INDUSTRIAL HYGIENE - SAMPLING OF GASES AND VAPORS

Slides:



Advertisements
Similar presentations
EH Terminology Presented by QBE Loss Control Services.
Advertisements

INDUSTRIAL HYGIENE - SAMPLING OF GASES AND VAPORS

INDUSTRIAL HYGIENE - AIR SAMPLING
Sampling and Measurement for Volatile Organic Compounds
Instrumental Analysis
IH&S 725 Dr. Myers, C.I.H. Compliance statistics Lecture Notes.
Air Monitoring Presented By: Etech Environmental & Safety Solutions, Inc.
World Health Organization
How long will the test last?
Dr Samah Kotb Lecturer of Biochemistry 1 CLS 432 Dr. Samah Kotb Nasr El-deen Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Nasr.
Selection of measuring instruments
Quality Assurance.
1 Seventh Lecture Error Analysis Instrumentation and Product Testing.
UNIVERSITY OF HOUSTON - CLEAR LAKE Quality product (or service) as one that is free of defects and performs those functions for which it was designed.
Classification of Instruments :
Vincent J. Giblin, General President 1293 Airport Road Beaver, WV Phone: (304) Fax: (304) Air Monitoring.
Quality Assurance in the clinical laboratory
Chemometrics Method comparison
Lesson 5 What devices do we use to measure radon?.
Energy Facility Contractors Group Safety Working Group Industrial Hygiene / Industrial Safety Technical Team Dina Siegel, Los Alamos National Laboratory.
UNIVERSITY OF HOUSTON - CLEAR LAKE SPRING Known low concentrations of air contaminants are required for: -testing and validation of analytical methods;
Quality Assurance.
Validation of Analytical Method
1 Occupational Air Sampling Strategies – who, when, how…. Lecture Notes.
Air Chemistry GISAT 112. Scientific and Technical Concepts Phases of airborne matter- gases, particles Inorganic and organic chemicals Balancing chemical.
Chapter 6 - Chromatography
Quality assurance of sampling and analytical instruments
Industrial Air Quality Monitoring. Sampling Protocols Grab vs. Integrated Personal vs. Area.
The following minimum specified ranges should be considered: Drug substance or a finished (drug) product 80 to 120 % of the test concentration Content.
Chromatography Chapter 6.
Analytical considerations
Analytical chemistry MLAB 243 Level 4 Lecture time: every WED 8 -10
Personal Samplers Instructional Goal The participants will see the various methods used to collect personal samples and how that may impact them on the.
1 Saxony-Anhalt EU Twinning RO 04/IB/EN/09 State Environmental Protection Agency Wolfgang GarcheWorkshop European Standards Requirements of.
Quality Control Lecture 5
Quality Assurance How do you know your results are correct? How confident are you?
5. Quality Assurance and Calibration Quality assurance is We do to get the right answer for our purpose. Have Sufficient accuracy and precision to support.
N 2 O-Measurement Methods Folie 1 Nitrous Oxide/Laughing Gas Molar mass: 44,013 kg/kmol Gas density at 0°C and 1,013 bar: 1,9781 kg/m³ Gas density in relation.
Biochemistry Clinical practice CLS 432 Dr. Samah Kotb Lecturer of Biochemistry 2015 Introduction to Quality Control.
1 Setting Action Levels and Controlling exposure with Air Monitoring A review...
Validation Defination Establishing documentary evidence which provides a high degree of assurance that specification process will consistently produce.
Emission source sampling and monitoring Topic 6 Ms Sherina Kamal May
IH&S 725 Dr. Myers, C.I.H. Quality assurance of sampling and analytical instruments Lecture Notes.
Quality Control Internal QC External QC. -Monitors a test's method precision and analytical bias. -Preparation of quality control samples and their interpretation.
LECTURE 13 QUALITY ASSURANCE METHOD VALIDATION
Lecture 10 ANALYTICAL METHOD DEVELOPMENT AND VALIDATION IN HPLC AND GC. Lecture 10 – Chromatography, Dr. Rasha Hanafi 1© Dr. Rasha Hanafi, GUC.
Chemical Risk Assessment & Exposure Monitoring Quantitative Risk Assessment Revision December Information provided subject to the 'Conditions for.
Ambient air sampling and monitoring Topic 5 Ms. Sherina Kamal.
Chapter 1: The Nature of Analytical Chemistry
SKC Inc. The Basics of Air Sampling Step-by-Step.
Aerosol & Particulate Research Lab
SEMINAR ON PRESENTED BY BRAHMABHATT BANSARI K. M. PHARM PART DEPARTMENT OF PHARMACEUTICS AND PHARMACEUTICAL TECHNOLGY L. M. COLLEGE OF PHARMACY.
Motivation Chemiluminescence Method Heteropoly Blue Results Due to the harmful health effects of respirable crystaline silica, the National Institute of.
1 Emergency and Disaster Response to Chemical Releases Monitoring Module 4.
EQUIPMENT and METHOD VALIDATION
Misuse of Combustible Gas Meters Supervisor Training 1/11/08.
이 장 우. 1. Introduction  HPLC-MS/MS methodology achieved its preferred status -Highly selective and effectively eliminated interference -Without.
CHAPTER 6 Gases and Vapors. Learning Objectives Describe the various properties of the chemicals that are most important to the practice of industrial.
Fundamentals of Analytical Chemistry Eighth Edition.
Quality Assurance in the clinical laboratory
Presented by Harry C. Elinsky, Jr. Filtech, Inc.
Chapter 5 Quality Assurance and Calibration Methods
ANALYTICAL METHOD VALIDATION
Introduction To Medical Technology
▪Internal quality control:
Quality Assessment The goal of laboratory analysis is to provide the accurate, reliable and timeliness result Quality assurance The overall program that.
Measurements & Error Analysis
Detector Tubes.
High-Performance Liquid Chromatography
Presentation transcript:

INDUSTRIAL HYGIENE - SAMPLING OF GASES AND VAPORS UNIVERSITY OF HOUSTON - CLEAR LAKE SPRING 2015

PURPOSE Introduce techniques to evaluate EXPOSURES to gases and vapors arising in or from the workplace. Additionally: - be aware of technology for assessment of environments; - indoor and ambient air; and, - capabilities and limitations of various referenced methods.

SAMPLING GASES AND VAPORS When developing a sampling strategy, review available sampling and analytical methods for contaminants of interest. Select most suitable! e.g. OSHA, NIOSH i.e. published and validated methods. EPA methods used for lower level indoor air pollutants and toxics in ambient air.

SAMPLING METHOD Select a method that meets sampling and analytical ACCURACY and PRECISION requirements of the referenced standard in unique field conditions. Usually stipulate measurement at the PEL within +/- 25% of the “true” value at a 95% confidence level. EPA – related to ambient air.

ANALYTICAL LABORATORY Select and consult with qualified analytical laboratory, e.g. AIHA for participation in appropriate Laboratory Accreditation Programs. Labs assist with method choice that meet sensitivity and specificity criteria for environment under evaluation. Choose sampling media and strategy compatible with method and review special handling. Two key factors: knowledge of occupational environment AND overall perspective of the limitation of the chemistry of sampling/analysis parameters.

SENSITIVITY Exercise caution when using traditional workplace sampling method for measuring contaminants in indoor or ambient air because expected concentrations may be below the working range. To obtain sensitivity, NIOSH recommends exceeding recommended air volume while observing the suggested maximum flow rate. The conservative value protects against break-through (e.g. primary vs. back-up sections of sampler) under “worst-case” conditions of high %RH and/or concentrations.

GAS/VAPOR For IH purposes, a substance is a GAS if this is normal physical state at room temperature (25 degrees C) and one-atmosphere pressure. Examples: CO, Cl, Oxygen, and Nitrogen. If substance is normally a liquid at normal temperature and pressure, then the gaseous component in equilibrium with liquid state is a VAPOR. Examples: CCl4, HCOH, and Benzene.

SAMPLING PLAN Designing a sampling plan involves consideration of the following: - location of samples, - number of workers, and - duration of sampling. Also other factors: - noise, - equipment, - size, - flow rate, and - security of equipment/sample integrity.

SAMPLE TYPES Two basic types of samples used to assess employee exposure: - Integrated - Grab.

INTEGRATED SAMPLING For gases and vapors, passage of a known volume of air through an ABsorbing or ADsorbing medium to remove the desired air contaminants during defined time period. Absorb – penetration into material; physically dissolve or chemically react with sample medium. Adsorb – traps airborne chemicals onto surface. Contaminants of interest are collected and concentrated over time to obtain the average exposure levels during the entire sampling period.

GRAB SAMPLING Direct collection of an air-contaminant mixture into a device (i.e. sampling bag, syringe, or evacuated flask) over a short interval of a few seconds or minutes. Represents the atmospheric concentrations at the sampling site at a given point in time.

WHOLE AIR SAMPLING Collection of air into a sealable container (e.g. stainless steel canister or sampling bag) for subsequent analysis. Example: Figure 11.10 [insert – follows] Obtained over a short period of time as grab samples OR integrated over a longer period of time to obtain Time-Weighted Average (TWA).

INTEGRATED SAMPLING Cover the entire period of exposure required because airborne contaminant concentrations during a typical work shift vary with time and activity! Grab samples do not reflect average exposures. Most integrated sampling is done to determine the 8-hour TWA and/or STELs for comparison with OSHA PELs, ACGIH TLVs and NIOSH RELs.

Personal Pocket Sampling Pump

INTEGRATED SAMPLING CONSIDERATIONS Appropriate sample duration and flow rate chosen relative to the purpose of sampling, sensitivity of analytical method, and expected concentration of contaminant. Accurate measurement of flow rate and time which depends on precise determination of the mass of contaminant collected and volume of air sampled.

ACTIVE SAMPLING Means of collecting an airborne substance with a mechanical device such as air sampling pump to draw the air/contaminant mixture into or through the sampling device. Examples: sorbent tube, treated filter, or impinger containing a liquid media. A key element is calibration that reliably measures the pump flow rate, thus allowing for an accurate determination of air volume.

ACTIVE SAMPLING Uses mechanical device to collect airborne substance such as air sampling pump to draw the air/contaminant mixture into or through the sampling device.

AIR SAMPLING PUMPS Integrated methods require relatively constant source of suction calibrated to recommended flow rate (within +/- 5% with collection media in-line).

AIR SAMPLING WITH PUMPS Personal sampling (i.e. breathing zone) OR area sampling for potential source(s). Features – constant flow capabilities/back pressure; intrinsically safe; etc. Maintain desired flow rate over the entire sampling period with in-line sample collection device. Pressure drop; constant flow vs. constant pressure

CALIBRATION Pump flow must be calibrated with the entire sampling train assembled as used. Good IH practice requires both pre- and post-pump calibration on the same day under pressure and temperature conditions similar to those at site. Not addressed with built-in rotameters (not precision devices and do not provide a quantitative measure of the airflow).

CALIBRATION STANDARDS Two Terms: Primary – direct and measurable linear dimensions (length and diameter of cylinder). Examples: spirometers and bubble meters Secondary – flowmeters that trace calibration to primary standards and maintain accuracy with reasonable care and handling in operation. Examples: precision rotameters, wet test meters, and dry gas meters Refer to instructions from manufacturers.

SAMPLE COLLECTION MEDIA Consult published air sampling methods to determine the appropriate collection media for a specific chemical contaminant. Review methods to determine applicability relative to field conditions. Such as: VP, BP, reactivity; interferences, also humidity/temperature effects, proper measuring range; physical state of the contaminant sampled; multiple phases (i.e. particulate and vapor, etc.).

ACTIVE SAMPLING – ADVANTAGES - Select method used by compliance personnel during OSHA inspection. - Offers calibrated, measured airflow for assurance in accuracy of sample volume. - Sorbent tube samples have a secondary layer for back-up indicating break-through. - Multiple phases can be assessed by a series of samplers.

ACTIVE SAMPLING – DISADVANTAGES - Cumbersome equipment and may interfere with job tasks during shift for personnel. - Pump calibration is time consuming and requires technical training. - Pump may become somewhat less reliable at maintaining constant flow over the entire sampling period, and more frequent calibration may be necessary.

PASSIVE SAMPLERS Passive sampling is the collection of airborne gases and vapors at a rate controlled by a physical process such as diffusion through a static air layer or permeation through a membrane without the active movement of air through sampler. Operate on principle of diffusion.

Organic Vapor Monitoring Badges

PASSIVE SAMPLERS Performance characteristics evaluated in the NIOSH Validation Protocol include: - Analytical recovery - Sampling rate and capacity - Reverse diffusion - Storage stability - Analyte concentration - Exposure time - Face velocity

PASSIVE SAMPLERS Performance characteristics evaluated in the NIOSH Validation Protocol include: - Relative humidity - Interferences - Monitor orientation - Temperature - Accuracy and precision - Shelf life - Behavior in field

DIFFUSIVE SAMPLERS Rely on movement of contaminant molecules across a concentration gradient, which for steady-state conditions can be defined by: Fick’s first law of diffusion. Consist of diffusion gap between external air and a sorbing medium to collect chemicals of interest, but also to maintain the concentration as close to zero as possible at the end of the diffusion path.

DIFFUSIVE SAMPLERS Each gas/vapor sampled has a specified diffusion coefficient (D). Uptake rates can vary under various field conditions. Validation! See Equations and Units on Page 274 (Third Edition).

OSHA ISSUES Report (1998) attempted to determine sampling rate variation of specific passive sampler designs. Concept of passive sampling equated to active sampling with pump error of +/- 5%. Significant - use of sampling rate variation for passive sampler along with the analytical error component allowed calculation of the overall Sampling and Analytical Error (SAE). SAE must be used by OSHA inspectors along with sample results to determine if PEL exceeded. Therefore, passive sampling methods can be used by OSHA.

PASSIVE SAMPLERS Various types are commercially available. Some samplers designed to collect broad range of compounds, whereas others due to collection media preferentially obtain a single chemical or family of chemicals. Examples: activated charcoal sorbent – organic vapors and GC analysis; chemical treated sorbents or filter paper for HPLC analysis. Direct-reading passive samplers based on colorimetric techniques; may not be as accurate as lab analytical methods.

PASSIVE SAMPLING – ADVANTAGES - Easy to use, samples can be collected by personnel with less technical training. - Less expensive. - Less obtrusive for wearer. - For most applications, the mass of contaminant collected by passive samplers is not significantly affected by temperature or pressure.

PASSIVE SAMPLING – DISADVANTAGES - May not be OSHA/NIOSH methods referenced in order to insure reliability of data. - Sampling rate, if theoretically calculated, may not prove valid under field conditions. - Reverse diffusion may be a factor. - Environmental parameters may influence the collection efficiency of passive samplers. Examples: stagnant air; high face velocities. - Low uptake rates may not provide sensitivity required for low-level determinations, and extended sampling times (> 24 hours) may enhance effects of reverse diffusion.

GRAB SAMPLING Collected sample measures gas and vapor concentrations AT A POINT IN TIME and used to evaluate “PEAK” exposures for comparison to “Ceiling” limits. Identify unknown contaminants, evaluate contaminant sources, or measure contaminant levels from intermittent processes/other sources. Collected using syringes, canisters, or bags, etc. Instantaneous (as well as integrated) measurements of gases/vapors also may be performed using detector tubes or direct-reading instruments.

Colorimetric Tubes and Sampling Pump

Colorimetric Gas Detection Tubes

GRAB SAMPLING – ADVANTAGES - After collection, frequently analyzed immediately by GC or direct-reading instruments. - Therefore, quick decisions made in field/site about source of leaks; Confined Space Entry (CSE); PPE, etc.

GRAB SAMPLING – DISADVANTAGES - For most applications, contaminants are collected but not integrated over time. Only some devices will allow use of a metering device to collect sample(s) at or near constant flow over period of time for TWA. - For low contaminant concentrations, analytical instrument may not be sensitive for detection. - Multiple grab samples to assess full-shift exposures is time-consuming process and subject to error.

OPERATIONAL LIMITS OF SAMPLING AND ANALYSIS Inherent limitations of method: Sampler capacity Limit of Detection (LOD) Limit of Quantification (LOQ) Upper measurement limits define useful range of referenced method. These factors determine the minimum, maximum, or optimum volume of air to be sampled and also confidence that can be placed in the results. Discuss with lab before sampling!

SAMPLER CAPACITY Predetermined conservative estimate of total mass of contaminant that can be collected on medium without loss or overloading. NIOSH definition of 2/3 of experimental break- through capacity of the solid sorbent [i.e. 67% of mass of contaminant on sorbent at the break-through volume]. Break-through volume is: volume of an atmosphere containing two times the PEL for the contaminant that can be sampled at the recommended flow rate before the efficiency of the sampler degrades to 95%.

LIMIT OF DETECTION (LOD) Lowest concentration level determined to be statistically different from a blank sample. Recommended value of the LOD is the amount of analyte that will give rise to a signal that is three times the standard deviation of the signal derived from the media blank.

LIMIT OF QUANTIFICATION (LOQ) Concentration level above which quantitative results may be obtained with a certain confidence. Recommended value of the LOQ is the amount of analyte that will give rise to a signal that is ten times the standard deviation of the signal from a series of media blanks. Corresponds to a relative uncertainty in the measurement of +/- 30% at 99% confidence level.

DETECTION LIMIT (DL) Described both in terms of: - DL of the analytical procedure, and - DL of the overall procedure. OSHA reports, in general, that detection limits (DL) are defined as: amount of analyte that gives a response that is significantly different (three SD) from the background response.

UPPER MEASUREMENT LIMIT Useful limit of the analytical instrument (mg of analyte per sample). Sample above Upper Measurement Limit, then re- dilute and re-analyze. Review and discuss with analytical lab.

TARGET CONCENTRATION Preliminary estimate of airborne contaminant concentration relative to the purpose of testing. Parameter can be used to determine the minimum and maximum air volumes. Can be estimated by: using previous sampling data, use of direct-reading instruments, or by relying on the professional judgment of the IH.

CALCULATIONS Sample volumes – minimum and maximum. Working range is range of contaminant concentration that may be quantitated at a specified air volume. Lower boundary of working range is defined by a sample that has a mass of contaminant equal to the LOQ. Upper boundary is defined by sampler capacity. Refer to Page 278 (Third Edition) for formulas.

SOLID SORBENT MEDIA Adsorb onto surface; Effectiveness determined by: Trap and retain nearly all contaminants from air Amenable to desorption from sorbent Sufficient capacity to retain contaminant quantity for analysis without large media pressure drop Not cause chemical change of contaminant except by analytical method as needed Absorb contaminant of interest in presence of other substances, possibly in higher concentrations than necessary for compound.

COLLECTION EFFICIENCY OF SOLID SORBENTS Various Factors: Temperature Humidity Sampling Rate Other Contaminants Sample Break-through - 25% - Migration

DESORPTION - Solvent extraction - extract contaminants of interest from adsorbent materials. Examples: carbon disulfide; mixtures. - Thermal desorption – drive contaminant off sorbent by high temperature; entire mass of contaminant introduced directly into analytical instrument without dilution; and, measure low airborne concentrations.

DESORPTION EFFICIENCY Measure of how much analyte can be recovered from the sorbent tube; determined typically at 0.1, 0.5, 1, and 2 times target concentration based on the recommended air volume and expressed as a percentage of analyte spike on media. To be determined for each lot number of solid sorbent used for sampling and also done in the concentration range of interest.

TYPES OF SORBENT MATERIALS Inorganic Sorbents – silica gel (polar; %RH); less reactive than charcoal Elemental Carbon – charcoal types; organics; high adsorptive capacity; stable compounds; high humidity parameters Carbonized or Graphitized Sorbents – low to moderate surface area; intermediate to high volatility; stable compounds for thermal desorption

Sorbent Tubes

Sorbent Tubes

TYPES OF SORBENT MATERIALS Organic Polymers - selectivity to particular applications, and stability at high temps enables thermal desorption (i.e. Tenax – broad range of organics can be collected). Other Sorbent Materials – sorbents with PUF; sorbent combinations; sorbent/filter combinations – OVS, etc.

Sorbent Tubes

OTHER MEDIA Chemically treated filters – derivatize/desorb Liquid absorbers Gas washing bottles – e.g. impingers Fritted glass bubblers

Filter Cassettes

Filter Cassettes

OTHER MEDIA Sampling bags/evacuated containers/canisters Situations: direct-reading instruments; leaks/emergencies; peaks; highly volatile compounds Precautions – storage time; reaction; diffusion.

CALCULATIONS Total Mass of Contaminant Airborne Concentration by sample volume (mass over volume) Air Volume (flow rate x sample time) Unit Conversions – mg/M3 to/from ppm Temperature/Pressure Corrections Time-Weighted Averages (TWA) Potential Work Shift Adjustments See Formulas and other Handouts/PPTs.

CALCULATIONS Unit Conversions – mg/M3 to/from ppm Temperature/Pressure Corrections Time-Weighted Averages (TWA) Potential Work Shift Adjustments Formulas on Page 286-287 (Third Edition).

REMEMBER CALCULATIONS Range of temperature and pressure changes tolerated before corrections are applied to the volume of air sampled. All OELs and environmental exposure standards and limits are expressed at 25 degrees C and 1 atmosphere (760 mm Hg), defined as normal temperature and pressure (NTP). Therefore, corrections needed for meaningful comparisons to published OELs.