 2005 James H. Philip, MEE, MD, CCE © Copyright 2005, James H Philip,
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U.
PART 3: Breathing Circuit
Parts of the Anesthetic Machine
The Anesthesia Gas Machine
CHAPTER 3 ISOFLURANE AND SEVOFLURANE (HALOGENATED COMPOUNDS) NITROUS OXIDE AND DESFLURANE ENFLURANE HALOTHANE METHOXYFLURANE DIETHYL ETHER Inhalation Anesthetics.
Review Lung Volumes Tidal Volume (V t )  volume moved during either an inspiratory or expiratory phase of each breath (L)
Anesthesia Machine VAPORIZERS. Vaporizers Convert liquid anesthetic into a volatile inhalation agent Based on laws of physics You must memorize the chemical.
Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics PhD(physiology) Mahatma gandhi medical college and research institute,
Anesthesia Breathing Systems
Inhaled anesthetic Delivery Systems
Scavenging System A&A Pg 356  To reduce exposure to waste anesthetic gases  Can be active or passive scavenging system Active Scavenging System  Connected.
+ Surgical Procedures 7.01 Implement techniques to prepare and monitor patients for surgery.
Anesthetic Breathing Circuits
By Dr. Ahmed Mostafa Assist. Prof. of anesthesia and I.C.U
To Menu.
Non-Rebreathing System  Used for patients weighing < 7 kg  Attach the clear, plastic hose to the fresh gas inlet (this tube comes from the vaporizer.
Ventilators All you need to know is….
SureVent 2 - SV2® Gas Powered Automatic Ventilation Resuscitator
Cont Sophisticated computerized controls are able to provide advanced types of ventilatory support such as synchronized intermittent mandatory ventilation.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 2 of 4.
Humidity Therapy and Humidifiers
Introductory Lecture Series: The Anesthesia Machine
Waste Anesthetic Gases. The anesthetic gas and vapors that leak out into the surrounding room during medical and surgical procedures are considered waste.
Anesthetic Equipment.
PRESENTED BY-DR. POOJA MODERATOR – DR.DARA SINGH CIRCLE SYSTEM AND LOW FLOW ANESTHESIA.
Without reference, identify principles about Anesthesia Units with at least 70 percent accuracy.
Oxygenation And Ventilation
ANAESTHESIA BREATHING CIRCUITS
Understanding Mapleson Circuits
PART 3: Breathing Circuit
To Menu.
Introduction of low- flow anesthesia 麻醉科 R2 楊美惠 2004/11/08.
VNTE Prep Anesthetic Review.
1 Copyright © 2011, 2003, 2000, 1994 by Mosby, Inc., an affiliate of Elsevier Inc. Veterinary technicians must be familiar with the human safety considerations.
Anesthesia Machine Circuits
Parts of the Anesthetic Machine A&A Pages
PART 3: Breathing Circuit
Vaporizers – principles Dr. S. Parthasarathy MD., DA., DNB, MD (Acu), Dip. Diab. DCA, Dip. Software statistics- PhD ( physiology)
Inhaled anesthetics By: Israa Omar.
Mechanical Ventilator
Introductory Lecture Series: The Anesthesia Machine
The Anesthesia Machine: Introduction
BREATHING SYSTEMS.
AMIR SALAH MODERN ANAESTHETIC MACHINE MODERN ANAESTHETIC MACHINE 3 of 4.
PART 3: Breathing System Parts
Anesthesia Equipment Session 3
MACS CPAP System Self Guided Tour.
Non-Rebreathing System
Vapor pressure, partial pressure, and concentration of inhaled anesthetics: The impact of atmospheric pressure on vaporizer output. Top. Variable bypass.
Parts of the Anesthetic Machine Parts 1 & 2: A&A pages
Diagram of a T piece (Mapleson E circuit)
Vapor pressure, partial pressure, and concentration of inhaled anesthetics: The impact of atmospheric pressure on vaporizer output. Top. Variable bypass.
Non-Rebreathing System
Breathing Systems Tom Williams.
Presentation transcript:

A). When the carrier gas is quickly switched to 100% nitrous oxide (B). the halothane concentration decreases to 3% within 8 to 10 seconds A new steady-state concentration of approximately 3.5% is then attained within 1 minute

 Viscosity  Density  relative solubility of the carrier gas  the flow-splitting characteristics of the specific vaporizer  the concentration control dial setting

 Safety Features: › Agent-specific, keyed filling devices › Overfilling is minimized › Firmly secured to a vaporizer manifold (Prevent vaporizer tipping ) › Contemporary interlock systems (prevent the administration of more than one inhaled anesthetic)  Hazards: Misfilling Contamination Tipping (transport “T” dial setting) Overfilling Underfilling (sevoflurane) Simultaneous Administration of Inhaled Anesthetics Leaks Environmental Considerations (MRI)

Boils at 22.8°C MAC: 6% to 7% (amount of desflurane vaporized over a given period is considerably greater than other anesthetic drugs that cause excessive cooling of the vaporizer and significantly reduce its output) Vapor pressures at 20°C enflurane 172, isoflurane 240, halothane 244, and desflurane 669 mm Hg At 1 atm and 20°C, 100 mL/min passing through the vaporizing chamber would entrain 735 mL/min of desflurane versus 29, 46, and 47 mL/min of enflurane, isoflurane, and halothane ((to produce a 1% desflurane output)) bypass flow would be approximately 73 L/min versus 5 L/min or less for the other three anesthetics

Simplified schematic of the Tec 6 desflurane vaporizer

The heart of the vaporizer is the electronically controlled flow control valve located in the outlet of the vaporizing chamber

Anesthetic Breathing Circuits Mapleson Systems Circle Breathing Systems

prevention of rebreathing:  spontaneous ventilation: A > DFE > CB  controlled ventilation: DFE > BC > A

 The integrity of the inner tube can be assessed as described by Pethick

 The most popular breathing system in the United States  “universal F” or “single-limb circuit,” (One version of the traditional circle system), has increased in popularity in recent years  A circle system depending on the amount of fresh gas inflow can be: Semi open, semi closed, or closed  A semi open (no rebreathing), requires a very high flow of fresh gas.  A semi closed system (some rebreathing of exhaled gases), is the most common in the United States.

 (1) a fresh gas inflow source  (2) inspiratory and expiratory unidirectional valves  (3) inspiratory and expiratory corrugated tubes  (4) a Y-piece connector  (5) an overflow or pop-off valve referred to as the APL valve  (6) a reservoir bag  (7) a canister containing a carbon dioxide absorbent.

 (1) a unidirectional valve must be located between the patient and the reservoir bag on both the inspiratory and expiratory limbs  (2) the fresh gas inflow cannot enter the circuit between the expiratory valve and the patient  (3) the overflow (pop-off) valve cannot be located between the patient and the inspiratory valve.

 (1) maintenance of relatively stable inspired gas concentrations  (2) conservation of respiratory moisture and heat  (3) prevention of operating room pollution.

 Complex design  In a recent closed-claim analysis: more than a third (25/72) of malpractice claims resulted from breathing circuit misconnections or disconnections.  valves stick: Rebreathing, barotrauma or volutrauma  Obstructed filters located in the expiratory limb: have caused increased airway pressure, hemodynamic collapse, and bilateral tension pneumothorax