For midterm. Topics to be covered  Repeat Analysis  Analysis Of Reject Versus Repeated Films  Determination Of Rejection Rate  Distribution Of Rejected.

Slides:



Advertisements
Similar presentations
PROCESSOR QA.
Advertisements

Preparing for a F&A Proposal Base Year 2011 NCURA Region VI & VII Conference Presented By: Adrienne Clifton and Ginger Baker, MAXIMUS.
Quality Management in Diagnostic Imaging
Artifact analysis.
Dr. Hamda Qotba, M.D,MFPH,FFPH
QUALITY MANAGEMENT University of Sharjah College of Health Sciences
Quality assurance and Quality control in medical radiography
Obj: Discuss imaging careers and technology.. Radiography or diagnostic imaging Radiography: Making film records of internal structures by passing radiographs.
OverviewOverview – Preparation – Day in the Life – Earnings – Employment – Career Path Forecast – ResourcesPreparationDay in the LifeEarningsEmploymentCareer.
Primary Exposure Factors IV
RAD309 Patient Dose.
FOCUS Voluntary Community, May 2011 IMAGING IN FOCUS.
ICASAS305A Provide Advice to Clients
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Laboratory Quality Control
Essentials of Dental Radiography for Dental Assistants and Hygienists CHAPTER Copyright ©2012 by Pearson Education, Inc. All rights reserved. Essentials.
Laboratory Management - 1
Prelim 11/14/11. 2 nd Semester, 3 rd Year Level  The Early 1900’s, With The Work Of An Industrial Engineer Named Frederick Winslow Taylor.  “Father.
a judgment of what constitutes good or bad Audit a systematic and critical examination to examine or verify.
Purpose of the Standards
Radiation Exposure and Risk John Hamaty. Evaluation of Radiation Exposure Levels in Cine Cardiac Catheterization Laboratories American Association of.
Quality Assurance/Quality Control Chapter 10. Quality Assurance Definition A system of activities whose purpose is to provide assurance that the overall.
Unit IV Analyzing the Image. Unit IV Analyzing the Image.
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
How to Effectively Implement Set-up Reduction in Any Organization.
Quality Assessment 2 Quality Control.
CLINICAL ENGINEERING part(3) Dr. Dalia H. Elkamchouchi.
9 Closing the Project Teaching Strategies
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.03: Optimization of protection in Mammography.
Lecture #9 Project Quality Management Quality Processes- Quality Assurance and Quality Control Ghazala Amin.
© 2015 Cengage Learning. All Rights Reserved. May not be scanned, copied or duplicated, or posted to a publicly accessible website, in whole or in part.
Evaluation of Internal Control System
Organization and Implementation of a National Regulatory Program for the Control of Radiation Sources Inspection Part II.
 The Facilities and Administrative Rate (F&A Rate) is the mechanism used to reimburse the University for the infrastructure support costs associated.
© 2012 Cengage Learning. All Rights Reserved. This edition is intended for use outside of the U.S. only, with content that may be different from the U.S.
4.3 Document control 4.4 Review of requests, tenders and contracts
Quality of Health and the use of protective devices by radiological technologists to minimize the effects of ionizing radiation among Northeast Tennessee.
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
21/4/2008 Evaluation of control measures 1. 21/4/2008 Evaluation of control measures 2 Family and Community Medicine Department.
SMS Planning.  Safety management addresses all of the operational activities of the entire organization.  The four (4) components of an SMS are: 1)
EXPERIMENTAL EPIDEMIOLOGY
Project quality management. Introduction Project quality management includes the process required to ensure that the project satisfies the needs for which.
Radiographic Quality Visibility and Sharpness
©2013 Delmar, Cengage Learning. All Rights Reserved. May not be scanned, copied, duplicated, or posted to a publicly accessible website, in whole or in.
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
COSTING AND THE VALUE CHAIN CHAPTER 18 PAGE# 794 Faisal
Quality Assurance in the Dental Office
Diagnostic reference levels in Medical Imaging. Concept and practice
Radiation Sources in medicine diagnostic Radiology
RADIOGRAPHIC TECHNIQUES IN CHILDREN Maha AL-SARHEED.
Quality Assurance. Define Quality (product & service) Exceeds the requirements of the customer. General excellence of standard or level. A product which.
Copyright © 2012, 2006, 2000, 1996 by Saunders, an imprint of Elsevier Inc. 1 Chapter 11 Dental Radiographs and the Dental Radiographer.
Analog X-ray Imaging Recommended Book: Walter Huda, REVIEW OF RADIOLOGIC PHYSICS By: Maisa Alhassoun
 Pharmaceutical Care is a patient-centered, outcomes oriented pharmacy practice that requires the pharmacist to work in concert with the patient and.
Quality assurance for diagnostic imaging equipment.
Quality Assurance in Dental Radiography
Accuracy, sensitivity and specificity analysis
RADIATION PROTECTION 04/12/2016.
QUALITY MANAGEMENT Suzanne Kamel-Mohamed PhD, MBA, MT (ASCP) Associate
Pediatric Radiology By Dr. Wambani, J.S. Chief Radiologist
PROJECT QUALITY MANAGEMENT Teknik Elektro FT UNDIP
Factors Effecting the Production
Quality Management Systems – Requirements
Practical clinical chemistry
Accuracy, sensitivity and specificity analysis
Conduct of Film Analysis
Shaimah Al-Failakawi Al Amiri Hospital Laboratory Quality Manager
Processing Radiographs
Pima Medical Institute Online Education
Quality Control Lecture 3
Presentation transcript:

For midterm

Topics to be covered  Repeat Analysis  Analysis Of Reject Versus Repeated Films  Determination Of Rejection Rate  Distribution Of Rejected Film  Guide To Good Practice

The desired outcome of a diagnostic x-ray study is the creation of an acceptable diagnostic image, a correct diagnosis, and the satisfaction of all internal and external customer.

 Quality control testing can only minimize the risk of obtaining sub- quality images but cannot prevent them entirely.  Quality management program must also include looking at the final outcome of a diagnostic x-ray procedure and determining the quality of outcome to see if further improvement can be achieved.

Outcome assessment  Repeat analysis of images  Artifact analysis of images  Accuracy, sensitivity and specificity analysis of diagnosis.

Repeat analysis It is a systematic process of cataloging rejected films and determining the nature of the repeat so that repeat images can be minimized or eliminated in the future

Importance of repeat analysis  It provides knowledge of data about: Equipment and accessory performance Departmental procedures Skill level of the technical staff  With this knowledge, solutions can be found to minimize repeats and also document the effectiveness of quality control and quality assurance protocols

Advantages  Repeat analysis can show how many repeats in the depart w/c in turns address the party concerns of what are the problems.  With lower department repeat rates: Improved department efficiency Lower department cost Lower patient doses

Improved department efficiency  With number of repeats are kept low, the amount of time that patient must spend undergoing diagnostic procedures decreases.  This in turns increases patient satisfaction and allows the department to service more patients in the same period.

Lower department costs  When the number of repeats are reduced, the costs associated with film, processing, labor and depreciation of the equipment decrease significantly.

Lower patient doses  Images that are unacceptable results in the repeat of the particular view, which means that the patient must be re-exposed to ionizing radiation.

Causal Repeat rate  In repeat analysis studies performed on departments without quality control procedures for the darkroom, processor and equipment, 75% of all repeats were caused by improper optical density of film.  With quality control protocol in place, studies shows that most repeats are results of positioning errors.

National average of repeat causes for departments with quality control procedures Category% Repeat Positioning30 Light films14 Misc. (artifacts)14 Darkrooms11 Black films9 Tomography scouts8 Fog5 Patient motion5 Mechanical problems4

Repeat analysis  For a study to be done, a worksheet for radiographic imaging departments should be used, so that the proper statistical information is obtained and recorded.

Worksheet  A worksheet should include the radiographic procedures performed in the department, along with the possible causes of rejection, such as positioning, overexposure, under exposure, motion, artifacts, and miscellaneous causes

Causal repeat rate  The causal repeat rate is the percentage of repeats from a specific cause and is calculated with the following equation: Causal repeat = rate Number of repeats for a specific cause Total Number of repeats X 100

Example  A total repeat of a department has 200 films in a month, and 90 out of 200 films are the result of positioning errors.  So the percent of repeats caused by positioning error is 45% of the total repeat films.

Total repeat rate  The total department repeat rate is determined with the following equation: Total repeat rate = Number of repeat films Total number of views taken X 100

Example  If a department has a total of 2300 views during a month and has incurred a 300 rejected film.  A 13% percent the department has incurred in a month

 Department repeat rates should not exceed 4% to 6% and should be less that 2% for mammographic procedures.  Any department exceeding 10% to 12% should be examined seriously, because of inefficiency and contribute to high patient dose.

 For a department to evaluate its efficiency a worksheet should be used for any repeat film.  As a QA/QC officers a worksheet can aide in identifying problems if it is mechanical/technical or by human errors.

Laboratory Exercise  Instruction Each group are assigned with a number of films. Analyze the films. After analyzing the films, the film should be rotated to other groups for analysis.

Survey period to location repeat category examinati on PositionOver- exposed Under- expose MotionArtifactsOtherTotal% UGI % Ba. Enema % IVP % OTHER % TOTAL %29%6%7%23%28%8% Total films viewed 1320Total repeat 144TotalRepeatRate11%