Technology transfer from HEP computing to the medical field

Slides:



Advertisements
Similar presentations
F. Foppiano3, S. Guatelli2, J. Moscicki1, M.G. Pia2 CERN1 INFN Genova2
Advertisements

A selection of Geant4 medical physics applications
Low Energy Electromagnetic Physics PART II
The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
The application of GEANT4 simulation code to the verification of a brachytherapy calibration procedure F. Foppiano 1, S. Garelli 1, P. Moresco 1, G. Paoli.
Maria Grazia Pia, INFN Genova 1 Part IV Geant4 results.
Maria Grazia Pia, INFN Genova Test & Analysis Project Maria Grazia Pia, INFN Genova on behalf of the T&A team
1 COMPARISON BETWEEN PLATO ISODOSE DISTRIBUTION OF A 192 IR SOURCE AND THOSE SIMULATED WITH GEANT4 TOOLKIT F. Foppiano 1, S. Agostinelli 1, S. Garelli.
Monte Carlo Based Implementation of an Energy Modulation System for Proton Therapy G.A.P. Cirrone Qualified Medical Physicist PhD Laboratori Nazionali.
Giorgio Russo National Research Council, Institute of Bioimaging and Molecular Imaging (IBFM) Fondazione Istituto San Raffaele G. Giglio di Cefalù Istituto.
F. Foppiano, M.G. Pia, M. Piergentili Medical Linac IEEE NSS, October 2004, Rome, Italy
Energy deposition and neutron background studies for a low energy proton therapy facility Roxana Rata*, Roger Barlow* * International Institute for Accelerator.
Low Energy Electromagnetic Physics
Monte Carlo simulation for radiotherapy in a distributed computing environment S. Chauvie 2,3, S. Guatelli 2, A. Mantero 2, J. Moscicki 1, M.G. Pia 2 CERN.
Other GEANT4 capabilities Event biasing Parameterisation (fast simulation) Persistency Parallelisation and integration in a distributed computing environment.
A general purpose dosimetric system for brachytherapy
Maria Grazia Pia, INFN Genova Distributed Processing, Monte Carlo and CT interface for Medical Treatment Plans F. Foppiano 3, S. Guatelli 2, J. Moscicki.
Maria Grazia Pia, INFN Genova CERN, 26 July 2004 Background of the Project.
1 M.G. Pia et al. The application of GEANT4 simulation code for brachytherapy treatment Maria Grazia Pia INFN Genova, Italy and CERN/IT
Maria Grazia Pia, INFN Genova Low Energy Electromagnetic Physics Maria Grazia Pia INFN Genova
The brachytherapy advanced example Susanna Guatelli (CERN/INFN)
Maria Grazia Pia Systematic validation of Geant4 electromagnetic and hadronic models against proton data Systematic validation of Geant4 electromagnetic.
Budker Inst. of Physics IHEP Protvino MEPHI Moscow Pittsburg University.
Maria Grazia Pia, INFN Genova - Como 2001 From HEP computing to bio-medical research and vice versa Simulation software: applications and results in the.
Dose Distribution and Scatter Analysis
Particle Physics Software aids Medicine Accurate geometry and material modeling in the fight against cancer Precise physics models for radiation interactions.
Научно-практический центр протонной лучевой терапии и радиохирургии (Москва-Дубна) A SYSTEM FOR MEASUREMENT OF A THERAPEUTIC PROTON BEAM DOSE DISTRIBUTION.
Test of the proposed method Introduction CCD Controller CCD Illuminator gel Filter 585nm Assembling the phantom before its irradiation. The phantom, ready.
Maria Grazia Pia, INFN Genova Particle Physics Software and the Fight against Cancer Maria Grazia Pia INFN Genova Seminar at DESY 9 February 2004
M.G. Pia et al. Brachytherapy at IST Results from an atypical Comparison Project Stefano Agostinelli 1,2, Franca Foppiano 1, Stefania Garelli 1, Matteo.
F. Foppiano, B. Mascialino, M. G. Pia, M. Piergentili Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy Monte Carlo 2005 Topical.
Maria Grazia Pia, INFN Genova Low Energy Electromagnetic Physics Maria Grazia Pia INFN Genova on behalf of the Low Energy Electromagnetic.
OOAD… LowE Electrons From HEP computing to medical research and vice versa Bidirectional From HEP computing to medical research and vice versa Bidirectional.
A General Purpose Brachytherapy Software Simulation + Analysis (isodose calculation) 2/10/2002 Geant4 Workshop CERN Susanna Guatelli Univ. and INFN Genova.
S. Guatelli, M.G Pia, INFN Genova G. Cosmo, S. Guatelli, M.G Pia Salamanca, July 2002
Medical Accelerator F. Foppiano, M.G. Pia, M. Piergentili
Geant4 DICOM Interface Susanna Guatelli INFN Genova, Italy 8 th March 2004 How to model human anatomy in a Geant4 application.
Technological Transfer from HEP to Medical Physics How precise Brachytherapy MonteCarlo simulations can be applied in Clinics Reality Problem: How to achieve.
S. Guatelli, A. Mantero, J. Moscicki, M. G. Pia Geant4 medical simulations in a distributed computing environment 4th Workshop on Geant4 Bio-medical Developments.
Precision Validation of Geant4 Electromagnetic Physics Geant4 DNA Project Meeting 26 July 2004, CERN Michela.
TPS & Simulations within PARTNER D. Bertrand, D. Prieels Valencia, SPAIN 19 JUNE 2009.
IRCC & Mauriziano Hospital & INFN & S Croce e Carle Hospital
Geant4 Training 2006 Short Course Katsuya Amako (KEK) Gabriele Cosmo (CERN) Susanna Guatelli (INFN Genova) Aatos Heikkinen (Helsinki Institute of Physics)
F. Foppiano, M.G. Pia, M. Piergentili
N 0 primary photons generated N d primary photons detected Determination of the photon mass attenuation coefficients Check on ParentID( ) Energy value.
Maria Grazia Pia, INFN Genova Low Energy Electromagnetic Physics Maria Grazia Pia, INFN Genova on behalf of the LowE WG
P. Rodrigues, A. Trindade, L.Peralta, J. Varela GEANT4 Medical Applications at LIP GEANT4 Workshop, September – 4 October LIP – Lisbon.
Susanna Guatelli Geant4 in a Distributed Computing Environment S. Guatelli 1, P. Mendez Lorenzo 2, J. Moscicki 2, M.G. Pia 1 1. INFN Genova, Italy, 2.
Geant4 Training 2004 Short Course Katsuya Amako (KEK) Gabriele Cosmo (CERN) Giuseppe Daquino (CERN) Susanna Guatelli (INFN Genova) Aatos Heikkinen (Helsinki.
Maria Grazia Pia, INFN Genova and CERN1 Geant4 highlights of relevance for medical physics applications Maria Grazia Pia INFN Genova and CERN.
AIDA Workshop, 3/6/2002S. Guatelli Overview of AIDA in Geant4 bio-medical applications A collection of contributions from various user groups AIDA Workshop.
Implementation of a New Monte Carlo Simulation Tool for the Development of a Proton Therapy Beam Line and Verification of the related Dose Distributions.
MCS overview in radiation therapy
F. Foppiano3, S. Guatelli2, J. Moscicki1, M.G. Pia2 CERN1 INFN Genova2
Simulation in a Distributed Computing Environment
Geant4 REMSIM application
A Brachytherapy Treatment Planning Software Based on Monte Carlo Simulations and Artificial Neural Network Algorithm Amir Moghadam.
F. Foppiano, S. Guatelli, B. Mascialino, M. G. Pia, M. Piergentili
Geant4 for Brachytherapy Simulation
Other GEANT4 capabilities
Introductory Course PTB, Braunschweig, June 2009
Simulation in a Distributed Computing Environment
Short Course Siena, 5-6 October 2006
The Hadrontherapy Geant4 advanced example
Introductory Course ORNL, May 2008
Geant4 at IST Applications in Brachytherapy
Short Course IEEE NSS/MIC 2003 Katsuya Amako (KEK) Makoto Asai (SLAC)
P. Rodrigues, A. Trindade, L.Peralta, J. Varela
G. A. P. Cirrone1, G. Cuttone1, F. Di Rosa1, S. Guatelli1, A
The Geant4 Hadrontherapy Advanced Example
Presentation transcript:

Technology transfer from HEP computing to the medical field http://www.ge.infn.it/geant4/talks F. Foppiano3, S. Guatelli2, J. Moscicki1, M.G. Pia2, M. Piergentili2 CERN1 INFN Genova2 National Institute for Cancer Research, IST Genova3 Topical Seminar on Innovative Radiation Detectors Siena, 23-26 May 2004 Including contributions from: S. Agostinelli, S. Garelli (IST Genova) L. Archambault, L. Beaulieu, J.-F. Carrier, V.-H. Tremblay (Univ. Laval) M.C. Lopes, L. Peralta, P. Rodrigues, A. Trindade (LIP Lisbon) G. Ghiso (S. Paolo Hospital, Savona)

A dosimetric system for brachytherapy derived from HEP computing Technology transfer A real life case A dosimetric system for brachytherapy derived from HEP computing (but all the developments and applications presented in this talk are general) Activity initiated at IST Genova, Natl. Inst. for Cancer Research (F. Foppiano et al.) hosted at San Martino Hospital in Genova (the largest hospital in Europe) Collaboration with San Paolo Hospital, Savona (G. Ghiso et al.) a small hospital in a small town

The goal of radiotherapy Delivering the required therapeutic dose to the tumor area with high precision, while preserving the surrounding healthy tissue Accurate dosimetry is at the basis of radiotherapy treatment planning Dosimetry system precision accurate model of the real configuration (from CT) speed adequate for clinical use user-friendly interface for hospital usage Calculate the dose released to the patient by the radiotherapy system

The reality Open issues Precision Cost Treatment planning is performed by means of commercial software The software calculates the dose distribution delivered to the patient in a given source configuration Open issues Precision Cost Commercial systems are based on approximated analytical methods, because of speed constraints Approximation in geometry modeling Approximation in material modeling Each treatment planning software is specific to one technique and one type of source Treatment planning software is expensive

Commercial factors Commercial treatment planning systems are governed by commercial rules (as any other commercial product...) i.e., they are produced and marketed by a company only if the investment for development is profitable Treatment planning systems for hadrontherapy are quite primitive not commercially convenient so far No commercial treatment planning systems are available for non-conventional radiotherapy techniques such as hadrontherapy or for niche applications such as superficial brachytherapy

Monte Carlo methods in radiotherapy Monte Carlo methods have been explored for years as a tool for precise dosimetry, in alternative to analytical methods de facto, Monte Carlo simulation is not used in clinical practice (only side studies) The limiting factor is the speed Other limitations: reliable? for “software specialists only”, not user-friendly for general practice requires ad hoc modeling

Comparison with commercial treatment planning systems Central-Axis depth dose Comparison with commercial treatment planning systems M. C. Lopes IPOFG-CROC Coimbra Oncological Regional Center L. Peralta, P. Rodrigues, A. Trindade LIP - Lisbon CT-simulation with a Rando phantom Experimental data with TLD LiF dosimeter Profile curves at 9.8 cm depth PLATO overestimates the dose at ~ 5% level CT images used to define the geometry: a thorax slice from a Rando anthropomorphic phantom

In some tumours sites (ex: larynx T2/T3-stage) Beam plane Skull bone A more complex set-up Tumor M. C. Lopes1, L. Peralta2, P. Rodrigues2, A. Trindade2 1 IPOFG-CROC Coimbra Oncological Regional Center - 2 LIP - Lisbon Air Bone Head and neck with two opposed beams for a 5x5 and 10x10 field size An off-axis depth dose taken at one of the slices near the isocenter PLATO fails on the air cavities and bone structures and cannot predict accurately the dose to tissue that is surrounded by air Deviations are up to 25-30% In some tumours sites (ex: larynx T2/T3-stage) a 5% underdosage will decrease local tumour control probability from ~75% to ~50%

The challenge

dosimetric system precise general purpose realistic geometry Develop a general purpose realistic geometry and material modeling with the capability of interface to CT images with a user-friendly interface low cost at adequate speed for clinical usage performing at

Accurate model of the real experimental set-up Requirements Calculation of 3-D dose distribution in tissue Determination of isodose curves Based on Monte Carlo methods Accurate description of physics interactions Experimental validation of physics involved Precision Accurate model of the real experimental set-up Realistic description of geometry and tissue Possibility to interface to CT images Simple user interface + Graphic visualisation Elaboration of dose distributions and isodoses Easy configuration for hospital usage Parallelisation Access to distributed computing resources Speed Transparent Open to extension and new functionality Publicly accessible Other requirements

Precision Based on Monte Carlo methods Accurate description of physics interactions Extension of electromagnetic interactions down to low energies (< 1 keV) Experimental validation of physics involved Microscopic validation of the physics models Comparison with experimental data specific to the brachytherapic practice

Run, Event and Track management PDG-compliant Particle management Geometry and Materials Tracking Detector response User Interface Visualisation Persistency Physics Processes Code and documentation publicly distributed from web 1st production release: end 1998 2 new releases/year since then Developed and maintained by an international collaboration of physicists and computer scientists

Based on EPDL97, EEDL and EADL evaluated data libraries Fe lines GaAs lines Atomic relaxation Fluorescence Auger effect shell effects e, down to 250 eV EGS4, ITS to 1 keV Geant3 to 10 keV Based on EPDL97, EEDL and EADL evaluated data libraries Based on Penelope analytical models Hadron and ion models based on Ziegler and ICRU data and parameterisations Barkas effect (charge dependence) models for negative hadrons antiprotons protons ions Bragg peak

Validation Microscopic validation: Dosimetric validation: verification of Geant4 physics Dosimetric validation: in the experimental context

Microscopic validation 2N-L=13.1 – =20 - p=0.87 NIST Geant4-LowE Geant4-Standard Photon attenuation coefficient Al 2N-S=23.2 – =15 - p=0.08 Microscopic validation Al NIST Geant4-LowE Geant4-Standard Stopping power many more validation results available! proton straggling ions e-, Sandia database

Dosimetric validation Comparison to manufacturer data, protocol data, original experimental data Distance along Z (mm) Simulation Nucletron Data F. Foppiano et al., IST Genova experimental mesurements G. Ghiso, S. Guatelli S. Paolo Hospital Savona Ir-192 I-125

General purpose system For any brachytherapy technique Object Oriented technology Software system designed in terms of Abstract Interfaces For any source type Abstract Factory design pattern Source spectrum and geometry transparently interchangeable

Flexibility of modeling Configuration of any brachytherapy technique any source type through an Abstract Factory to define geometry, primary spectrum Abstract Factory CT DICOM interface through Geant4 parameterised volumes parameterisation function: material Phantom various materials water, soft tissue, bone, muscle etc. General purpose software system for brachytherapy No commercial general software exists!

of the experimental set-up Realistic model of the experimental set-up Radioactive source Spectrum (192Ir, 125I) Geometry Patient Phantom with realistic material model Possibility to interface the system to CT images

Modeling the source geometry Precise geometry and material model of any type of source Iodium core Air Titanium capsule tip Titanium tube Iodium core I-125 source for interstitial brachytherapy Iodium core: Inner radius :0 Outer radius: 0.30mm Half length:1.75mm Titanium tube: Outer radius:0.40mm Half length:1.84mm Air: Outer radius:0.35mm half length:1.84mm Titanium capsule tip: Box Side :0.80mm Ir-192 source + applicator for superficial brachytherapy

Effects of source anisotropy Plato-BPS treatment planning algorithm makes some crude approximation ( dependence, no radial dependence) Distance along X (mm) Simulation Plato Data Rely on simulation for better accuracy than conventional treatment planning software Distance along Z (mm) Effects of source anisotropy Simulation Plato Transverse axis of the source Comparison with experimental data Longitudinal axis of the source Difficult to make direct measurements

Modeling the patient file Modeling geometry and materials from CT data Modeling a phantom 3D patient anatomy Acquisition of CT image source DICOM is the universal standard for sharing resources between heterogeneous and multi-vendor equipment file of any material (water, tissue, bone, muscle etc.) thanks to the flexibility of Geant4 materials package 3-D view Geant4-DICOM interface developed by L. Archambault, L. Beaulieu, V.-H. Tremblay (Univ. Laval and l'Hôtel-Dieu, Québec)

User-friendly interface to facilitate the usage in hospitals Dosimetric analysis Graphic visualisation of dose distributions Elaboration of isodose curves Web interface Application configuration Job submission

Dosimetry Simulation of energy deposit through Geant4 Low Energy Electromagnetic package to obtain accurate dose distribution Production threshold: 100 mm 2-D histogram with energy deposit in the plane containing the source Analysis of the energy deposit in the phantom resulting from the simulation Dose distribution Isodose curves AIDA + Anaphe Python for analysis for interactivity Abstract Interfaces for Data Analysis + any AIDA-compliant analysis system

MicroSelectron-HDR source Dosimetry Endocavitary brachytherapy MicroSelectron-HDR source Dosimetry Interstitial brachytherapy Dosimetry Superficial brachytherapy Bebig Isoseed I-125 source Leipzig applicator

Application configuration Fully configurable from the web Run modes: demo parallel on a cluster (under test) on the GRID (under development) Type of source Phantom configuration # events

Speed adequate for clinic use Parallelisation Transparent configuration in sequential or parallel mode Access to distributed computing resources Transparent access to the GRID through an intermediate software layer

Performance Endocavitary brachytherapy Superficial brachytherapy 1M events 61 minutes Superficial brachytherapy 1M events 65 minutes Interstitial brachytherapy 1M events 67 minutes on an “average” PIII machine, as an “average” hospital may own Monte Carlo simulation is not practically conceivable for clinical application, even if more precise

DIANE DIstributed ANalysis Environment Parallelisation Access to the GRID Transparent access to a distributed computing environment Hide complex details of underlying technology prototype for an intermediate layer between applications and the GRID R&D in progress for Large Scale Master-Worker Computing http://cern.ch/DIANE Developed by J. Moscicki, CERN

Performance: parallel mode on a local cluster preliminary: further optimisation in progress Performance: parallel mode on a local cluster Endocavitary brachytherapy 1M events 4 minutes 34’’ Superficial brachytherapy 1M events 4 minutes 25’’ Interstitial brachytherapy 5M events 4 minutes 36’’ on up to 50 workers, LSF at CERN, PIII machine, 500-1000 MHz Performance adequate for clinical application, but… it is not realistic to expect any hospital to own and maintain a PC farm

advanced software technologies and tools for radiotherapy Running on the GRID Via DIANE Same application code as running on a sequential machine or on a dedicated cluster completely transparent to the user A hospital is not required to own and maintain extensive computing resources to exploit the scientific advantages of Monte Carlo simulation for radiotherapy Any hospital – even small ones, or in less wealthy countries, that cannot afford expensive commercial software systems – may have access to advanced software technologies and tools for radiotherapy

Traceback from a run on CrossGrid testbed Resource broker running in Portugal Current #Grid setup (computing elements): 5000 events, 2 workers, 10 tasks (500 events each) - aocegrid.uab.es:2119/jobmanager-pbs-workq - bee001.ific.uv.es:2119/jobmanager-pbs-qgrid - cgnode00.di.uoa.gr:2119/jobmanager-pbs-workq - cms.fuw.edu.pl:2119/jobmanager-pbs-workq - grid01.physics.auth.gr:2119/jobmanager-pbs-workq - xg001.inp.demokritos.gr:2119/jobmanager-pbs-workq - xgrid.icm.edu.pl:2119/jobmanager-pbs-workq - zeus24.cyf-kr.edu.pl:2119/jobmanager-pbs-infinite - zeus24.cyf-kr.edu.pl:2119/jobmanager-pbs-long - zeus24.cyf-kr.edu.pl:2119/jobmanager-pbs-medium - zeus24.cyf-kr.edu.pl:2119/jobmanager-pbs-short - ce01.lip.pt:2119/jobmanager-pbs-qgrid Spain Poland Greece Portugal matchmaking CrossGrid computing elements

Extension and evolution Configuration of any brachytherapy technique any source type System extensible to any source configuration without changing the existing code General dosimetry system for radiotherapy extensible to other techniques plug-ins for external beams (factories for beam, geometry, physics...) treatment head hadrontherapy ... Plug-ins in progress

A medical accelerator for IMRT Build a simulation tool which determines the dose distributions given in a phantom by the head of a linear accelerator used for IMRT. Many algorithms were developed to estimate dose distributions, but even the most sophisticated ones resort to some approximations. These approximations might affect the outcome of dose calculation, especially in a complex treatment planning as IMRT. step and shoot The goal of this project is to build a simulation tool which determines the dose distributions given in a phantom by the head of a linear accelerator used for IMRT. We consider the step and shoot technics, in which the MLC stands still while the beam is on. In medical applications, dose distributions are useful because they are indicative of the quantity of radiation absorbed by tissues. Many algorithms were developed to estimate dose distributions, but the most sophisticated ones resort to some approximations too. These approximations might affect the outcome of dose calculation, especially in a complex treatment planning as IMRT. Treatment planning that implement such algorithms are expensive and not very accurate but quick and could be therefore used to have a on-line dose evaluation. IMRT generates tightly conforming dose distributions. This microscopic control allows IMRT to produce dose distribution patterns that are much closer to the desired patterns than possible previously

Work in progress... The user can choose the energy and standard deviation of the primary particles energy distribution (Gaussian) The primary particles (e-) leave from a point source with random direction (0˚< θ < 0.3˚) and a gaussian distribution The head components modeled include: target, primary and secondary collimators, vacuum window, flattening filter, ion chamber, mirror, vacuum and air Each pair of jaws can be rotated through an axis that is perpendicular to the beam axis The actual analysis produces some histograms from which the user can calculate the Percent Depth Dose (PDD) and the flatness at the following depths in the phantom: 15 mm, 50 mm, 100 mm and 200 mm. Rigorous software engineering is important in the medical physics domain to be maintainable over a large time scale to be extensible, to accommodate new user requirements (thanks to the OO technology)

Design

(very) Preliminary results Percent Depth Dose Flatness

talk by P. Cirrone on Monday CATANA hadrontherapy talk by P. Cirrone on Monday GEANT4 simulation Real hadron-therapy beam line

Dosimetry in interplanetary missions Aurora Programme vehicle concept Dose in astronaut resulting from Galactic Cosmic Rays

Conclusions precise, versatile, fast, user-friendly, low-cost dosimetry Geant4 + AIDA/Anaphe/PI + WWW + DIANE + GRID = Physics & software technology from HEP have a potential to address key issues in medical physics The social impact of technology transfer from HEP computing may be significant What is the support of HEP to technology transfer?

Thanks! This project has fostered a collaborative aggregation of contributions from many groups all over the world G. Cosmo (CERN, Geant4) L. Moneta, A. Pfeiffer(Anaphe/PI, CERN) J. Knobloch (CERN/IT) S. Agostinelli, S. Garelli (IST Genova) G. Ghiso, R. Martinelli (S. Paolo Hospital, Savona) G.A.P. Cirrone, G. Cuttone (INFN LNS, CATANA project) M.C. Lopes, L. Peralta, P. Rodrigues, A. Trindade (LIP Lisbon) L. Archambault, J.F. Carrier, L. Beaulieu, V.H. Tremblay (Univ. Laval) the authors F. Foppiano (IST) – medical physicist S. Guatelli, M. Piergentili (Univ. and INFN Genova) – students J. Moscicki (CERN) – computer scientist M.G. Pia (INFN Genova) – particle physicist