THE WMO FIELD INTERCOMPARISON OF RAINFALL INTENSITY (RI) GAUGES in Vigna di Valle (ITALY), October 2007- April 2009: relevant aspects and results. WMO-TECO,

Slides:



Advertisements
Similar presentations
WMO - TECO 2008 St, Petersburg, Russia, November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING.
Advertisements

19/5/2011 Cg-XVI, Doc. 3.1(4) World Meteorological Organization Working together in weather, climate and water WWW Programme Instruments and Methods of.
Meteorological Observatory Lindenberg – Richard Assmann Observatory The GCOS Reference Upper Air Network.
World Meteorological Organization Working together in weather, climate and water Snowfall Measurement Challenges WMO SPICE Solid Precipitation Intercomparison.
1 COPS Workshop 2008 University of Hohenheim, Stuttgart; 27 to 29 February 2008 IMGI‘s contribution to the COPS 2007 field experiment Simon Hölzl & Alexander.
World Health Organization
Operational Quality Control in Helsinki Testbed Mesoscale Atmospheric Network Workshop University of Helsinki, 13 February 2007 Hannu Lahtela & Heikki.
Earth and Atmospheric Sciences EAS535 Atmospheric Measurements and Observations II EAS 535
1 Seventh Lecture Error Analysis Instrumentation and Product Testing.
Classification of Instruments :
Surface-Based Instrument Intercomparisons and Calibration Methods Activity Report Draft recommendations Michel Leroy Chair of the ET-SBII&CM Commission.
Chapter 6 Random Error The Nature of Random Errors
V. Rouillard  Introduction to measurement and statistical analysis ASSESSING EXPERIMENTAL DATA : ERRORS Remember: no measurement is perfect – errors.
CARPE DIEM Centre for Water Resources Research NUID-UCD Contribution to Area-3 Dusseldorf meeting 26th to 28th May 2003.
Meteorological Observatory Lindenberg – Richard Assmann Observatory The GCOS Reference Upper Air Network.
1 GOES-R AWG Hydrology Algorithm Team: Rainfall Probability June 14, 2011 Presented By: Bob Kuligowski NOAA/NESDIS/STAR.
Proposed revision of ITU-R Recommendation providing EESS (active) Performance and Interference Criteria May 13, 2015 Committee On Radio Frequencies Thomas.
Results of the WMO Laboratory Intercomparison of rainfall intensity gauges Luca G. Lanza University of Genoa WMO (Project Leader) DIAM UNIGE September.
6. Conclusions and further work An analysis of storm dew-point temperatures, using all available dew-point estimates was carried out for 10 significant.
INTRODUCTION TO MEASUREMENT
Documentation of surface observation. Classification for siting and performance characteristics Michel Leroy, Météo-France.
April nd IBTrACS Workshop 1 Operational Procedures How can we build consistent, homogeneous, well- documented climate quality data?
Performance characteristics for measurement and instrumentation system
LECTURER PROF.Dr. DEMIR BAYKA AUTOMOTIVE ENGINEERING LABORATORY I.
1 Methods for detection of hidden changes in the EEG H. Hinrikus*, M.Bachmann*, J.Kalda**, M.Säkki**, J.Lass*, R.Tomson* *Biomedical Engineering Center.
Climate Monitoring of Precipitation: The GPCC - Status and plans Global Precipitation Climatology Centre U. Schneider, A. Meyer-Christoffer, B. Rudolf.
Counseling Research: Quantitative, Qualitative, and Mixed Methods, 1e © 2010 Pearson Education, Inc. All rights reserved. Basic Statistical Concepts Sang.
WMO Technical Conference on Meteorological and Environmental Instruments and Methods of Observation TECO-2012 Brussels, Belgium, 16 – 18 October 2012 Inter-comparison.
European Conference on Quality in Official Statistics Session 26: Quality Issues in Census « Rome, 10 July 2008 « Quality Assurance and Control Programme.
Laboratory QA/QC An Overview.
Piezoelectric precipitation sensor from VAISALA
CIMO Survey National Summaries of Methods and Instruments Related to Solid Precipitation Measurement at Automatic Weather Stations - Very Preliminary results.
Measures of central tendency are statistics that express the most typical or average scores in a distribution These measures are: The Mode The Median.
WE Surface-based Global Observing System for Weather.
Instruments and Methods of Observation Programme.
Quality Assurance How do you know your results are correct? How confident are you?
Ping Zhu, AHC5 234, Office Hours: M/W/F 10AM - 12 PM, or by appointment M/W/F,
Juan Ignacio Pérez Soladana RAIN GAUGE BASED ON A MOBILE PLATFORM WMO TECHNICAL CONFERENCE ON METEOROLOGICAL AND ENVIRONMENTAL INSTRUMENTS.
Status and Plans of the Global Precipitation Climatology Centre (GPCC) Bruno Rudolf, Tobias Fuchs and Udo Schneider (GPCC) Overview: Introduction to the.
EVALUATION OF THE RADAR PRECIPITATION MEASUREMENT ACCURACY USING RAIN GAUGE DATA Aurel Apostu Mariana Bogdan Coralia Dreve Silvia Radulescu.
Quality management, calibration, testing and comparison of instruments and observing systems M. Leroy, CIMO ET on SBII&CM.
Meteorological Observatory Lindenberg Results of the Measurement Strategy of the GCOS Reference Upper Air Network (GRUAN) Holger Vömel, GRUAN.
Module 1: Measurements & Error Analysis Measurement usually takes one of the following forms especially in industries: Physical dimension of an object.
1 The WMO Technical Conference on Meteorological and Environmental Instruments and Methods of Observation November 2008, St. Petersburg, Russia INTERCOMPARISON.
1 Module One: Measurements and Uncertainties No measurement can perfectly determine the value of the quantity being measured. The uncertainty of a measurement.
Chapter 7 Measuring of data Reliability of measuring instruments The reliability* of instrument is the consistency with which it measures the target attribute.
Chapter 7 Estimates, Confidence Intervals, and Sample Sizes
BME 353 – BIOMEDICAL MEASUREMENTS AND INSTRUMENTATION MEASUREMENT PRINCIPLES.
Measurements Measurements and errors : - Here, the goal is to have some understanding of the operation and behavior of electrical test instruments. Also,
WMO CIMO Survey National Summaries of Methods and Instruments for Solid Precipitation Measurement - Preliminary results - R Nitu Meteorological Service.
WMO TECHNICAL CONFERENCE ON METEOROLOGICAL AND ENVIRONMENTAL INSTRUMENTS AND METHODS OF OBSERVATION TECO-2012 Brussels, Belgium, October 2012 ENVIRONMENTAL.
Sarah Callaghan British Atmospheric Data Centre, UK, The effects of climate change on rain The consensus in the climate change.
 Remember Chemistry panel Quality Control:-  In a medical laboratory, it is a statistical process used to monitor and evaluate the analytical process.
Instrument Characteristics  Scientific Instrument: l A device for making a measurement.  Measurement: l An action intended to assign a number as the.
WE Surface-based GOS for Weather May 2008 Up-date.
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Diagnostic clinical chemistry
ITALIAN AIR FORCE METEOROLOGICAL SERVICE AND MÉTÉO FRANCE UPDATING AND DEVELOPMENT OF METHODS FOR WORLDWIDE ACCURATE MEASUREMENTS OF SUNSHINE DURATION.
Establishing by the laboratory of the functional requirements for uncertainty of measurements of each examination procedure Ioannis Sitaras.
8th INTERNATIONAL WORKSHOP ON PRECIPITATION IN URBAN AREAS
Eckhard Lanzinger Manfred Theel Herbert Windolph
Radar/Surface Quantitative Precipitation Estimation
Practical clinical chemistry
INSTRUMENTS POSITIONS
Introduction to Electrical Measurement
On the accuracy of precipitation measurements in the Arctic
Measurements Measurements and errors :
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Measurements & Error Analysis
Presentation transcript:

THE WMO FIELD INTERCOMPARISON OF RAINFALL INTENSITY (RI) GAUGES in Vigna di Valle (ITALY), October April 2009: relevant aspects and results. WMO-TECO, Helsinki, 31 Aug – 1 Sept 2010 Vuerich E. (ITALY) Sestola (Italy), 23 June 2009

CONTENTS BACKGROUND AND OBJECTIVES PROCEDURES AND METHODS DATA ANALYSIS AND RESULTS CONCLUSIONS AND DEVELOPMENTS RECOMMENDATIONS THE INTERCOMPARISON TEST SITE ITALIAN MET SERVICE – AIR FORCE (IMS) CENTRE OF MET EXPERIMENTATIONS (RESMA) VIGNA DI VALLE ITALY

BACKGROUND Following the increased recognition of scientific and practical issues related to the assessment of possible climatic trends, the mitigation of natural disasters (e.g. storms and floods), the hindering of desertification, the attention paid to accuracy and reliability in rainfall measurements is currently increasing The WMO recognized these emerging needs and promoted a couple of mile-stone meetings: - (1) Bratislava, Slovak Republic, April 2001: calibration of rain gauges and general aspects of measurements (I phase: WMO Laboratory Intercomparison of RI gauges, ) - (2) Geneva, Switzerland, 5-9 December 2005: operational aspects of field measurements and achievable accuracy (II phase: WMO Field Intercomparison of RI gauges, )

BACKGROUND Previous WMO intercomparisons: accumulated amounts, on low RI or on snowfall, on qualitative information The WMO FI-RI gauges started in Vigna di Valle (Italy) at the IMS Centre of Met Experimentations on the 1 st of October 2007 and, after a 6-months extension, it was concluded in April 2009 (first intercomparison for quantitative 1MIN-RI) Main objective: to assess and compare quantification and catching errors of both catching and non-catching type of RI gauges with the emphasis on high rainfall intensity in order to complain with the WIGOS basic concepts of improving standardization, data homogeneity, data quality, compatibility, interoperability. Other tasks: providing draft recommendations for CIMO and guidance material for inclusion into the CIMO guide.

PROCEDURES AND METHODS 30 RI gauges (4 working references) capable to measure RI up to 200mm/h at 1min resolution were selected, randomly distributed (avoid clustering of large gauges) and installed on dedicated ground platforms at 1 m height (no windshields) 13 additional meteorological sensors were installed in later positions for monitoring environmental conditions.

PROCEDURES AND METHODS #RAIN GAUGESMEAS. PRINCIPLE 1RIMCO 7499Tipping bucket 2Paar AP23Tipping bucket 3Precis-MecaniqueTipping bucket 4Thies PTTipping bucket 5/27ETG R102Tipping bucket 6LSI-LASTEM DQ031Tipping bucket 7SIAP-MICROS UM7525/I Tipping bucket 8/28CAE PMB2Tipping bucket 9Davis Rain Collector II Tipping bucket 10Lambrecht 15188Tipping bucket 11MTX PP040Tipping bucket 12Env. Meas. Ltd ARG100 Tipping bucket 13/29Meteoservis MRW500 Weighing gauge #RAIN GAUGESMEAS. PRINCIPLE 14Vaisala VRG101Weighing gauge 15OTT PluvioWeighing gauge 16EWS PG200Weighing gauge 17/30GEONOR T-200BWeighing gauge 18MPS TRwSWeighing gauge 19SA „MIRRAD“ MPA- 1M Not Participating 20Vaisala PWD22Optical Disdrometer 21OTT ParsivelOptical Disdrometer 22Thies LPTOptical Disdrometer 23Vaisala WXT510Acoustic impact 24Eigenbrodt ANS 410Water pressure 25KNMI electric raingauge Water level 26PVK ATTEX “DROP”Doppler Radar List of participating rain gauges (26+4 working references)

PROCEDURES AND METHODS 4 Reference gauges, the “Composite Working Reference ” (C.W.R.) were inserted in Reference Rain Gauge Pits (R.R.G.P.) at the centre of the Intercomparison site (collectors at ground level) – Minimization of weather related catching errors (e.g. Jevons effect, 1861) Standard adopted: ISO/EN-13798: “Specification for a reference rain gauge pit”, recently revised in 2010 CWR (Recom. 3 CIMO- XIV): corrected tipping bucket rain gauges (TBRG) and weighing gauges (WG) with the shortest step response and the lowest uncertainty according to Results of WMO Laboratory Intercomparison of RI ETG-R102 GEONOR T200B CAE-PMB2 Meteoservis MRW500

PROCEDURES AND METHODS 20 catching type gauges were calibrated in the laboratory before the Field Intercomparison. The WMO recognized laboratory at the University of Genoa performed the calibration based on the generation of a constant water flow from a reference hydraulic device (tests according to Recom.2 CIMO-XIV + statistics based on 1 min evaluation of relative errors ) Objectives: to single out the quantification errors associated with each instrument; measurement uncertainty and understanding of field results Reference hydraulic device The Qualification Module for RI Measurement Instruments developed at the University of Genova (Ur(95%)=0,45%)

PROCEDURES AND METHODS Lab tests: reference flow rates at 2, 20, 50, 90, 130, 170, 200 mm/h at 1MIN resolution for a variable duration, arithmetic mean of relative errors and spreading of data around it, providing correction curves (to be used beyond the intercomp. time) Results: the constant flow response plot, gauge relative error plotted versus Lab reference RI (superimposed box-plots)

PROCEDURES AND METHODS Further lab tests: investigation of the dynamic performance gauges at 1MIN resolution through the evaluation of the step response to a step input (time constant) which is a measure of the instrument stability and ability to detect rapid changes of the input signal. Results: step response plot, ratio of measured RI / Lab Reference RI versus time.

PROCEDURES AND METHODS Quality assurance: Raw data are processed by the Automatic Quality Control (AQC) to provide quality checked 1-minute data Periodic field tests: Field Calibration Device (developed at DICAT Laboratory – Univ. of Genoa) – Metrological confirmation Periodic maintenance, daily visual checks, cleaning of instruments, calibration status checks Meeting of Participants and local staff (Vigna di Valle, MAY 2008) – HMEI participated and reported QA reports by the Site Manager to the ET. Meeting of Participants Field Tests-Calibration Device

DATA ANALYSIS AND RESULTS Analysis dataset: 1MIN-QC checked RI data obtained through the reduction of the full FI dataset (162) by considering only synchronized events and events with 2 consecutive minutes and RI>12mm/h (totally 43 events, approx. 740 data). No. 2 events with RI > 150mm/h; No. 5 events with RI>100mm/h; No. 15 events with RI>60mm/h ….. Total availability of 1-min data (rain/no rain) 1-min valid data (rain/no rain): percentage of available 1min data that are valid according to QC Total numbers of precipitation daily events Hail and Mixed Rain/Hail events T.A. = 95.4%98.2% of T.A. 162 (Full FI Dataset) 6 events: 13 th Jan, 4 th Feb, 7 th May, 30 th Oct 2008; 1 st Jan, 5 th Mar 2009 Numbers of synchronized events Numbers of events for reference RI calculations Number of events for data analysis of rain gauges Rainfall accumulated over the intercomparison period 85 (Reduced FI Dataset) 79 ( min data) 43 (740 1-min RI data) 1325 mm

DATA ANALYSIS AND RESULTS Main objective: performing the best inter-comparison of rain gauges in high RI field conditions through the determination of a reference value (1MIN RIref) RIref is the best estimation of the 1 min RI true value and it must be derived through a RI Composite Working Reference value. 4 Nov 2008

DATA ANALYSIS AND RESULTS Method: weighted average of 1-min RI i measured by the 4 reference gauges Weights (µi, working reference gauge i) were the most challenging issue, since they must take into account effects related both to dynamical characteristics and possible lack of synchronization on 1 min time base Si are 3 statistical parameters calculated throughout the database of all precipitation events F = 0,1: “gross” parameter, for each working reference and for each event (F=1 if the working ref is not affected by 1-min lack of synchronization or high dynamic oscillation) (k≠i) F=0,1 µ R102 µ PMB2 µ MRW500 µ T200B

DATA ANALYSIS AND RESULTS CAE-PMB2 M-S MRW500 GEONOR- T200B ETG-R102

DATA ANALYSIS AND RESULTS Synchronization: to compare the 1-min RI data of all instruments, a synchronization of ±10 s was required, in other words the internal clock of the instrument should be within ±10 s compared to the DAQ system timestamp (nominal timestamp). If the difference/delay between the instrument’s data output time and the nominal timestamp exceeds the required ±10 s time interval, the result cannot be correctly compared to synchronized gauges. Gauge A and C are synchronized with the DAQ system clock; gauge B has a delay exceeding 10 seconds. Arrows indicate sample points of B with large difference due to non-synchronized data points of gauge B.

DATA ANALYSIS AND RESULTS Uncertainty of the reference: assuming a normal distribution of the deviations of pit gauges RI, the standard deviation of the distribution with respect to the reference intensity is calculated according (σ=[∑(RI- RIref)2/N]1/2); thus U95(RIref)=2 σ=4.3mm/h Relative uncertainty: ur(RIref)= (U(RIref) / RIref )∙100 [%] (green in plot)

DATA ANALYSIS AND RESULTS To show the general results of the intercomparison, the intensities measured by the rain gauges are plotted versus 1MIN Riref (CWR) and data are fitted with a power law trend curve RI=a x RIref b (corresponding bet fit equation reported on graphs) To assess the accuracy of field measurements compared to the reference, a tolerance region (composed of the WMO required 5% uncertainty and of the uncertainty of the reference, calculated as [urel(RIref) ] 1/2 [%] ) is represented in dashed lines on each plot. For easier comparison, the instruments have been divided into seven groups according to the measurement principle, as indicated in the title of each plot. WG instruments are split in two groups for easier presentation of results. Specific results and plots: Data Sheets included in the Final Report To assess the impact of wind losses (Jevons effect) on catching errors, the ratio RIi OUT / RIi is plotted versus wind speed (being RIi OUT and RIi the measured intensities by two identical gauges, one in the pit and the other one in the corresponding open field platform)

DATA ANALYSIS AND RESULTS TBRG

DATA ANALYSIS AND RESULTS TBRG-SC

DATA ANALYSIS AND RESULTS TBRG-PC

DATA ANALYSIS AND RESULTS TBRG-MC & LRG

DATA ANALYSIS AND RESULTS WG (1)

DATA ANALYSIS AND RESULTS WG (2)

DATA ANALYSIS AND RESULTS NON CATCHING TYPE RAIN GAUGES

DATA ANALYSIS AND RESULTS Measurement accuracy of all gauges by relative deviations from RIref

DATA ANALYSIS AND RESULTS A relevant effect of the wind did not appear in this intercomparison, we are able to affirm that the wind is not affecting in a significant way the outer instruments compared to those installed in the pit.

CONCLUSIONS AND DEVELOPMENTS This comparison at 1MIN time scale in field conditions demonstrates the possibility to evaluate the performance of RI gauges, identifying directions where research and technological development should be oriented. The achievable accuracy of WG can be improved in field conditions by means of the reduction of the response time below 1-minute and by appropriate filtering methods With regard to TBRG, the method applied by TBRG-SC confirms the possibility to improve the 1-min RI resolution and to provide accurate field measurements With regard to TBRG, the method applied by TBRG-PC revealed the possibility to provide accurate field measurements at higher RI, even if the performance is limited by their resolution at lower RI. With regard to non catching type rain gauges, despite the need of very low maintenance and the possibility to determine the type of precipitation (not for all), none of them agreed well with the reference. We are confident that their performance will be improved in the next future, considering the possibility of improving calibration methods to reduce systematic errors.

CONCLUSIONS AND DEVELOPMENTS Disdrometers tended to overestimate the RI. Despite their very different calibration procedures, they agreed better to each other than to the reference, as indication of a good degree of precision but they were not as accurate as conventional gauges For the best quality instruments, the achievable measurement uncertainty in laboratory, under constant flow conditions, was found to be 5% above 2 mm/h and 2% above 10 mm/h. One of the most challenging aspects of the Intercomparison was the definition of a 1-minute field reference intensity. The procedure adopted confirmed the suitability of R102-ETG, PMB2-CAE and T200B-GEONOR for the calculation of the reference. The rainfall intensity is highly variable from minute to minute. Therefore, the time synchronization of the instruments is crucial to inter-compare their measurements and to design the measurement systems. The results of the intercomparison confirmed the feasibility to measure 1MIN RI and provided information on the achievable measurement uncertainties, responding to the need of accuracy, homogeneity and standardization of RI measurements, as required by WIGOS.

CONCLUSIONS AND DEVELOPMENTS In field conditions, the uncertainty of the RI CWR was evaluated to be 4.3 mm/h. Consequently, the relative uncertainty of the reference was found to be below 5% only for intensities above 90 mm/h. Below 90 mm/h, the relative uncertainty of the reference values was higher than the 5% required measurement uncertainty provided in the CIMO Guide. A number of standardization activities at national and European level (CEN- TC “Hydrometry”)) has been promoted by the Italian Met Service, such as a national standard on the classification of rain gauges according to accuracy, the revision of the standard “Specification for a reference rain gauge” (ISO/EN13798:2010), a pre-standard (CEN technical report) on RI measurements (a possible WMO/ISO common standard in future) The IMS proposed the establishment of a CIMO Lead Centre on precipitation intensity (Univ. Genova, Vigna di Valle, M.Cimone).

RECOMMENDATIONS (3 of 18) The 1-min RI is highly variable from minute to minute. Therefore, it is recommended that 1-min RI should only be measured in a station and used for further analysis if the following conditions are met: (a) All 1-min data must be transmitted and used (a single 1-min RI value is not representative of a longer period of time; (b) A very good time synchronization, better than 10 s, must be achieved, both between the reference time and the different instruments of the observing station. Changes be made to the CIMO Guide Table: Precipitation intensity (liquid): Achievable measurement uncertainties: -Under constant flow conditions in laboratory, 5% above 2 mm/h, 2% above 10 mm/h. -In field conditions, 5 mm/h, and 5% above 100 mm/h It is recommended that RI measurements be further standardized based on the advances in knowledge obtained from this intercomparison to allow the users to obtain homogeneous data sets (based on the achievable RI measurement performance (accuracy) rather than on measuring principle or gauge design/technical solutions)

ACKNOLEDGEMENTS The ITALIAN MET SERVICE – AIR FORCE The WMO-CIMO (Dr Ondras, Dr Ruedi) The colleagues, local staff of the intercomparison site (Vigna di Valle) Dr Claudia Monesi – Univ. of Rome TRE (data analyst) Dr Eckhard Lanzinger – DWD (The project Leader) Dr Bruce Baker –USA (ET member) Dr Michel Leroy – Meteo France (ET chair) And Prof. Luca Lanza and Eng. Luigi Stagi – Univ. of Genova