Operation of the S4700 FESEM

Slides:



Advertisements
Similar presentations
Using Microsoft PowerPoint in the Classroom
Advertisements

Use lenses to direct light in order to enlarge images
Click on the rectangle on the lower right-hand side of the screen. This will quickly clear your desktop.
Introduction to Operation of the XL-30 Scanning Electron Microscope Tutorial # 1 Overview of XL-30 Microscope Control Program.
Microsoft Office XP Microsoft Excel
Microscope Basics T. Trimpe
Smart Cart Multimedia Lectern Technology Resource Center.
Scanning Electron Microscope (SEM)
DNR-322L & DNR-326.
1Rugby 55 Presentation – January 2007 The Rugby 55 User Presentation.
1. Pump down the chamber from the atmosphere pressure 2. Using RHEED to obtain diffraction images 3. Open the RHEED chamber or venting from high vacuum.
Tool bar and button bar (Many SEM functions are mouse-controlled)
LEO Jon Paul Johnson’s Montage Ductile Iron Images.
EDAX Operation Operation of the Phoenix EDAX system
Flatbed scanning in the HoA Computer Room How to use the Epson scanner.
Introduction to Operation of the XL- 30 Scanning Electron Microscope Tutorial # 2 Introduction to Operation and Imaging.
ActEyes IP Cam Training Part 2: Basic Camera Features.
Google Earth How to create a Google Earth Tour and place it in your Wiki.
Adobe Forms THE FORM ELEMENT PANEL. Creating a form using the Adobe FormsCentral is a quick and easy way to distribute a variety of forms including surveys.
Camera Functions Using Your Digital Camera. 1. What happens when you press the shutter button down halfway? What does macro mode allow you to do? Pressing.
Introducing the LEO 1400 Series
TEM Prep using FIB and Kleindiek Micro-Manipulator General Reminders, Tips and Tricks When in FIB view, work quickly to do what is needed and then freeze.
A stereo dissecting microscope is normally used to view relatively large specimens at magnifications from about 5X to about 50X. The scope provides a very.
MDS Alignment All 3 modes have their own alignment.
Copyright ⓒ VATECH Global Co., Ltd. All rights reserved EasyStartGuide Please refer to the included User Manual for more detailed functions on EzDent-i.
DO NOW: 1.Hand in the field trip brainstorming you wrote on Friday. 2.Take out your microscope Green packets from last week. 3.In your NOTES section of.
Videographer Basic Tutorials For the Internet. Materials you should have At least a 150 GB external drive Your camera The proper cables needed to connect.
Shooting. Initial Camera Settings Cameras have default settings for picture quality. The school’s cameras are no exception. Camera resets to default settings.
 Understanding Aperture – Shutter Speed - ISO.
MICROSCOPE NOTES. Types of Microscopes  1. Compound Light Microscope: allows light to pass through an image and use two lenses to magnify object  2.
Color Correct and Remove Keystoning A minimalist approach to photographing your art By Paul Marley.
DIC Differential Interference Contrast Kim Peifley 08/14/15.
Gauge Operation and Software by Scott A. Ager. Computer Recommendations 750 MHz Pentium III 64 Meg SRAM 40 Gig Hard Drive 1024 x 768 graphics CD Writer.
Scanning Electron Microscope (SEM)
PowerPoint Basics Tutorial 3: Graphics In this tutorial we’ll be looking at graphics, and the various types of illustrations that can be included in a.
Microscopes SC.912.L.14.4 Compare and contrast structure and function of various types of microscopes.
Scanning Electron Microscope (SEM)
MAJ Kris Paolino, MD Infectious Disease Staff Chief, Clinical Trials Center Walter Reed Army Institute of Research.
Nikon Spinning Disk Imaging & Training Support. Find your cells with DIC in Live Preview 1.The green DIA Lamp, used for DIC, should be on. 2. Condenser.
Flatbed scanning in the HoA Visual Resources Centre How to use the Epson scanner.
VISTA IMAGING RADIOLOGY VIEWER. 2 The focus of this document is on the VistA Imaging Display Radiology Viewer. Other Display changes will be discussed.
Microscope Basics.
Setting up a trinocular dissecting microscope for best viewing
PHOTOGRAPHY BASICS PARTS OF THE CAMERA AND HOW IT WORKS.
Understanding Microscopes
Microscopes Biology 11. The History 4 Many people experimented with making microscopes 4 Was the microscope originally made by accident? (Most people.
Chapter 3 Device Monitor Screen Otasuke GP-EX! Chapter 3 Device Monitor Screen Chapter 3 Device Monitor Screen.
Microscopes Biology I. Magnification 1. Power a microscope has to increase an objects size.
1 DVR SYSTEM User’s Manual Manual Revision : Version 1.0 Release Date : Nov. 2, 2002 ■ SERVER ■ CLIENT.
JMARS THEMIS Targeting Software An Overview for Mars Student Scientists Revised: 19 September 2006.
1 Setting Laser Power and PMT Gain Begin with laser power at 20% Set PMT gain to 6.00.
COMPUTER LITERACY 1 TRANSITION PLUS BASIC COMPUTER USE PART 1.
The Camera -Using a 35mm SLR Camera - An Introduction to.
5/14/2003Sprint TekNet IP Train the Trainer1 Open TekNet Software If working at a client station, enter the IP address of the server and mark page as a.
General Biology lab Microscope.
Scanning Electron Microscope Eee-Jay Rodriguez. The Structure of the Microscope Inside Outside.
Mapping EP4.
The Microscope.
How to SWIVl: Using a Gen 2 SWIVL RObot
Visual Aid.
Standard Operation Procedure
Feb 23, 2010 Pick up a sheet from the front Read both sides
DPP Shell Adjustments Short Procedure
University of Warith AL-Anbiya’a
Harvard Medical School Neurobiology Imaging Facility
Quanta FIB Basic Training xT Microscope Control
Basic Operational Instructions
Edge Lights Tutorial Making the Light Shows.
Quanta FIB Basic Training Eucentric, Pt Deposition, and Trenching
Apreo FE-SEM Basic Training xT Microscope Control & Operation
Presentation transcript:

Operation of the S4700 FESEM Quality SEM analysis is a matter of three factors 60% Sample- Preparation or type of sample 30% Knowledge of the operator 10% Type of instrument

Hitachi S4700 Instrument was installed in August of 1999 Has EDS, Backscatter, EBSD capabilities High resolution imaging, up to 500,000X Low voltage imaging, down to 0.5kV

Basic Operation of the S4700 Each time you use the scope you need to fill the LN2 trap to the side of the scope.

Degas not Degauss If you are the first to use the scope then you must “Degas” the lower aperture This is to clean the lower aperture. Flip the switch located in the left of the instrument panel on the column down from “Heating” to “DEGAS”

Software Software You will be given a user account and you will set up a password that will be changed each time you use the scope. The initial logo should appear once you log in If it does not then open the SEM software from the FE-SEM icon on the desktop

This is the operating window of the 4700 software. All operation can be done with the mouse or with the help of the small control panel.

Log Book The microscope operating conditions will be written in a log book. These include the readings of the ion pumps, Ip1, Ip2, Ip3, the Vext, Ie, Vacc, date, user name, advisor and flash information if applicable

Set the HV using the HV control window and the drop down Vacc menu Set the HV using the HV control window and the drop down Vacc menu. HV’s between 0.5-30 kV are available

kV selection for a sample should be based on the type of analysis Samples that are non-conductors such as ceramics, glass, some oxides should be run at lower kV due to charging effects, which can cause problems with imaging. Samples that are conductors such as metals can be run at higher kVs but might not be necessary. Check the sample to see if the desired results are there at lower kVs. There is no need to blast the sample if the desired results can be obtained at a lower voltage. Remember the 2X rule for doing EDS on samples.

Insert Sample into Sample Exchange chamber Once the sample has been correctly prepped, it should be screwed to the end of the insertion rod. DO NOT BEND THE INSERTION ROD WHEN MOVING THE SAMPLE!

Chamber Scope On switch The S4700 is fitted with in infrared camera. This allows the user to see inside the sample chamber and avoid bumping into detectors when inserting samples and moving the stage Chamber scope is turned on prior to inserting the sample into the sample chamber On switch

Sample Chamber When the S.E.C. light on the front panel turns green open MV1 on the sample exchange door and allow the sample to be pulled into the sample chamber by the vacuum, watch on the chamber scope.

Column Setup As the sample chamber pumps down, use the Column Setup for the type of SEM examination The S4700 has 6 operation modes for the column Ultra High Resolution Analysis Normal Long WD Magnetic UHRA

Working distance – True working distance and can be set. Condenser lens 1 - Default is 5, smaller numbers are bigger spot size, range is 1-16. DeGauss- To degauss the column Select Detector – Mixed for analysis and imaging Upper for high resolution Lower for analysis and samples that have slight charging

HV On Turn on HV in the HV window when the sample chamber is at L X 10-3 Check Ip3, if it is 1 X 10-5 then DO NOT OPERATE Note the Vacc, Ie and Vext and put in Log book

Imaging Beam alignment Select scan speed Locate area of interest Go to High Mag mode Aperture alignment Stigmate Final Focus Brightness and contrast Capture image

Beam alignment Focus the image Click on “Beam alignment” in the “Alignment” window Use the left knobs on control panel to adjust the beam in the middle of the target. Will have to been done multiple times throughout operation.

Alignment window Beam align button Beam Move the beam to the center of the target Can be done in the grid window using the mouse

Select scan speed 6 scan speeds Fast 1 and 2 – Frame averaging Can be set in the “Signal Processing Window” Slow 1-4 – are raster scan in seconds and are preset 5 is a reduced area scan used to aid in focusing in a smaller area

Signal Processing window Scan Speed buttons 1 2 1 2 3 4 5 Signal Processing window

Locate area of interest Orient the sample using 5 axis stage X, Y, Z, Tilt, and rotation Move the sample to locate an area of interest, with trackball, click and drag, stage arrows, change x, y, z, tilt, and rotate Trackball Mouse

Stage Control Change x and y positions Rotate Z position Tilt Click and drag either the stage or the beam. This button for the beam. Blue hand cursor Stage arrows This button for the stage. Yellow hand cursor Stage Control Change x and y positions Rotate Z position Tilt

Go to High Mag mode Use the Low Mag mode to locate an area of interest Low Mag mode is 30-10,000x and can not be used for EDS Use High Mag mode for EDS and imaging High Mag mode is 250x to 500,000x Not all samples can be imaged at 500,000x Typical magnification is 25,000x to 250,000x

Aperture alignment Check the beam alignment Increase the mag, best done at the highest resolution Click on the “Aperture Align” in the “Alignment” window Use X and Y alignment knobs on control panel Good alignment has little or no X or Y movement. Image should pulse or rotate

Low Mag/High Mag mode Aperture Alignment

Stigmate Refocus the image after Aperture alignment If the image skews out of focus it is out of sigmation. Use knobs on the left of the control panel Y Stigmate first, X Stigmate, fine focus then check Y and X stigmation again Gross movement of the image indicates the sigmators need alignment Use the Alignment window and proceed with sigmation alignment in the same manner as aperture alignment.

Stigmation alignment

Final Focus Select a slow scan speed to increase resolution Refocus image

Brightness and contrast After focus do an “Auto Brightness and Contrast” Brightness and Contrast can also be adjusted manually with the “Brightness and Contrast” monitor button

Auto Brightness and Contrast Button Brightness and Contrast Monitor Button

Capture Image Select image capture options in the “Capture Image Window” Capture an image with the “M” button on the main window. Image will be temporarily saved to a “Capture Image Window” “Capture Image Window” will appear, use the “Disk Icon” to save the image. All Images are to be saved on the D drive At completion images should be moved to the internet or recorded on a CD.

Capture Resolution-Change the size of the captured image Capture speed of the image- Frame averaging or slow scanning speed

Capture Image Window Red “Saved”, image has been saved Red number in upper left corner is the active image Yellow number in the upper left corner is the next window to be occupied Disk Icon for saving the image

Use the SEM Data Manager to view your images

Blurred Images Beam Alignement Aperture Alignement Stigmantion Coarse Focus Fine Focus Building Vibrations Stray Electric Fields Tip Noise

Hints on Stigmation Check Stigmation Alignment Focus best possible image with fine focus with no “Skewing” Y stigmate X stigmate Fine Focus

Tip Noise Flash Tip Open the HV Window Select “Flash” Click on Flash and watch the Ie number quickly appear in the window Note this number in the Log Book DO NOT operate if the Vext has changed 1.2 more than the most recent Flash Ie

Ie and Vext Ie changes rapidly after flashing “SET” a new Vext as the Ie drops Best working conditions are at an Ie of >9 If Ie goes up to 12 STOP operation and flash

Ie can be changed, most common is 10 micro Amps The “On” button becomes a “SET” Button when HV is on Flashing button Ie can be changed, most common is 10 micro Amps