Centrifuge Theory and Safety

Slides:



Advertisements
Similar presentations
Section 13.3 & 13.4 Centripetal and Centrifugal Forces
Advertisements

Hydrodynamic Techniques
Introduction Study of cells, tissues or organs in vitro More than 24 hours Behaviour of animal cell without variations in the animal.
Centrifuges. Topics  Principle of Operation  Diagrams  Applications  Types of Centrifuges  History  Safety  Operation  Preventive Maintenance.
CentrifugeSafety. Warning Basically very simple Basically very simple Potentially very dangerous Potentially very dangerous.
Good Laboratory Safety Practices. Buddy System Be Safe! Don’t be alone when working with hazardous materials or in dangerous work conditions. Your life.
PLASMID ISOLATION AND ANALYSIS Part II Plasmid Purification and Isolation.
Plasmid Minipreps Kits….
If you use your computer more than a few hours per day, we suggest you not to turn off your computer at the end of the day. Instead turn off only the.
Cloning a DNA segment from bacteriophage Recombinant DNA transformed into bacterial cells Last week we plated cells onto agar plates + ampicillin + X-gal.
Cloning a DNA segment from sheep Recombinant DNA transformed into bacterial cells Last week we plated cells onto agar plates + ampicillin + X-gal Controls:
Safety Practices in Healthcare. Safety Standards A. Defined: set of rules designed to protect both the patient and the health care worker B. Established.
1 T.Elsarnagawy MLI-mde 210, 311 Laboratory Centrifuges.
Microhematocrit.
Filtering and Centrifugation Physical Separation of Solids from Liquids.
The Airtex Educational Partnership (AEP) was started several years ago by Airtex Automotive Division as a way to merge corporate automotive expertise with.
1 MLAB 2401: Clinical Chemistry Chapter 3: Basic Principles and Practice of Clinical Chemistry, part 2.
Annual Refresher Centrifugation Hazards. Centrifugation 101 Every time you use a centrifuge, you make series of choices. Which centrifuge, which rotor,
Centrifuge Safety Department of Cell Physiology andPharmacology Tim Walton Departmental Safety Officer.
Centrifuges
Module 11: Use and Care of Equipment At the HIV Rapid Testing Site.
Clinical Laboratory equipment. Clinical laboratory is equipped with various biomedical instruments, equipments, materials and reagents for performing.
Experiment. No 2 Common materials, apparatuses and equipment used in Serology laboratory Done By: Thamer Hamdan M.Sc. Clinical Microbiology and Immunology.
Laboratory Hardware Part 1 Bioscience, Equipment Unit 6 Centrifuge, Water Baths, and Incubators 8/29/
Pharos university Faculty of Allied Medical SCIENCE Clinical Laboratory Instrumentation (MELI-201) Dr. Tarek El Sewedy.
BY: AARTI SHINDE JYOTI SHUKLA MANISH PILLAI SARANYA NAIR.
Proteomics Module Day 1 Tech talk. Experiment: Yeast protein expression changes caused by H 2 O 2 exposure. ► 2 Control groups (A and B): nothing added.
Proteomics Module Day 1 Tech talk 10 students in 5 groups of 2.
CAVEAT!!! Usually we do these procedures with a pure culture ◦ Grow bacteria in broth ◦ Streak on plate ◦ Grow single colony on broth.
CLINICAL APPLICATIONS AND MAINTENANCE PROCEEDURES FOR LABORATORY EQUIPMENT.
Dr Gihan Gawish 1. Hydrodynamic methods  Common experimental manipulations in biochemistry  centrifugation, dialysis and filtration are strongly influenced.
CENTRIFUGE WITH ADJUSTABLE SPEED AND DIRECTION OF ROTATION
Cell fractionation and centrifugation
Pharos university Faculty of Allied Medical SCIENCE Clinical Laboratory Instrumentation (MELI-201) Dr. Tarek El Sewedy.
Plasmid Isolation Transfection Transfect plasmid into cells
LAB RULES Your life depends on it. Rule # 1 Wear safety goggles at all times.
CAPE Biology Workshop on Concepts in Biotechnology & Genetic Engineering Prepared and presented by Dr. Marcia E. Roye.
CENTRIFUGATION. OUTLINE Basic theory Applications Instrumentation.
Sedimentaion.
1 Part 2 Directions: As your instructor goes through the safety rules for the lab, write down the missing word or words in the blank spaces on your student.
CENTRIFUGE GUIDANCE. Background  Most common centrifuge incidents are related to the centrifuge rotors.  All centrifuges are at risk of rotor-related.
The Cell Joseph. Plant Cells Plant cells are cells that are in plants. Plant cells are like animal cells, but they have a cell wall and chloroplasts.
HVACR416 - Design Pressure Measurement Part 1. Why Measure? Need to Make Sure: All combustion appliances vent properly All HVAC systems are installed.
T.Elsarnagawy 11 Laboratory Centrifuges MLI-mde 210, 311.
Lab SAFETY. Laboratory Safety Rap! eo.php?video_id=
Microhematocrit Centrifuge
1 T i r e s a n d W h e e l s 2 What Wheels Are Made Of  Made  Made of stamped or pressed steel that are either riveted or welded together. or of aluminum.
© SSER Ltd..
 “High fire” glazes are fired in the large brick kiln to a temperature of 2380° Fahrenheit.  It takes about 24 hours to reach that temperature.
KAU-Faculty of Science- Biochemistry department Analytical biochemistry lab (Bioc 343) 2012 T.A Nouf Alshareef Lab (8):
By-pooja kamble m.Sc-i. UNIT I:MICROTOMY,MICROSCOPY,CENTRI FUGATION 1.3-PRINCIPALS AND APPLICATIONS OF CENTRIFUGATION.
Cell fractionation and centrifugation
Biotechniques 410 Centrifugation.
Chapter 11. 원심분리를 이용한 마우스 혈액 면역 세포 분리
B.T. INSTITUTE OF RESERCH & TECHNOLOGY SAGAR (M.P.)
Veterinary Practice Laboratory Unit 1 Chapter 2 General Laboratory Equipment Copyright © 2015 by Mosby, an imprint of Elsevier Inc. All rights reserved.
Centrifugation.
Chapter 5: Equations Equations are a way to describe relationships using mathematical symbols. V = IR voltagecurrent resistance.
Bioseparation I Centrifugation. What is Bioseparation?  Purification or separation of a specific material of interest from contaminants in a manner that.
Centrifuge The centrifuge used to separate solids suspended in a liquid by sedimentation. The rotational movements allow forces much greater than gravity.
Laboratory Equipment.
Laboratory instruments
TOPIC - CENTRIFUGATION PRESENTED BY – SHIVALI VERMA.
Laboratory instruments
Protein Production Jackpot!
# A mechanical method of seperating immiscible
Hydrodynamic Techniques
Click anywhere to get started…
Introduction To Medical Technology
Centrifuges.
Presentation transcript:

Centrifuge Theory and Safety

Centrifuge A mechanical device that uses the principle of centrifugal force to separate materials of different densities. In biotechnology research, these materials are usually cells, organelles, or large molecules, such as DNA. Centrifugal force- An apparent force that acts outward on a body moving around a center, arising from the body's inertia (whereas centripetal force is inward)

Centrifuge at NASA Ames Research Center. Featured in movie “Space Cowboys,” 2000

Urania Observatory, The Netherlands

Benchtop Centrifuge Also known as multipurpose centrifuge. Not always found on a benchtop. Uses: Pellet cells and bacteria Fixed angle rotor: 17,000 g/14,000 rpm Swinging bucket rotor: 3,800 g/ 4,800 rpm

Clinical Centrifuge Uses: Serum, urine and blood sedimentation Fixed angle rotor and swinging bucket: 4600 g/ 6000 rpm

Microfuge Uses: Cells (at low speeds) and to spin down a reagent in a tube Fixed angle (rare swinging bucket): 16,000 g/ 13,000 rpm

High Speed Centrifuge Uses: Large volume, pelleting bacteria, spin columns and protein precipitations Fixed angle rotor and swinging buckets: 82,000 g / 26,000 rpm

Ultracentrifuge Uses: Virus concentration, membrane and subcellular fraction isolation, DNA and RNA isolation Fixed angle rotor and swinging bucket: 800,000 g / 100,000 rpm

Benchtop Ultracentrifuge Uses: Membrane preps, virus isolation, subcellular fractionation, CsCl DNA plasmid isolations in 30 minutes Fixed angle rotor and swinging bucket: 625,000 g/ 120,000 rpm

Rotors The internal rotating unit within a centrifuge which holds tubes and plates. Fixed angle rotor Swinging bucket rotor

Primary hazards Mechanical failure Dispersion of aerosols (hazardous aerosols)

Mechanical Failure Is caused by age and by improper use or inadequate care of centrifuge or rotor. Especially the rotor.

Care and Attention Safe high-speed spin requires nearly perfectly balanced load. Age, use, and misuse contribute to rotor flaws. A rotor which comes apart at high speed can be deadly.

Tiny flaws are …NOT so tiny at 80,000 rpm These are micro-fissures and cracks caused by stress and corrosion in the bottom of a tube cavity.

(Purdue 2003) This rotor came apart while coming up to speed (not yet spinning at full speed) The 6 inch (15 cm) long chunk damaged (ruined) but did not completely penetrate the centrifuge lid and housing. The motor and spindle were also destroyed.

To reduce wear and chance of failure Follow instructions in manual and rotor care guide Use only compatible rotors Check routinely for rotor damage Don't overfill tubes Cap tubes Balance load carefully Check that rotor correctly seated on drive spindle Run at < max safe speed continued……

To reduce wear and chance of failure, continued Stay right there until full speed is reached. Stop the centrifuge if anything seems unusual Never open until stopped (never ‘manual brake’) Clean rotor gently Dry completely Store rotor upside down (why?) Maintain a careful rotor log Derate and retire rotors for age/use

(MIT 1999) This rotor split in half at 55,000 rpm after 3 h of what was supposed to be a much longer spin. Grad student report “Rotor was manufactured in 1986 and was covered for 10,000 hours of operation or 5 years, whichever comes first.” No surprise that it flew apart 13 years later. Fortunately the centrifuge housing contained it.

(Cornell, 1998) The rotor failure was not contained. Much other damage done, equipment ruined, chemicals spilled. Nobody was near, or someone would have been badly hurt or killed.