Thermal Imaging Comparison of Longitudinal versus Torsional Phacoemulsification in INFINITI Young Keun Han, M.D., Kevin M. Miller, M.D. Department of Ophthalmology,

Slides:



Advertisements
Similar presentations
A prospective, randomized clinical trial
Advertisements

INTRODUCTION. Seminar on IMPROVEMENT OF THERMAL EFFICIENCY BY RECOVERY OF HEAT FROM IC ENGINE EXHAUST.
TESTUPLOAD. TORSIONAL PHACOEMULSIFICATION In January 2006 Alcon Surgical incorporated Ozil torsional into the Infiniti Vision System. Unlike the conventional.
A Deperturbation Method to Aid in the Interpretation of Infrared Isotopic Spectra G. Garcia and C. M. L. Rittby Texas Christian University Fort Worth,
Electromagnetic Waves
OHM Comparative quantification of ingress of trypan blue into anterior chamber following microcoaxial phacoemulsification with torsional or longitudinal.
David Allen Sunderland UK Financial disclosure A & E The author has had travel and lodging costs paid by as well as occasional honoraria from Alcon Surgical.
Mitra Nejad, BS, 1 and Kevin Miller, MD 1,2 1 David Geffen School of Medicine at UCLA, Los Angeles, CA 2 Jules Stein Eye Institute and Department of Ophthalmology,
Comparison of surgically induced astigmatism after phacoemulsification trough 3.2, 2.2 and 1.8 clear corneal incision. Luis Izquierdo Jr MD. PhD. Maria.
Mars’ North and South Polar Hood Clouds Jennifer L. Benson Jet Propulsion Laboratory, California Institute of Technology July 22, 2010 Copyright 2010 California.
MESF593 Finite Element Methods HW #4 11:00pm, 10 May, 2010.
Non-Invasive Keratograph Tear Break-Up Time versus Invasive Fluorescein-Associated Tear Stability Test: Experimental Validation Study Nizar S Abdelfattah,
Quantification of hydroxyl radical during phacoemulsification. Department of Ophthalmology, Dokkyo Medical University (JAPAN) Norihito Gotoh MD PhD, Hiroyuki.
Microcoaxial phaco using 1
So-Hyang Chung, MD, PhD, Choun-Ki Joo, MD, PhD Department of Ophthalmology and Visual Science, College of Medicine, The Catholic University of Korea, Seoul,
Hyun Seung Kim, M.D. Department of Ophthalmology, St. Mary’s hospital, The Catholic University of Korea Changes in Astigmatism After Clear Corneal Temporal.
Transparency of Transition from 2.75 mm to 1.8 mm Microincision Surgery Jay McDonald II, MD Adjunct Clinical Professor University of Arkansas School of.
Introduction A lightweight film known as Kapton with 12 piezoelectric transducers attached was subjected to a frequency range of 1 kHz to 15 MHz of current.
Kendall R.B. Dobbins, MD Geisinger Medical Center
Evaluation of Corneal Parameters and Spherical Aberration After DSAEK Measured with Pentacam System Orkun Muftuoglu, Pawan Prasher, R. Wayne Bowman, Steven.
Clinical Observations of the Single-Use Irrigation/Aspiration Tip
Phacoemulsification in eyes with previous anterior chamber phakic IOL surgery Walton Nosé, MD, PhD 1,2 Adriana dos Santos Forseto, MD 1 Mariana Ávila,
Free Radical Generation During Phacoemulsification Using Different Consoles Professor Steven D. Aust, PhD, Thomas Hebdon, Jordan Humbert, Scott Terry,
Energy and Change CHAPTER the BIG idea CHAPTER OUTLINE Waves and heat flow transfer energy. Change occurs as energy is transferred. 1.1 Radiation transfers.
New Phaco Technology Mark Packer, MD, FACS Clinical Associate Professor of Ophthalmology Oregon Health & Sciences University Drs. Fine, Hoffman & Packer,
Sleeping Bag Comparison Which sleeping bag will keep you warmer? By: Kyle Layton and Ben Salazar.
2-D Heat Transfer Model of A Horizontal U-Tube M. S. Islam 1, A. Fujimoto 2, A. Saida 2 and T. Fukuhara 2 2-D Heat Transfer Model of A Horizontal U-Tube.
Astigmatism management with toric intraocular lenses in cataract patients Adriano Guarnieri 1-2, Luis W. Lu 3-4, Alfonso Arias- Puente INCIVI, Madrid,
BY KEVIN STUBBS AND BRAD IVEY Peruvian Water Heater.
An integro-partial differential equation This model comes courtesy of Daniel Smith and Ali Shajii of MKS Instruments, Wilmington, Massachusetts, USA.
Energy in the Atmosphere Heat Transfer Global Winds Local Winds
Jorge L. Alio MD PhD Ma. Cecilia C. Agdeppa MD VISSUM-Instituto Oftalmologico de Alicante ASCRS Convention 2010 Boston Efficiency of MicroIncision Cataract.
Mitsui Memorial Hospital Takayuki Akahoshi, MD The author has no financial interest in the products introduced in this presentation.
Daniele Tognetto MD, Pia Leon MD, Paolo Cecchini MD, Silvia Rinaldi MD, Giuseppe Ravalico MD The Authors have no proprietary interest in any products.
Analysis of Clear Corneal Incision Architecture with Anterior Segment Spectral-Domain OCT Theodore Leng, MD, Jianhua Wang, MD, PhD, Sonia H. Yoo, MD, Brandon.
GLOBAL FIT ANALYSIS OF THE FOUR LOWEST VIBRATIONAL STATES OF ETHANE: THE 12  9 BAND L. Borvayeh and N. Moazzen-Ahmadi Department of Physics and Astronomy.
Small Wound Leaks are Associated with Myopic Surprises in Cataract Surgery John A. Hovanesian, M.D. Clinical Instructor, UCLA Jules Stein Eye Institute.
The authors have no financial interest in the subject matter of this e-poster M. K. Kummelil, S. Nagappa, A. Shetty, A. Braganza Cataract and Refractive.
1 Non-contact Specular Microscopy for Evaluation of Corneal Endothelium in Early Fuchs’ Endothelial Corneal Dystrophy Jianyan Huang 1, MD, PhD; Tudor Tepelus.
Thermal Study of Longitudinal and Torsional Ultrasound Phacoemulsification : Tracking the Temperature of Corneal Surface, Incision and Handpiece Bokkwan.
Surgeon Generated Tip Travel Comparing Longitudinal vs. Torsional Ultrasound with Interjected Longitudinal Segments James A. Davison, MD FACS Wolfe Eye.
The Effect of Corneal Anterior Surface Eccentricity on Astigmatism after Cataract Surgery Choul Yong Park MD 1 Sung Jun Lee MD 1 Prabjot Channa MD 2 Roy.
COMPARISON OF OCULAR RESULTS OF MECHANICAL CHOPER VS ULTRACHOPER FERNANDO AGUILERA MD. INSTITUTO DE OJOS, MEXICO NO FINANCIAL INTEREST.
Wound Architecture and Wound Healing after Torsional and Longitudinal Phaco in a Rabbit Model Carolina Eyecare Physicians, LLC Research Assistant Professor.
No author has any financial or proprietary interest in any materials or methods mentioned Seung Hyun Kim M.D. ; Tae Hoon Oh M.D. Department of Ophthalmology.
Corneal Wound Architecture and Integrity after OZil and Mixed Phacoemulsification: Evaluation of Standard and Microincisional Coaxial Techniques Bokkwan.
THERMOGRAPHY Presented By: Nagaraj S Patil. Contents Introduction What it is? Where it is used? What makes it useful? Principle Application Advantages.
In this presentation you will: explore the impact of the Sun’s energy Impact of the Sun’s Energy Next >
THE OUTCOMES OF MICS WITH CRUISE CONTROL SYSTEM VS MICS WITH WHITESTAR ICE AND CASE SETTINGS IN HARD CATARACTS HELVACIOGLU Firat, MD, SENCAN Sadik, MD,
Comparing Factors Affecting Surgically Induced Astigmatism
Kinetic and Thermal Energy. Energy ***ENERGY DURING PHASES CHANGES – Energy must be lost to the environment or gained from the environment in order for.
Corneal Endothelial Cell Loss Results in a Comparison of Longitudinal vs. Torsional with Vacuum Demand Interjected Longitudinal (IP) Phacoemulsification.
Torsional Phaco with a straight needle and “Spade” tip A bench test and clinical examination comparing it with the standard bent “Kelman” needle… Nigel.
Liquifaction Method and Extent of Posterior Capsule Opacification: Two-Year Follow-up Marie Kalfertova, Mariya Burova, Pavel Rozsival, Nada Jiraskova Nada.
Assessment of Anterior Segment Measurements with Swept Source Anterior Segment OCT Before and After Trabectome Surgery Handan Akil, Mayss Al-Sheikh, SriniVas.
Comparison of two differents ultrasound mode in BMICS technique : Pulse and Continuous US Danielle DEIDIER M.D. Clinique Saint Vincent Toulon - France.
Roughness and Electrical Resistivity of Thin Films Spencer Twining, Marion Titze, Ozgur Yavuzcetin University of Wisconsin – Whitewater, Department of.
Evidence of anisotropy of small scale turbulence in the laboratory model of an atmospheric cloud P.M. Korczyk, T.A. Kowalewski, S. P. Malinowski IPPT PAN,
Evidence of anisotropy of small scale turbulence in the laboratory model of an atmospheric cloud P.M. Korczyk, T.A. Kowalewski, S. P. Malinowski IPPT PAN,
This study processes the optimization of heat extraction under the varied pressure and flow rate. Based on the validated model, two kinds of test tube.
Ethylene Furnace High Temperature Cameras
超臨界CO2在增強型地熱系統儲集層中取熱之研究-子計畫三 CO2在增強型地熱系統取熱模型之建構及效能分析
Prospective Study Comparing Outcomes of Torsional versus Traditional Phacoemulsification Systems on Dense Cataracts Bonnie An Henderson MD, Kelly J Grimes.
Ethylene Furnace High Temperature Cameras
Sherleen Chen, M.D. and Roberto Pineda II, M.D.
New 3-Dimensional Scheimpflug Analysis of Cataract Density
성모병원 안센터 CHANGES IN ASTIGMATISM RELATIVE TO IOL HAPTIC INSERTION AXIS IN WITH-THE-RULE AND AGAINST-THE-RULE ASTIGMATISM PATIENTS Hyun Seung Kim, M.D.
Authors have no financial interests
A Simple and Easy Procedure
Young Jeung Park, M.D. Ph.D. Won Suk Choi, M.D.
Presentation transcript:

Thermal Imaging Comparison of Longitudinal versus Torsional Phacoemulsification in INFINITI Young Keun Han, M.D., Kevin M. Miller, M.D. Department of Ophthalmology, David Geffen School of Medicine at UCLA and the Jules Stein Eye Institute, Los Angeles, California, USA Authors have no financial interest

Introduction   Heat production during phacoemulsification - Friction between vibrating needle and silicone sleeve   Factor leading to thermal damage - Constriction of the silicone sleeve - Occlusion of aspiration - Prolonged phaco time   Complications as a result of thermal damage - Delayed wound healing - Fistula formation - Damage to the corneal stroma and endothelium - Inability to close the incision - Increased surgically induced astigmatism Olson MD, Miller KM JCRS 2005;31(8):1640-7

Introduction  In torsional phacoemulsification, small-angle rotational movement of the shaft is translated into a nearly horizontal stroke at the tip of the needle. Longitudinal phacoemulsification motion of tip’s end = motion of tip’s shaft Torsional phacoemulsification motion of tip’s end > motion of tip’s shaft  For equivalent longitudinal and horizontal strokes at the tip, torsional phacoemulsification should generate less frictional heat at the incision because of reduced movement at that location.  The purpose of this study was to compare heat generated at the incision in longitudinal versus torsional phacoemulsification.

Materials and Methods   A straight 30  non-ABS MicroTip on a INFINITI ultrasound handpiece and a 45  Kelman Mini-Flared ABS tip on a OZil Torsional handpiece (all from Alcon Laboratories, Inc., Fort Worth, Texas) were each capped with a BSS-filled silicone test chambers (volume 0.85 cc).   25.3-gram weights were suspended using elastomer bands from the silicone tube to simulate tight corneal incisions.   The aspiration line was clamped using needle nose pliers to simulate complete occlusion of the tip.

Materials and Methods   The outside of the chamber was imaged using an FLIR model P60 ThermaCAM TM (FLIR Systems Inc., North Billerica, Massachusetts).   Camera calibration was verified by measuring warm and cold water baths using the infrared camera and comparing against a calibrated thermocouple at temperatures across the experimental range.

Materials and Methods   The thermal camera captured still images of the probes before power application (at 0 seconds), and then 10, 30, 60, and 120 seconds after power application. It determined the highest surface temperature inside a circle drawn around the point of contact of the sleeve and elastomer band. These thermal images were obtained with the handpiece operating in longitudinal mode at 100% power.

Materials and Methods   We performed 3 different types of experiments to simulate 3 different types of “equivalent” power applications. - Experiment 1 Same displayed power on all 3 instrument consoles 40, 60, 70, 100% - Experiment 2 Same stroke lengths at the cutting edge 1.55, 2.40, 2.80, 3.45 mils - Experiment 3 - Fairest comparison Same applied energy (energy = frequency X stroke length) Longitudinal: 37 kHz, torsional: 32 kHz 57.0, 89.0, 104, 129 KHz*mils

Materials and Methods   In Experiment 2, we compared heat production under conditions of identical stroke at the distal end of the tip. To measure the stroke, sleeveless phaco tips were placed inside a fluid tank with a viewing window and imaged using a video microscope (Meiji Techno, Japan).   When ultrasound power was applied, the distance of motion was measured using a VIA-150 video image marker-measurement system (Boeckeler, Arizona, USA). → Video clip

Results Experiment 1: Time vs. Temperature comparing torsional (red) and longitudinal (blue) modalities where the similarity is the displayed power on instrument console: 40, 60, 70, and 100%

Results Experiment 2: Time vs. Temperature comparing torsional (red) and longitudinal (blue) modalities where the similarity is the stroke length as measured at the tip’s cutting edge: 1.55, 2.40, 2.80, and 3.45 mils.

Results Experiment 3: Time vs. Temperature comparing torsional (red) and longitudinal (blue) modalities where the similarity is the calculated applied energy (frequency * stroke length): 57.0, 89.0, 104, and 129 kHz*mils.

Conclusion   These three approaches comparing the thermal differences of longitudinal and torsional phacoemulsification all suggest that the torsional mode is preferable. This may be especially beneficial for surgeons who are less experienced, working with harder nuclei, having longer case times, or utilizing elevated power levels.   In summary this 3-pronged study, where temperatures were measured to determine heat generation behavior, found that torsional phacoemulsification produced less heat than longitudinal phacoemulsification in all comparative test approaches.