CT Quality Control for CT Scanners. Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends.

Slides:



Advertisements
Similar presentations
ARTIFACTS IN COMPUTED TOMOGRAPHY
Advertisements

IMAGE GENERATION IN CT.
Image Reconstruction.
Improved Conspicuity of Abdominal Lesions with Single-Source Dual-Energy MDCT Hadassah Hebrew University Medical Center Jerusalem, Israel Ruth Eliahou.
Quality Control in Diagnostic Radiology
Image Quality Bushong Chapter 7.
IMAGE QUALITY.
Computed Tomography Physics, Instrumentation, and Imaging
Digital Radiography.
Advanced Biomedical Imaging Dr. Azza Helal A. Prof. of Medical Physics Faculty of Medicine Alexandria University Lecture 6 Basic physical principles of.
Computed Tomography II
Seeram Chapter 13: Single Slice Spiral - Helical CT
RAD309 Patient Dose.
NC HPS Meeting 10/18-19/2001 Boone, NC Recent Advances in CT Technology and Issues of CT Dosimetry T. Yoshizumi 1,2, M. Sarder 1, R. Reiman 1,2, E. Paulson.
MULTISLICE CT.
1 Physics Testing for Performance Based Protocol version 3.0 (Now based on ACRIN 6678) QIBA Group 1C.
David E. Hintenlang, Ph.D., DABR
Digital Image Manipulation I By Professor Stelmark.
IMAGE QUALITY NOISE LINEARITY CROSS-FIELD UNIFORMITY IMAGE ARTIFACTS.
Computed Tomography III
Current Topics in Medical Physics Research Xiaoming Zheng, PhD. School of Dentistry and Health Science Chengdu, China, 2009.
The Origins of X-Rays. The X-Ray Spectrum The X-Ray Spectrum (Changes in Voltage) The characteristic lines are a result of electrons ejecting orbital.
X-Rays and CT Scans.
Computed Tomography
Basic quality control tests in medical radiography
 QC testing of screen speed should occur on acceptance and then yearly.  Evaluate first whether similar cassettes marked with the same relative speed.
Conventional and Computed Tomography
…….CT Physics - Continued V.G.WimalasenaPrincipal School of radiography.
Seeram Chapter 11: Image Quality
Basic principles Geometry and historical development
COMPUTED TOMOGRAPHY - I RAD 365 CT - Scan
Quality Control Rad T 110.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) May 6, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) December 23, 2009 CT Cross-platform Sizing of Phantom Lesions Quantitative Imaging Biomarker Alliance.
SPECIAL IMAGING/ADVANCED IMAGING INTRODUCTION TO DIGITAL RADIOGRAPHY
RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY
LASER AND ADVANCES IN METROLOGY
Seeram Chapter 7: Image Reconstruction
Seeram Chapter #3: Digital Imaging
George David Associate Professor of Radiology Medical College of Georgia.
Computed Tomography Q & A
PRINCIPLES OF CT Dr Mohamed El Safwany, MD. Intended learning outcome The student should learn at the end of this lecture principles of CT.
CT Instrumentation and X-ray system
Factors affecting CT image RAD
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) September 2, 2009 Interclinical Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
IAEA International Atomic Energy Agency Optimization of Protection in Computed Tomography (CT)-What can radiographers do? IAEA Regional Training Course.
QIBA CT Volumetrics - Cross-Platform Study (Group 1C) March 18, 2009 Interclinic Comparison of CT Volumetry Quantitative Imaging Biomarker Alliance.
Radiographic Equipment
COMPUTED TOMOGRAPHY - I RAD 365 CT - Scan
The production of X-Rays X-Rays are produced whenever charged particles are slowed down or stopped. X-rays are generated via interactions of the accelerated.
CS TC 22 CT Basics CT Principle Preprocessing. 2 CT Basics CT principle preprocessing CS TC 22 Blockdiagram image processor.
Image Display. But first a review Remember the 3 main steps 1. Data Acquisition 2. Image Reconstruction 3. Image Display.
CT QUALITY MANAGEMENT. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Part No...., Module No....Lesson No
Quality Assurance.
Modulation Transfer Function (MTF)
Assumptions: Cockcroft-Walton photomultiplier bases are the same for all ECAL sections Digital to analog converters are installed on the distribution boards.
1 Physics Testing for Performance Based Protocol version 2.1 QIBA Group 1C.
Quality Control Bushong Chapter 10.
What are the dose quantities in CT ? IAEA/RCA Kampala Kampala.
IAEA International Atomic Energy Agency RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY Part 19.04: Optimization of protection in Mammography.
Computed Tomography Computed Tomography is the most significant development in radiology in the past 40 years. MRI and Ultrasound are also significant.
Computed tomography. Formation of a CT image Data acquisitionImage reconstruction Image display, manipulation Storage, communication And recording.
IMAGE QUALITY. SPATIAL RESOLUTION CONTRAST RESOLUTION NOISE IMAGE ARTIFACTS RADIATION DOSE.
Single Slice Spiral - Helical CT
CT Multi-Slice CT.
Physics Testing for ACRIN 6678 Protocol version 3.0 QIBA Group 1C
Basic principles Geometry and historical development
Computed Tomography (C.T)
Presentation transcript:

CT Quality Control for CT Scanners

Quality Control in CT A good idea? Yes Required for accreditation? Sometimes Improves image quality? Sometimes Depends on starting point Reduces downtime? Sometimes Depends on starting point Possible recognition of problems before clinical manifestations total failure may reduce unscheduled downtime

High Contrast Resolution Baseline established when scanner in good operating condition Subsequent test compared to baseline

CT Quality Control Standard protocol required same kVp, mAs, scan diameter, slice width, algorithm, focal spot, filter, etc When performed At regular intervals After major service tube change

CT # of Water What do we check? Average (ideally 0 by definition) Standard Deviation (ideally 0) Flatness ROI’s at various locations in phantom Results depend upon protocol kVp, mA, scan time slice width phantom size ROI position Problem causes Mis-calibration Noise / drift in detector system

CT # of Water Can repeat for various Slice thicknesses Reconstruction algorithms Centering positions Phantom diameters Fields of view

CT Number Accuracy Phantom with several materials of known CT #’s Do ROI’s at each and check CT # accuracy

High Contrast (or Limiting) Resolution Phantom rows of equally spaced high contrast objects holes get smaller from row to row Bars of decreasing size & separation Determine smallest row of holes where all holes can be clearly seen smaller holes = better resolution Visual scoring

Low Contrast Resolution Phantom holes get smaller from column to column Holes have less contrast from row to row Visual scoring

Low Contrast Resolution Noise limited less noise = more rows visualized Standardize mAs more mAs  less noise  better low contrast resolution reconstruction algorithm / filter smoothing reduces noise improves low contrast resolution

Low Contrast Resolution Possible sources of failure: anything that can increase noise decreased dose to detectors reduced tube output electronic noise detectors amplifiers A - D convertor

Distance-Measuring Accuracy Phantom object with precisely known object distances Use distance-measuring software Check vertical, horizontal & diagonal Check monitor & hard copy

Monitor Performance / Hard Copy Output Standard gray scale image needed computer generated (SMPTE pattern) Check monitor & hard copy 5% patches Contrast patches Resolution bars Hard copy Measure optical density of steps with densitometer

Slice Localization / Thickness Accuracy Phantom with slanted wire Measure length Regularly spaced (in z direction) objects count Align phantom slice with laser Make single scan with known slice thickness

Slice Localization / Thickness Accuracy

Table Indexing X-ray cassette on table Load table with weight Select multiple scans small beam width large mm table increment

Table Indexing Image should show series of exposed bands Measure distance between bands Possible causes of failure slippage of table drive mechanism table mis-calibration

Table Backlash Does table return to same position from both directions? Test Apply masking tape to table Load table with weight Note numerical position Mark tape at starting position using laser position Move table away from starting position & then back to same numerical location Check laser vs. mark on tape

Laser Accuracy Align phantom on table matching outer laser to external marks on phantom “Zero” table Scan phantom “0” z-location should show correct phantom position on image Repeat for internal laser

CT Noise Characteristics Water phantom Multiple scans changing only mAs Measure standard deviation of CT #’s using identical ROI’s Noise proportional to standard deviation

CT Noise Characteristics Excessive noise can be caused by detector problems electronic noise in detector amplifier circuits reduced output per mAs