Dose Distribution and Scatter Analysis

Slides:



Advertisements
Similar presentations
X-Ray & γ-Ray Interactions with Matter
Advertisements

Formation Characteristics
X-Ray Interaction with Matter & Human Biology
Photon Beam Monitor-Unit Calculations
Radiation Interaction Q&A
BME 560 Medical Imaging: X-ray, CT, and Nuclear Methods
Radioactivity – review of laboratory results For presentation on May 2, 2008 by Dr. Brian Davies, WIU Physics Dept.
Dose. Energy Gained Particles lose energy in matter. Eventually energy loss is due to ionization. An important measure is the amount of energy gained.
Measurement of Absorbed Dose (6)
Radioactivity – inverse square law, absorption, and rates presentation for Apr. 30, 2008 by Dr. Brian Davies, WIU Physics Dept.
Sec.3 Practical application 临床处方剂量的计算.  What we should calculate?  Time or mu required to delivery proper dose for tumor  Dose distribution at the.
Radiation Dosimetry Dose Calculations D, LET & H can frequently be obtained reliably by calculations: Alpha & low – Energy Beta Emitters Distributed in.
8.1 PRODUCTION AND CHARACTERISTICS OF X-RAYS
X-Ray Production & Emission
Alexander Brandl ERHS 630 Exposure and Dose Environmental and Radiological Health Sciences.
Radiology is concerned with the application of radiation to the human body for diagnostically and therapeutically purposes. This requires an understanding.
At the position d max of maximum energy loss of radiation, the number of secondary ionizations products peaks which in turn maximizes the dose at that.
Electron Beams: Physical Principles and Dosimetry
X-Ray Production & Emission
Radiographic Grids I By Professor Stelmark.
Beam Therapy Equipment 3 Patient Treatment and Accessories.
1 Photon Interactions  When a photon beam enters matter, it undergoes an interaction at random and is removed from the beam.
MECHANISMS OF HEAT TRANSFER
Electron Beams: Physical Principles and Dosimetry
5.4.1 X-Rays. (a) describe the nature of X-rays Stowmarket Physics X-rays - nature Forms of electromagnetic radiation Short wavelength High frequency.
Radiation therapy is based on the exposure of malign tumor cells to significant but well localized doses of radiation to destroy the tumor cells. The.
Stopping Power The linear stopping power S for charged particles in a given absorber is simply defined as the differential energy loss for that particle.
Linac Beam.
Dosimetric evaluation of a new design MOSFET detector Per H. Halvorsen* & Stephanie Parker University of North Carolina.
External Beam Radiotherapy
Radiographic Dosimetry
Resident Physics Lectures
Fundamental Dosimetry Quantities and Concepts: Review Introduction to Medical Physics III: Therapy Steve Kirsner, MS Department of Radiation Physics.
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
1 Dosimetry part 1: X-ray interactions with matter G.Haddadi, PhD, MSc Associated prof. of Medical Physics Fassa University of Medical Sciences.
APHY398C 6/4/ Dosimetry   Quantifying the incidence of various biological changes as a function of the radiation dose.   Exposure Ratio of total.
Photon Tissue Interactions
Chapter 5 Interactions of Ionizing Radiation. Ionization The process by which a neutral atom acquires a positive or a negative charge Directly ionizing.
Interactions of EM Radiation with Matter
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
Part VIII:Medical Exposures in Radiotherapy
Appendix G1: Chapter 12: X-Ray Interaction with Matter
Determining Radiation Intensity
Attenuation As x-rays pays through matter, the exit beam will contain less photons than the entrance beam. This reduction in the quantity of photons is.
Radiographic Grids I By Professor Stelmark.
The Radiotherapy Application NMT 232 L 8 1. Radiotherapy Generators Kilovoltage X-rays are generally used in the treatment of skin cancers and superficial.
Interaction of x-ray photons (and gamma ray photons) with matter.
1 A protocol for the determination of absorbed dose from high energy photon and electron beams AAPM TG-21 Protocol (Med Phys 10: , 1983) 和信治癌中心醫院.
AAPM TG-51 Protocol (Med Phys 26: , 1999)
Interactions of Ionizing Radiation
Interaction of Radiation with Matter
Chapter 5 Central Axis Depth Dose Calculations. 2 Definition of Beam Geometry The accurate delivery of a radiation dose to patient depends on the precise.
Chp 2 : Interaction of radiation with matter Radiation Contrast
CHAPTER 3 DOSE DETERMINATION FOR EXTERNAL BEAMS
INTERACTION OF PARTICLES WITH MATTER
Electron Beam Therapy.
Khan, The Physics of Radiation Therapy, Ch7
Part No...., Module No....Lesson No
Absolute Dose Measurement
A system of dosimetric calculations
Chapter-9 Dose Distribution And Scatter Analysis
Interactions of Radiation With Matter
X-ray Interactions with Matter
Resident Physics Lectures (Year 1)
Ch 10. A System of Dosimetric Calculations
Chapter 8 (Part 1) Measurement of Absorbed Dose
Resident Physics Lectures (Year 1)
Hot and cold spots are common problems associated with planning:
AL305-Physics II Meeting #2 Photon Beam Dosimetry
Principles and Practice of Radiation Therapy
Presentation transcript:

Dose Distribution and Scatter Analysis Phantoms Depth Dose Distribution Percentage Depth Dose Tissue-Air Ratio Scatter-Air Ratio

Phantoms Water phantom: closely approximates the radiation absorption and scattering properties of muscle and other soft tissues; universally available with reproducible

PHANTOMS Basic dose distribution data are usually measured in a water phantom, which closely approximates the radiation absorption and scattering properties of muscle and other soft tissue Another reason for the choice of water as a phantom material is that it is universally available with reproducible radiation properties.

Compton effect is the main interaction PHANTOMS Solid dry phantoms tissue or water equivalent, it must have the same effective atomic number number of electrons per gram mass density For megavoltage photon beams in the clinical range, the necessary condition for water equivalence same electron density (number of electrons per cubic centimeter) Compton effect is the main interaction

Solid dry phantoms

Solid dry (Slab) phantoms

Alderson Rando Phantom anthropomorphic phantom Frequently used for clinical dosimetry Incorporates materials to simulate various body tissues, muscle, bone, lung, and air cavities

RANDO phantom CT slice through lung Head with TLD holes

Depth Dose Distribution The absorbed dose in the patient varies with depth The variation depends on depth, field size, distance from source, beam energy and beam collimation Percentage depth dose, tissue-air ratios, tissue-phantom ratios and tissue-maximum ratios---measurements made in water phantoms using small ionization chambers

Percentage Depth Dose Absorbed dose at any depth: d Absorbed dose at a fixed reference depth: d0 collimator surface phantom D d0 D d d d0

PERCENTAGE DEPTH DOSE For orthovoltage (up to about 400 kVp) and lower-energy x-rays, the reference depth is usually the surface (do = 0). For higher energies, the reference depth is taken at the position of the peak absorbed dose (do = dm).

Percentage Depth Dose For higher energies, the reference depth is at the peak absorbed dose ( d 0= d m) D max : maximum dose, the dose maximum, the given dose collimator surface phantom D max D d d dm

Percentage Depth Dose (a)Dependence on beam quality and depth (b)Effect of field size and shape (c)Dependence on SSD

Percentage Depth Dose (a)Dependence on beam quality and depth Kerma— (1) kinetic energy released per mass in the medium; (2) the energy transferred from photons to directly ionizing electron; (3) maximum at the surface and decreases with depth due to decreased in the photon energy fluence; (4) the production of electrons also decreases with depth

Percentage Depth Dose (a)Dependence on beam quality and depth Absorbed dose: (1) depends on the electron fluence; (2) high-speed electrons are ejected from the surface and subsequent layers; (3) theses electrons deposit their energy a significant distance away from their site of origin

Fig. 9.3 central axis depth dose distribution for different quality photon beams

Percentage Depth Dose (b)Effect of field size and shape Geometrical field size: the projection, on a plane perpendicular to the beam axis, of the distal end of the collimator as seen from the front center of the source Dosimetric ( Physical ) field size: the distance intercepted by a given isodose curve (usually 50% isodose ) on a plane perpendicular to the beam axis

PDD - Effect of Field Size and Shape Geometrical Dosimetrical or physical SAD FS

Percentage Depth Dose (b)Effect of field size and shape As the field size is increased, the contribution of the scattered radiation to the absorbed dose increases This increase in scattered dose is greater at larger depths than at the depth of D max , the percent depth dose increases with increasing field size Dd Dmax Scatter dose

Percentage Depth Dose (b)Effect of field size and shape Depends on beam quality The scattering probability or cross-section decreases with energy increase and the higher-energy photons are scattered more predominantly in the forward direction, the field size dependence of PDD is less pronounced for the higher-energy than for the lower-energy beams

Percentage Depth Dose (b)Effect of field size and shape PDD data for radiotherapy beams are usually tabulated for square fields In clinical practice require rectangular and irregularly shaped fields A system of equating square fields to different field shapes is required: equivalent square Quick calculation of the equivalent

rectangular field c A square field B c A x B c = 2 x A + B

Percentage Depth Dose (b)Effect of field size and shape Quick calculation of the equivalent field parameters: for rectangular fields For square fields, since a = b, the side of an equivalent square of a rectangular field is a b

Percentage Depth Dose(3)--(b)Effect of field size and shape Equivalent circle has the same area as the equivalent square a b r

Percentage Depth Dose (c) dependence on SSD Photon fluence emitted by a point source of radiation varies inversely as a square of the distance from the source The actual dose rate at a point decreases with increase in distance from the source, the percent depth dose, which is a relative dose, increases with SSD Mayneord F factor

PDD - Dependence on Source-Surface Distance Dose rate in free space from a point source varies inversely as the square of the distance. (IVSL) scattering material in the beam may cause deviation from the inverse square law. PDD increases with SSD IVSL dm d SSD SSD’

Percentage Depth Dose (c) dependence on SSD F1+dm F2+dm F1+d F2+d Fig. 9.5 Plot of relative dose rate as inverse square law function of distance from a point source. Reference distance = 80 cm

d dm f1 r d dm r f2

the Mayneord F Factor ( without considering changes in scattering ) dm f1 r f2 r PDD increases with SSD the Mayneord F Factor ( without considering changes in scattering )

PDD - Dependence on Source-Surface Distance PDD increases with SSD

Example The PDD for a 15×15 field size, 10-cm depth, and 80-cm SSD is 58.4-Gy (C0-60 Beam). Find the PDD for the same field size and depth for a 100-cm SSD Assuming dm=0.5-cm for (C0-60 Gamma Rays). F=1.043 P= 58.4*1.043=60.9

Percentage Depth Dose (c) dependence on SSD Under extreme conditions such as lower energy, large field (the proportion of scattered radiation is relatively greater), large depth, and large SSD, the Mayneord F factor is significant errors In general, the Mayneord F factor overestimates the increase in PDD with increase in SSD

PDD - Dependence on Source-Surface Distance PDD increases with SSD the Mayneord F Factor works reasonably well for small fields since the scattering is minimal under these conditions. However, the method can give rise to significant errors under extreme conditions such as lower energy, large field, large depth, and large SSD change.

Tissue-Air ratio The ratio of the dose ( D d ) at a given point in the phantom to the dose in free space ( D f s ) TAR depends on depth d and field size rd at the depth: (BSF) Equilibrium mass phantom d rd rd Dd D f s

Tissue-Air ratio ( a ) Effect of Distance Independent of the distance from the source The TAR represents modification of the dose at a point owing only to attenuation and scattering of the beam in the phantom compared with the dose at the same point in the miniphantom ( or equilibrium phantom ) placed in free air

Tissue-Air ratio ( b ) Variation with energy, depth, and field size For the megavoltage beams, the TAR builds up to a maximum at the d m and then decreases with depth As the field size is increased, the scattered component of the dose increases and the variation of TAR with depth becomes more complex

Tissue-Air ratio ( b ) Variation with energy, depth, and field size: BSF Backscatter factor (BSF) depends only on the beam quality and field size Above 8 MV, the scatter at the depth of Dmax becomes negligibly small and the BSF approaches its minimum value of unity

Fig. 9.8 Variation of backscatter factors with beam quality

The meaning of Backscatter factor For example, BSF for a 10x10 cm field for 60Co is 1.036 means that D max will be 3.6% higher than the dose in free space This increase in dose is the result of radiation scatter reaching the point of D max from the overlying and underlying tissues

( c ) relationship between TAR and PDD Tissue-Air ratio ( c ) relationship between TAR and PDD

Tissue-Air ratio ( c ) relationship between TAR and PDD-- Conversion of PDD from one SSD to another : The TAR method Burns’s equation:

( d ) calculation of dose in rotation therapy Tissue-Air ratio ( d ) calculation of dose in rotation therapy d=16.6

Scatter-Air Ratio(SAR) Calculating scattered dose in the medium The ratio of the scattered dose at a given point in the phantom to the dose in free space at the same point TAR(d,0): the primary component of the beam d Dd rd D f s phantom Equilibrium mass

Scatter-Air Ratio--Dose calculation in irregular fields: Clarkson’s Method Based on the principle that the scattered component of the depth dose can be calculated separately from the primary component Average tissue-air ratio Average scatter-air ratio TAR ( 0 ) = tissue-air ratio for 0 x 0 field